Preamble Reg. 1 Citation and scope. Schedule I Reg. 2 Definitions. Schedule Ii Reg. 3 Applicability to sub-assemblies. Schedule Iii Reg. 4 Penalties. Schedule Iv Reg. 5 Subject matter. Schedule V Reg. 6 Optionality. Schedule Vi Reg. 7 Essential requirements. Schedule Vii Reg. 8 Making available on the market and putting into use. Schedule Viii Reg. 9 Obligations of manufacturers. Schedule Ix Reg. 10 Compliance of a measuring instrument. Schedule X Reg. 11 Authorised representatives. Schedule Xi Reg. 12 Obligations of importers. Schedule Xii Reg. 13 Obligations of distributors. Schedule Xiii Reg. 14 Cases in which obligations of manufacturers apply to importers and distributors. Reg. 15 Identification of economic operators. Reg. 16 Presumption of conformity of measuring instruments. Reg. 17 Conformity assessment procedures. Reg. 18 Technical documentation. Reg. 19 EU declaration of conformity. Reg. 20 Conformity marking. Reg. 21 General principles of the CE marking and of the supplementary metrology marking. Reg. 22 Rules and conditions for affixing the CE marking and the supplementary metrology marking. Reg. 23 Notification. Reg. 24 Notifying authorities. Reg. 25 Requirements relating to notifying authorities. Reg. 26 Information obligation on notifying authorities. Reg. 27 Requirements relating to notified bodies. Reg. 28 Presumption of conformity of notified bodies. Reg. 29 Subsidiaries of and subcontracting by notified bodies. Reg. 30 Accredited in- house bodies Reg. 31 Application for notification. Reg. 32 Notification procedure. Reg. 33 Changes to notifications. Reg. 34 Challenge of the competence of notified bodies. Reg. 35 Operational obligations of notified bodies. Reg. 36 Appeal against decisions of notified bodies. Reg. 37 Information obligation on notified bodies. Reg. 38 Coordination of notified bodies. Reg. 39 Union market surveillance and control of measuring instruments entering the Union market. Reg. 40 Procedure for dealing with measuring instruments presenting a risk at national level. Reg. 41 Union safeguard procedure. Reg. 42 Compliant measuring instruments which present a risk. Reg. 43 Formal non- compliance. Reg. 44 Transitory provisions. Reg. 45 Committee procedure.

Statutory Text (59 Articles)

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

MEASURING INSTRUMENTS

SUBSIDIARY LEGISLATION 454.15

MEASURING INSTRUMENTS REGULATIONS

20th April, 2016 *

LEGAL NOTICE 122 of 2016.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) The title of these regulations is the Measuring

Instruments Regulations.

(2) These regulations implement Directive 2014/32/EU of the

European Parliament and of the Council of 26 February 2014 on the

harmonisation of the laws of the Member States relating to the

making available on the market of measuring instruments (recast).

(3) These regulations apply to the measuring instruments

defined in the instrument-specific Schedules III to XII (hereinafter

referred to as "instrument-specific Schedules") concerning water

meters (MI-001), gas meters and volume conversion devices (MI-

002), active electrical energy meters (MI-003), thermal energy

meters (MI-004), measuring systems for continuous and dynamic

measurement of quantities of liquids other than water (MI-005),

automatic weighing instruments (MI-006), taxi-meters (MI-007),

material measures (MI-008), dimensional measuring instruments

(MI-009) and exhaust gas analysers (MI-010).

(4) These regulations shall apply with respect to requirements

for electromagnetic immunity within the meaning of Article 2(3) of

Directive 2014/30/EU of the European Parliament and of the

Council (1) which provisions continue to apply with regard to

emission requirements.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

ESSENTIAL REQUIREMENTS

A measuring instrument shall provide a high level of metrological protection

in order that any party affected can have confidence in the result of measurement,

and shall be designed and manufactured to a high level of quality in respect of the

measurement technology and security of the measurement data.

The essential requirements that shall be met by measuring instruments are

set out below and are supplemented, where appropriate, by specific instrument

requirements in Schedules III to XII that provide more detail on certain aspects of

the general requirements.

The solutions adopted in the pursuit of the essential requirements shall take

account of the intended use of the instrument and any foreseeable misuse thereof.

Definitions

Measurand

The measurand is the particular quantity subject to

measurement.

Influence quantity

An influence quantity is a quantity that is not the

measurand but that affects the result of measurement.

Rated Operating Conditions The rated operating conditions are the values for the

measurand and influence quantities making up the

normal working conditions of an instrument.

Disturbance

An influence quantity having a value within the limits

specified in the appropriate requirement but outside

the specified rated operating conditions of the

measuring instrument. An influence quantity is a

disturbance if for that influence quantity the rated

operating conditions are not specified.

Critical change value

The critical change value is the value at which the

change in t he m easurem ent result is considered

undesirable.

Material Measure

A material measure is a device intended to reproduce

or supply in a permanent manner during its use one or

more known values of a given quantity.

Direct sales

A trading transaction is direct sales if

-

the measurement result serves as the basis for

the price to pay; and

-

at least one of the parties involved in the

transaction related to measurement is a

consumer or any other party requiring a

similar level of protection; and

-

all the parties in the transaction accept the

measurement result at that time and place.

Climatic environments

Climatic environments are the conditions in which

measuring instruments may be used. To cope with

climatic differences between the Member States, a

range of temperature limits has been defined.

Utility

A utility is regarded as a supplier of electricity, gas,

thermal energy or water.

ESSENTIAL REQUIREMENTS

1. Allowable Errors

(1) Under rated operating conditions and in the absence of a disturbance, the

error of measurement shall not exceed the maximum permissible error (MPE) value

as laid down in the appropriate instrument-specific requirements.

Unless stated otherwise in the instrument-specific schedules, MPE is

expressed as a bilateral value of the deviation from the true measurement value.

(2) Under rated operating conditions and in the presence of a disturbance, the

performance requirement shall be as laid down in the appropriate instrument-specific

requirements.

Where the instrument is intended to be used in a specified permanent

continuous electromagnetic field the permitted performance during the radiated

electromagnetic field-amplitude modulated test shall be within MPE.

(3) The manufacturer shall specify the climatic, mechanical and electromagnetic

environments in which the instrument is intended to be used, power supply and other

influence quantities likely to affect its accuracy, taking account of the requirements

laid down in the appropriate instrument-specific schedules.

(3.1) Climatic environments

The manufacturer shall specify the upper temperature limit and the lower

temperature limit from any of the values in Table 1 unless otherwise specified in the

Schedules III to XII, and indicate whether the instrument is designed for condensing

or non-condensing humidity as well as the intended location for the instrument, i.e.

open or closed.

Table 1

Temperature Limits

Upper temperature limit

30°C

40°C

55°C

70°C

Lower temperature limit

5°C

- 10°C

- 25°C

- 40°C

(3.2). (a) Mechanical environments are classified into classes M1 to M3 as

described below.

This class applies to instruments used in locations with vibration and

shocks of low significance, e.g. for instruments fastened to light

supporting structures subject to negligible vibrations and shocks

transmitted from local blasting or pile-driving activities, slamming

doors, etc.

This class applies to instruments used in locations with significant or

high levels of vibration and shock, e.g. transmitted from machines and

passing vehicles in the vicinity or adjacent to heavy machines,

conveyor belts, etc.

This class applies to instruments used in locations where the level of

vibration and shock is high and very high, e.g. for instruments mounted

directly on machines, conveyor belts, etc.

(b) The following influence quantities shall be considered in relation with

mechanical environments

-

Vibration;

-

Mechanical shock.

(3.3). (a) Electromagnetic environments are classified into classes E1, E2 or E3 as

described below, unless otherwise laid down in the appropriate instrument-specific

schedules.

This class applies to instruments used in locations with electromagnetic

disturbances corresponding to those likely to be found in residential,

commercial and light industrial buildings.

This class applies to instruments used in locations with electromagnetic

disturbances corresponding to those likely to be found in other

industrial buildings.

This class applies to instruments supplied by the battery of a vehicle.

Such instruments shall comply with the requirements of E2 and the

following additional requirements

-

voltage reductions caused by energising the starter-motor

circuits of internal combustion engines,

-

load dump transients occurring in the event of a discharged

battery being disconnected while the engine is running.

(b) The following influence quantities shall be considered in relation with

electromagnetic environments

-

voltage interruptions;

-

short voltage reductions;

-

voltage transients on supply lines and/or signal lines;

-

electrostatic discharges

-

radio frequency electromagnetic fields;

-

conducted radio frequency electromagnetic fields on supply lines and/or

signal lines;

-

surges on supply lines and/or signal lines.

(3.4) Other influence quantities to be considered, where appropriate, are

-

voltage variation;

-

mains frequency variation;

-

power frequency magnetic fields;

-

any other quantity likely to influence in a significant way the accuracy

of the instrument.

(4) When carrying out the tests as envisaged in these regulations, the following

points shall apply

(4.1). Basic rules for testing and the determination of errors

Essential requirements specified in points 1.1 and 1.2 shall be verified for

each relevant influence quantity. Unless otherwise specified in the appropriate

instrument-specific schedules, these essential requirements apply when each

influence quantity is applied and its effect evaluated separately, all other influence

quantities being kept relatively constant at their reference value.

Metrological tests shall be carried out during or after the application of the

influence quantity, whichever condition corresponds to the normal operational status

of the instrument when that influence quantity is likely to occur.

(4.2) Ambient humidity

(a) According to the climatic operating environment in which the

instrument is intended to be used either the damp heat-steady state (non-

condensing) or damp heat cyclic (condensing) test may be appropriate.

(b) The damp heat cyclic test is appropriate where condensation is

important or when penetration of vapour will be accelerated by the

effect of breathing. In conditions where non-condensing humidity is a

factor the damp- heat steady state is appropriate.

2. Reproducibility

The application of the same measurand in a different location or by a

different user, all other conditions being the same, shall result in the close agreement

of successive measurements. The difference between the measurement results shall

be small when compared with the MPE.

3. Repeatability

The application of the same measurand under the same conditions of

measurement shall result in the close agreement of successive measurements. The

difference between the measurement results shall be small when compared with the

MPE.

4. Discrimination and Sensitivity

A measuring instrument shall be sufficiently sensitive and the

discrimination threshold shall be sufficiently low for the intended measurement task.

5. Durability

A measuring instrument shall be designed to maintain an adequate stability

of i t s m e t r o l og i c a l c h a r a c t e r i s t i c s o v e r a p e r i o d o f t im e es t i ma t e d b y th e

manufacturer, provided that it is properly installed, maintained and used according

to the manufacturer’s instruction when in the environmental conditions for which it

is intended.

6. Reliability

A measuring instrument shall be designed to reduce as far as possible the

effect of a defect that would lead to an inaccurate measurement result, unless the

presence of such a defect is obvious.

7. Suitability

(1) A measuring instrument shall have no feature likely to facilitate fraudulent

use, whereas possibilities for unintentional misuse shall be minimal.

(2) A measuring instrument shall be suitable for its intended use taking account

of the practical working conditions and shall not require unreasonable demands of

the user in order to obtain a correct measurement result.

(3) The errors of a utility measuring instrument at flows or currents outside the

controlled range shall not be unduly biased.

(4) Where a measuring instrument is designed for the measurement of values of

the measurand that are constant over time, the measuring instrument shall be

insensitive to small fluctuations of the value of the measurand, or shall take

appropriate action.

(5) A measuring instrument shall be robust and its materials of construction

shall be suitable for the conditions in which it is intended to be used.

(6) A measuring instrument shall be designed so as to allow the control of the

measuring tasks after the instrument has been placed on the market and put into use.

If necessary, special equipment or software for this control shall be part of the

instrument. The test procedure shall be described in the operation manual.

When a measuring instrument has associated software which provides other

functions besides the measuring function, the software that is critical for the

metrological characteristics shall be identifiable and shall not be inadmissibly

influenced by the associated software.

8. Protection against corruption

(1) The metrological characteristics of a measuring instrument shall not be

influenced in any inadmissible way by the connection to it of another device, by any

feature of the connected device itself or by any remote device that communicates

with the measuring instrument.

(2) A hardware component that is critical for metrological characteristics shall

be designed so that it can be secured.

Security measures foreseen shall provide for evidence of an intervention.

(3) Software that is critical for metrological characteristics shall be identified as

such and shall be secured.

Software identification shall be easily provided by the measuring

instrument. Evidence of an intervention shall be available for a reasonable period of

time.

(4) Measurement data, software that is critical for measurement characteristics

and metrologically important parameters stored or transmitted shall be adequately

protected against accidental or intentional corruption.

(5) For utility measuring instruments the display of the total quantity supplied

or the displays from which the total quantity supplied can be derived, whole or

partial reference to which is the basis for payment, shall not be able to be reset

during use.

9. Information to be borne by and to accompany the instrument

(1)

A measuring instrument shall bear the following inscriptions

(a) manufacturer’s name, registered trade name or registered trade mark;

(b) information in respect of its accuracy; and, where applicable;

(c) information in respect of the conditions of use;

(d) measuring capacity;

(e) measuring range;

(f)

identity marking;

(g) number of the EU-type examination certificate or the EU design

examination certificate;

(h) information whether or not additional devices providing metrological

results comply with the provisions of these regulations on legal

metrological control.

(2) An instrument of dimensions too small or of too sensitive a composition to

allow it to bear the relevant information shall have its packaging, if any, and the

accompanying documents required by the provisions of these regulations suitably

marked.

(3) The instrument shall be accompanied by information on its operation, unless

the simplicity of the measuring instrument makes this unnecessary. Information shall

be easily understandable and shall include where relevant

(a) rated operating conditions;

(b) mechanical and electromagnetic environment classes;

(c) the upper and lower temperature limit, whether condensation is possible

or not, open or closed location;

(d) instructions for

installation,

maintenance,

repairs,

permissible

adjustments;

(e) instructions for correct operation and any special conditions of use;

(f)

conditions for compatibility with interfaces, sub-assemblies or

measuring instruments.

(4) Groups of identical measuring instruments used in the same location or used

for utility measurements do not necessarily require individual instruction manuals.

(5) Unless specified otherwise in an instrument-specific schedule, the scale

interval for a measured value shall be in the form 1 × 10n, 2 × 10n, or 5 × 10n, where

n is any integer or zero. The unit of measurement or its symbol shall be shown close

to the numerical value.

(6) A material measure shall be marked with a nominal value or a scale,

accompanied by the unit of measurement used.

(7) The units of measurement used and their symbols shall be in accordance

with the provisions of Union legislation on units of measurement and their symbols.

(8) All marks and inscriptions required under any requirement shall be clear,

non-erasable, unambiguous and non- transferable.

10. Indication of result

(1)

Indication of the result shall be by means of a display or hard copy.

(2) The indication of any result shall be clear and unambiguous and

accompanied by such marks and inscriptions necessary to inform the user of the

significance of the result. Easy reading of the presented result shall be permitted

under normal conditions of use. Additional indications may be shown provided they

cannot be confused with the metrologically controlled indications.

(3) In the case of hard copy the print or record shall also be easily legible and

non-erasable.

(4) A measuring instrument for direct sales trading transactions shall be

designed to present the measurement result to both parties in the transaction when

installed as intended. When critical in case of direct sales, any ticket provided to the

consumer by an ancillary device not complying with the appropriate requirements of

these regulations shall bear appropriate restrictive information.

(5) Whether or not a measuring instrument intended for utility measurement

purposes can be remotely read it shall in any case be fitted with a metrological

controlled display accessible without tools to the consumer. The reading of this

display is the measurement result that serves as the basis for the price to pay.

11. Further processing of data to conclude the trading transaction

(1) A measuring instrument other than a utility measuring instrument shall

record by a durable means the measurement result accompanied by information to

identify the particular transaction, when

(a) the measurement is non-repeatable; and

(b) the measuring instrument is normally intended for use in the absence of

one of the trading parties.

(2) Additionally, a durable proof of the measurement result and the information

to identify the transaction shall be available on request at the time the measurement

is concluded.

12. Conformity evaluation

A measuring instrument shall be designed so as to allow ready evaluation of

its conformity with the appropriate requirements of these regulations.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

In these regulations, unless the context otherwise requires

"accreditation" means accreditation as defined in Article 2 of

Regulation (EC) no 765/2008;

"authorised representative" means any natural or legal person

established within the Union who has received a written mandate

from a manufacturer to act on his behalf in relation to specified

tasks;

"Authority" means the Standards and Metrology Institute of the

Malta Competition and Consumer Affairs Authority;

"CE marking" means a marking by which the manufacturer

indicates that the measuring instrument is in conformity with the

applicable requirements set out in Union harmonisation legislation

providing for its affixing;

"conformity assessment" means the process demonstrating

whether the essential requirements of these regulations relating to a

measuring instrument have been fulfilled;

"conformity assessment body" means a body that performs

conformity assessment activities including calibration, testing,

*

see regulation 1(2) of these Regulations, as originally promulgated.

MEASURING INSTRUMENTS

certification and inspection.

"distributor" means any natural or legal person in the supply

chain, other than the manufacturer or the importer, who makes a

measuring instrument available on the market;

"economic operators" means the manufacturer, the authorised

representative, the importer and the distributor;

"harmonised standard" means harmonised standard as defined in

regulation 2 of Regulation (EU) No 1025/2012;

"importer" means any natural or legal person established within

the Union who places a measuring instrument from a third country

on the Union market;

"legal metrological control" means the control of the

measurement tasks intended for the field of application of a

measuring instrument, for reasons of public interest, public health,

public safety, public order, protection of the environment, levying

of taxes and duties, protection of the consumers and fair trading;

"making available on the m arket " m eans any supply of a

measuring instrument for distribution or use on the Union market in

the course of a commercial activity, whether in return for payment

or free of charge;

"manufacturer" means any natural or legal person who

manufactures a measuring instrument or has a measuring

instrument designed or manufactured, and markets that measuring

instrument under his name or trade mark or puts it into use for his

own purposes;

"measuring instrument" means any device or system with a

measurement according to the provisions of these regulations;

"Member State" means a Member State of the European Union;

"Minister" means the Minister responsible for standards and

metrology and includes, to the extent of the authority given, any

person authorised by the Minister to act in that behalf for any

purpose of these regulations;

"National Accreditation Board (Malta)" means the National

Accreditation Board (Malta) established in terms of regulation 3 of

the National Accreditation Board (Malta) (Establishment)

Regulations;

"normative document" means a document containing technical

specifications adopted by the International Organisation of Legal

Metrology * ;

"placing on the market" means the first making available of a

measuring instrument on the Union market;

"putting into use" means the first use of a measuring instrument

intended for the end-user for t he purposes for which it was

intended;

"recall" means any measure aimed at achieving the return of a

*

https://www.oiml.org/en

MEASURING INSTRUMENTS

measuring instrument that has already been made available to the

end-user;

"sub-assembly" means a hardware device, mentioned as such in

the instrument-specific schedules that functions independently and

m a k e s up a m e a s u r in g i n s t ru m e n t t o ge th e r wi t h o th e r su b-

assemblies with which it is compatible, or with a measuring

instrument with which it is compatible;

"technical specification" means a document that prescribes

technical requirements to be fulfilled by a measuring instrument;

"the Union" means the European Union as referred to article 2 of

the European Union Act;

"Union harmonisation legislation" means any Union legislation

harmonising the conditions for the marketing of products; and

"withdrawal" means any measure aimed at preventing a

measuring instrument in the supply chain from being made

available on the market.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

MODULE A: INTERNAL PRODUCTION CONTROL

1. ‘Internal production control’ is the conformity assessment procedure

whereby the manufacturer fulfils the obligations laid down in points 2, 3 and 4, and

ensures and declares on his sole responsibility that the measuring instruments

concerned satisfy the requirements of these regulations that apply to them.

2. Technical documentation

The manufacturer shall establish the technical documentation as described in

regulation 18. The documentation shall make it possible to assess the instrument’s

conformity to the relevant requirements, and shall include an adequate analysis and

assessment of the risk(s). The technical documentation shall specify the applicable

requi rements and cover, as far as r el ev an t for the assessment, the desig n,

manufacture and operation of the instrument.

3. Manufacturing

The manufacturer shall take all measures necessary so that the

manufacturing process and its monitoring ensure compliance of the manufactured

instruments with the technical documentation referred to in point 2 and with the

requirements of these regulations that apply to them.

4. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking and the supplementary

metrology marking set out in these regulations to each individual measuring

instrument that satisfies the applicable requirements of these regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

an instrument model and keep it together with the technical documentation at the

disposal of the national authorities for 10 years after the measuring instrument has

been placed on the market. The EU declaration of conformity shall identify the

instrument for which it was drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market. However,

this requirement may be interpreted as applying to a batch or consignment rather

than individual instruments in those cases where a large number of instruments is

delivered to a single user.

5. Authorised representative

The manufacturer’s obligations set out in point 4 may be fulfilled by his

authorised representative, on his behalf and under his responsibility, provided that

they are specified in the mandate.

MODULE A2: INTERNAL PRODUCTION CONTROL

PLUS

SUPERVISED INSTRUMENT CHECKS AT RANDOM INTERVALS

Internal production control plus supervised instrument checks at random

intervals is the conformity assessment procedure whereby the manufacturer fulfils

the obligations laid down in points 2, 3, 4, and 5, and ensures and declares on his

s o l e r e s p o n s i b i l i t y t h a t t h e m e a s u r i n g i n s t r u m e n t s c o n c e r n e d s at i s f y t h e

requirements of these regulations that apply to them.

2. Technical documentation

The manufacturer shall establish the technical documentation as described in

regulation 18. The documentation shall make it possible to assess the instrument’s

conformity with the relevant requirements, and shall include an adequate analysis

and assessment of the risk(s). The technical documentation shall specify the

applicable requirements and cover, as far as relevant for the assessment, the design,

manufacture and operation of the instrument.

3. Manufacturing

The manufacturer shall take all measures necessary so that the

manufacturing process and its monitoring ensure compliance of the manufactured

instruments with the technical documentation referred to in point 2 and with the

requirements of these regulations that apply to them.

4. Instrument checks

At the choice of the manufacturer, either an accredited in-house body or a

notified body, chosen by the manufacturer, shall carry out instrument checks or have

them carried out at random intervals determined by the body, in order to verify the

quality of the internal checks of the instrument, taking into account, inter alia, the

technological complexity of the instruments and the quantity of production. An

adequate sample of the final measuring instruments, taken on site by the body before

the placing on the market, shall be examined and appropriate tests as identified by

the relevant parts of the harmonised standard, and/or normative document, and/or

equivalent tests set out in other relevant technical specifications, shall be carried out

to verify the conformity of the instruments with the relevant requirements of these

regulations. In the absence of a relevant harmonised standard or normative

document, the accredited in-house body or notified body concerned shall decide on

the appropriate tests to be carried out.

In those cases where a relevant number of instruments in the sample do not

conform to an acceptable quality level, the accredited in-house body or notified body

shall take appropriate measures.

Where the tests are carried out by a notified body, the manufacturer shall,

under the responsibility of the notified body, affix the notified body’s identification

number during the manufacturing process.

5. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking and the supplementary

metrology marking set out in these regulations to each individual instrument that

satisfies the applicable requirements of these regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

an instrument model and keep it together with the technical documentation at the

disposal of the national authorities for 10 years after the instrument has been placed

MEASURING INSTRUMENTS

on the market. The EU declaration of conformity shall identify the instrument for

which it was drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market. However,

this requirement may be interpreted as applying to a batch or consignment rather

than individual instruments in those cases where a large number of instruments is

delivered to a single user.

6. Authorised representative

The manufacturer’s obligations set out in point 5 may be fulfilled by his

authorised representative, on his behalf and under his responsibility provided that

they are specified in the mandate.

MODULE B: EU- TYPE EXAMINATION

1. ‘EU-type examination’ is the part of a conformity assessment procedure in

which a notified body examines the technical design of an instrument and verifies

and attests that the technical design of the instrument meets the requirements of

these regulations that apply to it.

2. EU-type examination may be carried out in either of the following manners

(a) examination of a specimen, representative of the production envisaged,

of the complete measuring instrument (production type),

(b) assessment of the adequacy of the technical design of the instrument

through examination of the technical documentation and supporting

evidence referred to in point 3, plus examination of specimens,

representative of the production envisaged, of one or more critical parts

of the instrument (combination of production type and design type);

(c) assessment of the adequacy of the technical design of the instrument

through examination of the technical documentation and supporting

evidence referred to in point 3, without examination of a specimen

(design type).

The notified body decides on the appropriate manner and the specimens

3. The manufacturer shall lodge an application for EU-type examination with a

single notified body of his choice.

The application shall include

(a) the name and address of the manufacturer and, if the application is

lodged by the authorised representative, his name and address as well;

(b) a written declaration that the same application has not been lodged with

any other notified body;

(c) the technical documentation as described in regulation 16. The technical

documentation shall make it possible to assess the instrument’s

conformity with the applicable requirements of these regulations and

shall include an adequate analysis and assessment of the risk(s). The

technical documentation shall specify the applicable requirements and

cover, as far as relevant for the assessment, the design, manufacture and

operation of the instrument.

The application shall in addition contain, wherever applicable;

(d) the specimens, representative of the production envisaged. The notified

body may request further specimens if needed for carrying out the test

programme;

(e) the supporting evidence for the adequacy of the technical design

solution. This supporting evidence shall mention any documents that

have been used, in particular where the relevant harmonised standards,

and/or normative documents have not been applied in full. The

supporting evidence shall include, where necessary, the results of tests

carried out in accordance with other relevant technical specifications by

the appropriate laboratory of the manufacturer, or by another testing

laboratory on his behalf and under his responsibility.

4. The notified body shall: For the instrument

(1) examine the technical documentation and supporting evidence to assess the

adequacy of the technical design of the instrument;

For the specimen(s)

(2) verify that the specimen(s) have been manufactured in conformity with the

technical documentation and identify the elements which have been designed in

accordance with the applicable provisions of the relevant harmonised standards and/

or normative documents, as well as the elements which have been designed in

accordance with other relevant technical specifications;

(3) carry out appropriate examinations and tests, or have them carried out, to

check whether, where the manufacturer has chosen to apply the solutions in the

relevant harmonised standards and normative documents, these have been applied

correctly;

(4) carry out appropriate examinations and tests, or have them carried out, to

check whether, where the solutions in the relevant harmonised standards, and/or

normative docum en ts have not b een applied, the solu tions ado pted by the

manufacturer applying other relevant technical specifications meet the

corresponding essential requirements of these regulations

(5) agree with the manufacturer on the location where the examinations and

tests will be carried out.

For the other parts of the measuring instrument

(6) examine the technical documentation and supporting evidence to assess the

adequacy of the technical design of the other parts of the measuring instrument.

5. The notified body shall draw up an evaluation report that records the

activities undertaken in accordance with point 4 and their outcomes. Without

prejudice to its obligations vis-à-vis, the notifying authorities, the notified body

shall release the content of that report, in full or in part, only with the agreement of

the manufacturer.

6. Where the type meets the requirements of these regulations, the notified

body shall issue an EU-type examination certificate to the manufacturer. That

certificate shall contain the name and address of the manufacturer, the conclusions of

the examination, the conditions (if any) for its validity and the necessary data for

identification of the approved type. The EU-type examination certificate may have

one or more schedules attached.

The EU-type examination certificate and its schedules shall contain all

relevant information to allow the conformity of manufactured measuring instruments

with the examined type to be evaluated and to allow for in-service control. In

particular, to allow the conformity of manufactured instruments to be evaluated with

the examined type regarding the reproducibility of their metrological performances,

when they are properly adjusted using appropriate means, content shall include

-

the metrological characteristics of the type of instrument

-

measures required for ensuring the integrity of the instruments (sealing,

identification of software, etc.);

-

information on other elements necessary for the identification of the

instruments and to check their visual external conformity to type;

-

if appropriate, any specific information necessary to verify the

characteristics of manufactured instruments;

-

in the case of a sub-assembly, all necessary information to ensure the

compatibility with other sub-assemblies or measuring instruments.

The EU-type examination certificate shall have a validity of 10 years from

the date of its issue, and may be renewed for subsequent periods of 10 years each.

Where the type does not satisfy the applicable requirements of these

regulations, the notified body shall refuse to issue an EU-type examination

certificate and shall inform the applicant accordingly, giving detailed reasons for its

refusal.

7. The notified body shall keep itself apprised of any changes in the generally

acknowledged state of the art which indicate that the approved type may no longer

comply with the applicable requirements of these regulations, and shall determine

whether such changes require further investigation. If so, the notified body shall

inform the manufacturer accordingly.

8. The manufacturer shall inform the notified body that holds the technical

documentation relating to the EU-type examination certificate of all modifications to

the approved type that may affect the conformity of the instrument with the essential

requirements of these regulations or the conditions for validity of that certificate

Such modifications shall require additional approval in the form of an addition to the

original EU-type examination certificate.

9. Each notified body shall inform the Authority concerning the EU-type

examination certificates and/or any additions thereto which it has issued or

withdrawn, and shall, periodically or upon request, make available to the Authority

the list of such certificates and/or any additions thereto refused, suspended or

otherwise restricted.

The Commission, the Member States and the other notified bodies may, on

request, obtain a copy of the EU-type examination certificates and/or additions

thereto. On request, the Commission and the Member States may obtain a copy of

the technical documentation and the results of the examinations carried out by the

notified body.

The notified body shall keep a copy of the EU-type examination certificate,

its schedules and additions, as well as the technical file including the documentation

submitted by the manufacturer until the expiry of the validity of that certificate.

10. The manufacturer shall keep a copy of the EU-type examination certificate,

its schedules and additions together with the technical documentation at the disposal

of the national authorities for 10 years after the instrument has been placed on the

market.

11. The manufacturer’s authorised representative may lodge the application

referred to in point 3 and fulfil the obligations set out in points 8 and 10, provided

that they are specified in the mandate.

MODULE C: CONFORMITY TO TYPE BASED ON INTERNAL

PRODUCTION CONTROL

1. Conformity to type based on internal production control is the part of a

conformity assessment procedure whereby the manufacturer fulfils the obligations

laid down in points 2 and 3, and ensures and declares that the measuring instruments

concerned are in conformity with the type described in the EU-type examination

certificate and satisfy the requirements of these regulations that apply to them.

2. Manufacturing

The manufacturer shall take all measures necessary so that the

manufacturing process and its monitoring ensure conformity of the manufactured

measuring instruments with the approved type described in the EU-type examination

certificate and with the requirements of these regulations that apply to them.

3. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking and the supplementary

metrology marking set out in these regulations to each individual instrument that is

in conformity with the type described in the EU-type examination certificate and

satisfies the applicable requirements of these regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

each instrument model and keep it at the disposal of the national authorities for 10

years after the instrument has been placed on the market. The EU declaration of

conformity shall identify the instrument model for which it was drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market. However,

this requirement may be interpreted as applying to a batch or consignment rather

than individual instruments in those cases where a large number of instruments is

delivered to a single user.

4. Authorised representative

The manufacturer’s obligations set out in point 3 may be fulfilled by his

authorised representative, on his behalf and under his responsibility, provided that

they are specified in the mandate.

MODULE C2: CONFORMITY TO TYPE BASED ON INTERNAL

PRODUCTION CONTROL PLUS SUPERVISED INSTRUMENT CHECKS AT

RANDOM INTERVALS

1. Conformity to type based on internal production control plus supervised

instrument checks at random intervals is the part of a conformity assessment

procedure whereby the manufacturer fulfils the obligations laid down in points 2, 3

and 4, and ensures and declares on his sole responsibility that the measuring

instruments concerned are in conformity with the type described in the EU-type

examination certificate and satisfy the requirements of these regulations that apply

to them.

2. Manufacturing

The manufacturer shall take all measures necessary so that the

manufacturing process and its monitoring ensure conformity of the manufactured

measuring instruments with the type described in the EU-type examination

certificate and with the requirements of these regulations that apply to them.

3. Instrument checks

At the choice of the manufacturer, either an accredited in-house body or a

notified body, chosen by the manufacturer, shall carry out instrument checks or have

them carried out at random intervals determined by the body, in order to verify the

quality of the internal checks on the instrument, taking into account, inter alia, the

technological complexity of the measuring instruments and the quantity of

production. An adequate sample of the final measuring instrument, taken on site by

the accredited in-house body or by the notified body before the placing on the

market, shall be examined and appropriate tests, as identified by the relevant parts of

the harmonised standards, and/or normative documents, and/or equivalent tests set

out in other relevant technical specifications, shall be carried out to verify the

conformity of the instrument with the type described in the EU-type examination

certificate and with the relevant requirements of these regulations.

Where a sample does not conform to an acceptable quality level, the

accredited in-house body or notified body shall take appropriate measures.

The acceptance sampling procedure to be applied is intended to determine

whether the manufacturing process of the instrument performs within acceptable

limits, with a view to ensuring conformity of the instrument.

Where the tests are carried out by a notified body, the manufacturer shall,

under the responsibility of the notified body, affix the notified body’s identification

number during the manufacturing process.

4. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking, and the supplementary

metrology marking set out in these regulations to each individual measuring

instrument that is in conformity with the type described in the EU-type examination

certificate and satisfies the applicable requirements of these regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

each instrument model and keep it at the disposal of the national authorities for 10

years after the instrument has been placed on the market. The EU declaration of

conformity shall identify the instrument model for which it has been drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market. However,

this requirement may be interpreted as applying to a batch or consignment rather

than individual instruments in those cases where a large number of instruments is

delivered to a single user.

5. Authorised representative

The manufacturer’s obligations set out in point 4 may be fulfilled by his

authorised representative, on his behalf and under his responsibility, provided that

they are specified in the mandate.

MODULE D: CONFORMITY TO TYPE

BASED

ON

QUALITY

ASSURANCE OF THE PRODUCTION PROCESS

1. Conformity to type based on quality assurance of the production process is

the part of a conformity assessment procedure whereby the manufacturer fulfils the

obligations laid down in points 2 and 5, and ensures and declares on his sole

responsibility that the measuring instruments concerned are in conformity with the

type described in the EU-type examination certificate and satisfy the requirements of

these regulations that apply to them.

2. Manufacturing

The manufacturer shall operate an approved quality system for production,

final product inspection and testing of the measuring instruments concerned as

specified in point 3 and shall be subject to surveillance as specified in point 4.

3. Quality system

(1) The manufacturer shall lodge an application for assessment of his quality

system with a notified body of his choice, for the measuring instruments concerned.

The application shall include

(a) the name and address of the manufacturer and, if the application is

lodged by the authorised representative, his name and address as well,

(b) a written declaration that the same application has not been lodged with

any other notified body,

(c) all relevant information for the instrument category envisaged;

(d) the documentation concerning the quality system;

(e) the technical documentation of the approved type and a copy of the EU-

type examination certificate.

(2) The quality system shall ensure that the measuring instruments are in

conformity with the type described in the EU- type examination certificate and

comply with the requirements of these regulations that apply to them.

All the elements, requirements and provisions adopted by the manufacturer

shall be documented in a systematic and orderly manner in the form of written

policies, procedures and instructions. This quality system documentation shall

permit a consistent interpretation of the quality programmes, plans, manuals and

records.

It shall, in particular, contain an adequate description of

(a) the quality objectives and the organisational structure, responsibilities

and powers of the management with regard to product quality;

(b) the corresponding manufacturing, quality control and quality assurance

techniques, processes and systematic actions that will be used;

(c) the examinations and tests that will be carried out before, during, and

after manufacture, and the frequency with which they will be carried

out;

(d) the quality records, such as inspection reports and test data, calibration

data, qualification reports on the personnel concerned;

(e) the means of monitoring the achievement of the required product

quality and the effective operation of the quality system.

(3) The notified body shall assess the quality system to determine whether it

satisfies the requirements referred to in point 3.2.

It shall presume conformity with those requirements in respect of the

elements of the quality system that comply with the corresponding specifications of

the relevant harmonised standard.

In addition to experience in quality management systems, the auditing team

shall have at least one member with experience of evaluation in the relevant

instrument field and instrument technology concerned, and knowledge of the

applicable requirements of these regulations. The audit shall include an assessment

visit to the manufacturer’s premises.

The auditing team shall review the technical documentation referred to in

point (e) of point 3.1, to verify the manufacturer’s ability to identify the relevant

requirements of these regulations and to carry out the necessary examinations with a

view to ensuring compliance of the instrument with those requirements.

The decision shall be notified to the manufacturer. The notification shall

contain the conclusions of the audit and the reasoned assessment decision.

(4) The manufacturer shall undertake to fulfil the obligations arising out of the

quality system as approved and to maintain it so that it remains adequate and

efficient.

(5) The manufacturer shall keep the notified body that has approved the quality

system informed of any intended change of the quality system.

The notified body shall evaluate any proposed changes and decide whether

the modified quality system will continue to satisfy the requirements referred to in

point 3.2 or whether a re-assessment is necessary.

It shall notify the manufacturer of its decision. The notification shall contain

the conclusions of the examination and the reasoned assessment decision.

4. Surveillance under the responsibility of the notified body

(1) The purpose of surveillance is to make sure that the manufacturer duly

fulfils the obligations arising out of the approved quality system.

(2) The manufacturer shall, for assessment purposes, allow the notified body

access to the manufacture, inspection, testing and storage sites, and shall provide it

with all necessary information, in particular

(a) the quality system documentation;

(b) the quality records, such as inspection reports and test data, calibration

data, qualification reports on the personnel concerned.

(3) The notified body shall carry out periodic audits to make sure that the

manufacturer maintains and applies the quality system and shall provide the

manufacturer with an audit report.

(4) In addition, the notified body may pay unexpected visits to the

manufacturer. During such visits the notified body may, if necessary, carry out

instrument tests, or have them carried out, in order to verify that the quality system

is functioning correctly. The notified body shall provide the manufacturer with a

visit report and, if tests have been carried out, with a test report.

5. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking and the supplementary

metrology marking set out in these regulations, and, under the responsibility of the

notified body referred to in point 3.1, the latter ’s identification number to each

individual measuring instrument that is in conformity with the type described in the

EU-type examination certificate and satisfies the applicable requirements of these

regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

each instrument model and keep it at the disposal of the national authorities for 10

years after the instrument has been placed on the market. The EU declaration of

conformity shall identify the instrument model for which it has been drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market. However,

this requirement may be interpreted as applying to a batch or consignment rather

than individual instruments in those cases where a large number of instruments are

delivered to a single user.

6. The manufacturer shall, for a period ending 10 years after the instrument has

been placed on the market, keep at the disposal of the national authorities

(a) the documentation referred to in point 3.1,

(b) the information relating to the change referred to in point 3.5, as

approved;

(c) the decisions and reports from the notified body referred to in points

3.5, 4.3 and 4.4.

7. Each notified body shall inform the Authority of quality system approvals

issued or withdrawn, and shall, periodically or upon request, make available to the

Authority the list of quality system approvals refused, suspended or otherwise

restricted.

8. Authorised representative

The manufacturer’s obligations set out in points 3.1, 3.5, 5 and 6 may be

fulfilled by his authorised representative, on his behalf and under his responsibility,

provided that they are specified in the mandate.

MODULE D1: QUALITY ASSURANCE OF THE PRODUCTION

PROCESS

1. Quality assurance of the production process is the conformity assessment

procedure whereby the manufacturer fulfils the obligations laid down in points 2, 4

and 7, and ensures and declares on his sole responsibility that the measuring

instruments concerned satisfy the requirements of these regulations that apply to

them.

2. Technical documentation

The manufacturer shall establish the technical documentation as described in

regulation 18. The documentation shall make it possible to assess the instrument’s

conformity with the relevant requirements, and shall include an adequate analysis

and assessment of the risk(s). The technical documentation shall specify the

applicable requirements and cover, as far as relevant for the assessment, the design,

manufacture and operation of the instrument.

3. The manufacturer shall keep the technical documentation at the disposal of

the national authorities for 10 years after the instrument has been placed on the

market.

4. Manufacturing

The manufacturer shall operate an approved quality system for production,

final product inspection and testing of the measuring instruments concerned as

specified in point 5 and shall be subject to surveillance as specified in point 6.

5. Quality system

(1) The manufacturer shall lodge an application for assessment of his quality

system with a notified body of his choice, for the measuring instruments concerned.

The application shall include

(a) the name and address of the manufacturer and, if the application is

lodged by the authorised representative, his name and address as well;

(b) a written declaration that the same application has not been lodged with

any other notified body;

(c) all relevant information for the instrument category envisaged;

(d) the documentation concerning the quality system;

(e) the technical documentation referred to in point 2.

(2) The quality system shall ensure compliance of the measuring instruments

with the requirements of these regulations that apply to them.

All the elements, requirements and provisions adopted by the manufacturer

shall be documented in a systematic and orderly manner in the form of written

policies, procedures and instructions. This quality system documentation shall

permit a consistent interpretation of the quality programmes, plans, manuals and

records.

It shall, in particular, contain an adequate description of

(a) the quality objectives and the organisational structure, responsibilities

and powers of the management with regard to product quality;

(b) the corresponding manufacturing, quality control and quality assurance

techniques, processes and systematic actions that will be used;

(c) the examinations and tests that will be carried out before, during, and

after manufacture, and the frequency with which they will be carried

out;

(d) the quality records, such as inspection reports and test data, calibration

data, qualification reports on the personnel concerned;

(e) the means of monitoring the achievement of the required product

quality and the effective operation of the quality system.

(3) The notified body shall assess the quality system to determine whether it

satisfies the requirements referred to in point 5.2.

It shall presume conformity with those requirements in respect of the

elements of the quality system that comply with the corresponding specifications of

the relevant harmonised standard.

In addition to experience in quality management systems, the auditing team

shall have at least one member with experience of evaluation in the relevant

instrument field and instrument technology concerned, and knowledge of the

applicable requirements of these regulations. The audit shall include an assessment

visit to the manufacturer’s premises.

The auditing team shall review the technical documentation referred to in

point 2 in order to verify the manufacturer ’s ability to identify the relevant

requirements of these regulations and to carry out the necessary examinations with a

view to ensuring compliance of the instrument with those requirements.

The decision shall be notified to the manufacturer. The notification shall

contain the conclusions of the audit and the reasoned assessment decision.

(4) The manufacturer shall undertake to fulfil the obligations arising out of the

quality system as approved and to maintain it so that it remains adequate and

efficient.

(5) The manufacturer shall keep the notified body that has approved the quality

system informed of any intended change of the quality system.

The notified body shall evaluate any proposed changes and decide whether

the modified quality system will continue to satisfy the requirements referred to in

point 5.2 or whether are assessment is necessary.

It shall notify the manufacturer of its decision. The notification shall contain

the conclusions of the examination and the reasoned assessment decision.

6. Surveillance under the responsibility of the notified body

(1) The purpose of surveillance is to make sure that the manufacturer duly

fulfils the obligations arising out of the approved quality system.

(2) The manufacturer shall, for assessment purposes, allow the notified body

access to the manufacture, inspection, testing and storage sites, and shall provide it

with all necessary information, in particular

(a) the quality system documentation;

(b) the technical documentation referred to in point 2;

(c) the quality records, such as inspection reports and test data, calibration

data, qualification reports on the personnel concerned.

(3) The notified body shall carry out periodic audits to make sure that the

manufacturer maintains and applies the quality system and shall provide the

manufacturer with an audit report.

(4) In addition, the notified body may pay unexpected visits to the

manufacturer. During such visits the notified body may, if necessary, carry out

instrument tests, or have them carried out, in order to verify that the quality system

is functioning correctly. The notified body shall provide the manufacturer with a

visit report and, if tests have been carried out, with a test report.

7. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking, the supplementary metrology

marking set out in these regulations, and, under the responsibility of the notified

body referred to in point 5.1, the latter’s identification number to each individual

measuring instrument that satisfies the applicable requirements of these regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

each instrument model and keep it at the disposal of the national authorities for 10

years after the instrument has been placed on the market. The EU declaration of

conformity shall identify the instrument model for which it has been drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request.

A copy of the EU declaration of conformity shall be supplied with each

measuring instrument that is placed on the market. However, this requirement may

be interpreted as applying to a batch or consignment rather than individual

instruments in those cases where a large number of instruments is delivered to a

single user.

8. The manufacturer shall, for a period ending 10 years after the instrument has

been placed on the market, keep at the disposal of the national authorities

(a) the documentation referred to in point 5.1;

(b) the information relating to the change referred to in point 5.5, as

approved;

(c) the decisions and reports of the notified body referred to in points 5.5,

6.3 and 6.4.

9. Each notified body shall inform the Authority of quality system approvals

issued or withdrawn, and shall, periodically or upon request, make available to the

Authority the list of quality system approvals refused, suspended or otherwise

restricted.

10. Authorised representative

The manufacturer’s obligations set out in points 3, 5.1, 5.5, 7 and 8 may be

fulfilled by his authorised representative, on his behalf and under his responsibility,

provided that they are specified in the mandate.

MODULE E: CONFORMITY TO TYPE BASED ON INSTRUMENT

QUALITY ASSURANCE

1. Conformity to type based on instrument quality assurance is that part of a

conformity assessment procedure whereby the manufacturer fulfils the obligations

laid down in points 2 and 5, and ensures and declares on his sole responsibility that

the measuring instruments concerned are in conformity with the type described in

the EU-type examination certificate and satisfy the requirements of these regulations

that apply to them.

2. Manufacturing

The manufacturer shall operate an approved quality system for final product

inspection and testing of the measuring instruments concerned as specified in point 3

and shall be subject to surveillance, as specified in point 4.

3. Quality system

(1) The manufacturer shall lodge an application for assessment of his quality

system with a notified body of his choice, for the measuring instruments concerned.

The application shall include

(a) the name and address of the manufacturer and, if the application is

lodged by the authorised representative, his name and address as well;

(b) a written declaration that the same application has not been lodged with

any other notified body;

(c) all relevant information for the instrument category envisaged;

(d) the documentation concerning the quality system;

(e) the technical documentation of the approved type and a copy of the EU-

type examination certificate.

(2) The quality system shall ensure compliance of the measuring instruments

with the type described in the EU-type examination certificate and with the

applicable requirements of these regulations.

All the elements, requirements and provisions adopted by the manufacturer

shall be documented in a systematic and orderly manner in the form of written

policies, procedures and instructions. This quality system documentation shall

permit a consistent interpretation of the quality programmes, plans, manuals and

records.

It shall, in particular, contain an adequate description of

(a) the quality objectives and the organisational structure, responsibilities

and powers of the management with regard to product quality;

(b) the examinations and tests that will be carried out after manufacture;

(c) the quality records, such as inspection reports and test data, calibration

data, qualification reports on the personnel concerned;

(d) the means of monitoring the effective operation of the quality system.

(3)

The notified body shall assess the quality system to determine whether it

satisfies the requirements referred to in point 3.2.

It shall presume conformity with those requirements in respect of the

elements of the quality system that comply with the corresponding specifications of

the relevant harmonised standard.

In addition to experience in quality management systems, the auditing team

shall have at least one member with experience of evaluation in the relevant

instrument field and instrument technology concerned, and knowledge of the

applicable requirements of these regulations. The audit shall include an assessment

visit to the manufacturer’s premises.

The auditing team shall review the technical documentation referred to in

point (e) of point 3.1, in order to verify the manufacturer’s ability to identify the

re le va nt r e q u ir e m e nt s of th e s e r e g u la ti on s an d t o c a r r y o u t th e n eces sary

examinations with a view to ensuring compliance of the instrument with those

requirements.

The decision shall be notified to the manufacturer. The notification shall

contain the conclusions of audit and the reasoned assessment decision.

(4) The manufacturer shall undertake to fulfil the obligations arising out of the

quality system as approved and to maintain it so that it remains adequate and

efficient.

(5) The manufacturer shall keep the notified body that has approved the quality

system informed of any intended change to the quality system.

The notified body shall evaluate any proposed changes and decide whether

the modified quality system will continue to satisfy the requirements referred to in

point 3.2 or whether a re-assessment is necessary.

It shall notify the manufacturer of its decision. The notification shall contain

the conclusions of the examination and the reasoned assessment decision.

4. Surveillance under the responsibility of the notified body

(1) The purpose of surveillance is to make sure that the manufacturer duly

fulfils the obligations arising out of the approved quality system.

(2) The manufacturer shall, for assessment purposes, allow the notified body

access to the manufacture, inspection, testing and storage sites, and shall provide it

with all necessary information, in particular

(a) the quality system documentation;

(b) the quality records, such as inspection reports and test data, calibration

data, qualification reports on the personnel concerned.

(3) The notified body shall carry out periodic audits to make sure that the

manufacturer maintains and applies the quality system and shall provide the

manufacturer with an audit report.

(4) In addition, the notified body may pay unexpected visits to the

manufacturer. During such visits the notified body may, if necessary, carry out

instrument tests, or have them carried out, in order to verify that the quality system

is functioning correctly. The notified body shall provide the manufacturer with a

visit report and, if tests have been carried out, with a test report.

5. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking, the supplementary metrology

marking set out in these regulations, and, under the responsibility of the notified

body referred to in point 3.1, the latter’s identification number to each individual

instrument that is in conformity with the type described in the EU-type examination

certificate and satisfies the applicable requirements of these regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

each instrument model and keep it at the disposal of the national authorities for 10

years after the instrument has been placed on the market. The EU declaration of

conformity shall identify the instrument model for which it has been drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market. However,

this requirement may be interpreted as applying to a batch or consignment rather

than individual instruments in those cases where a large number of instruments are

delivered to a single user.

6. The manufacturer shall, for a period ending 10 years after the instrument has

been placed on the market, keep at the disposal of the national authorities

(a) the documentation referred to in point 3.1;

(b) the information relating to the change referred to in point 3.5, as

approved;

(c) the decisions and reports of the notified body referred to in points 3.5,

4.3 and 4.4.

7. Each notified body shall inform the Authority of quality system approvals

issued or withdrawn, and shall, periodically or upon request, make available to the

Authority the list of quality system approvals refused, suspended or otherwise

restricted.

8. Authorised representative

The manufacturer’s obligations set out in points 3.1, 3.5, 5 and 6 may be

fulfilled by his authorised representative, on his behalf and under his responsibility,

provided that they are specified in the mandate.

MODULE E1: QUALITY ASSURANCE OF FINAL INSTRUMENT

INSPECTION AND TESTING

1. Quality assurance of final instrument inspection and testing is the

conformity assessment procedure whereby the manufacturer fulfils the obligations

laid down in points 2, 4 and 7, and ensures and declares on his sole responsibility

that the measuring instruments concerned satisfy the requirements of these

regulations that apply to them.

2. Technical documentation

The manufacturer shall establish the technical documentation as described in

regulation 18. The documentation shall make it possible to assess the instrument’s

conformity with the relevant requirements, and shall include an adequate analysis

and assessment of the risk(s). The technical documentation shall specify the

applicable requirements and cover, as far as relevant for the assessment, the design,

manufacture and operation of the instrument.

3. The manufacturer shall keep the technical documentation at the disposal of

the relevant national authorities for 10 years after the instrument has been placed on

the market.

4. Manufacturing

The manufacturer shall operate an approved quality system for final product

inspection and testing of the measuring instruments concerned as specified in point 5

and shall be subject to surveillance as specified in point 6.

5. Quality system

5(1) The manufacturer shall lodge an application for assessment of his quality

system with the notified body of his choice, for the measuring instruments

concerned.

The application shall include

(a) the name and address of the manufacturer and, if the application is

lodged by the authorised representative, his name and address as well;

(b) a written declaration that the same application has not been lodged with

any other notified body;

(c) all relevant information for the instrument category envisaged;

(d) the documentation concerning the quality system;

(e) the technical documentation referred to in point 2.

(2) The quality system shall ensure compliance of the measuring instruments

with the requirements of these regulations that apply to them.

All the elements, requirements and provisions adopted by the manufacturer

shall be documented in a systematic and orderly manner in the form of written

policies, procedures and instructions. The quality system documentation shall permit

a consistent interpretation of the quality programmes, plans, manuals and records.

It shall, in particular, contain an adequate description of

(a) the quality objectives and the organisational structure, responsibilities

and powers of the management with regard to product quality;

(b) the examinations and tests that will be carried out after manufacture;

(c) the quality records, such as inspection reports and test data, calibration

data, qualification reports on the personnel concerned;

(d) the means of monitoring the effective operation of the quality system.

(3) The notified body shall assess the quality system to determine whether it

satisfies the requirements referred to in point 5.2.

It shall presume conformity with those requirements in respect of the

elements of the quality system that comply with the corresponding specifications of

the relevant harmonised standard.

In addition to experience in quality management systems, the auditing team

shall have at least one member with experience of evaluation in the relevant

instrument field and instrument technology concerned, and knowledge of the

applicable requirements of these regulations. The audit shall include an assessment

visit to the manufacturer’s premises.

The auditing team shall review the technical documentation referred to in

point 2 in order to verify the manufacturer ’s ability to identify the relevant

requirements of these regulations and to carry out the necessary examinations with a

view to ensuring compliance of the instrument with those requirements.

The decision shall be notified to the manufacturer. The notification shall

contain the conclusions of the audit and the reasoned assessment decision.

(4) The manufacturer shall undertake to fulfil the obligations arising out of the

quality system as approved and to maintain it so that it remains adequate and

efficient.

(5) The manufacturer shall keep the notified body that has approved the quality

system informed of any intended change to the quality system.

The notified body shall evaluate any proposed changes and decide whether

the modified quality system will continue to satisfy the requirements referred to in

point 5.2 or whether a re-assessment is necessary.

It shall notify the manufacturer of its decision. The notification shall contain

the conclusions of the examination and the reasoned assessment decision.

6. Surveillance under the responsibility of the notified body

(1) The purpose of surveillance is to make sure that the manufacturer duly

fulfils the obligations arising out of the approved quality system.

(2) The manufacturer shall, for assessment purposes, allow the notified body

access to the manufacture, inspection, testing and storage sites, and shall provide it

with all necessary information, in particular

(a) the quality system documentation;

(b) the technical documentation referred to in point 2;

(c) the quality records, such as inspection reports and test data, calibration

data, qualification reports on the personnel concerned.

(3) The notified body shall carry out periodic audits to make sure that the

manufacturer maintains and applies the quality system and shall provide the

manufacturer with an audit report.

(4) In addition, the notified body may pay unexpected visits to the manufacturer.

During such visits the notified body may, if necessary, carry out instrument tests, or

have them carried out, in order to verify that the quality system is functioning

correctly. The notified body shall provide the manufacturer with a visit report and, if

tests have been carried out, with a test report.

7. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking, the supplementary metrology

marking set out in these regulations, and, under the responsibility of the notified

body referred to in point 5.1, the latter’s identification number to each individual

measuring instrument that satisfies the applicable requirements of these regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

each instrument model and keep it at the disposal of the national authorities for 10

years after the instrument has been placed on the market. The EU declaration of

conformity shall identify the instrument model for which it has been drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market.

However, this requirement may be interpreted as applying to a batch or

consignment rather than individual instruments in those cases where a large number

of instruments is delivered to a single user.

8. The manufacturer shall, for a period ending 10 years after the instrument has

been placed on the market, keep at the disposal of the national authorities

(a) the documentation referred to in point 5.1,

(b) the information relating to the change referred to in point 5.5, as

approved;

(c) the decisions and reports from the notified body referred to in points

5.5, 6.3 and 6.4.

9. Each notified body shall inform the Authority of quality system approvals

issued or withdrawn, and shall, periodically or upon request, make available to the

Authority the list of quality system approvals refused, suspended or otherwise

restricted.

10. Authorised representative

The manufacturer’s obligations set out in points 3, 5.1, 5.5, 7 and 8 may be

fulfilled his authorised representative, on his behalf and under his responsibility,

provided that they are specified in the mandate.

MODULE F: CONFORMITY TO TYPE BASED ON PRODUCT

VERIFICATION

1. Conformity to type based on product verification is the part of a conformity

assessment procedure whereby the manufacturer fulfils the obligations laid down in

points 2, 5.1 and 6, and ensures and declares on his sole responsibility that the

measuring instruments concerned, which have been subject to the provisions of point

3, are in conformity with the type described in the EU-type examination certificate

and satisfy the requirements of these regulations that apply to them.

2. Manufacturing

The manufacturer shall take all measures necessary so that the

manufacturing process and its monitoring ensure conformity of the manufactured

measuring instruments with the approved type described in the EU-type examination

certificate and with the requirements of these regulations that apply to them.

3. Verification

A notified body chosen by the manufacturer shall carry out the appropriate

examinations and tests, or have them carried out, to verify the conformity of the

instruments with the type as described in the EU-type examination certificate and the

appropriate requirements of these regulations.

The examinations and tests to verify the conformity of the measuring

instruments with the appropriate requirements shall be carried out, at the choice of

the manufacturer, either by examination and testing of every instrument as specified

in point 4, or by examination and testing of the measuring instruments on a statistical

basis as specified in point 5.

4. Verification of conformity by examination and testing of every instrument

(1) All measuring instruments shall be individually examined and appropriate

tests set out in the relevant harmonised standard(s) and/or normative documents,

and/or equivalent tests set out in other relevant technical specifications, shall be

carried out in order to verify their conformity with the approved type described in

the EU-type examination certificate and with the appropriate requirements of these

regulations.

In the absence of a harmonised standard or normative document, the notified

body concerned shall decide on the appropriate tests to be carried out.

(2) The notified body shall issue a certificate of conformity in respect of the

examinations and tests carried out, and shall affix its identification number to each

approved instrument or have it affixed under its responsibility.

The manufacturer shall keep the certificates of conformity available for

inspection by the national authorities for 10 years after the instrument has been

placed on the market.

5. Statistical verification of conformity

(1) The manufacturer shall take all measures necessary so that the

manufacturing process and its monitoring ensure the homogeneity of each lot

produced, and shall present his measuring instruments for verification in the form of

homogeneous lots.

(2) A random sample shall be taken from each lot according to the requirements

of point 5.3. All measuring instruments in a sample shall be individually examined

and appropriate tests set out in the relevant harmonised standard(s) and/or normative

document(s), and/or equivalent tests set out in other relevant technical

specifications, shall be carried out in order to verify their conformity with the type

d e s c r i b e d i n t he E U - t y pe ex a m i n a t i o n c e r t i f i c a t e a n d w i t h t he ap p l i c a b le

requirements of these regulations, and to determine whether the lot is accepted or

rejected. In the absence of such harmonised standard or normative document, the

notified body concerned shall decide on the appropriate tests to be carried out.

(3)

The statistical procedure shall meet the following requirements

The statistical control will be based on attributes. The sampling system shall

ensure

(a) a level of quality corresponding to a probability of acceptance of 95%,

with a non-conformity of less than 1%;

(b) a limit quality corresponding to a probability of acceptance of 5%, with

a non-conformity of less than 7%.

(4) If a lot is accepted, all measuring instruments of the lot shall be considered

approved, except for those measuring instruments from the sample that have been

found not to satisfy the tests.

The notified body shall issue a certificate of conformity in respect of the

examinations and tests carried out, and shall affix its identification number to each

approved instrument or have it affixed under its responsibility.

The manufacturer shall keep the certificates of conformity at the disposal of

the national authorities for 10 years after the instrument has been placed on the

market.

(5) If a lot is rejected, the notified body shall take appropriate measures to

prevent the placing on the market of that lot.

In the event of frequent rejection of lots the notified body may suspend the

statistical verification and take appropriate measures.

6. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking and the supplementary

metrology marking set out in these regulations, and, under the responsibility of the

notified body referred to in point 3, the latter ’s identification number to each

individual instrument that is in conformity with the approved type described in the

EU-type examination certificate and satisfies the applicable requirements of these

regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

each instrument model and keep it at the disposal of the national authorities for 10

years after the instrument has been placed on the market. The EU declaration of

conformity shall identify the instrument model for which it has been drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market. However,

this requirement may be interpreted as applying to a batch or consignment rather

than individual instruments in those cases where a large number of instruments is

delivered to a single user.

If the notified body referred to in point 3 agrees and under its responsibility,

the manufacturer may also affix the notified body’s identification number to the

measuring instruments.

7. If the notified body agrees and under its responsibility, the manufacturer

may affix the notified body’s identification number to the measuring instruments

during the manufacturing process.

8. Authorised representative

The manufacturer’s obligations may be fulfilled by his authorised

representative, on his behalf and under his responsibility, provided that they are

s pe c if i e d in th e m a n d a te . A n aut h or is e d r e pr e s e nt a ti v e m a y no t fu l fi l t he

manufacturer’s obligations set out in points 2 and 5.1.

MODULE F1: CONFORMITY BASED ON PRODUCT VERIFICATION

1. Conformity based on product verification is the conformity assessment

procedure whereby the manufacturer fulfils the obligations laid down in points 2, 3,

6.1 and 7 and ensures and declares on his sole responsibility that the measuring

instruments concerned which have been subject to the provisions of point 4, are in

conformity with the requirements of these regulations that apply to them.

2. Technical documentation

The manufacturer shall establish the technical documentation as described in

regulation 18. The documentation shall make it possible to assess the instrument’s

conformity with the relevant requirements, and shall include an adequate analysis

and assessment of the risk(s). The technical documentation shall specify the

applicable requirements and cover, as far as relevant for the assessment, the design,

manufacture and operation of the instrument.

The manufacturer shall keep the technical documentation at the disposal of

the relevant national authorities for 10 years after the instrument has been placed on

the market.

3. Manufacturing

The manufacturer shall take all measures necessary so that the

manufacturing process and its monitoring ensure conformity of the manufactured

measuring instruments with the applicable requirements of these regulations.

4. Verification

A notified body chosen by the manufacturer shall carry out the appropriate

examinations and tests, or have them carried out, to verify the conformity of the

measuring instruments with the applicable requirements of these regulations.

The examinations and tests to verify the conformity with the requirements

shall be carried out, at the choice of the manufacturer, either by examination and

testing of every instrument as specified in point 5, or by examination and testing of

the measuring instruments on a statistical basis as specified in point 6.

5. Verification of conformity by examination and testing of every instrument

(1) All measuring instruments shall be individually examined and appropriate

tests, set out in the relevant harmonized standards and/or normative documents, and/

or equivalent tests set out in other relevant technical specifications, shall be carried

out to verify their conformity with the requirements that apply to them. In the

absence of such a harmonised standard, or normative document, the notified body

concerned shall decide on the appropriate tests to be carried out.

(2) The notified body shall issue a certificate of conformity in respect of the

examinations and tests carried out, and shall affix its identification number to each

approved instrument or have it affixed under its responsibility.

The manufacturer shall keep the certificates of conformity at the disposal of

the national authorities for 10 years after the instrument has been placed on the

market.

6. Statistical verification of conformity

(1) The manufacturer shall take all measures necessary so that the

manufacturing process ensures the homogeneity of each lot produced, and shall

present his measuring instruments for verification in the form of homogeneous lots.

(2) A random sample shall be taken from each lot according to the requirements

of point 6.4.

(3) All measuring instruments in the sample shall be individually examined and

appropriate tests set out in the relevant harmonised standards and/or normative

documents, and/or equivalent tests set out in other relevant technical specifications,

shall be carried out in order to verify their conformity with the applicable

requirements of these regulations and to determine whether the lot is accepted or

rejected. In the absence of such harmonised standard, or normative document, the

notified body concerned shall decide on the appropriate tests to be carried out.

(4)

The statistical procedure shall meet the following requirements

The statistical control will be based on attributes. The sampling system shall

ensure

(a) a level of quality corresponding to a probability of acceptance of 95%,

with a non-conformity of less than 1%;

(b) a limit quality corresponding to a probability of acceptance of 5%, with

a non-conformity of less than 7%.

(5) If a lot is accepted, all measuring instruments of the lot shall be considered

approved, except for those measuring instruments from the sample that have been

found not to satisfy the tests.

The notified body shall issue a certificate of conformity in respect of the

examinations and tests carried out, and shall affix its identification number to each

approved instrument or have it affixed under its responsibility.

The manufacturer shall keep the certificates of conformity at the disposal of

the national authorities for 10 years after the instrument has been placed on the

market.

If a lot is rejected, the notified body shall take appropriate measures to

prevent that lot from being placed on the market. In the event of frequent rejection of

lots the notified body may suspend the statistical verification and take appropriate

measures.

7. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking and the supplementary

metrology marking set out in these regulations, and under the responsibility of the

notified body referred to in point 4, the latter ’s identification number to each

individual measuring instrument that satisfies the applicable requirements of these

regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

each instrument model and keep it at the disposal of the national authorities for 10

years after the instrument has been placed on the market. The EU declaration of

conformity shall identify the instrument model for which it has been drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market. However,

this requirement may be interpreted as applying to a batch or consignment rather

than individual measuring instruments in those cases where a large number of

instruments is delivered to a single user.

If the notified body referred to in point 5 agrees and under its responsibility,

the manufacturer may also affix the notified body’s identification number to the

measuring instruments.

8. If the notified body agrees and under its responsibility, the manufacturer

may affix the notified body’s identification number to the measuring instruments

during the manufacturing process.

9. Authorised representative

The manufacturer’s obligations may be fulfilled by his authorised

representative, on his behalf and under his responsibility, provided that they are

s pe c if i e d in th e m a n d a te . A n aut h or is e d r e pr e s e nt a ti v e m a y no t fu l fi l t he

manufacturer’s obligations set out in point 2, first paragraph, point 3 and point 6.1.

MODULE G: CONFORMITY BASED ON UNIT VERIFICATION

1. Conformity based on unit verification is the conformity assessment

procedure whereby the manufacturer fulfils the obligations laid down in points 2, 3

and 5 and ensures and declares on his sole responsibility that the instrument

concerned, which has been subject to the provisions of point 4, is in conformity with

the requirements of these regulation that apply to it.

2. Technical documentation

The manufacturer shall establish the technical documentation as described in

regulation 18 and make it available to the notified body referred to in point 4. The

documentation shall make it possible to assess the instrument’s conformity with the

relevant requirements, and shall include an adequate analysis and assessment of the

risk(s). The technical documentation shall specify the applicable requirements and

cover, as far as relevant for the assessment, the design, manufacture and operation of

the instrument.

The manufacturer shall keep the technical documentation at the disposal of

the relevant national authorities for 10 years after the instrument has been placed on

the market.

3. Manufacturing

The manufacturer shall take all measures necessary so that the

manufacturing process and its monitoring ensure conformity of the manufactured

instrument with the applicable requirements of these regulations.

4. Verification

A notified body chosen by the manufacturer shall carry out the appropriate

examinations and tests set out in the relevant harmonised standards, and/or

normative documents or equivalent tests set out in other relevant technical

specifications, to verify the conformity of the instrument with the applicable

requirements of these regulations, or have them carried out. In the absence of such a

harmonised standard, or normative document, the notified body concerned shall

decide on the appropriate tests to be carried out.

The notified body shall issue a certificate of conformity in respect of the

examinations and tests carried out and affix its identification number to the approved

instrument, or have it affixed under its responsibility.

The manufacturer shall keep the certificates of conformity at the disposal of

the national authorities for 10 years after the instrument has been placed on the

market.

5. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking and the supplementary

metrology marking set out in these regulations and, under the responsibility of the

notified body referred to in point 4, the latter ’s identification number to each

instrument that satisfies the applicable requirements of these regulations.

5.2. The manufacturer shall draw up a written EU declaration of conformity and

keep it at the disposal of the national authorities for 10 years after the instrument has

been placed on the market. The EU declaration of conformity shall identify the

instrument for which it has been drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with the measuring instrument.

6. Authorised representative

The manufacturer’s obligations set out in points 2 and 5 may be fulfilled by

his authorised representative, on his behalf and under his responsibility, provided

that they are specified in the mandate.

MODULE H: CONFORMITY BASED ON FULL QUALITY ASSURANCE

1. Conformity based on full quality assurance is the conformity assessment

procedure whereby the manufacturer fulfils the obligations laid down in points 2 and

5, and ensures and declares on his sole responsibility that the measuring instruments

concerned satisfy the requirements of these regulations that apply to them.

2. Manufacturing

The manufacturer shall operate an approved quality system for design,

manufacture and final product inspection and testing of the measuring instruments

concerned as specified in point 3, and shall be subject to surveillance as specified in

point 4.

3. Quality system

(1) The manufacturer shall lodge an application for assessment of his quality

system with the notified body of his choice, for the measuring instruments

concerned.

The application shall include

(a) the name and address of the manufacturer and, if the application is

lodged by the authorised representative, his name and address as well,

(b) the technical documentation, as described in regulation 18, for one

model of each category of measuring instruments intended to be

manufactured. The documentation shall make it possible to assess the

instrument’s conformity with the relevant requirements, and shall

include an adequate analysis and assessment of the risk(s). The

technical documentation shall specify the applicable requirements and

cover, as far as relevant for the assessment, the design, manufacture and

operation of the instrument,

(c) the documentation concerning the quality system, and

(d) a written declaration that the same application has not been lodged with

any other notified body.

(2) The quality system shall ensure compliance of the measuring instruments

with the requirements of these regulations that apply to them.

All the elements, requirements and provisions adopted by the manufacturer

shall be documented in a systematic and orderly manner in the form of written

policies, procedures and instructions. This quality system documentation shall

permit a consistent interpretation of the quality programmes, plans, manuals and

It shall, in particular, contain an adequate description of

(a) the quality objectives and the organisational structure, responsibilities

and powers of the management with regard to design and product

quality;

(b) the technical design specifications, including standards, that will be

applied and, where the relevant harmonised standards, and/or normative

documents will not be applied in full, the means that will be used to

ensure that the essential requirements of these regulations that apply to

the measuring instruments will be met applying other relevant technical

specifications;

(c) the design control and design verification techniques, processes and

systematic actions that will be used when designing the measuring

instruments pertaining to the instrument category covered;

(d) the corresponding manufacturing, quality control and quality assurance

techniques, processes and systematic actions that will be used;

(e) the examinations and tests that will be carried out before, during and

after manufacture, and the frequency with which they will be carried

out;

(f)

the quality records, such as inspection reports and test data, calibration

data, qualification reports on the personnel concerned;

(g) the means of monitoring the achievement of the required design and

product quality and the effective operation of the quality system.

(3) The notified body shall assess the quality system to determine whether it

satisfies the requirements referred to in point 3.2.

It shall presume conformity with those requirements in respect of the

elements of the quality system that comply with the corresponding specifications of

the relevant harmonised standard.

In addition to experience in quality management systems, the auditing team

shall have at least one member experienced as an assessor in the relevant instrument

field and instrument technology concerned, and knowledge of the applicable

requirements of these regulations. The audit shall include an assessment visit to the

manufacturer’s premises.

The auditing team shall review the technical documentation referred to in

point (b) of point 3.1 to verify the manufacturer’s ability to identify the applicable

requirements of these regulations and to carry out the necessary examinations with a

view to ensuring compliance of the instrument with those requirements.

The manufacturer or his authorised representative shall be notified of the

decision. The notification shall contain the conclusions of the audit and the reasoned

assessment decision.

(4) The manufacturer shall undertake to fulfil the obligations arising out of the

quality system as approved and to maintain it so that it remains adequate and

efficient.

(5) The manufacturer shall keep the notified body that has approved the quality

system informed of any intended change to the quality system.

The notified body shall evaluate any proposed changes and decide whether

the modified quality system will continue to satisfy the requirements referred to in

point 3.2 or whether a re-assessment is necessary.

It shall notify the manufacturer of its decision. The notification shall contain

the conclusions of the examination and the reasoned assessment decision.

4. Surveillance under the responsibility of the notified body

(1) The purpose of surveillance is to make sure that the manufacturer duly

fulfils the obligations arising out of the approved quality system.

(2) The manufacturer shall, for assessment purposes, allow the notified body

access to the design, manufacture, inspection, testing and storage sites, and shall

provide it with all necessary information, in particular

(a) the quality system documentation;

(b) the quality records as provided for by the design part of the quality

system, such as results of analyses, calculations, tests.;

(c) the quality records as provided for by the manufacturing part of the

quality system, such as inspection reports and test data, calibration data,

qualification reports on the personnel concerned.

(3) The notified body shall carry out periodic audits to make sure that the

manufacturer maintains and applies the quality system and shall provide the

manufacturer with an audit report.

(4) In addition, the notified body may pay unexpected visits to the

manufacturer. During such visits the notified body may, if necessary, carry out

instrument tests, or have them carried out, in order to check the proper functioning

of the quality system. It shall provide the manufacturer with a visit report and, if

tests have been carried out, with a test report.

5. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking, the supplementary metrology

marking set out in these regulations and, under the responsibility of the notified body

referred to in point 3.1, the latter ’s identification number to each individual

instrument that satisfies the applicable requirements of these regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

each instrument model and keep it at the disposal of the national authorities for 10

years after the instrument has been placed on the market. The EU declaration of

conformity shall identify the instrument model for which it has been drawn up.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market. However,

this requirement may be interpreted as applying to a batch or consignment rather

than individual instruments in those cases where a large number of instruments are

delivered to a single user.

6. The manufacturer shall, for a period ending 10 years after the instrument has

been placed on the market, keep at the disposal of the national authorities

(a) the technical documentation referred to in point 3.1,

(b) the documentation concerning the quality system referred to in point

3.1,

(c) the information relating to the change referred to in point 3.5, as

approved;

(d) the decisions and reports of the notified body referred to in points 3.5,

4.3 and 4.4.

7. Each notified body shall inform the Authority of quality system approvals

issued or withdrawn, and shall, periodically or upon request, make available to the

Authority the list of quality system approvals refused, suspended or otherwise

restricted.

8. Authorised representative

The manufacturer’s obligations set out in points 3.1, 3.5, 5 and 6 may be

fulfilled by his authorised representative, on his behalf and under his responsibility,

provided that they are specified in the mandate.

MODULE H1: CONFORMITY BASED

ON

FULL

QUALITY

ASSURANCE PLUS DESIGN EXAMINATION

1. Conformity based on full quality assurance plus design examination is the

conformity assessment procedure whereby the manufacturer fulfils the obligations

laid down in points 2 and 6, and ensures and declares on his sole responsibility that

the measuring instruments concerned satisfy the requirements of these regulations

that apply to them.

2. Manufacturing

The manufacturer shall operate an approved quality system for design,

manufacture and final product inspection and testing of the measuring instruments

concerned as specified in point 3, and shall be subject to surveillance as specified in

point 5.

The adequacy of the technical design of the measuring instruments shall

have been examined in accordance with point 4.

3. Quality system

(1) The manufacturer shall lodge an application for assessment of the quality

system with the notified body of his choice for the measuring instruments concerned.

The application shall include

(a) the name and address of the manufacturer and, if the application is

lodged by the authorised representative, his name and address as well;

(b) all relevant information for the instrument category envisaged;

(c) the documentation concerning the quality system;

(d) a written declaration that the same application has not been lodged with

any other notified body.

(2)

The quality system shall ensure compliance of the measuring instruments

with the requirements of these regulations that apply to them.

All the elements, requirements and provisions adopted by the manufacturer

shall be documented in a systematic and orderly manner in the form of written

policies, procedures and instructions. This quality system documentation shall

permit a consistent interpretation of the quality programmes, plans, manuals and

records.

It shall, in particular, contain an adequate description of

(a) the quality objectives and the organisational structure, responsibilities

and powers of the management with regard to design and product

quality;

(b) the technical design specifications, including standards, that will be

applied and, where the relevant harmonised standards and/or normative

documents will not be applied in full, the means that will be used to

ensure that the essential requirements of these regulations that apply to

the measuring instruments will be met, applying other relevant technical

specifications;

(c) the design control and design verification techniques, processes and

systematic actions that will be used when designing the measuring

instruments pertaining to the instrument category covered;

(d) the corresponding manufacturing, quality control and quality assurance

techniques, processes and systematic actions that will be used;

(e) the examinations and tests that will be carried out before, during and

after manufacture, and the frequency with which they will be carried

out;

(f)

the quality records, such as inspection reports and test data, calibration

data, qualification reports on the personnel concerned;

(g) the means of monitoring the achievement of the required design and

product quality and the effective operation of the quality system.

(3) The notified body shall assess the quality system to determine whether it

satisfies the requirements referred to in point 3.2. It shall presume conformity with

those requirements in respect of the elements of the quality system that comply with

the corresponding specifications of the relevant harmonised standard.

In addition to experience in quality management systems, the auditing team

shall have at least one member experienced as an assessor in the relevant instrument

field and instrument technology concerned, and knowledge of the applicable

requirements of these regulations. The audit shall include an assessment visit to the

manufacturer’s premises.

The manufacturer or his authorised representative shall be notified of the

decision. The notification shall contain the conclusions of the audit and the reasoned

assessment decision.

(4) The manufacturer shall undertake to fulfil the obligations arising out of the

quality system as approved and to maintain it so that it remains adequate and

efficient.

(5) The manufacturer shall keep the notified body that has approved the quality

system informed of any intended change to the quality system.

The notified body shall evaluate any proposed changes and decide whether

the modified quality system will continue to satisfy the requirements referred to in

point 3.2 or whether a re-assessment is necessary.

It shall notify the manufacturer or his authorised representative of its

decision. The notification shall contain the conclusions of the examination and the

reasoned assessment decision.

(6) Each notified body shall inform the Authority of quality system approvals

issued or withdrawn, and shall periodically or upon request, make available to the

Authority the list of quality system approvals refused, suspended or otherwise

restricted.

4. Design examination

(1) The manufacturer shall lodge an application for examination of the design

with the notified body referred to in point 3.1.

(2) The application shall make it possible to understand the design, manufacture

and operation of the instrument, and to assess the conformity with the requirements

of these regulations that apply to it.

It shall include

(a) the name and address of the manufacturer;

(b) a written declaration that the same application has not been lodged with

any other notified body;

(c) the technical documentation as described in regulation 18. The

documentation shall make it possible to assess the instrument’s

conformity with the relevant requirements, and shall include an

adequate analysis and assessment of the risk(s). It shall, as far as

relevant for such assessment, cover the design and operation of the

instrument;

(d) the supporting evidence for the adequacy of the technical design. This

supporting evidence shall mention any documents that have been used,

in particular where the relevant harmonised standards and/or normative

documents have not been applied in full, and shall include, where

necessary, the results of tests carried out in accordance with other

relevant technical specifications, by the appropriate laboratory of the

manufacturer, or by another testing laboratory on his behalf and under

his responsibility.

(3) The notified body shall examine the application, and where the design meets

the requirements of these regulations that apply to the instrument it shall issue an EU

design examination certificate to the manufacturer. That certificate shall give the

name and address of the manufacturer, the conclusions of the examination, the

conditions (if any) for its validity and the data necessary for identification of the

approved design. That certificate may have one or more schedules attached.

That certificate and its schedules shall contain all relevant information to

allow the conformity of manufactured measuring instruments with the examined

design to be evaluated and to allow for in-service control. It shall allow the

evaluation of conformity of the manufactured instruments with the examined design

regarding the reproducibility of their metrological performances, when they are

properly adjusted using appropriate means, including

(a) the metrological characteristics of the design of the instrument;

(b) measures required for ensuring the integrity of the instruments (sealing,

identification of software, etc.);

(c) information on other elements necessary for the identification of the

instrument and to check its visual external conformity to the design;

(d) if appropriate, any specific information necessary to verify the

characteristics of manufactured instruments;

(e) in the case of a sub-assembly, all necessary information to ensure the

compatibility with other sub-assemblies or measuring instruments.

The notified body shall establish an evaluation report in this regard and keep

it at the disposal of the Member State that designated it. Without prejudice to Article

27(10), the notified body shall release the content of this report, in full or in part,

only with the agreement of the manufacturer.

The certificate shall have a validity of 10 years from the date of its issue,

and may be renewed for subsequent periods of 10 years each.

Where the design does not satisfy the applicable requirements of these

regulations, the notified body shall refuse to issue an EU design examination

certificate and shall inform the applicant accordingly, giving detailed reasons for its

refusal.

(4) The notified body shall keep itself appraised of any changes in the generally

acknowledged state of the art which indicate that the approved design may no longer

comply with the applicable requirements of these regulations, and shall determine

whether such changes require further investigation. If so, the notified body shall

inform the manufacturer accordingly.

The manufacturer shall keep the notified body that has issued the EU design

examination certificate informed of any modification to the approved design that

may affect the conformity with the essential requirements of these regulations or the

conditions for validity of the certificate. Such modifications shall require additional

approval – from the notified body that issued the EU design examination certificate

– in the form of an addition to the original EU design examination certificate.

(5) Each notified body shall inform the Authority of the EU design examination

certificates and/or any additions thereto which it has issued or withdrawn, and shall,

periodically or upon request, make available to the Authority the list of certificates

and/or any additions thereto refused, suspended or otherwise restricted.

The Commission, the Member States and the other notified bodies may, on

request, obtain a copy of the EU design examination certificates and/or additions

thereto. On request, the Commission and the Member States may obtain a copy of

the technical documentation and of the results of the examinations carried out by the

notified body.

The notified body shall keep a copy of the EU design examination

certificate, its schedules and additions, as well as the technical file including the

documentation submitted by the manufacturer until the expiry of the validity of the

certificate.

(6) The manufacturer shall keep a copy of the EU design examination

certificate, its schedules and additions with the technical documentation at the

disposal of the national authorities for 10 years after the instrument has been placed

on the market.

5. Surveillance under the responsibility of the notified body

The purpose of surveillance is to make sure that the manufacturer duly

fulfils the obligations arising out of the approved quality system.

(2) The manufacturer shall, for assessment purposes, allow the notified body

access to the design, manufacture, inspection, testing and storage sites, and shall

provide it with all necessary information, in particular

(a) the quality system documentation;

(b) the quality records as provided for by the design part of the quality

system, such as results of analyses, calculations, tests, etc.;

(c) the quality records as provided for by the manufacturing part of the

quality system, such as inspection reports and test data, calibration data,

qualification reports on the personnel concerned, etc.

(3) The notified body shall carry out periodic audits to make sure that the

manufacturer maintains and applies the quality system and shall provide the

manufacturer with an audit report.

(4) In addition, the notified body may pay unexpected visits to the

manufacturer. During such visits the notified body may, if necessary, carry out

instrument tests, or have them carried out, in order to check the proper functioning

of the quality system. It shall provide the manufacturer with a visit report and, if

tests have been carried out, with a test report.

6. Conformity marking and EU declaration of conformity

(1) The manufacturer shall affix the CE marking and the supplementary

metrology marking set out in these regulations, and, under the responsibility of the

notified body referred to in point 3.1, the latter’s identification number to each

individual instrument that satisfies the applicable requirements of these regulations.

(2) The manufacturer shall draw up a written EU declaration of conformity for

each instrument model and keep it at the disposal of the national authorities for 10

years after the instrument has been placed on the market. The EU declaration of

conformity shall identify the instrument model for which it has been drawn up and

shall mention the number of the design examination certificate.

A copy of the EU declaration of conformity shall be made available to the

relevant authorities upon request. A copy of the EU declaration of conformity shall

be supplied with each measuring instrument that is placed on the market. However,

this requirement may be interpreted as applying to a batch or consignment rather

than individual instruments in those cases where a large number of instruments is

delivered to a single user.

7. The manufacturer shall, for a period ending 10 years after the instrument has

been placed on the market, keep at the disposal of the national authorities

(a) the documentation concerning the quality system referred to in point

3.1,

(b) the information relating to the change referred to in point 3.5, as

approved;

(c) the decisions and reports of the notified body referred to in points 3.5,

5.3 and 5.4.

8. Authorised representative

The manufacturer’s authorised representative may lodge the application

referred to in points 4.1 and 4.2 and fulfil the obligations set out in points 3.1, 3.5,

4.4, 4.6, 6 and 7, on his behalf and under his responsibility, provided that they are

specified in the mandate.

MEASURING INSTRUMENTS

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Where instrument-specific schedules lay down the

essential requirements for sub-assemblies, these regulations shall

mutatis mutandis apply to such sub-assemblies.

(2) Sub-assemblies and measuring instruments may be assessed

independently and separately for the purpose of establishing

conformity.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

WATER METERS (MI-001)

The relevant requirements of Schedule I, the specific requirements of this

Schedule and the conformity assessment clean, cold or heated water in residential,

commercial and light industrial use procedures listed in this Schedule, apply to water

meters intended for the measurement of volumes of

DEFINITIONS

Water Meter

An instrument designed to measure, memorise and

display the volume at metering conditions of water

passing through the measurement transducer.

Minimum Flowrate (Q1)

The lowest flowrate at which the water meter provides

indications that satisfy the requirements concerning

the maximum permissible errors (MPEs.)

Transitional Flowrate (Q2) T h e t r a n s i t i o n a l f l o w r a t e i s t h e f l o w r a t e v a l u e

occurring between the perm anent and min imum

flowrates, at which the flowrate range is divided into

two zones, the ‘upper zone’ and the ‘lower zone’.

Each zone has a characteristic MPE.

Permanent Flowrate (Q3)

The highest flowrate at which the water meter

operates in a satisfactory ma nner under normal

conditions of use, i.e. under steady or intermittent

flow conditions.

Overload Flowrate (Q4)

The overload flowrate is the highest flowrate at which

the meter operates in a satisfactory manner for a short

period of time without deteriorating.

SPECIFIC REQUIREMENTS

Rated Operating Conditions

The manufacturer shall specify the rated operating conditions for the

instrument, in particular

1. The flowrate range of the water.

The values for the flowrate range shall fulfil the following conditions

Q3/Q1 ≥ 40

Q2/Q1 = 1,6

Q4/Q3 = 1,25

2. The temperature range of the water.

The values for the temperature range shall fulfil the following conditions

0,1°C to at least 30°C, or

30°C to at least 90°C.

The meter may be designed to operate over both ranges.

3. The relative pressure range of the water, the range being 0,3 bar to at least

10 bar at Q3.

4. For the power supply: the nominal value of the AC voltage supply and/or the

limits of DC supply.

5. The MPE, positive or negative, on volumes delivered at flowrates between

the transitional flowrate (Q2) (included) and the overload flowrate (Q4) is

2% for water having a temperature ≤ 30°C,

3% for water having a temperature > 30°C.

The meter shall not exploit the MPE or systematically favour any party.

6. The MPE, positive or negative, on volumes delivered at flowrates between

the minimum flowrate (Q1) and the transitional flowrate (Q2) (excluded) is 5% for

water having any temperature.

The meter shall not exploit the MPE or systematically favour any party.

Permissible Effect of Disturbances

Electromagnetic immunity

7(1)(a)The effect of an electromagnetic disturbance on a water meter shall be such

the change in the measurement result is no greater than the critical

change value as defined in point 7(1)(c), or

the indication of the measurement result is such that it cannot be

interpreted as a valid result, such as a momentary variation that cannot

be interpreted, memorised or transmitted as a measuring result.

7(1)(b) After undergoing an electromagnetic disturbance the water meter shall

recover to operate within MPE, and

have all measurement functions safeguarded, and

allow recovery of all measurement data present just before the

disturbance.

7(1)(c)The critical change value is the smaller of the two following

values

the volume corresponding to half of the magnitude of the MPE in the

upper zone on the measured volume;

the volume corresponding to the MPE on the volume corresponding to

one minute at flowrate Q3.

(2)

Durability

After an appropriate test, taking into account the period of time estimated by

the manufacturer, has been performed, the following criteria shall be satisfied

(2.1) The variation of the measurement result after the durability test, when

compared with the initial measurement result, shall not exceed

3% of the metered volume between Q1 included and Q2 excluded;

1,5% of the metered volume between Q2 included and Q4 included.

(2.2) The error of indication for the volume metered after the durability test shall

not exceed

± 6% of the metered volume between Q1 included and Q2 excluded;

± 2,5% of the metered volume between Q2 included and Q4 included for

water meters intended to meter water with a temperature between 0,1°C

and 30°C,

-

± 3,5% of the metered volume between Q2 included and Q4 included for

water meters intended to meter water with a temperature between 30°C

and 90°C.

8. Suitability

(1) The meter shall be able to be installed to operate in any position unless

clearly marked otherwise.

(2) The manufacturer shall specify whether the meter is designed to measure

reverse flow. In such a case, the reverse flow volume shall either be subtracted from

the cumulated volume or shall be separately recorded. The same MPE shall apply to

both forward and reverse flow.

Water meters not designed to measure reverse flow shall either prevent

reverse flow or shall withstand an accidental reverse flow without any deterioration

or change in metrological properties.

Units of Measurement

9. Metered volume shall be displayed in cubic metres.

Putting into Use

10. The Authority shall ensure that the requirements under points 1, 2 and 3 are

determined by the utility or the person legally designated for installing the meter, so

that the meter is appropriate for the accurate measurement of consumption that is

foreseen or foreseeable.

CONFORMITY ASSESSMENT

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are: B + F or B + D or H1.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

The penalties applicable for the infringement of the

provisions of Regulation (EU) No 167/2013 shall be those provided

for in Part IV of the Product Safety Act.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

GAS METERS AND VOLUME CONVERSION DEVICES (MI-002)

The relevant requirements of Schedule I, the specific requirements of this

Schedule and the conformity assessment procedures listed in this Schedule, apply to

gas meters and volume conversion devices defined below, intended for residential,

commercial and light industrial use.

DEFINITIONS

Gas meter

An instrument designed to measure, memorise and

display the quantity of fuel gas (volume or mass) that

has passed it.

Conversion device

A device fitted to a gas meter that automatically

converts the quantity measured at metering conditions

into a quantity at base conditions.

Minimum flowrate (Q min )

The lowest flowrate at which the gas meter provides

indications that satisfy the requirements regarding

maximum permissible error (MPE).

Maximum flowrate (Q max )

The highest flowrate at which the gas meter provides

indications that satisfy the requirements regarding

MPE.

Transitional flowrate (Q t )

The transitional flowrate is the flowrate occurring

between the maximum and minimum flowrates at

which the flowrate range is divided into two zones,

the ‘upper zone’ and the ‘lower zone’. Each zone has a

characteristic MPE.

Overload Flowrate (Q r )

The overload flowrate is the highest flowrate at which

the meter operates for a short period of time without

deteriorating.

Base conditions

The specified conditions to which the measured

quantity of fluid is converted.

PART I

SPECIFIC REQUIREMENTS GAS METERS

1. Rated operating conditions

The manufacturer shall specify the rated operating conditions of the gas

meter, taking into account

The flowrate range of the gas shall fulfil at least the following conditions

Class

Q max /Q min

Q max /Q t

Q r /Q max

1,5

≥ 150

≥ 10

1,2

1,0

≥ 20

≥ 5

1,2

(2)

The temperature range of the gas, with a minimum range of 40°C.

(3)

The fuel/gas related conditions

The gas meter shall be designed for the range of gases and supply pressures

of the country of destination. In particular the manufacturer shall indicate

the gas family or group;

the maximum operating pressure.

(4)

A minimum temperature range of 50°C for the climatic environment.

(5)

The nominal value of the AC voltage supply and/or the limits of DC supply.

2. Maximum permissible error (MPEs)

(1)

Gas meter indicating the volume at metering conditions or mass

Table 1

Class

1,5

1,0

Q min ≤ Q < Qt

3%

2%

Qt ≤ Q ≤ Q max

1,5%

1%

The gas meter shall not exploit the MPEs or systematically favour any party.

(2) For a gas meter with temperature conversion, which only indicates the

converted volume, the MPE of the meter is increased by 0,5% in a range of 30°C

extending symmetrically around the temperature specified by the manufacturer that

lies between 15°C and 25°C. Outside this range, an additional increase of 0,5% is

permitted in each interval of 10°C.

3. Permissible effect of disturbances

(1)

Electromagnetic immunity

(1.1) The effect of an electromagnetic disturbance on a gas meter or volume

conversion device shall be such that

-

the change in the measurement result is no greater than the critical

change value as defined in point 3.1.3, or

-

the indication of the measurement result is such that it cannot be

interpreted as a valid result, such as a momentary variation that cannot

be interpreted, memorised or transmitted as a measuring result.

(1.2) After undergoing a disturbance, the gas meter shall

-

recover to operate within MPE, and

-

have all measurement functions safeguarded, and

-

allow recovery of all measurement data present just before the

disturbance.

(1.3) The critical change value is the smaller of the two following values

-

the quantity corresponding to half of the magnitude of the MPE in the

upper zone on the measured volume;

-

the quantity corresponding to the MPE on the quantity corresponding to

one minute at maximum flowrate.

(2)

Effect of upstream-downstream flow disturbances

Under installation conditions specified by the manufacturer, the effect of the

flow disturbances shall not exceed one third of the MPE.

4. Durability

After an appropriate test, taking into account the period of time estimated by

the manufacturer, has been performed, the following criteria shall be satisfied

(1)

Class 1,5 $3

(1.1) The variation of the measurement result after the durability test when

compared with the initial measurement result for the flow rates in the range Qt to

Q max - shall not exceed the measurement result by more than 2%.

(1.2) The error of indication after the durability test shall not exceed twice the

MPE in point 2.

(2)

Class 1,0 $3

(2.1) The variation of the measurement result after the durability test when

compared with the initial measurement result shall not exceed one-third of the MPE

in point 2.

(2.2) The error of indication after the durability test shall not exceed the MPE in

point 2.

5. Suitability

A gas meter powered from the mains (AC or DC) shall be provided with an

emergency power supply device or other means to ensure, during a failure of the

principal power source, that all measuring functions are safeguarded.

(2) A dedicated power source shall have a lifetime of at least five years. After

90% of its lifetime an appropriate warning shall be shown.

(3) An indicating device shall have a sufficient number of digits to ensure that

the quantity passed during 8 000 hours at Q max does not return the digits to their

initial values.

(4)

The gas meter shall be able to be installed to operate in any position

declared by the manufacturer in its installation instruction.

(5) The gas meter shall have a test element, which shall enable tests to be

carried out in a reasonable time.

(6) The gas meter shall respect the MPE in any flow direction or only in one

flow direction clearly marked.

6. Units

Metered quantity shall be displayed in cubic metre, or in kilogram.

PART II

SPECIFIC REQUIREMENTS VOLUME CONVERSION DEVICES

A volume conversion device constitutes a sub-assembly when it is together

with a measuring instrument with which it is compatible.

For a volume conversion device, the essential requirements for the gas meter

shall apply, if applicable. In addition, the following requirements shall apply

7. Base conditions for converted quantities

The manufacturer shall specify the base conditions for converted quantities.

8. MPE

-

0,5% at ambient temperature 20°C ± 3°C, ambient humidity 60% ±

15%, nominal values for power supply;

-

0,7% for temperature conversion devices at rated operating conditions;

-

1% for other conversion devices at rated operating conditions. Note

The error of the gas meter is not taken into account.

The volume conversion device shall not exploit the MPEs or

systematically favour any party.

9. Suitability

(1) An electronic conversion device shall be capable of detecting when it is

operating outside the operating range(s) stated by the manufacturer for parameters

that are relevant for measurement accuracy. In such a case, the conversion device

must stop integrating the converted quantity, and may totalise separately the

converted quantity for the time it is operating outside the operating range(s).

(2) An electronic conversion device shall be capable to display all relevant data

for the measurement without additional equipment.

PART III

PUTTING INTO USE AND CONFORMITY ASSESSMENT

Putting into use

10. (a) Where the Authority imposes measurement of residential use, it shall

allow such measurement to be performed by means of any Class 1,5 $3, and by Class

1,0 $3 which have a Q max /Q min ratio equal or greater than 150.

(b) Where the Authority imposes measurement of commercial and/or light

industrial use, it shall allow such measurement to be performed by any Class 1,5

USD 3.

MEASURING INSTRUMENTS

(c) As regards the requirements under points 1.2 and 1.3, the Authority shall

ensure that the properties be determined by the utility or the person legally

designated for installing the meter, so that the meter is appropriate for the accurate

measurement of consumption that is foreseen or foreseeable.

CONFORMITY ASSESSMENT

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are

B + F or B + D or H1.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

These regulations establish the requirements that measuring

instruments have to satisfy with a view to their being made

available on the market and, or put into use for the measuring tasks

referred to in these regulations.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

ACTIVE ELECTRICAL ENERGY METERS (MI-003)

The relevant requirements of Schedule I, the specific requirements of this

Schedule and the conformity assessment procedures listed in this Schedule, apply to

active electrical energy meters intended for residential, commercial and light

industrial use.

Note

Electrical energy meters may be used in combination with external

instrument transformers, depending upon the measurement technique applied.

However, this Schedule covers only electrical energy meters but not instrument

transformers.

DEFINITIONS

An active electrical energy meter is a device which measures the active

electrical energy consumed in a circuit.

the electrical current flowing through the meter;

the specified reference current for which the transformer operated

meter has been designed;

the lowest declared value of I at which the meter registers active

electrical energy at unity power factor (polyphase meters with

balanced load);

the value of I above which the error lies within maximum

permissible errors (MPEs) (polyphase meters with balanced load);

the value of I above which the error lies within the smallest MPE

corresponding to the class index of the meter;

the maximum value of I for which the error lies within the MPEs;

the voltage of the electricity supplied to the meter;

the specified reference voltage;

the frequency of the voltage supplied to the meter;

the specified reference frequency;

power factor = cosφ = the cosine of the phase difference φ between I

and U.

SPECIFIC REQUIREMENTS

1. Accuracy

The manufacturer shall specify the class index of the meter. The class

indices are defined as: Class A, B and C.

2. Rated operating conditions

The manufacturer shall specify the rated operating conditions of the meter;

in particular

The values of fn, Un, In, Ist, Imin, Itr and Imax that apply to the meter. For

the current values specified, the meter shall satisfy the conditions given in Table 1;

Table 1

Class A

Class B

Class C

For direct-connected meters

≤ 0,05 · I tr ≤ 0,04 · I tr ≤ 0,04 · I tr

≤ 0,5 · I tr

≤ 0,5 · I tr

≤ 0,3 · I tr

≥ 50 · I tr

≥ 50 · I tr

≥ 50 · I tr

For transformer-operated meters

≤ 0,06 · I tr ≤ 0,04 · I tr ≤ 0,02 · I tr

= 20 · I tr

= 20 · I tr

= 20 · I tr

≥ 1,2 · I n

≥ 1,2 · I n

≥ 1,2 · I n

(1) For Class B electromechanical meters I min ≤ 0,4 · I tr shall apply.

The voltage, frequency and power factor ranges within which the meter shall

satisfy the MPE requirements are specified in Table 2. These ranges shall recognise

the typical characteristics of electricity supplied by public distribution systems.

The voltage and frequency ranges shall be at least

0,9 · Un ≤ U ≤ 1,1 · Un

0,98 · fn ≤ f ≤ 1,02 · fn

power factor range at least from cosφ = 0,5 inductive to cosφ = 0,8

capacitive.

3. MPEs

The effects of the various measurands and influence quantities (a, b, c,…)

are evaluated separately, all other measurands and influence quantities being kept

relatively constant at their reference values. The error of measurement, that shall not

exceed the MPE stated in Table 2, is calculated as

Error of measurement

 ( a 2  b 2  c 2 ...)

When the meter is operating under varying-load current, the percentage

errors shall not exceed the limits given in Table 2.

Table 2

MPEs in percent at rated operating conditions and defined load current levels and operating

temperature

Operating

Operating

Operating

Operating

Temperatures

Temperatures

Temperatures

Temperatures

+ 5°C … + 30°C – 10°C … + 5°C – 25°C – 10°C – 40°C . – 25°C

or + 30°C … +

or + 40°C ..…. or + 55°C ....…

40°C

+ 55°C

+ 70°C

Meter Class

A

B

C

A

B

C

A

B

C

A

B

C

Single phase meter; polyphase meter if operating with balanced loads

I min < I < I tr

3,5

2,5

1,3

3,5 1,7

I tr < I < I max

3,5

0,7

4,5

2,5

3,5 1,3

1,5

Polyphase meter if operating with single phase load

I tr < I < Imax, see 4

2,5

1,3

4 1,7

4,5

exception below

For electromechanical polyphase meters the current range for single-phase load is limited to

5I tr < I < I max

When a meter operates in different temperature ranges the relevant MPE

values shall apply. The meter shall not exploit the MPEs or systematically favour

any party.

4. Permissible effect of disturbances

(1)

General

As electrical energy meters are directly connected to the mains supply and

as m ains current is also one of the measurands, a special electrom agnetic

environment is used for electricity meters.

The meter shall comply with the electromagnetic environment E2 and the

additional requirements in points 4.2 and 4.3.

The electromagnetic environment and permissible effects reflect the

situation that there are disturbances of long duration which shall not affect the

accuracy beyond the critical change values and transient disturbances, which may

cause a temporary degradation or loss of function or performance but from which the

meter shall recover and shall not affect the accuracy beyond the critical change

values.

When there is a foreseeable high risk due to lightning or where overhead

supply networks are predominant, the metrological characteristics of the meter shall

be protected.

4.2. Effect of disturbances of long duration

Table 3

Disturbance

Critical change values in

percent for meters of class

A

B

C

Reversed phase sequence

1.5

1.5

0.3

Voltage unbalance (only applicable to poly phase

meters)

Harmonic contents in the current circuits (1)

0.8

0.5

DC and harmonics in the current circuit (1)

1.5

Fast transient bursts

Magnetic fields; HF (radiated RF) electromagnetic

field; Conducted disturbances introduced by radio-

frequency fields; and Oscillatory waves immunity

(1) In the case of electromechanical electricity meters, no critical change values are

defined for harmonic contents in the current circuits and for DC and harmonics in

the current circuit.

Critical change values for disturbances of long duration

(3)

Permissible effect of transient electromagnetic phenomena

(3.1) The effect of an electromagnetic disturbance on an electrical energy meter

shall be such that during and immediately after a disturbance

any output intended for testing the accuracy of the meter does not

produce pulses or signals corresponding to an energy of more than the

critical change value, and in reasonable time after the disturbance the

meter shall

recover to operate within the MPE limits, and

have all measurement functions safeguarded, and

allow recovery of all measurement data present prior to the disturbance,

and

not indicate a change in the registered energy of more than the critical

change value. The critical change value in kWh is m · U n · I max · 10–6

(m being the number of measuring elements of the meter, U n in Volts and I max in

(3.2) For overcurrent the critical change value is 1,5%.

5. Suitability

(1) Below the rated operating voltage the positive error of the meter shall not

exceed 10%.

(2) The display of the total energy shall have a sufficient number of digits to

ensure that when the meter is operated for 4 000 hours at full load (I = I max , U = U n

and PF = 1) the indication does not return to its initial value and shall not be able to

be reset during use.

(3) In the event of loss of electricity in the circuit, the amounts of electrical

energy measured shall remain available for reading during a period of at least 4

(4)

Running with no load

When the voltage is applied with no current flowing in the current circuit

(current circuit shall be open circuit), the meter shall not register energy at any

voltage between 0,8 · U n and 1,1 U n .

(5)

Starting

The meter shall start and continue to register at U n , PF = 1 (polyphase meter

with balanced loads) and a current which is equal to I st .

6. Units

The electrical energy measured shall be displayed in kilowatt-hours or in

megawatt-hours.

7. Putting into use

(a) Where the Authority imposes measurement of residential use, it shall

allow such measurement to be performed by means of any Class A

meter. For specified purposes Malta is authorised to require any Class B

meter.

(b) Where the Authority imposes measurement of commercial and/or light

industrial use, it shall allow such measurement to be performed by any

Class B meter. For specified purposes Malta is authorised to require any

Class C meter.

(c) The Authority shall ensure that the current range be determined by the

utility or the person legally designated for installing the meter, so that

MEASURING INSTRUMENTS

the meter is appropriate for the accurate measurement of consumption

that is foreseen or foreseeable.

CONFORMITY ASSESSMENT

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are: B + F or B + D or H1.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) The Authority may prescribe the use of measuring

instruments for measuring tasks, where they consider it justified for

reasons of public interest, public health, public safety, public order,

protection of the environment, protection of consumers, levying of

taxes and duties and fair trading.

(2) If the Authority does not prescribe such use, it shall

communicate the reasons therefore to the Commission and the

other Member States.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

THERMAL ENERGY METERS (MI-004)

The relevant requirements of Schedule I, the specific requirements and the

conformity assessment procedures listed in this Schedule, apply to thermal energy

meters defined below, intended for residential, commercial and light industrial use.

DEFINITIONS

A thermal energy meter is an instrument designed to measure the thermal

energy which, in a thermal energy exchange circuit, is given up by a liquid called the

thermal energy-conveying liquid.

A thermal energy meter is either a complete instrument or a combined

instrument consisting of the sub-assemblies, flow sensor, temperature sensor pair,

and calculator, as defined in regulation 3(2), or a combination thereof

the temperature of the thermal energy-conveying liquid;

the value of θ at the inlet of the thermal energy exchange circuit;

the value of θ at the outlet of the thermal energy exchange circuit;

the temperature difference θ in - θ οut with Δ θ ≥ 0;

the upper limit of θ for the thermal energy meter to function correctly

within the MPEs;

the lower limit of θ for the thermal energy meter to function correctly

within the MPEs;

the upper limit of Δθ for the thermal energy meter to function

correctly within the MPEs;

the lower limit of Δθ for the thermal energy meter to function

correctly within the MPEs;

the flow rate of the thermal energy conveying liquid;

the highest value of q that is permitted for short periods of time for

the thermal energy meter to function correctly;

the highest value of q that is permitted permanently for the thermal

energy meter to function correctly;

the lowest value of q that is permitted for the thermal energy meter to

function correctly;

the thermal power of the thermal energy exchange;

the upper limit of P that is permitted for the thermal energy meter to

function correctly.

SPECIFIC REQUIREMENTS

1. Rated operating conditions

The values of the rated operating conditions shall be specified by the

manufacturer as follows

(1)

For the temperature of the liquid: θ max , θ min ,

-

for the temperature differences: Δθ max , Δθ min ,

subject to the following restrictions: Δθ max /Δθ min ≥ 10; Δθ min = 3 K or 5 K

or 10 K.

(2) For the pressure of the liquid: The maximum positive internal pressure that

the thermal energy meter can withstand permanently at the upper limit of the

temperature.

(3) For the flow rates of the liquid: q s , q p , q i , where the values of qp and qi are

subject to the following restriction: q p /q i ≥ 10.

(4)

For the thermal power: P s .

2. Accuracy classes

The following accuracy classes are defined for thermal energy meters: 1, 2,

3.

3. MPEs applicable to complete thermal energy meters

The maximum permissible relative errors applicable to a complete thermal

energy meter, expressed in percent of the true value for each accuracy class, are

For class 1: E = Ef + Et + Ec, with Ef, Et, Ec according to points 7.1 to 7.3.

For class 2: E = Ef + Et + Ec, with Ef, Et, Ec according to points 7.1 to 7.3.

For class 3: E = Ef + Et + Ec, with Ef, Et, Ec according to points 7.1 to 7.3.

The complete thermal energy meter shall not exploit the MPEs or

systematically favour any party.

4. Permissible influences of electromagnetic disturbances

(1) The instrument shall not be influenced by static magnetic fields and by

electromagnetic fields at mains frequency.

(2) The influence of an electromagnetic disturbance shall be such that the

change in the measurement result is not greater than the critical change value as laid

down in requirement 4.3 or the indication of the measurement result is such that it

cannot be interpreted as a valid result.

(3) The critical change value for a complete thermal energy meter is equal to the

absolute value of the MPE applicable to that thermal energy meter (see point (3).

5. Durability

After an appropriate test, taking into account the period of time estimated by

the manufacturer, has been performed, the following criteria shall be satisfied

(1) Flow sensors: The variation of the measurement result after the durability

test, when compared with the initial measurement result, shall not exceed the critical

change value.

(2) Temperature sensors: The variation of the measurement result after the

durability test, when compared with the initial measurement result, shall not exceed

6. Inscriptions on a thermal energy meter

Accuracy class

-

Limits of flow rate

-

Limits of temperature

-

Limits of temperature difference

-

Place of the flow sensor installation: flow or return

-

Indication of the direction of flow

7. Sub-assemblies

The provisions for sub-assemblies may apply to sub-assemblies

manufactured by the same or different manufacturers. Where a thermal energy meter

consists of sub-assemblies, the essential requirements for the thermal energy meter

apply to the sub-assemblies as relevant. In addition, the following apply

(1)

The relative MPE of the flow sensor, expressed in%, for accuracy classes

-

Class 1: Ef = (1 + 0,01 q p /q), but not more than 5%,

-

Class 2: Ef = (2 + 0,02 q p /q), but not more than 5%,

-

Class 3: Ef = (3 + 0,05 q p /q), but not more than 5%,

where the error Ef relates the indicated value to the true value of the

relationship between flow sensor output signal and the mass or the volume.

(2)

The relative MPE of the temperature sensor pair, expressed in%

-

Et = (0,5 + 3 · Δθ min /Δθ),

where the error Et relates the indicated value to the true value of the relationship

between temperature sensor pair output and temperature difference.

(3)

The relative MPE of the calculator, expressed in%

-

Ec = (0,5 + Δθ min /Δθ),

where the error Ec relates the value of the thermal energy indicated to the true value

of the thermal energy.

(4) The critical change value for a sub-assembly of a thermal energy meter is

equal to the respective absolute value of the MPE applicable to the sub-assembly

(see points 7.1, 7.2 or 7.3).

(5)

Inscriptions on the sub-assemblies

Flow sensor

Accuracy class

Limits of flow rate

Limits of temperature

Nominal meter factor (e.g. litres/pulse) or corresponding output

signal

Indication of the direction of flow

Temperature

Type identification (e.g. P t 100)

sensor pair

Limits of temperature

Limits of temperature difference

Type of temperature sensors

- Limits of temperature

- Limits of temperature difference

- Required nominal meter factor (e.g. litres/pulse) or

corresponding input signal coming from the flow sensor

- Place of the flow sensor installation: flow or return

8. PUTTING INTO USE

(a) Where the Authority imposes measurement of residential use, it shall

allow such measurement to be performed by means of any Class 3

meter.

(b) Where the Authority imposes measurement of commercial and/or light

industrial use, it is authorised to require any Class 2 meter.

(c) As regards the requirements under points 1.1 to 1.4, the Authority shall

ensure that the properties be determined by the utility or the person

legally designated for installing the meter, so that the meter is

appropriate for the accurate measurement of consumption that is

foreseen or foreseeable.

CONFORMITY ASSESSMENT

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are: B + F or B + D or H1.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) A measuring instrument shall meet the essential

requirements set out in Schedule I and in the relevant instrument-

specific Schedules.

(2) For the correct use of the instrument, the information

referred to in point 9 of Schedule I or in the relevant instrument-

specific Schedules may be provided in at least English and, or

Maltese.

MEASURING INSTRUMENTS

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

MEASURING SYSTEMS FOR THE CONTINUOUS AND DYNAMIC

MEASUREMENT OF QUANTITIES OF LIQUIDS OTHER THAN WATER (MI-

The relevant essential requirements of Schedule I, the specific requirements

of this Schedule and the conformity assessment procedures listed in this Schedule,

apply to measuring systems intended for the continuous and dynamic measurement

of quantities (volumes or masses) of liquids other than water. If appropriate, the

terms ‘volume and L’ in this Schedule can be read as: ‘mass and kg’.

DEFINITIONS

An instrument designed to measure continuously, memorise and

display the quantity at metering conditions of liquid flowing

through the measurement transducer in a closed, fully charged

conduit.

A part of a meter that receives the output signals from the

m e a s u r e m e n t t r a n s d u c e r ( s ) a n d p o s s i b l y, f r o m a s s o c i a t e d

measuring instruments and displays the measurement results.

An instrument connected to the calculator for measuring certain

quantities which are characteristic of the liquid, with a view to

make a correction and/or conversion.

MEASURING INSTRUMENTS

A part of the calculator which by taking account of the

characteristics of the liquid (temperature, density, etc.) measured

using associated measuring instruments, or stored in a memory,

automatically converts

-

the volume of the liquid measured at metering conditions into

a volume at base conditions and/or into mass, or

-

the mass of the liquid measured at metering conditions into a

volume at metering conditions and/or into a volume at base

conditions

Note

A conversion device includes the relevant associated measuring

instruments.

The specified conditions to which the measured quantity of liquid

at metering conditions is converted.

A system that comprises the meter itself and all devices required to

ensure correct measurement or intended to facilitate the measuring

operations.

Fuel dispenser A measuring system intended for the refuelling of motor vehicles,

small boats and small aircraft.

An arrangement that allows the customer to use a measuring

system for the purpose of obtaining liquid for his own use.

A specific device that is part of a self-service arrangement and

which allows one of more measuring systems to perform in this

self-service arrangement.

The smallest quantity of liquid for which the measurement is

metrological acceptable for the measuring system.

The indication, either volume or mass, corresponding to the

measure and that the meter is physically capable of measuring.

Note: The direct indication may be converted into another quantity

using a conversion device.

Interruptible/ A m e a s u r i n g s y s t e m i s c o n s i d e r e d a s i n t e r r u p t i b l e / n o n -

interruptible when the liquid flow can/cannot be stopped easily and

rapidly.

Flowrate range The range between the minimum flowrate (Q min ) and maximum

flowrate (Q max ).

SPECIFIC REQUIREMENTS

1. Rated operating conditions

The manufacturer shall specify the rated operating conditions for the

instrument, in particular;

(1)

The flowrate range

The flowrate range is subject to the following conditions

(i)

the flowrate range of a measuring system shall be within the flowrate

range of each of its elements, in particular the meter.

(ii) meter and measuring system

Table 1

Specific measuring

Characteristic of liquid

Minimum ratio of Q max : Q min

system

Fuel dispensers

Not Liquefied gases

10: 1

Liquefied gases

5: 1

Measuring system

Cryogenic liquids

5:1

Measuring systems on

All liquids

Suitable for use

pipeline and systems for

loading ships

All other measuring

All liquids

4:1

systems

(2) The properties of the liquid to be measured by the instrument by specifying

the name or type of the liquid or its relevant characteristics, for example

Temperature range;

Pressure range;

Density range;

Viscosity range.

(3) The nominal value of the AC voltage supply and/or limits of the DC voltage

supply.

(4)

The base conditions for converted values.

Note

Point 1.4 is without prejudice to the Authority’s obligations to require use of

a temperature of either 15°C in accordance with Article 12(2) of Council Regulation

2003/96/EC of 27 October 2003 restructuring the Community framework for the

taxation of energy products and electricity (1).

2. Accuracy classification and maximum permissible errors (MPEs)

(1)

For quantities equal to or greater than 2 litres the MPE on indications is

Table 2

Accuracy Class

0,3

0,5

1,0

1,5

2,5

Measuring systems (1)

0,3%

0,5%

1,0%

1,5%

2,5%

Meters (2)

0,2%

0,3%

0,6%

1,0%

1,5%

(1)

OJ L 283, 31.10.2003, p. 51.

(2)

For quantities less than two litres the MPE on indications is

Table 3

Measured volume V

MPE

V < 0,1 l

4 × value in Table 2, applied to 0,1 L

0,1 l ≤ V < 0,2 l

4 × value in Table 2

0,2 l ≤ V < 0,4 l

2 × value in Table 2, applied to 0,4 L

0,4 l ≤ V < 1 l

2 × value in Table 2

1 l ≤ V < 2 l

Value in Table 2, applied to 2 L

MEASURING INSTRUMENTS

(3) However, no matter what the measured quantity may be, the magnitude of

the MPE is given by the greater of the following two values

the absolute value of the MPE given in Table 2 or Table 3,

the absolute value of the MPE for the minimum measured quantity (E min ).

(4.1) For minimum measured quantities greater than or equal to 2 litres the

following conditions apply

Condition 1

E min shall fulfil the condition: E min ≥ 2 R, where R is the smallest scale

interval of the indication device.

Condition 2

E min is given by the formula: E min = (2MMQ) × (A/100), where

MMQ is the minimum measured quantity,

A is the numerical value specified in line A of Table 2.

(4.2) For minimum measured quantities of less than two litres, the above

mentioned condition 1 applies and Emin is twice the value specified in Table 3, and

related to line A of Table 2.

(5)

Converted indication

In the case of a converted indication the MPEs are as in line A of Table 2.

(6) Conversion devices

MPEs on converted indications due to a conversion device are equal to ± (A-

B), A and B being the values specified in Table 2.

Parts of conversion devices that can be tested separately

(a)

Calculator

MPEs on quantities of liquid indications applicable to calculation, positive

or negative, are equal to one-tenth of the MPEs as defined in line A of Table 2.

(b)

Associated measuring instruments

Associated measuring instruments shall have accuracy at least as good as the

values in Table 4

Table 4

MPE on Measurements

Accuracy classes of the measuring system

0,3

0,5

1,0

1,5

2,5

Temperature

±0,3°C

± 0,5°C

± 1,0°C

MPE on Measurements

Accuracy classes of the measuring system

0,3

0,5

1,0

1,5

2,5

Pressure

Less than 1 MPa: ± 50 kPa

From 1 to 4 MPa: ± 5%

Over 4 MPa: ± 200 kPa

± 1 kg/m 3

± 2 kg/m 3

±5 kg/m 3

Density

These values apply to the indication of the characteristic quantities of the

liquid displayed by the conversion device.

(c)

Accuracy for calculating function

The MPE for the calculation of each characteristic quantity of the liquid,

positive or negative is equal to two fifths of the value fixed in (b).

(7) The requirement (a) in point 2.6 applies to any calculation, not only

conversion.

(8) The measuring system shall not exploit the MPEs or systematically favour

3. Maximum permissible effect of disturbances

(1) The effect of an electromagnetic disturbance on a measuring system shall be

one of the following

the change in the measurement result is not greater than the critical

change value as defined in point 3.2, or

the indication of the measurement result shows a momentary variation

that cannot be interpreted, memorised or transmitted as a measuring

result. Furthermore, in the case of an interruptible system, this can also

mean the impossibility to perform any measurement, or

the change in the measurement result is greater than the critical change

value, in which case the measuring system shall permit the retrieval of

the measuring result just before the critical change value occurred and

cut off the flow.

(2) The critical change value is the greater of MPE/5 for a particular measured

quantity or Emin.

4. Durability

After an appropriate test, taking into account the period of time estimated by

the manufacturer, has been performed, the following criterion shall be satisfied

The variation of the measurement result after the durability test, when

compared with the initial measurement result, shall not exceed the value for meters

specified in line B of table 2.

5. Suitability

(1) For any measured quantity relating to the same measurement, the indications

provided by various devices shall not deviate one from another by more than one

scale interval where devices have the same scale interval. In the case where the

devices have different scale intervals, the deviation shall not be more than that of the

greatest scale interval.

However, in the case of a self-service arrangement the scale intervals of the

main indicating device on the measuring system and the scale intervals of the self-

service device shall be the same and results of measurement shall not deviate one

from another.

(2) It shall not be possible to divert the measured quantity in normal conditions of

use unless it is readily apparent.

(3) Any percentage of air or gas not easily detectable in the liquid shall not lead

to a variation of error greater than

0,5% for liquids other than potable liquids and for liquids of a viscosity

not exceeding 1 mPa.s, or

1% for potable liquids and for liquids of a viscosity exceeding 1 mPa.s.

However, the allowed variation shall never be smaller than 1% of MMQ.

This value applies in the case of air or gas pockets.

(4) Instruments for direct sales

(4.1) A measuring system for direct sales shall be provided with means for

resetting the display to zero.

It shall not be possible to divert the measured quantity.

(4.2) The display of the quantity on which the transaction is based shall be

permanent until all parties in the transaction have accepted the measurement result.

(4.3) Measuring systems for direct sales shall be interruptible.

(4.4) Any percentage of air or gas in the liquid shall not lead to a variation of

error greater than the values specified in point 5.3.

(5)

Fuel Dispensers

(5.1) Displays on fuel dispensers shall not be capable of being reset to zero during

a measurement.

(5.2) The start of a new measurement shall be inhibited until the display has been

reset to zero.

(5.3) Where a measuring system is fitted with a price display, the difference

between the indicated price and the price calculated from the unit price and the

indicated quantity shall not exceed the price corresponding to Emin. However this

difference need not be less than the smallest monetary value.

6. Power supply failure

A measuring system shall either be provided with an emergency power

supply device that will safeguard all measuring functions during the failure of the

main power supply device or be equipped with means to save and display the data

present in order to permit the conclusion of the transaction in progress and with

means to stop the flow at the moment of the failure of the main power supply device.

7. Putting into use

Table 5

Accuracy Class

Types of Measuring system

0,3

Measuring systems on pipeline

0,5

All measuring systems if not differently stated elsewhere in this

Table, in particular

-

fuel dispensers (not for liquefied gases),

-

measuring systems on road tankers for liquids of low

viscosity (< 20 mPa.s)

-

measuring systems for (un)loading ships and rail and

road tankers (1)

-

measuring systems for milk - measuring systems for

refuelling aircraft

Measuring systems for liquefied gases under pressure measured

at a temperature equal to or above – 10°C

Measuring systems normally in class 0,3 or 0,5 but used for

liquids

-

whose temperature is less than – 10°C or greater than

50°C

-

whose dynamic viscosity is higher than 1 000 mPa.s

-

whose maximum volumetric flowrate is not higher than

20 L/h

Measuring systems for liquefied carbon dioxide

Measuring systems for liquefied gases under pressure measured

at a temperature below – 10°C (other than cryogenic liquids)

measuring systems for cryogenic liquids (temperature below –

153°C)

(1) However, the Authority may require measuring systems of accuracy class 0,3 or

0,5 when used for the levying of duties on mineral oils when (un)loading ships and

rail and road tankers. Note: However, the manufacturer may specify a better

accuracy for a certain type of measuring system.

8. Units of measurement

The metered quantity shall be displayed in millilitres, cubic centimetres,

litres, cubic metres, grams, kilograms or tonnes.

CONFORMITY ASSESSMENT

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are: B + F or B + D or H1 or G.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) The making available on the market and, or putting into

use of any measuring instrument that satisfies the requirements of

these regulations shall not be impeded for reasons covered by these

regulations.

(2) The Authority shall take all appropriate measures to ensure

that measuring instruments are made available on the market and/or

put into use only if they satisfy the requirements of these

regulations.

(3) A measuring instrument may be required to satisfy the

provisions governing its putting into use that are justified by local

climatic conditions. In such a case, the Authority shall choose

appropriate upper and lower temperature limits from Table 1 of

Schedule I and may specify humidity conditions (condensing or

non-condensing) and whether the intended location of use is open

or closed.

(4) When different accuracy classes are defined for a

measuring instrument

(a) the instrument-specific Schedules under the heading

‘Putting into use’ may indicate the accuracy classes to

be used for specific applications;

(b) in all other cases the Authority may determine the

accuracy classes to be used for specific applications

within the classes defined, subject to allowing the use

of all accuracy classes on its territory;

(c) for the purposes of paragraph (a) or paragraph (b),

measuring instruments of a better accuracy class may

be used if the owner so chooses.

(5) At trade fairs, exhibitions, demonstrations or similar

events, the showing of measuring instruments not in conformity

with these regulations shall not be prevented

Provided that a visible sign clearly indicates their non-

conformity and their non-availability for making available on the

market and/or putting into use until they are brought into

conformity.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

AUTOMATIC WEIGHING INSTRUMENTS (MI-006)

The relevant essential requirements of Schedule I, the specific requirements

of this Schedule and the conformity assessment procedures listed in Chapter I of this

Schedule, apply to automatic weighing instruments defined below, intended to

determine the mass of a body by using the action of gravity on that body.

An instrument that determines the mass of a product without the

intervention of an operator and follows a predetermined

programme of automatic processes characteristic of the

instrument.

An automatic weighing instrument that determines the mass of

pre-assembled discrete loads (for example prepackages) or

single loads of loose material.

An automatic catchweigher that subdivides articles of different

mass into two or more subgroups according to the value of the

difference of their mass and a nominal set-point.

An automatic catchweigher that labels individual articles with

the weight value.

MEASURING INSTRUMENTS

An automatic catchweigher that labels individual articles with

the weight value, and price information.

An automatic weighing instrument that fills containers with a

predetermined and virtually constant mass of product from bulk.

filling instrument

An automatic weighing instrument that determines the mass of a

bulk product by dividing it into discrete loads. The mass of each

(totalising hopper discrete load is determined in sequence and summed. Each

discrete load is then delivered to bulk.

An automatic weighing instrument that continuously determines

the mass of a bulk product on a conveyor belt, without

systematic subdivision of the product and without interrupting

the movement of the conveyor belt.

Rail-weighbridge An automatic weighing instrument having a load receptor

inclusive of rails for conveying railway vehicles.

SPECIFIC REQUIREMENTS

Requirements common to all types of automatic weighing instruments

1. Rated Operating Conditions

The manufacturer shall specify the rated operating conditions for the

instrument as follows

(1)

For the measurand

The measuring range in terms of its maximum and minimum capacity.

(2)

For the electrical supply influence quantities

the nominal AC voltage supply, or the AC voltage limits.

the nominal and minimum DC voltage supply, or the DC voltage

limits.

(3)

For the mechanical and climatic influence quantities

The minimum temperature range is 30°C unless specified otherwise in the

following chapters of this Schedule. The mechanical environment classes according

to Schedule I, point 1.3.2 are not applicable. For instruments which are used under

special mechanical strain, e.g. instruments incorporated into vehicles, the

manufacturer shall define the mechanical conditions of use.

(4)

For other influence quantities (if applicable): The rate(s) of operation.

The characteristics of the product(s) to be weighed.

2. Permissible effect of disturbances - Electromagnetic environment

The required performance and the critical change value are given in the

relevant Chapter of this Schedule for each type of instrument.

3. Suitability

(1) Means shall be provided to limit the effects of tilt, loading and rate of

operation such that maximum permissible errors (MPEs) are not exceeded in normal

(2)

Adequate material handling facilities shall be provided to enable the

instrument to respect the MPEs during normal operation.

(3)

Any operator control interface shall be clear and effective.

(4) The integrity of the display (where present) shall be verifiable by the

operator.

(5) Adequate zero setting capability shall be provided to enable the instrument

to respect the MPEs during normal operation.

(6) Any result outside the measurement range shall be identified as such, where

a printout is possible.

4. Conformity assessment

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are: For mechanical systems

B + D or B + E or B + F or D1 or F1 or G or H1. For electromechanical

instruments

B + D or B + E or B + F or G or H1.

For electronic systems or systems containing software: B + D or B + F or G

or H1.

CHAPTER II

Automatic Catchweighers

1. Accuracy Classes

(1) Instruments are divided into primary categories designated by: X or Y as

specified by the manufacturer.

(2) These primary categories are further divided into four accuracy classes: XI,

XII, XIII & XIIII and Y(I), Y(II), Y(a) & Y(b) which shall be specified by the

manufacturer.

2. Category X Instruments

(1) Category X applies to instruments used to check prepackages made up in

accordance with the requirements of Council Regulation 76/211/EEC of 20 January

1976 on the approximation of the laws of Malta relating to the making-up by weight

or by volume of certain prepackaged products (1) applicable to prepackages.

(2) The accuracy classes are supplemented by a factor (x) that quantifies the

maximum permissible standard deviation as specified in point 4.2.

The manufacturer shall specify the factor (x), where (x) shall be ≤ 2 and in

the form 1 × 10k, 2 × 10k or 5 × 10k, where k is a negative whole number or zero.

3. Category Y Instruments

Category Y applies to all other automatic catchweighers

4. MPE

(1)

Mean error Category X/MPE Category Y instruments

Table1

Maximum

Maximum

Net Load (m) in verification scale intervals (e)

permissible

permissible

mean error

error

XI

Y(I)

XII

Y(II)

XIII

Y(a)

XIIII

Y(b)

X

Y

0 < m ≤ 50 000

0 < m ≤ 5 000

0 < m ≤ 500

0 < m ≤ 50

± 0,5 e

± 1 e

50 000 < m

5 000 < m

500 < m

50 < m ≤ 200

± 1,0 e

± 1,5 e

≤ 200 000

≤ 20 000

≤ 2 000

200 000 < m

20 000 < m

2 000 < m

200 < m ≤ 1 000

± 1,5 e

± 2 e

≤ 100 000

≤ 10 000

(2)

Standard deviation

Maximum permissible value for the standard deviation of a class X (x)

instrument is the result of the multiplication of the factor (x) by the value in Table 2

below.

Table 2

Net Load (m)

Maximum permissible standard deviation for class X(1)

m ≤ 50 g

0,48%

50 g < m ≤ 100 g

0,24 g

100 g < m ≤ 200 g

0,24%

200 g < m ≤ 300 g

0,48 g

300 g < m ≤ 500 g

0,16%

500 g < m ≤ 1 000 g

0,8 g

1 000 g < m ≤ 10 000 g

0,08%

10 000 g < m ≤ 15 000 g

8 g

15 000 g < m

0,053%

For class XI and XII (x) shall be less than 1.

For class XIII (x) shall be not greater than 1.

(1)

OJ L 46, 21.2.1976, p. 1.

For class XIIII (x) shall be greater than 1.

(3)

Verification scale interval -single interval instruments

Table 3

Accuracy classes

Verification scale

Number of verification scale

interval

intervals n = Max/e

Minimum

Maximum

XI

Y(I)

0,001 g ≤ e

50 000

-

XII

Y(II)

0,001 g ≤ e ≤ 0,05 g

100 000

0,1 g ≤ e

5 000

100 000

XIII

Y(a)

0,1 g ≤ e ≤ 2 g

10 000

5 g ≤ e

10 000

XIIII

Y(b)

5 g ≤ e

1 000

(4)

Verification scale interval - multi-interval instruments

Table 4

Accuracy classes

Verification scale

Number of verification scale

interval

intervals n = Max/e

Minimum

Maximum

value (1)

n = Maxi / e i

n = Maxi / e

(i=1)

XI

Y(I)

0,001 g ≤ e i

50 000

-

XII

Y(II)

0,001 g ≤ e i ≤ 0,05 g

5 000

100 000

0,1 g ≤ e i

5 000

100 000

XIII

Y(a)

0,1 g ≤ e i

10 000

XIIII

Y(b)

5 g ≤ e i

1 000

(1) For i = r the corresponding column of Table 3 applies with e replaced by e r.

Where

i = 1, 2, … r

i = partial weighing range

r = total number of partial ranges

5. Measurement Range

In specifying the measurement range for class Y instruments the

manufacturer shall take account that the minimum capacity shall not be less than

class Y(I)

100 e

class Y(II)

20 e for 0,001 g ≤ e ≤ 0,05 g, and 50 e for 0,1 g ≤ e

class Y(a)

20 e

class Y(b)

10 e

Scales used for grading,

5 e

e.g. postal scales and

garbage weighers

6. Dynamic Setting

(1) The dynamic setting facility shall operate within a load range specified by

the manufacturer.

(2) When fitted, a dynamic setting facility that compensates for the dynamic

effects of the load in motion shall be inhibited from operating outside the load range,

and shall be capable of being secured.

7. Performance under Influence Factors and Electromagnetic Disturbances

(1)

The MPEs due to influence factors are

(1.1) For category X instruments

-

For automatic operation; as specified in Tables 1 and 2,

-

For static weighing in non-automatic operation; as specified in Table 1.

(1.2) For category Y instruments

-

For each load in automatic operation; as specified in Table 1,

-

For static weighing in non-automatic operation; as specified for

category X in Table 1.

(2) The critical change value due to a disturbance is one verification scale

interval.

(3)

Temperature range

-

For class XI and Y(I) the minimum range is 5°C,

-

For class XII and Y(II) the minimum range is 15°C.

CHAPTER III

Automatic Gravimetric Filling Instruments

1. Accuracy classes

(1) The manufacturer shall specify both the reference accuracy class Ref(x) and

the operational accuracy class(es) X(x).

(2) An instrument type is designated a reference accuracy class, Ref(x),

corresponding to the best possible accuracy for instruments of the type. After

installation, individual instruments are designated for one or more operational

accuracy classes, X(x), having taken account of the specific products to be weighed.

The class designation factor (x) shall be ≤ 2, and in the form 1 × 10k, 2 × 10k or 5 ×

10k where k is a negative whole number or zero.

(3)

The reference accuracy class, Ref(x) is applicable for static loads.

(4) For the operational accuracy class X(x), X is a regime relating accuracy to

load weight and (x) is a multiplier for the limits of error specified for class X(1) in

point 2.2.

2. MPE

(1)

Static weighing error

(1.1) For static loads under rated operating conditions, the MPE for reference

accuracy class Ref(x), shall be 0,312 of the maximum permissible deviation of each

fill from the average; as specified in Table 5; multiplied by the class designation

factor (x).

(1.2) For instruments where the fill may be made up from more than one load (e.g.

cumulative or selective combination weighers) the MPE for static loads shall be the

accuracy required for the fill as specified in point 2.2 (i.e. not the sum of the

maximum permissible deviation for the individual loads).

(2)

Deviation from average fill

Table 5

Value of the mass, m (g), of the fills

Maximum permissible deviation of each

fill from the average for class X(1)

m ≤ 50

7,2%

50 < m ≤ 100

3,6 g

100 < m ≤ 200

3,6%

200 < m ≤ 300

7,2 g

300 < m ≤ 500

2,4%

500 < m ≤ 1 000

12 g

1 000 < m ≤ 10 000

1,2%

10 000 < m ≤ 15 000

120 g

15 000 < m

0,8%

Note

The calculated deviation of each fill from the average may be adjusted to

take account for the effect of material particle size.

(3)

Error relative to pre-set value (setting error)

For instruments where it is possible to pre-set a fill weight; the maximum

difference between the pre-set value and the average mass of the fills shall not

exceed 0,312 of the maximum permissible deviation of each fill from the average, as

specified in Table 5.

3. Performance Under Influence Factor And Electromagnetic Disturbance

(1)

The MPE due to influence factors shall be as specified in point 2.1.

(2) The critical change value due to a disturbance is a change of the static

weight indication equal to the MPE as specified in point 2.1 calculated for the rated

minimum fill, or a change that would give equivalent effect on the fill in the case of

instruments where the fill consists of multiple loads. The calculated critical change

value shall be rounded to the next higher scale interval (d).

(3)

The manufacturer shall specify the value of the rated minimum fill.

CHAPTER IV

Discontinuous Totalisers

1. Accuracy Classes

Instruments are divided into four accuracy classes as follows: 0,2; 0,5; 1; 2.

2. MPEs

Table 6

Accuracy class

MPE of totalised load

0,2

± 0,10%

0,5

± 0,25%

± 0,50%

± 1,00%

3. Totalisation scale interval

The totalisation scale interval (dt) shall be in the range

0,01% Max ≤ dt ≤ 0,2% Max

4. Minimum Totalised Load (Σmin)

The minimum totalised load (Σmin) shall be not less than the load at which

the MPE is equal to the totalisation scale interval (dt) and not less than the minimum

load as specified by the manufacturer.

5. Zero Setting

Instruments that do not tare weigh after each discharge shall have a zero

setting device. Automatic operation shall be inhibited if zero indication varies by

-

1 dt on instruments with automatic zero setting device;

-

0,5 dt on instruments with a semi-automatic, or non-automatic, zero

setting device.

6. Operator Interface

Operator adjustments and reset function shall be inhibited during automatic

operation.

7. Printout

On instruments equipped with a printing device, the reset of the total shall

be inhibited until the total is printed. The printout of the total shall occur if

automatic operation is interrupted.

MEASURING INSTRUMENTS

8. Performance under influence factors and electromagnetic disturbances

(1)

The MPEs due to influence factors shall be as specified in Table 7.

Table 7

Load (m) in totalisation scale intervals (dt) MPE

0 < m ≤ 500

± 0,5 dt

500 < m ≤ 2 000

± 1,0 dt

2 000 < m ≤ 10 000

± 1,5 dt

(2) The critical change value due to a disturbance is one totalisation scale

interval for any weight indication and any stored total.

CHAPTER V

Continuous Totalisers

1. Accuracy classes

Instruments are divided into three accuracy classes as follows: 0,5; 1; 2.

2. Measurement Range

(1) The manufacturer shall specify the measurement range, the ratio between the

minimum net load on the weighing unit and the maximum capacity, and the

minimum totalised load.

(2)

The minimum totalised load Σmin shall not be less than

800 d for class 0.5,

400 d for class 1,

200 d for class 2.

Where d is the totalisation scale interval of the general totalisation device.

3. MPE

Table 8

Accuracy class

MPE for totalised load

0,5

± 0,25%

± 0,5%

± 1,0%

4. Speed of the belt

The speed of the belt shall be specified by the manufacturer. For single-

speed beltweighers, and variable-speed beltweighers having a manual speed setting

control, the speed shall not vary by more than 5% of the nominal value. The product

shall not have a different speed than the speed of the belt.

5. General Totalisation Device

It shall not be possible to reset the general totalisation device to zero.

6. Performance under influence factors and electromagnetic disturbances

(1) The MPE due to influence factor, for a load not less than the Σmin, shall be

0,7 times the appropriate value specified in Table 8, rounded to the nearest

totalisation scale interval (d).

(2) The critical change value due to a disturbance shall be 0,7 times the

appropriate value specified in Table 8, for a load equal to Σmin, for the designated

class of the beltweigher; rounded up to the next higher totalisation scale interval (d).

CHAPTER VI

Automatic Rail Weighbridges

1. Accuracy classes

Instruments are divided into four accuracy classes as follows

0,2; 0,5; 1; 2.

2. MPE

(1) The MPEs for weighing-in-motion of a single wagon or a total train are

shown in Table 9.

Accuracy class

MPE

0,2

± 0,1%

0,5

± 0,25%

± 0,5%

± 1,0%

(2) The MPEs for the weight of coupled or uncoupled wagons weighing-in-

motion shall be one of the following values, whichever is the greatest

the value calculated according to Table 9, rounded to the nearest scale

interval;

the value calculated according to Table 9, rounded to the nearest scale

interval for a weight equal to 35% of the maximum wagon weight (as

inscribed on the descriptive markings);

one scale interval (d).

(3) The MPEs for the weight of train weighing-in-motion shall be one of the

following values, whichever is the greatest

the value calculated according to Table 9, rounded to the nearest scale

interval;

the value calculated according to Table 9, for the weight of a single

wagon equal to 35% of the maximum wagon weight (as inscribed on the

descriptive markings) multiplied by the number of reference wagons

(not exceeding 10) in the train, and rounded to the nearest scale

interval;

one scale interval (d) for each wagon in the train, but not exceeding 10

d.

(4) When weighing coupled wagons; the errors of not more than 10% of the

weighing results taken from one or more passes of the train may exceed the

appropriate MPE given in point 2.2, but shall not exceed twice the MPE.

3. Scale interval (d)

The relationship between the accuracy class and the scale interval shall be as

specified in Table 10.

MEASURING INSTRUMENTS

Table 10

Accuracy class

Scale interval (d)

0,2

d ≤ 50 kg

0,5

d ≤ 100 kg

d ≤ 200 kg

d ≤ 500 kg

4. Measurement range

(1) The minimum capacity shall not be less than 1 t, and not greater than the

value of the result of the minimum wagon weight divided by the number of partial

weighings.

The minimum wagon weight shall not be less than 50 d.

5. Performance under influence factor and electromagnetic disturbance

(1)

The MPE due to an influence factor shall be as specified in Table 11.

Table 11

Load (m) in verification scale intervals (d)

MPE

0 < m ≤ 500

± 0,5 d

500 < m ≤ 2 000

± 1,0 d

2 000 < m ≤ 10 000

± 1,5 d

The critical change value due to a disturbance is one scale interval.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) When placing their measuring instruments on the

market and, or putting them into use, manufacturers shall ensure

that they have been designed and manufactured in accordance with

the essential requirements set out in Schedule I and in the relevant

instrument-specific Schedules.

(2) Manufacturers shall draw up the technical documentation

referred to in regulation 10 and carry out the relevant conformity

assessment procedure referred to in regulation 17 or have it carried

out.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

TAXIMETERS (MI-007)

The relevant requirements of Schedule I, the specific requirements of this

Schedule and the conformity assessment procedures listed in this Schedule apply to

taximeters.

DEFINITIONS

Taximeter

A device that works together with a signal generator ( * ) to make a measuring

instrument.

This device measures duration, calculates distance on the basis of a signal

delivered by the distance signal generator. Additionally, it calculates and displays

the fare to be paid for a trip on the basis of the calculated distance and/or the

measured duration of the trip.

Fare

The total amount of money due for a trip based on a fixed initial hire fee

and/or the length and/or the duration of the trip. The fare does not include a

supplement charged for extra services.

Cross-over speed

*The distance signal generator is outside the scope of this regulation.

The speed value found by division of a time tariff value by a distance tariff

Normal calculation mode S (single application of tariff)

Fare calculation based on application of the time tariff below the cross-over

speed and application of the distance tariff above the cross-over speed.

Normal calculation mode D (double application of tariff)

Fare calculation based on simultaneous application of time tariff and

distance tariff over the whole trip.

Operating position

The different modes in which a taximeter fulfils the different parts of its

functioning. The operating positions are distinguished by the following indications

‘For Hire’

The operating position in which the fare calculation

is disabled

‘Hired’

The operating position in which the fare calculation

takes place on the basis of a possible initial charge

and a tariff for distance travelled and/or time of the

trip

‘Stopped’

The operating position in which the fare due for the

trip is indicated and at least the fare calculation

based on time is disabled.

DESIGN REQUIREMENTS

1. The taximeter shall be designed to calculate the distance and to measure the

duration of a trip.

2. The taximeter shall be designed to calculate and display the fare,

incrementing in steps equal to the resolution fixed by MCCAA in the operation

position ‘Hired’. The taximeter shall also be designed to display the final value for

the trip in the operating position ‘Stopped’.

3. A taximeter shall be able to apply the normal calculation modes S and D. It

shall be possible to choose between these calculation modes by a secured setting.

4. A taximeter shall be able to supply the following data through an

appropriate secured interface(s)

-

operation position: ‘For Hire’, ‘Hired’ or ‘Stopped’;

-

totaliser data according to point 15.1;

-

general information: constant of the distance signal generator, date of

securing, taxi identifier, real time, identification of the tariff;

-

fare information for a trip: total charged, fare, calculation of the fare,

supplement charge, date, start time, finish time, distance travelled;

-

tariff(s) information: parameters of tariff(s).

National legislation may require certain devices to be connected to the

interface(s) of a taximeter. Where such a device is required; it shall be possible, by

secured setting, to inhibit automatically the operation of the taximeter for reasons of

the non-presence or improper functioning of the required device.

5. If relevant, it shall be possible to adjust a taximeter for the constant of the

distance signal generator to which it is to be connected and to secure the adjustment.

6. RATED OPERATING CONDITIONS

(1)

The mechanical environment class that applies is M3.

(2) The manufacturer shall specify the rated operating conditions for the

instrument, in particular

-

a minimum temperature range of 80°C for the climatic environment;

-

the limits of the DC power supply for which the instrument has been

designed. MAXIMUM PERMISSIBLE ERRORS (MPEs)

7. The MPE, excluding any errors due to application of the taximeter in a taxi,

are

-

For the time elapsed: ± 0,1%

-

minimum value of mpe: 0,2 s;

-

For the distance travelled: ± 0,2%

-

minimum value of mpe: 4 m;

-

For the calculation of the fare: ± 0,1%

-

minimum, including rounding: corresponding to the least significant

digit of the fare indication.

PERMISSIBLE EFFECT OF DISTURBANCES

8. Electromagnetic immunity

(1)

The electromagnetic class that applies is E3.

(2) The MPE laid down in point 7 shall also be respected in the presence of an

electromagnetic disturbance.

POWER SUPPLY FAILURE

9. In case of a reduction of the voltage supply to a value below the lower

operating limit as specified by the manufacturer, the taximeter shall

-

continue to work correctly or resume its correct functioning without

loss of data available before the voltage drop if the voltage drop is

temporary, i.e. due to restarting the engine;

-

abort an existing measurement and return to the position ‘For Hire’ if

the voltage drop is for a longer period.

OTHER REQUIREMENTS

10. The conditions for the compatibility between the taximeter and the distance

signal generator shall be specified by the manufacturer of the taximeter.

11. If there is a supplement charge for an extra service, entered by the driver on

manual command, this shall be excluded from the fare displayed. However, in that

case a taximeter may display temporarily the value of the fare including the

supplementary charge.

12. If the fare is calculated according to calculation mode D a taximeter may

have an additional display mode in which only the total distance and duration of the

trip are displayed in real time.

13. All values displayed for the passenger shall be suitably identified. These

values as well as their identification shall be clearly readable under daylight and

night conditions.

14. (1)If the fare to be paid or the measures to be taken against fraudulent use

can be affected by the choice of functionality from a pre-programmed setting or by

free data setting, it shall be possible to secure the instrument settings and data

entered.

(2) The securing possibilities available in a taximeter shall be such that separate

securing of the settings is possible.

(3)

The provisions in point 8.3 of Schedule I apply also to the tariffs.

15. (1) A taximeter shall be fitted with non-resettable totalisers for all of the

following values

-

The total distance travelled by the taxi;

-

The total distance travelled when hired;

-

The total number of hirings;

-

The total amount of money charged as supplements;

-

The total amount of money charged as fare.

-

The totalised values shall include the values saved according to point 9

under conditions of loss of power supply.

(2) If disconnected from power, a taximeter shall allow the totalised values to be

stored for one year for the purpose of reading out the values from the taximeter to

another medium.

(3) Adequate measures shall be taken to prevent the display of totalised values

from being used to deceive passengers.

16. Automatic change of tariffs is allowed due to the

-

distance of the trip;

-

duration of the trip;

-

time of the day;

-

date;

-

day of the week.

17. If properties of the taxi are important for the correctness of the taximeter,

the taximeter shall provide means to secure the connection of the taximeter to the

taxi in which it is installed.

18. For the purpose of testing after installation, the taximeter shall be equipped

with the possibility to test separately the accuracy of time and distance measurement

and the accuracy of the calculation.

19. A taximeter and its installation instructions specified by the manufacturer

shall be such that, if installed according to the manufacturer ’s instructions,

fraudulent alterations of the measurement signal representing the distance travelled

are sufficiently excluded.

20. The general essential requirement dealing with fraudulent use shall be

fulfilled in such a way that the interests of the customer, the driver, the driver’s

employer and the fiscal authorities are protected.

21. A taximeter shall be designed so that it can respect the MPEs without

adjustment during a period of one year of normal use.

22. The taximeter shall be equipped with a real-timeclock by means of which

the time of the day and the date are kept, one or both can be used for automatic

change of tariffs. The requirements for the real-time clock are

-

the timekeeping shall have an accuracy of 0,02%;

-

the correction possibility of the clock shall be not more than 2 minutes

per week. Correction for summer and wintertime shall be performed

automatically;

-

correction, automatic or manually, during a trip shall be prevented.

23. The values of distance travelled and time elapsed, when displayed or printed

in accordance with this regulations, shall use the following units

Distance travelled

-

kilometres;

-

miles, in those Member States to which Article (1)(b) of Regulation 80/

181/EEC applies.

Time elapsed

-

seconds, minutes or hours, as may be suitable; keeping in mind the

necessary resolution and the need to prevent misunderstandings.

CONFORMITY ASSESSMENT

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are: B + F or B + D or H1.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Where compliance of a measuring instrument with the

applicable requirements of these regulations has been demonstrated

by that conformity assessment procedure, manufacturers shall draw

up an EU declaration of conformity and affix the CE marking and

the supplementary metrology marking.

MEASURING INSTRUMENTS

(2) Manufacturers shall keep the technical documentation and

the EU declaration of conformity for ten years after the measuring

instrument has been placed on the market.

(3) Manufacturers shall ensure that procedures are in place for

series production to remain in conformity with these regulations.

Changes in measuring instrument design or characteristics and

changes in the harmonised standards, normative documents or in

other technical specifications by reference to which conformity of a

measuring instrument is declared shall be adequately taken into

account.

(4) When deemed appropriate with regard to the performance

of a measuring instrument, manufacturers shall carry out sample

testing of measuring instruments made available on the market,

investigate and, if necessary, keep a register of complaints, of non-

conforming measuring instruments and measuring instrument

recalls, and shall keep distributors informed of any such

monitoring.

(5) Manufacturers shall ensure that measuring instruments

which they have placed on the market bear a type, batch or serial

number or other element allowing their identification, or, where the

size or nature of the measuring instrument does not allow it, that

the required information is provided in a document accompanying

t h e m e a s u r i n g i n s t r u m e n t a n d o n t h e p a c k ag i n g , i f a n y, i n

accordance with point 9.2 of Schedule I.

(6) Manufacturers shall indicate on the measuring instrument

their name, registered trade name or registered trade mark and the

postal address at which they can be contacted or, where that is not

possible, in a document accompanying the measuring instrument

and on the packaging, if any, in accordance with point 9.2 of

Schedule I. The address shall indicate a single point at which the

manufacturer can be contacted. The contact details shall be in at

least English and, or Maltese.

(7) Manufacturers shall ensure that the measuring instrument

which they have placed on the market is accompanied by a copy of

the EU Declaration of conformity and by instructions and

information in accordance with point 9.3 of Schedule I, in at least

English and, or Maltese. Such instructions and information, as well

as any labelling, shall be clear, understandable and intelligible.

(8) Manufacturers who consider or have reason to believe that

a measuring instrument which they have placed on the market is not

in conformity with these regulations shall immediately take the

corrective measures necessary to bring that measuring instrument

into conformity, to withdraw i t or r ecall it, if a ppropriate.

Furthermore, where the measuring instrument presents a risk,

manufacturers shall immediately inform the Authority in which

they made the measuring instrument available on the market to that

effect, giving details, in particular, of the non-compliance and of

any corrective measures taken.

(9) Manufacturers shall, further to a reasoned request from the

Authority, provide it with all the information and documentation in

paper or electronic form necessary to demonstrate the conformity

MEASURING INSTRUMENTS

of the measuring instrument with these regulations, in at least

English and, or Maltese. They shall cooperate with that authority, at

its request, on any action taken to eliminate the risks posed by

measuring instruments which they have placed on the market.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

MATERIAL MEASURES (MI-008) CHAPTER I

Material measures of length

The relevant essential requirements of Schedule I, the specific requirements

of this Schedule and the conformity assessment procedures listed in this chapter,

apply to material measures of length defined below. However, the requirement for

the supply of a copy of declarations of conformity may be interpreted as applying to

a batch or consignment rather than each individual instrument.

DEFINITIONS

Material measure of length

An instrument comprising scale marks whose

distances are given in legal units of length.

SPECIFIC REQUIREMENTS

1. Reference Conditions

(1) For tapes of length equal to or greater than 5 metres, the maximum

permissible errors (MPEs) are to be met when a tractive force of fifty newtons or

other force values as specified by the manufacturer and marked on the tape

accordingly, or in the case of rigid or semi-rigid measures no tractive force is

needed, is applied.

(2) The reference temperature is 20°C unless otherwise specified by the

manufacturer and marked on the measure accordingly.

MPEs

2. The MPE, positive or negative in mm, between two non-consecutive scale

marks is (a + bL), where

-

L is the value of the length rounded up to the next whole metre; and

-

a and b are given in Table 1 below.

When a terminal interval is bounded by a surface, the MPE for any

distance beginning at this point is increased by the value c given in

Table 1.

Table 1

Accuracy Class

a (mm)

b

c (mm)

I

0,1

0,1

0,1

II

0,3

0,2

0,2

III

0,6

0,4

0,3

D - special class for dipping tapes (1)

1,5

zero

zero

Up to and including 30 m (2)

S - special class for tank strapping tapes

1,5

zero

zero

For each 30 m length when the tape is supported

on a flat surface

(1) Applies to the tape/dip weight combinations.

(2) If the nominal tape length exceeds 30 m, an additional mpe of 0,75 mm shall be

permitted for each 30 m of tape length.

Dip tapes may also be of Classes I or II in which case for any length between

two scale marks, one of which is on the sinker and the other on the tape, the MPE is

± 0,6 mm when application of the formula gives a value of less than 0,6 mm.

The MPE for the length between consecutive scale marks, and the maximum

permissible difference between two consecutive intervals, are given in Table 2

below.

Table 2

Length i of the interval

MPE or difference in millimetres

according to accuracy class

I

II

III

i ≤ 1 mm

0,1

0,2

0,3

1 mm < i ≤ 1 cm

0,2

0,4

0,6

Where a rule is of the folding type, the jointing shall be such as not to cause

any errors, supplementary to those above, exceeding: 0,3 mm for Class II, and 0,5

mm for Class III.

3. Materials

(1) Materials used for material measures shall be such that length variations due

to temperature excursions up to ± 8°C about the reference temperature do not exceed

the MPE. This does not apply to Class S and Class D m easures wh ere the

manufacturer intends that thermal expansion corrections shall be applied to observe

readings where necessary.

(2) Measures made from material whose dimensions may alter materially when

subjected to a wide range of relative humidity, may only be included in Classes II or

III.

4. Markings

The nominal value shall be marked on the measure. Millimetre scales shall

be numbered every centimetre and measures with a scale interval greater than 2 cm

shall have all scale marks numbered.

CONFORMITY ASSESSMENT

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are: F1 or D1 or B + D or H or G.

CHAPTER II

Capacity serving measures

The relevant essential requirements of Schedule I, and the specific

requirements and the conformity assessment procedures listed in this chapter, apply

to capacity serving measures defined below. However, the requirement for the

supply of a copy of declarations of conformity may be interpreted as applying to a

batch or consignment rather than each individual instrument. Also, the requirement

for the instrument to bear information in respect of its accuracy shall not apply.

DEFINITIONS

Capacity serving measure

A capacity measure (such as a drinking glass, jug or

thimble measure) designed to determine a specified

volume of a liquid (other than a pharmaceutical

product) which is sold for immediate consumption.

Line measure

A capacity serving measure marked with a line to

indicate nominal capacity.

Brim measure

A capacity serving measure for which the internal

volume is equal to the nominal capacity.

Transfer measure

A capacity serving measure from which it is intended

that the liquid is decanted prior to consumption.

Capacity

The capacity is the internal volume for brim

measures or internal volume to a filling mark for line

measures.

SPECIFIC REQUIREMENTS

1. Reference Conditions

(1)

Temperature: the reference temperature for measurement of capacity is

20°C.

(2)

Position for correct indication: free standing on a level surface.

2. MPEs

Table 1

Line

Brim

Transfer measures

< 100 ml

± 2 ml

– 0

+ 4 ml

≥ 100 ml

± 3%

– 0

+ 6%

Serving measures

< 200 ml

± 5%

– 0

+ 10%

≥ 200 ml

± (5 ml + 2,5%)

– 0

+ 10 ml + 5%

3. Materials

Capacity serving measures shall be made of material which is sufficiently

rigid and dimensionally stable to maintain capacity within the MPE.

4. Shape

(1) Transfer measures shall be designed so that a change of contents equal to the

MPE causes a change in level of at least 2 mm at the brim or filling mark.

(2) Transfer measures shall be designed so that the complete discharge of the

liquid being measured will not be impeded.

5. Marking

(1) The nominal capacity declared shall be clearly and indelibly marked on the

measure.

(2) Capacity serving measures may also be marked with up to three clearly

distinguishable capacities, none of which shall lead to confusion one to the other.

(3) All filling marks shall be sufficiently clear and durable to ensure that MPEs

are not exceeded in use.

CONFORMITY ASSESSMENT

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are: A2 or F1 or D1 or E1 or B + E or B + D or H.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) A manufacturer may, by a written mandate, appoint an

authorised representative.

(2) The obligations laid down in regulation 9(1) and the

obligation to draw up technical documentation referred to in

regulation 9(2) shall not form part of the authorised

representative’s mandate.

(3) An authorised representative shall perform the tasks

specified in the mandate received from the manufacturer. The

mandate shall allow the authorised representative to do at least the

following

(a) keep the EU declaration of conformity and the

technical documentation at the disposal of the

Authority for ten years after the measuring instrument

has been placed on the market;

(b) further to a reasoned request from a competent

national authority, provide that authority with all the

information and documentation necessary to

demonstrate the conformity of a measuring instrument;

(c) cooperate with the competent national authorities, at

their request, on any action taken to eliminate the risks

posed by measuring instruments covered by their

mandate.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

DIMENSIONAL MEASURING INSTRUMENTS (MI-009)

The relevant essential requirements of Schedule I, the specific requirements

of this Schedule and the conformity assessment procedures listed in this Schedule,

apply to dimensional measuring instruments of the types defined below.

DEFINITIONS

Length measuring instrument A l e n g t h m e a s u r i n g i n s t r u m e n t s e r v e s f o r t h e

determination of the length of rope- type materials

(e.g. textiles, bands, cables) during feed motion of

the product to be measured.

Area Measuring Instruments A n a r e a m e a s u r i n g i n s t r u m e n t s e r v e s f o r t h e

determination of the area of irregular shaped objects,

e.g. for leather.

Multi dimensional

A multi-dimensional measuring instrument serves for

Measuring Instruments

the determination of the edge length (length, height,

width) of the smallest enclosing rectangular parallel

piped of a product.

CHAPTER I

Requirements common to all dimensional measuring instruments

Electromagnetic immunity

1. The effect of an electromagnetic disturbance on a dimensional measuring

instrument shall be such that

-

the change in measurement result is no greater than the critical change

value as defined in point 2; or

-

it is impossible to perform any measurement; or

-

there are momentary variations in the measurement result that cannot be

interpreted, memorised or transmitted as a measuring result; or

-

there are variations in the measurement result severe enough to be

noticed by all those interested in the measurement result.

2. The critical change value is equal to one scale interval.

CONFORMITY ASSESSMENT

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are: For mechanical or electromechanical

instruments

F1 or E1 or D1 or B + F or B + E or B + D or H or H1 or G. For electronic

instruments or instruments containing software

B + F or B + D or H1 or G.

CHAPTER II

Length measuring instruments

Characteristics of the product to be measured

1. Textiles are characterised by the characteristic factor K. This factor takes the

stretchability and force per unit area of the product measured into account and is

defined by the following formula

K

= ε · (GA + 2,2 N/m2), where

ε is the relative elongation of a cloth specimen 1 m wide at a tensile

force of 10 N,

GA is the weight force per unit area of a cloth specimen in N/m 2 .

2. Operating conditions

(1)

Range

Dimensions and K-factor, where applicable, within the range specified by

the manufacturer for the instrument. The ranges of K-factor are given in Table 1

Table 1

Group

Range of K

Product

0 < K < 2 × 10 –2 N/m 2

I

low stretchability

2 × 10 –2 N/m 2 < K < 8 × 10 –2 N/m 2

II

medium stretchability

8 × 10 –2 N/m 2 < K < 24 × 10 –2 N/m 2

III

high stretchability

24 × 10 –2 N/m 2 < K

IV

very high stretchability

(2) Where the measured object is not transported by the measuring instrument,

its speed must be within the range specified by the manufacturer for the instrument.

(3) If the measurement result depends on the thickness, the surface condition

and the kind of delivery (e.g. from a big roll or from a pile), corresponding

limitations are specified by the manufacturer.

MPEs

Instrument

Table 2

Accuracy class

MPE

I

0,125%, but not less than 0,005 Lm

II

0,25%, but not less than 0,01 Lm

III

0,5%, but not less than 0,02 Lm

Where Lm is the minimum measurable length, that is to say the smallest

length specified by the manufacturer for which the instrument is intended to be used.

The true length value of the different types of materials shall be measured

using suitable instruments (e.g. tapes of length). Thereby, the material which is

going to be measured shall be laid out on a suitable underlay (e.g. a suitable table)

straight and unstretched.

Other requirements

The instruments must ensure that the product is measured unstretched

according to the intended stretch ability for which the instrument is designed.

CHAPTER III

Area measuring instruments

1. Operating conditions

(1)

Range

Dimensions within the range specified by the manufacturer for the

instrument.

(2)

Condition of the product

The manufacturer shall specify the limitations of the instruments due to the

speed, and thickness of the surface conditions if relevant, of the product.

MPEs

2. Instrument

The MPE is 1,0%, but not less than 1 dm 2 .

Other requirements

3. Presentation of the product

In the case of pulling back or stopping the product, it shall not be possible to

have an error of measurement or the display must be blanked.

4. Scale interval

The instruments must have a scale interval of 1,0 dm0,5%, but not less than

0,02 Lm. In addition, it must be possible to have a scale interval of 0,1 dm 2 for

testing purposes.

CHAPTER IV

Multidimensional measuring instruments

MEASURING INSTRUMENTS

Operating conditions

Dimensions within the range specified by the manufacturer for the

The lower limit of the minimum dimension for all values of the scale

interval is given in Table 1.

Minimum dimension (min) (lower limit)

10 d

20 d

50 d

The speed must be within the range specified by the manufacturer for the

2. The MPE is ± 1,0 d.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Importers shall place only compliant measuring

instruments on the market.

(2) Before placing a measuring instrument on the market and,

or putting a measuring instrument into use importers shall ensure

that the appropriate conformity assessment procedure referred to in

regulation 17 has been carried out by the manufacturer. They shall

ensure that the manufacturer has drawn up the technical

documentation, that the measuring instrument bears the CE

marking and the supplementary metrology marking and is

accompanied by a copy of the EU declaration of conformity and the

required documents, and that the manufacturer has complied with

the requirements set out in regulation 10(5) and (6).

(3) Where an importer considers or has reason to believe that a

measuring instrument is not in conformity with the essential

requirements set out in Schedule I and in the relevant instrument-

specific Schedules, he shall not place the measuring instrument on

t he m ark et o r pu t it i nto u se unt il i t has be en brou gh t in to

conformity. Furthermore, where the measuring instrument presents

a risk, the importer shall inform the manufacturer and the Authority

to that effect.

(4) Importers shall indicate on the measuring instrument their

name, registered trade name or registered trade mark and the postal

MEASURING INSTRUMENTS

address at which they can be contacted or, where that is not

possible, in a document accompanying the measuring instrument

and on its packaging, if any, in accordance with point 9.2 of

Schedule I. The contact details shall be in at least English and, or

Maltese.

(5) Importers shall ensure that the measuring instrument is

accompanied by instructions and information in accordance with

point 9.3 of Schedule I, in at least English and, or Maltese.

(6) Importers shall ensure that, while a measuring instrument is

under their responsibility, its storage or transport conditions do not

jeopardise its compliance with the essential requirements set out in

Schedule I and in the relevant instrument-specific Schedules.

(7) When deemed appropriate with regard to the performance

of a measuring instrument, importers shall carry out sample testing

of measuring instruments made available on the market,

investigate, and, if necessary, keep a register of complaints, of non-

conforming measuring instruments and measuring instrument

recalls, and shall keep distributors informed of any such

monitoring.

(8) Importers who consider or have reason to believe that a

measuring instrument which they have placed on the market is not

in conformity with these regulations shall immediately take the

corrective measures necessary to bring that measuring instrument

into conformity, to withdraw i t or r ecall it, if a ppropriate.

Furthermore, where the measuring instrument presents a risk,

importers shall immediately inform the Authority in which they

made the measuring instrument available on the market to that

effect, giving details, in particular, of the non-compliance and of

any corrective measures taken.

(9) Importers shall, for ten years after the measuring

instrument has been placed on the market keep a copy of the EU

declaration of conformity at the disposal of the Authority and

ensure that the technical documentation can be made available to

those authorities, upon request.

(10) Importers shall, further to a reasoned request from the

Authority, provide it with all the information and documentation in

paper or electronic form necessary to demonstrate the conformity

of a measuring instrument in at least English and, or Maltese. They

shall cooperate with the Authority, at its request, on any action

taken to eliminate the risks posed by measuring instruments which

they have placed on the market.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

EXHAUST GAS ANALYSERS (MI-010)

The relevant requirements of Schedule I, the specific requirements of this

Schedule and the conformity assessment procedures listed in this Schedule, apply to

exhaust gas analysers defined below intended for inspection and professional

maintenance of motor vehicles in use.

An exhaust gas analyser is a measuring instrument that

serves to determine the volume fractions of specified

components of the exhaust gas of a motor vehicle engine

with spark ignition at the moisture level of the sample

analysed.

These gas components are carbon monoxide (CO), carbon

dioxide (CO 2 ), oxygen (O 2 ) and hydrocarbons (HC).

The content of hydrocarbons has to be expressed as

concentration of n-hexane (C 6 H 14 ), measured with near-

infrared absorption techniques.

The volume fractions of the gas components are expressed

as a percentage (% vol) for CO, CO 2 and O 2 and in parts

per million (ppm vol) for HC.

Moreover, an exhaust gas analyser calculates the lambda

value from the volume fractions of the components of the

exhaust gas.

Lambda is a dimensionless value representative of the

burning efficiency of an engine in terms of air/fuel ratio in

the exhaust gases. It is determined with a reference

standardised formula.

SPECIFIC REQUIREMENTS

Two classes (0 and I) are being defined for exhaust gas analysers. The

relevant minimum measuring ranges for these classes are shown in Table 1.

Classes and measuring ranges

Parameter

Classes 0 and I

CO fraction

from 0 to 5% vol

CO 2 fraction

from 0 to 16% vol

HC fraction

from 0 to 2 000 ppm vol

O fraction

from 0 to 21% vol

from 0,8 to 1,2

Rated operating conditions

2. The values of the operating conditions shall be specified by the

manufacturer as follows

(1)

For the climatic and mechanical influence quantities

a minimum temperature range of 35°C for the climatic environment;

the mechanical environment class that applies is M1.

For the electrical power influence quantities

the voltage and frequency range for the AC voltage supply;

the limits of the DC voltage supply.

(3)

For the ambient pressure

the minimum and the maximum values of the ambient pressure are for

both classes: pmin ≤ 860 hPa, Pmax ≥ 1 060 hPa.

Maximum permissible errors (MPEs)

3. The MPEs are defined as follows

(1) For each of the fractions measured, the maximum error value permitted

under rated operating conditions according to point 1.1 of Schedule I is the greater of

the two values shown in Table 2. Absolute values are expressed in% vol or ppm vol,

percentage values are percent of the true value.

Class 0

Class I

± 0,03% vol

± 0,06% vol

± 5%

± 5%

CO 2 fraction

± 0,5% vol

± 0,5% vol

±5%

± 5%

HC fraction

± 10 ppm vol

± 12 ppm vol

± 5%

± 5%

O 2 fraction

± 0,1% vol

± 0,1% vol

± 5%

± 5%

(2) The MPE on lambda calculation is 0,3%. The conventional true value is

calculated according to the formula set out in point 5.3.7.3 of Regulation No 83 of

the Economic Commission for Europe of the United Nations (UN/ECE) (1).

For this purpose, the values displayed by the instrument are used for

calculation.

Permissible effect of disturbances

4. For each of the volume fractions measured by the instrument, the critical

change value is equal to the MPE for the parameter concerned.

5. The effect of an electromagnetic disturbance shall be such that

-

either the change in the measurement result is not greater than the

critical change value laid down in point 4;

-

or the presentation of the measurement result is such that it cannot be

taken for a valid result.

Other requirements

The resolution shall be equal to or of one order of magnitude higher than the

values shown in Table 3.

Table 3

Resolution

CO

CO 2

HC

Class 0 and Class 1

0.01% vol

0.1% vol

(15)

1 ppm vol

(1)

0,01% vol for measurand values below or equal to 4% vol, otherwise 0,1%

vol.

The lambda value shall be displayed with a resolution of 0,001. see OJ L42,

15.2.2012, p.1

7. The standard deviation of 20 measurements shall not be greater than one

third of the modulus of the MPE for each applicable gas volume fraction.

8. For measuring CO, CO2 and HC, the instrument, including the specified gas

handling system, must indicate 95% of the final value as determined with calibration

gases within 15 seconds after changing from a gas with zero content, e.g. fresh air.

For measuring O2, the instrument under similar conditions must indicate a value

differing less than 0,1% vol from zero within 60 seconds after changing from fresh

air to an oxygen-free gas.

9. The components in the exhaust gas, other than the components whose values

are subject to the measurement, shall not affect the measurement results by more

than the half of the modulus of the MPEs when those components are present in the

following maximum volume fractions

6% vol CO,

16% vol CO2,

10% vol O2,

5% vol H2,

0,3% vol NO,

2 000 ppm vol HC (as n-hexane), water vapor up to saturation.

10. An exhaust gas analyser shall have an adjustment facility that provides

operations for zero-setting, gas calibration and internal adjustment. The adjustment

facility for zero-setting and internal adjustment shall be automatic.

11. For automatic or semi-automatic adjustment facilities, the instrument shall

be unable to make a measurement as long as the adjustments have not been made.

12. An exhaust gas analyser shall detect hydrocarbon residues in the gas

handling system. It shall not be possible to carry out a measurement if the

hydrocarbon residues, present before any measurement, exceed 20 ppm vol.

13. An exhaust gas analyser shall have a device for automatically recognising

any malfunctioning of the sensor of the oxygen channel due to wear or a break in the

connecting line.

14. If the exhaust gas analyser is capable to operate with different fuels (e.g.

petrol or liquefied gas), there shall be the possibility to select the suitable

coefficients for the Lambda calculation without ambiguity concerning the

appropriate formula.

CONFORMITY ASSESSMENT

The conformity assessment procedures referred to in Article 17 that the

manufacturer can choose between are: B + F or B + D or H1.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) When making a measuring instrument available on the

market and, or putting it into use, distributors shall act with due

care in relation to the requirements of these regulations.

(2) Before making a measuring instrument available on the

market and, or putting a measuring instrument into use distributors

shall verify that the measuring instrument bears the CE marking

and the supplementary metrology marking, that it is accompanied

by the EU declaration of conformity, by the required documents

and by instructions and information in accordance with point 9.3 of

MEASURING INSTRUMENTS

Schedule I, in at least English and, or Maltese, or put into use, and

that the manufacturer and the importer have complied with the

requirements set out in regulation 10(5) and (6) and regulation

12(3) respectively.

(3) Where a distributor considers or has reason to believe that a

measuring instrument is not in conformity with the essential

requirements set out in Schedule I and in the relevant instrument-

specific Schedules, he shall not make the measuring instrument

available on the market or put it into use, until it has been brought

into conformity. Furthermore, where the measuring instrument

presents a risk, the distributor shall inform the manufacturer or the

importer to that effect as well as the Authority.

(4) Distributors shall ensure that, while a measuring instrument

is under their responsibility, its storage or transport conditions do

not jeopardise its compliance with the essential requirements set

out in Schedule I and in the relevant instrument-specific Schedules.

(5) Distributors who consider or have reason to believe that a

measuring instrument which they have made available on the

market or put into use is not in conformity with these regulations

shall make sure that the corrective measures necessary to bring that

measuring instrument into conformity, to withdraw it or recall it, if

appropriate, are taken. Furthermore, where the measuring

instrument presents a risk, distributors shall immediately inform

the competent national authorities of the Member States in which

they made the measuring instrument available on the market to that

effect, giving details, in particular, of the non-compliance and of

any corrective measures taken.

(5) Distributors shall, further to a reasoned request from the

Authority, provide it with all the information and documentation in

paper or electronic form necessary to demonstrate the conformity

of a measuring instrument. They shall cooperate with the Authority,

at its request, on any action taken to eliminate the risks posed by

measuring instruments which they have made available on the

market.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

EU DECLARATION OF CONFORMITY (No XXXX) ( * )

1. Instrument model/Instrument (product, type, batch or serial number)

2. Name and address of the manufacturer and, where applicable, his authorised

representative

3. This declaration of conformity is issued under the sole responsibility of the

manufacturer.

4. Object of the declaration (identification of instrument allowing traceability;

it may, where necessary for the identification of the instrument, include an image)

5. The object of the declaration described above is in conformity with the

relevant Union harmonisation legislation

6. References to the relevant harmonised standards or normative documents

used or references to the other technical specifications in relation to which

conformity is declared

7. Where applicable, the notified body … (name, number) performed …

(description of intervention) and issued the certificate

8. Additional information

* It is optional for the manufacturer to assign a number to the declaration of conformity.

Signed for and on behalf of

(place and date of issue)

(name, function)

(signature)

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

An importer or distributor shall be considered a

manufacturer for the purposes of these regulations and he shall be

subject to the obligations of the manufacturer under regulation 9,

where he places a measuring instrument on the market under his

name or trade mark or modifies a measuring instrument already

placed on the market in such a way that compliance with these

regulations may be affected.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

Economic operators shall, on request, identify the

following to the Authority

(a) any economic operator who has supplied them with a

measuring instrument;

(b) any economic operator to whom they have supplied a

measuring instrument;

(c) economic operators shall be able to present the

information referred to in paragraph (a) for ten years

MEASURING INSTRUMENTS

after they have been supplied with the measuring

instrument and for ten years after they have supplied

the measuring instrument.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Measuring instruments which are in conformity with

harmonised standards or parts thereof shall be presumed to be in

conformity with the essential requirements set out in Schedule I

and in the relevant instrument-specific Schedules covered by those

standards or parts thereof.

(2) Measuring instruments which are in conformity with parts

of normative documents shall be presumed to be in conformity with

the essential requirements set out in Schedule I and in the relevant

instrument-specific Schedules covered by those parts of normative

documents.

(3) A manufacturer may choose to use any technical solution

that complies with the essential requirements set out in Schedule I

and in the relevant instrument-specific Schedules. In addition, to

benefit from the presumption of conformity, the manufacturer must

correctly apply solutions mentioned either in the relevant

harmonised standards or in the normative documents referred to in

sub-regulations (1) and (2).

(4) Compliance shall be presumed with the appropriate tests

mentioned in regulation 18(3) if the corresponding test programme

has been performed in accordance with the relevant documents

mentioned in sub-regulations (1), (2) and (3) and if the test results

ensure compliance with the essential requirements.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Conformity assessment of a measuring instrument with

the applicable essential requirements shall be carried out by the

application, at the choice of the manufacturer, of one of the

conformity assessment procedures listed in the relevant instrument-

specific Schedules.

(2) The conformity assessment procedures are set out in

Schedule II. Records and correspondence relating to conformity

assessment procedures shall be drawn up in the at least English

language where the notified body carrying out the conformity

assessment procedures is established, or in a language accepted by

that body.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) The technical documentation shall render the design,

manufacture and operation of the measuring instrument intelligible

and shall permit an assessment of its conformity with the applicable

requirements of these regulations.

(2) The technical documentation shall be sufficiently detailed

to ensure compliance with the following requirements

(a) the definition of the metrological characteristics;

(b) the reproducibility of the metrological performances

of produced measuring instruments when properly

adjusted using appropriate intended means;

(c) the integrity of the measuring instrument.

(3)

The technical documentation shall insofar as relevant for

MEASURING INSTRUMENTS

assessment and identification of the type and, or the measuring

instruments include the following information

(a) a general description of the measuring instrument;

(b) conceptual design and manufacturing drawings and

plans of components, sub-assemblies, circuits, etc.;

(c) manufacturing

procedures

to

ensure

consistent

production;

(d) if applicable, a description of the electronic devices

with drawings, diagrams, flow diagrams of the logic

and general software information explaining their

characteristics and operations;

(e) descriptions and explanations necessary for the

understanding of the information referred to in

paragraphs (b), (c) and (d), including the operation of

the measuring instrument;

(f)

a list of the harmonised standards and, or normative

documents referred to in regulation, applied in full or

in part, the references of which have been published in

the Official Journal of the European Union;

(g) descriptions of the solutions adopted to meet the

essential requirements where the harmonised standards

and/or normative documents referred to in regulation

16 have not been applied, including a list of other

relevant technical specifications applied;

(h) results of design calculations, examinations, etc.;

(i)

the appropriate test results, where necessary, to

demonstrate that the type and, or the measuring

instruments comply with the following

(i)

the requirements of these regulations under

declared rated operating conditions and under

specified environmental disturbances; and

(ii) the durability specifications for gas-, water-,

thermal energy-meters as well as for liquids

other than water;

(j)

the EU-type examination certificates or EU design

examination certificates in respect of measuring

instruments containing parts identical to those in the

design.

(4) The manufacturer shall specify where seals and markings

have been applied.

(5) The manufacturer shall indicate the conditions for

compatibility with interfaces and sub-assemblies, where relevant.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) The EU declaration of conformity shall state that the

fulfilment of the essential requirements set out in Schedule I and in

the relevant instrument-specific Schedules has been demonstrated.

(2) The EU declaration of conformity shall have the model

structure set out in Schedule XIII, shall contain the elements

specified in the relevant modules set out in Schedule II and shall be

MEASURING INSTRUMENTS

continuously updated. It shall be translated in at least English and,

or Maltese.

(3) Where a measuring instrument is subject to more than one

Union act requiring an EU declaration of conformity, a single EU

declaration of conformity shall be drawn up in respect of all such

Union acts. That declaration shall contain the identification of the

Union acts concerned, including their publication references.

(4) By drawing up the EU declaration of conformity, the

manufacturer shall assume responsibility for the compliance of the

measuring instrument with the requirements laid down in these

regulations.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

The conformity of a measuring instrument with these

regulations shall be indicated by the presence on it of the CE

marking and the supplementary metrology marking as specified in

regulation 21.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) The CE marking shall be subject to the general

principles set out in regulation 30 of Regulation (EC) No 765/2008.

(2) The supplementary metrology marking shall consist of the

capital letter ‘M’ and the last two digits of the year of its affixing,

surrounded by a rectangle. The height of the rectangle shall be

equal to the height of the CE marking.

(3) The general principles set out in regulation 30 of

Regulation (EC) No 765/2008 shall apply, mutatis mutandis, to the

supplementary metrology marking.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) The CE marking and the supplementary metrology

marking shall be affixed visibly, legibly and indelibly to the

measuring instrument or to its data plate. Where that is not possible

or not warranted on account of the nature of the measuring

instrument, they shall be affixed to the accompanying documents

and to the packaging, if any.

(2) When a measuring instrument consists of a set of devices,

not being sub-assemblies, operating together, the CE marking and

the supplementary metrology marking shall be affixed on the

instrument’s main device.

(3) The CE marking and the supplementary metrology marking

shall be affixed before the measuring instrument is placed on the

market.

(4) The CE marking and the supplementary metrology marking

may be affixed to the instrument during the fabrication process, if

justified.

(5) The supplementary metrology marking shall immediately

follow the CE marking.

(6) The CE marking and the supplementary metrology marking

shall be followed by the identification number of the notified body,

where that body is involved in the production control phase as set

out in Schedule II.

(7)

The identification number of the notified body shall be

MEASURING INSTRUMENTS

affi xed b y the body itself or, under its instructions, by the

manufacturer or his authorised representative.

(8) The identification number of the notified body concerned

shall be indelible or self destructive upon removal.

(9) The CE marking, the supplementary metrology marking

and, where applicable, the identification number of the notified

body may be followed by any other mark indicating a special risk

or use.

(10) The Authority shall build upon existing mechanisms to

ensure correct application of the regime governing the CE marking

and shall take appropriate action in the event of improper use of

that marking.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

The Authority shall notify the Commission and the other

Member States of bodies authorised to carry out third-party

conformity assessment tasks under these regulations.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) The Authority shall be responsible for setting up and

carrying out the necessary procedures for the assessment and

notification of conformity assessment bodies and the monitoring of

notified bodies, including compliance with the provisions of

regulation 29.

(2) The assessment and monitoring referred to in regulation

21(1) shall be carried out by the National Accreditation Board

(Malta).

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) The Authority shall be

(a) established in such a way that no conflict of interest

with conformity assessment bodies occurs;

(b) organised and operated so as to safeguard the

objectivity and impartiality of its activities;

(c) be organised in such a way that each decision relating

to notification of a conformity assessment body is

taken by competent persons different from those who

carried out the assessment;

(d) not offer or provide any activities that conformity

assessment bodies perform or consultancy services on

a commercial or competitive basis;

(e) safeguard the confidentiality of the information it

obtains;

(f)

have a sufficient number of competent personnel at its

disposal for the proper performance of its tasks.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

The Authority shall inform the Commission of their

procedures for the assessment and notification of conformity

assessment bodies and the monitoring of notified bodies, and of any

changes thereto.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) For the purposes of notification, a conformity

assessment body shall meet the requirements laid down in sub-

regulations (2) to (11).

MEASURING INSTRUMENTS

(2) A conformity assessment body shall be established under

national law of Malta and have legal personality.

(3) A conformity assessment body shall be a third-party body

independent of the organisation or the measuring instrument it

assesses.

A body belonging to a business association or professional

federation representing undertakings involved in the design,

manufacturing, provision, assembly, use or maintenance of

measuring instruments which it assesses, may, on condition that its

independence and the absence of any conflict of interest are

demonstrated, be considered such a body.

(4) A conformity assessment body, its top level management

and the personnel responsible for carrying out the conformity

assessment tasks shall not be the designer, manufacturer, supplier,

installer, purchaser, owner, user or maintainer of the measuring

instruments which they assess, nor the representative of any of

those parties. This shall not preclude the use of assessed measuring

instruments that are necessary for the operations of the conformity

assessment body or the use of such measuring instruments for

personal purposes.

A conformity assessment body, its top level management

and the personnel responsible for carrying out the conformity

assessment tasks shall not be directly involved in the design,

manufacture or construction, the marketing, installation, use or

maintenance of those measuring instruments, or represent the

parties engaged in those activities. They shall not engage in any

activity that may conflict with their independence of judgement or

integrity in relation to conformity assessment activities for which

they are notified. This shall in particular apply to consultancy

services.

The second paragraph does not, however, preclude the

possibility of exchanges of technical information between the

m a n u f a c t u r e r a n d t h e b od y f o r th e pu r p o s e s o f c o n f o r m i ty

assessment.

Conformity assessment bodies shall ensure that the

activities of their subsidiaries or subcontractors do not affect the

confidentiality, objectivity or impartiality of their conformity

assessment activities.

(5) Conformity assessment bodies and their personnel shall

carry out the conformity assessment activities with the highest

degree of professional integrity and the requisite technical

competence in the specific field and shall be free from all pressures

and inducements, particularly financial, which might influence

their judgement or the results of their conformity assessment

activities, especially as regards persons or groups of persons with

an interest in the results of those activities.

(6) A conformity assessment body shall be capable of carrying

out all the conformity assessment tasks assigned to it by Schedule

II and in relation to which it has been notified, whether those tasks

are carried out by the conformity assessment body itself or on its

behalf and under its responsibility. At all times and for each

MEASURING INSTRUMENTS

conformity assessment procedure and each kind or category of

measuring instruments in relation to which it has been notified, a

conformity assessment body shall have at its disposal the

necessary

(a) personnel with technical knowledge and sufficient and

appropriate experience to perform the conformity

assessment tasks;

(b) descriptions of procedures in accordance with which

conformity assessment is carried out, ensuring the

transparency and the ability of reproduction of those

procedures. It shall have appropriate policies and

procedures in place that distinguish between tasks it

carries out as a notified body and other activities;

(c) procedures for the performance of activities which

take due account of the size of an undertaking, the

sector in which it operates, its structure, the degree of

complexity of the measuring instrument technology in

question and the mass or serial nature of the

production process.

A conformity assessment body shall have the means

necessary to perform the technical and administrative tasks

c o n n e c t ed w i t h t h e c o n f o r m i t y as s e s s m en t a c t i v i t i e s i n a n

a p p r o p r i a t e m an n e r a n d sh al l h a v e a c c e s s t o a l l n e c es s a r y

equipment or facilities.

(7) The personnel responsible for carrying out conformity

assessment tasks shall have the following

(a) sound technical and vocational training covering all

the conformity assessment activities in relation to

which the conformity assessment body has been

notified;

(b) satisfactory knowledge of the requirements of the

assessments they carry out and adequate authority to

carry out those assessments;

(c) appropriate knowledge and understanding of the

essential requirements set out in Schedule I and in the

relevant instrument-specific Schedules, of the

applicable harmonised standards and normative

documents and of the relevant provisions of Union

harmonisation legislation and of national legislation;

(d) the ability to draw up certificates, records and reports

demonstrating that assessments have been carried out.

(8) The impartiality of the conformity assessment bodies, their

top level management and of the personnel responsible for carrying

out the conformity assessment tasks shall be guaranteed.

The remuneration of the top level management and

personnel responsible for carrying out the conformity assessment

tasks of a conformity assessment body shall not depend on the

number of assessments carried out or on the results of those

assessments.

MEASURING INSTRUMENTS

(9) Conformity assessment bodies shall take out liability

insurance unless liability is assumed by the Government of Malta

in accordance with national law, or the Authority itself is directly

responsible for the conformity assessment.

(10) The personnel of a conformity assessment body shall

observe professional secrecy with regard to all information

obtained in carrying out their tasks under Schedule II or any

provision of national law giving effect to it, except in relation to

the competent authorities of Malta in which its activities are carried

out. Proprietary rights shall be protected.

(11) Conformity assessment bodies shall participate in, or

ensure that their personnel responsible for carrying out the

c o n f o r m i t y a s s e s s m e n t t a s k s a re i n f o r m e d o f , t h e r el ev a n t

standardisation activities and the activities of the notified body

coordination group established under the relevant Union

harmonisation legislation and shall apply as general guidance the

administrative decisions and documents produced as a result of the

work of that group.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

Where a conformity assessment body demonstrates its

conformity with the criteria laid down in the relevant harmonised

standards or parts thereof, the references of which have been

published in the Official Journal of the European Union, it shall be

presumed to comply with the requirements set out in regulation 27

in so far as the applicable harmonised standards cover those

requirements.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Where a notified body subcontracts specific tasks

connected with conformity assessment or has recourse to a

subsidiary, it shall ensure that the subcontractor or the subsidiary

meets the requirements set out in regulation 27 and shall inform the

Authority accordingly.

(2) Notified bodies shall take full responsibility for the tasks

performed by subcontractors or subsidiaries wherever these are

established.

(3) Activities may be subcontracted or carried out by a

subsidiary only with the agreement of the client.

(4) Notified bodies shall keep at the disposal of the Authority

the relevant documents concerning the assessment of the

qualifications of the subcontractor or the subsidiary and the work

carried out by them under Schedule II.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) An accredited in-house body may be used to carry out

conformity assessment activities for the undertaking of which it

forms a part for the purpose of implementing the procedures set out

in point 2 (Module A2) and point 5 (Module C2) of Schedule II.

That body shall constitute a separate and distinct part of the

undertaking and shall not participate in the design, production,

s u p p l y, i n s t a l l a t i o n , u s e o r m a i n t e n a n c e o f t h e m e a s u r i n g

instruments it assesses.

(2) An accredited in-house body shall meet the following

requirements

MEASURING INSTRUMENTS

(a) it shall be accredited in accordance with Regulation

(EC) No 765/2008;

(b) the body and its personnel shall be organisationally

identifiable and have reporting methods within the

undertaking of which they form a part which ensure

their impartiality and demonstrate it to the relevant

national accreditation body;

(c) neither the body nor its personnel shall be responsible

for the design, manufacture, supply, installation,

operation or maintenance of the measuring instruments

they assess nor shall they engage in any activity that

might conflict with their independence of judgment or

integrity in relation to their assessment activities;

(d) it shall supply its services exclusively to the

undertaking of which it forms a part.

(3) An accredited in-house body shall not be notified the

Authority, but information concerning its accreditation shall be

given by the undertaking of which it forms a part or by the National

Accreditation Board (Malta) to the Authority at the request of the

Authority.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) A conformity assessment body established in Malta

shall submit an application for notification under these regulations

to the Authority.

(2) The application for notification shall be accompanied by a

description of the conformity assessment activities, the conformity

assessment module or modules and the measuring instrument or

measuring instruments for which that body claims to be competent,

as well as by an accreditation certificate, where one exists, issued

by National Accreditation Board (Malta) attesting that the

conformity assessment body fulfils the requirements laid down in

regulation 27.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) The Authority may notify only conformity assessment

b od ie s w hi c h ha ve s a t is f i e d th e req u ir e m e nt s lai d do wn in

regulation 27.

(2) They shall notify the Commission and the other Member

States using the electronic notification tool developed and managed

by the Commission.

(3) The notification shall include information on the kind(s) of

measuring instrument(s) for which each body has been designated

and, where relevant, the instrument accuracy classes, the measuring

range, the measurement technology, and any other instrument

characteristic limiting the scope of the notification. The

notification shall include full details of the conformity assessment

activities, the conformity assessment module or modules and

measuring instrument or measuring instruments concerned and the

relevant attestation of competence.

(4) The body concerned may perform the activities of a notified

body only where no objections are raised by the Commission or the

other Member States within two weeks of a notification where an

MEASURING INSTRUMENTS

accreditation certificate is used.

(5) Only such a body shall be considered a notified body for the

purposes of these regulations.

(6) The Authority shall notify the Commission and the other

M e m b e r St a t e s o f a n y s u b s e q u e n t r e l e v a n t c h a n g e s t o t h e

notification.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Where the Authority has ascertained or has been

informed that a notified body no longer meets the requirements laid

down in regulation 27, or that it is failing to fulfil its obligations,

the Authority shall restrict, suspend or withdraw notification as

appropriate, depending on the seriousness of the failure to meet

those requirements or fulfil those obligations. It shall immediately

inform the Commission and the other Member States accordingly.

(2) In the event of restriction, suspension or withdrawal of

notification, or where the notified body has ceased its activity, the

Authority shall take appropriate steps to ensure that the files of that

body are either processed by another notified body or kept

available for the Authority.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

The Authority shall provide the Commission, on request,

with all information relating to the basis for the notification or the

maintenance of the competence of the notified body concerned.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Notified bodies shall carry out conformity assessments

in accordance with the conformity assessment procedures provided

for in Schedule II.

(2) Conformity assessments shall be carried out in a

proportionate manner, avoiding unnecessary burdens for economic

operators. Conformity assessment bodies shall perform their

activities taking due account of the size of an undertaking, the

sector in which it operates, its structure, the degree of complexity

of the measuring instrument technology in question and the mass or

serial nature of the production process.

In so doing they shall nevertheless respect the degree of

rigour and the level of protection required for the compliance of the

measuring instrument with these regulations.

(3) Where a notified body finds that the essential requirements

set out in Schedule I and in the relevant instrument-specific

Schedule or corresponding harmonised standards, normative

documents or other technical specifications have not been met by a

manufacturer, it shall require that manufacturer to take appropriate

corrective measures and shall not issue a certificate of conformity.

(4) Where, in the course of the monitoring of conformity

following the issue of a certificate, a notified body finds that a

measuring instrument no longer complies, it shall require the

manufacturer to take appropriate corrective measures and shall

suspend or withdraw the certificate if necessary.

(5) Where corrective measures are not taken or do not have the

required effect, the notified body shall restrict, suspend or

withdraw any certificates, as appropriate.

MEASURING INSTRUMENTS

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

The Authority shall ensure that an appeal procedure against

decisions of the notified bodies is available.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Notified bodies shall inform the Authority of the

following

(a) any refusal, restriction, suspension or withdrawal of a

certificate;

(b) any circumstances affecting the scope of or conditions

for notification;

(c) any request for information which they have received

from market surveillance authorities regarding

conformity assessment activities;

(d) on request, conformity assessment activities

performed within the scope of their notification and

any other activity performed, including cross-border

activities and subcontracting.

(2) Notified bodies shall provide the other bodies notified

under these regulations carrying out similar conformity assessment

activities covering the same measuring instruments with relevant

information on issues relating to negative and, on request, positive

conformity assessment results.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

The Authority shall ensure that the bodies notified by them

participate in the work of that group or those groups, directly or by

means of designated representatives.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

Regulation 16(3) and Regulation 16 to 29 of Regulation

(EC) No 765/2008 shall apply to measuring instruments.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Where the Authority have sufficient reason to believe

that a measuring instrument covered by these regulations presents a

risk to aspects of public interest protection covered by these

regulations, they shall carry out an evaluation in relation to the

measuring instrument concerned covering all relevant requirements

laid down in these regulations. The relevant economic operators

shall cooperate as necessary with the Authority for that purpose.

Where, in the course of the evaluation referred to in the

first paragraph, the Authority finds that the measuring instrument

d o e s n o t c o m p l y w i t h t h e r e q u i r e m en t s l a i d d o w n i n t h e s e

regulations, it shall without delay require the relevant economic

operator to take all appropriate corrective actions to bring the

measuring instrument into compliance with those requirements, to

withdraw the measuring instrument from the market, or to recall it

within a reasonable period, commensurate with the nature of the

risk, as they may prescribe.

The Authority shall inform the relevant notified body

accordingly.

Regulation 21 of Regulation (EC) No 765/2008 shall apply

MEASURING INSTRUMENTS

to the measures referred to in the second paragraph.

(2) Where the Authority considers that non-compliance is not

restricted to Malta, it shall inform the Commission and the other

Member States of the results of the evaluation and of the actions

which they have required the economic operator to take.

(3) The economic operator shall ensure that all appropriate

corrective action is taken in respect of all the measuring

instruments concerned that it has made available on the market

throughout the Union.

(4) Where the relevant economic operator does not take

adequate corrective action within the period referred to in the

second paragraph of sub-regulation (1), the Authority shall take all

appropriate provisional measures to prohibit or restrict the

measuring instrument being made available on their national

market, to withdraw the measuring instrument from that market or

to recall it.

The Authority shall inform the Commission and the other

Member States, without delay, of those measures.

(5) The information referred to in the second paragraph of sub-

regulation (4) shall include all available details, in particular the

d a t a n e c e s sa r y f o r t h e i d e n t i f i ca t i o n o f t h e n o n - c o m p l i a n t

measuring instrument, the origin of the measuring instrument, the

nature of the non-compliance alleged and the risk involved, the

nature and duration of the national measures taken and the

arguments put forward by the relevant economic operator. In

particular, the Authority shall indicate whether the non-compliance

is due to either of the following

(a) failure of the measuring instrument to meet

requirements relating to aspects of public interest

protection laid down in these regulations; or

(b) shortcomings in the harmonised standards or

normative documents referred to in regulation 18

conferring a presumption of conformity.

(6) Member States other than Malta initiating the procedure

under this regulation shall without delay inform the Commission

and the other Member States of any measures adopted and of any

additional information at their disposal relating to the non-

compliance of the measuring instrument concerned, and, in the

event of disagreement with the adopted national measure, of their

objections.

(7) Where, within three months of receipt of the information

referred to in the second paragraph of sub-regulation (4), no

o b j e c t i o n h a s b e e n r a i s e d b y e i t h e r a M e m b e r St a t e o r t h e

Commission in respect of a provisional measure taken the

Authority, that measure shall be deemed justified.

(8) The Authority shall ensure that appropriate restrictive

measures, such as withdrawal of the measuring instrument from the

market, are taken in respect of the measuring instrument concerned,

without delay.

MEASURING INSTRUMENTS

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

If the national measure is considered justified, the

Authority shall take the necessary measures to ensure that the non-

compliant measuring instrument is withdrawn from their market,

and shall inform the Commission accordingly. If the national

measure is considered unjustified, the Authority shall withdraw

that measure.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Where, having carried out an evaluation under

regulation 40(1), the Authority finds that although a measuring

instrument is in compliance with these regulations, it presents a

risk to aspects of public interest protection, it shall require the

relevant economic operator to take all appropriate measures to

ensure that the measuring instrument concerned, when placed on

the market, no longer presents that risk, to withdraw the measuring

instrument from the market or to recall it within a reasonable

period, commensurate with the nature of the risk, as it may

prescribe.

(2) The economic operator shall ensure that corrective action is

taken in respect of all the measuring instruments concerned that he

has made available on the market throughout the Union.

(3) The Authority shall immediately inform the Commission

and the other Member States that information shall include all

available details, in particular the data necessary for the

identification of the measuring instrument concerned, the origin

and the supply chain of the measuring instrument, the nature of the

risk involved and the nature and duration of the national measures

taken.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Without prejudice to regulation 40, where the

Authority makes one of the following findings, it shall require the

relevant economic operator to put an end to the non-compliance

concerned

(a) the CE marking or the supplementary metrology

marking has been affixed in violation of Regulation 30

of Regulation (EC) No 765/2008 or of regulation 22 of

these regulations;

(b) the CE marking or the supplementary metrology

marking has not been affixed;

(c) the identification number of the notified body, where

that body is involved in the production control phase,

has been affixed in violation of regulation 22 or has

not been affixed;

(d) the EU declaration of conformity does not accompany

the measuring instrument;

(e) the EU declaration of conformity has not been drawn

up correctly;

(f)

technical documentation is either not available or not

complete;

(g) the information referred to in regulation 10(6) or

regulation 12(3) is absent, false or incomplete;

(h) any other administrative requirement provided for in

MEASURING INSTRUMENTS

regulation 9 or regulation 12 is not fulfilled.

(2) Where the non-compliance referred to in sub-regulation (1)

persists the Authority shall take all appropriate measures to restrict

or prohibit the measuring instrument being made available on the

market or ensure that it is recalled or withdrawn from the market.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) the Authority shall not impede the making available on

the market and/or the putting into use of measuring instruments

covered by Regulation 2004/22/EC which are in conformity with

that Regulation and which were placed on the market before 20

April 2016.

Certificates issued under Regulation 2004/22/EC shall be

valid under these regulations.

(2) The effects of regulation 23 of Regulation 2004/22/EC shall

continue until 30th October, 2016.

Laws of Malta — Subsidiary Legislation 454.15 — MEASURING INSTRUMENTS REGULATIONS

(1) Where reference is made to this sub-regulation,

Regulation 4 of Regulation (EU) No. 182/2011 shall apply.

(2) Where reference is made to this sub-regulation, Regulation

5 of Regulation (EU) No. 182/2011 shall apply.