IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

25 Articles / Provisions · Malta
Open official source Machine-readable Markdown (.md)

Table of Provisions (25)

Statutory Text (25 Articles)

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

SUBSIDIARY LEGISLATION 427.16

IN VITRO DIAGNOSTIC MEDICAL DEVICES

REGULATIONS

1st January, 2003

LEGAL NOTICE 61 of 2002, as amended by Legal Notices 30 of 2004,

21 of 2010, 371 of 2012 and 318 of 2020.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. The title of these regulations is the In Vitro Diagnostic

Medical Devices Regulations.

1.2. These regulations shall become operative on the 1st

January, 2003.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

Based on Annex I of the Directive

ESSENTIAL REQUIREMENTS

A.

GENERAL REQUIREMENTS

1. The devices must be designed and manufactured in such a way that, when

used under the conditions and for the purposes intended, they will not compromise,

direcdy or indirectly, the clinical condition or the safety of the patients, the safety or

health of users or, where applicable, other persons, or the safety of property. Any

risks, which may be associated with their use, must be acceptable when weighed

against the benefits to the patient and be compatible with a high level of protection

of health and safety.

2. The solutions adopted by the manufacturer for the design and construction

of the devices must conform to safety principles, taking account of the generally

acknowledged state of the art. In selecting the most appropriate solutions, the

manufacturer must apply the following principles in the following order

-

eliminate or reduce risks as far as possible (inherently safe design and

construction),

-

where appropriate take adequate protection measures in relation to risks

that cannot be eliminated,

-

inform users of the residual risks due to any shortcomings of the

protection measures adopted.

3. The devices must be designed and manufactured in such a way that they are

suitable for the purposes referred to in regulation 3.1.3, as specified by the

manufacturer, taking account of the generally acknowledged state of the art. They

must achieve the performances, in particular, where appropriate, in terms of

analytical sensitivity, diagnostic sensitivity, analytical specificity, diagnostic

specificity, accuracy, repeatability, reproducibility, including control of known

relevant interference, and limits of detection, stated by the manufacturer. The

traceability of values assigned to calibrators and, or control materials must be

assured through available reference measurement procedures and, or available

reference materials of a higher order

4. The characteristics and performances referred to in points 1 and 3 of this

section must not be adversely affected to such a degree that the health or the safety

of the patient or the user and, where applicable, of other persons, are compromised

during the lifetime of the device as indicated by the manufacturer, when the device is

subjected to the stresses which can occur during normal conditions of use. When no

lifetime is stated, the same applies for the lifetime reasonably to be expected of a

device of that kind, having regard to the intended purpose and the anticipated use of

the device.

5. The devices must be designed, manufactured and packed in such a way that

their characteristics and performances during their intended use will not be adversely

affected under storage and transport conditions (temperature, humidity, etc.) taking

account of the instructions and information provided by the manufacturer.

B.

DESIGN AND MANUFACTURING REQUIREMENTS

1. Chemical and physical properties

1.1. The devices must be designed and manufactured in such a way as to achieve

the characteristics and performances referred to in section A on the "General

requirements". Particular attention must be paid to the possibility of impairment of

IN VITRO DIAGNOSTIC MEDICAL DEVICES

analytical performance due to incompatibility between the materials used and the

specimens (such as biological tissues, cells, body fluids and micro-organisms)

intended to be used with the device, taking account of its intended purpose.

1.2. The devices must be designed, manufactured and packed in such a way as to

reduce as far as possible the risk posed by product leakage, contaminants and

residues to the persons involved in the transport, storage and use of the medical

devices, taking account of the intended purpose of the products.

2. Infection and microbial contamination

2.1. The devices and their manufacturing processes must be designed in such a

way as to eliminate or reduce as far as possible the risk of infection to the user or

other persons. The design must allow easy handling and, where necessary, reduce as

far as possible contamination of, and leakage from, the device during use and, in the

case of specimen receptacles, the risk of contamination of the specimen. The

manufacturing processes must be appropriate for these purposes.

2.2. Where a device incorporates biological substances, the risks of infection

must be reduced as far as possible by selecting appropriate donors and appropriate

substances and by using appropriate, validated inactivation, conservation, test and

control procedures.

2.3. Devices labelled either as "STERILE" or as having a special microbiological

state must be designed, manufactured and packed in an appropriate pack, according

to procedures suitable for ensuring that they remain in the appropriate

microbiological state indicated on the label when placed on the market, under the

storage and transport conditions specified by the manufacturer, until the protective

packaging is damaged or opened.

2.4. Devices labelled either as "STERILE" or as having a special microbiological

state must have been processed by an appropriate validated method.

2.5. Packaging systems for devices other than those referred to in point 2.3 of

this section must keep the product without deterioration at the level of cleanliness

indicated by the manufacturer and, if the devices are to be sterilised prior to use,

reduce as far as possible the risk of microbial contamination.

Steps must be taken to reduce as far as possible microbial contamination

du rin g sel ectio n an d h andli ng of raw m aterials, m anufactu re, storage and

distribution where the performance of the device can be adversely affected by such

contamination.

2.6. Devices intended to be sterilised must be manufactured in appropriately

controlled (e.g. environmental) conditions.

2.7. Packaging systems for non-sterile devices must keep the product without

deterioration at the level of cleanliness stipulated and, if the devices are to be

sterilised prior to use, minimise the risk of microbial contamination; the packaging

system must be suitable taking account of the method of sterilisation indicated by the

manufacturer.

3. Manufacturing and environmental properties

3.1. If the device is intended for use in combination with other devices or

equipment, the whole combination, including the connection system, must be safe

and must not impair the specified performances of the devices. Any restrictions on

use must be indicated on the label and, or in the instructions for use.

3.2. Devices must be designed and manufactured in such a way as to reduce as

far as possible the risks linked to their use in conjunction with materials, substances

and gases with which they may come into contact during normal conditions of use.

3.3. Devices must be designed and manufactured in such a way as to remove or

reduce as far as possible

-

the risk of injury linked to their physical features (in particular aspects

of volume x pressure, dimension and, where appropriate, ergonomic

features),

-

risks linked to reasonably foreseeable external influences, such as

magnetic fields, external electrical effects, electrostatic discharge,

pressure, humidity, temperature or variations in pressure or acceleration

or accidental penetration of substances into the medical device.

Devices must be designed and manufactured in such a way as to provide an

adequate level of intrinsic immunity of electromagnetic disturbance to enable them

to operate as intended.

3.4. Devices must be designed and manufactured in such a way as to reduce as

far as possible the risks of fire or explosion during normal use and in single fault

condition. Particular attention must be paid to devices whose intended use includes

exposure to or use in association with flammable substances or substances which

could cause combustion.

3.5. Devices must be designed and manufactured in such a way as to facilitate

the management of safe waste disposal.

3.6. The measuring, monitoring or display scale (including colour change and

other visual indicators) must be designed and manufactured in line with ergonomic

principles, taking account of the intended purpose of the device.

4. Devices which are instruments or apparatus with a measuring function

4.1. Devices which are instruments or apparatus having a primary analytical

measuring function must be designed and manufactured in such a way as to provide

adequate stability and accuracy of measurement within appropriate accuracy limits,

taking into account the intended purpose of the medical device and of available and

appropriate reference measurement procedures and materials. The accuracy limits

have to be specified by the manufacturer.

4.2. When values are expressed numerically, they must be given in legal units

conforming to the provisions of Council Directive 80/181/EEC of 20 December 1979

on the approximation of the laws of the Member States relating to units of

measurement.

5. Protection against radiation

5.1. Devices shall be designed, manufactured and packaged in such a way that

exposure of users and other persons to the emitted radiation is minimised.

5.2. When devices are intended to emit potentially hazardous, visible and, or

invisible radiation, they must as far as possible be

-

designed and manufactured in such a way as to ensure that the

characteristics and the quantity of radiation emitted can be controlled

and, or adjusted,

-

fitted with visual displays and, or audible warnings of such emissions.

5.3. The operating instructions for medical devices emitting radiation must give

detailed information as to the nature of the emitted radiation, means of protecting the

user, and on ways of avoiding misuse and of eliminating the risks inherent in

installation.

6. Requirements for devices connected to or equipped with an energy source

6.1. Devices incorporating electronic programmable systems, including

software, must be designed to ensure the repeatability, reliability and performance of

these systems according to the intended use.

6.2. Devices must be designed and manufactured in such a way as to minimise

the risks of creating electromagnetic perturbation which could impair the operation

of other devices or equipment in the usual environment.

6.3. Devices must be designed and manufactured in such a way as to avoid, as far

as possible, the risk of accidental electric shocks during normal use and in single

fault condition, provided the devices are installed and maintained correctly.

6.4. Protection against mechanical and thermal risks

Devices must be designed and manufactured in such a way as to

protect the user against mechanical risks. Devices must be

sufficiently stable under the foreseen operating conditions. They

must be suitable to withstand stresses inherent in the foreseen

working environment, and to retain this resistance during the

expected life of the devices, subject to any inspection and

maintenance requirements as indicated by the manufacturer.

Where there are risks due to the presence of moving parts, risks due

to break-up or detachment, or leakage of substances, then

appropriate protection means must be incorporated.

Any guards or other means included with the device to provide

protection, in particular against moving parts, must be secure and

must not interfere with access for the normal operation of the

medical device, or restrict routine maintenance of the device as

intended by the manufacturer.

Devices must be designed and manufactured in such a way as to

reduce to the lowest possible level the risks arising from vibration

generated by the devices, taking account of technical progress and of

the means available for limiting vibrations, particularly at source,

unless the vibrations are part of the specified performance.

Devices must be designed and manufactured in such a way as to

reduce as far as possible the risks arising from the noise emitted,

taking account of technical progress and of the means available to

reduce noise, particularly at source, unless the noise emitted is part

of the specified performance.

Terminals and connectors to electricity, gas or hydraulic and

pneumatic energy supplies which the user has to handle must be

designed and manufactured in such a way as to minimise all possible

risks.

Accessible parts of the devices (excluding the parts of areas

intended to supply heat or reach given temperatures) and their

surroundings must not attain potentially dangerous temperatures

under normal use.

7. Requirements for devices for self-testing

Devices for self-testing must be designed and manufactured in such a way

that they perform appropriately for their intended purpose taking into account the

skills and the means available to users and the influence resulting from variation that

can reasonably be anticipated in users’ technique and environment. The information

and instructions provided by the manufacturer should be easily understood and

applied by the user.

7.1. Devices for self-testing must be designed and manufactured in such a way as

to

-

ensure that the device is easy to use by the intended lay user at all stages

of the procedure, and

-

reduce as far as practicable the risk of user error in the handling of the

device and in the interpretation of the results.

7.2. Devices for self-testing must, where reasonably possible, include user

control, i.e. a procedure by which the user can verify that, at the time of use, the

product will perform as intended.

8. Information supplied by a manufacturer

8.1. Each device must be accompanied by the information needed to use it safely

and properly, taking account of the training and knowledge of the potential users,

and to identify the manufacturer. This information comprises the data on the label

and in the instructions for use.

As far as practicable and appropriate, the information needed to use the

device safely and properly must be set out on the device itself and, or, where

appropriate, on the sales packaging. If individual full labelling of each unit is not

practicable, the information must be set out on the packaging and, or in the

instructions for use supplied with one or more devices.

Instructions for use must accompany or be included in the packaging of one

or more devices.

In duly justified and exceptional cases no such instructions for use are

needed for a device if it can be used properly and safely without them.

The decision whether to translate the instructions for use and the label into

one or more languages of the European Union shall be left to the Member States,

except that, for devices for self-testing, the instructions for use and the label must

include a translation into the official language(s) of the Member State in which the

device for self testing reaches its final user.

8.2. Where appropriate, the information to be supplied should take the form of

symbols. Any symbol and identification colour used must conform to the harmonised

standards. In areas for which no standards exist, the symbols and colour used must

be described in the documentation supplied with the device.

8.3. In the case of devices containing a preparation which may be considered as

being dangerous, taking account of the nature and quantity of its constituents and the

form under which they are present, relevant danger symbols and labelling

requirements of Directive 671548/EEC * and Directive 88/379/EEC † shall apply.

Where there is insufficient space to put all the information on the device itself or on

its label, the relevant danger symbols shall be put on the label and the other

information required by those Directives shall be given in the instructions for use.

*Council Directive 67/548/EEC of 27 June 1967 on the approximation of laws, regulations and

administrative provisions relating to the classification, packaging and labeling of dangerous substances (OJ

No. L 196, 16.08.67, P. 1 ). Directive as last amended by Commission Directive 97/69/EC (OJ No. L 343,

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. These regulations shall apply to in vitro diagnostic medical

devices and their accessories.

2.1.1.

For the purposes of these regulations, accessories

shall be treated as medical devices in their own

right.

2.1.2.

Both in vitro diagnostic medical devices and

accessories shall hereinafter be termed "devices".

2.2. These regulations shall not apply to devices manufactured

and used only within the same health institution and on the

premises of their manufacture or used on premises in the immediate

vicinity without having been transferred to another legal entity.

2.3. These regulations shall not affect national laws which

provide for the supply of medical devices by a medical

prescription.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

L.N. 371 of 2012.

Based on Annex II of the Directive

LIST OF DEVICES REFERRED TO IN REGULATIONS 7.2 AND 7.3

1. List A

Reagents and reagent products, including related calibrators and control

materials, for determining the following blood groups: ABO system,

rhesus (C, c, D, E, e) anti-Kell,

reagents and reagent products, including related calibrators and control

materials, for the detection, confirmation and quantification in human

specimens of markers of HIV infection (HIV 1 and 2), HTLV I and II,

and hepatitis B, C and D,

Variant Creutzfeldt-Jakob disease (vCJD) assays for blood screening,

diagnosis and confirmation.

2. List B

Reagents and reagent products, including related calibrators and control

materials, for determining the following blood groups: anti-Duffy and

anti-Kidd,

reagents and reagent products, including related calibrators and control

materials, for determining irregular anti-erythrocytic antibodies,

reagents and reagent products, including related calibrators and control

materials, for the detection and quantification in human samples of the

following congenital infections: rubella, toxoplasmosis,

reagents and reagent products, including related calibrators and control

materials, for diagnosing the following hereditary disease

phenylketonuria,

reagents and reagent products, including related calibrators and control

materials, for determining the following human infections

cytomegalovirus, chlamydia,

reagents and reagent products, including related calibrators and control

materials, for determining the following HLA tissue groups: DR, A, B,

reagents and reagent products, including related calibrators and control

materials, for determining the following tumoral marker: PSA,

reagents and reagent products, including related calibrators, control

materials and software, designed specifically for evaluating the risk of

trisomy 21,

the following device for self-diagnosis, including its related calibrators

and control materials: device for the measurement of blood sugar.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. For the purposes of these regulations, the following

definitions shall apply

3.1.1.

"the Directive" means Council Directive 98/79/EC

of the European Parliament and of the Council on

in vitro diagnostic medical devices.

3.1.1.1.

This Directive is a specific directive

within the meaning of Article 2(2) of

Directive 89/336/EEC, which shall cease

to apply to medical devices which have

complied with this Directive.

3.1.2.

"medical device" means any instrument, apparatus,

appliance, material or other article, whether used

alone or in combination, including the software

necessary for its proper application intended by the

manufacturer to be used for human beings for the

purpose of

- diagnosis, prevention, monitoring, treatment

or alleviation of disease,

- diagnosis, monitoring, treatment, alleviation

of or compensation for an injury or handicap,

IN VITRO DIAGNOSTIC MEDICAL DEVICES

- investigation, replacement or modification of

the anatomy or of a physiological process,

- control of conception,

and which does not achieve its principal intended

action in or on the human body by

pharmacological, immunological or metabolic

means, but which may be assisted in its function by

such means;

3.1.3.

"in vitro diagnostic medical device" means a

medical device which -

3.1.3.1.

is a reagent, reagent product, calibrator,

control

material,

kit,

instrument,

apparatus, equipment, or system,

whether used alone or in combination;

and

3.1.3.2.

is intended by the manufacturer to be

used in vitro for the examination of

specimens, including blood and tissue

donations, derived from the human body,

solely or principally for the purpose of

providing information

- concerning a physiological or

pathological state,

- concerning

a

congenital

abnormality,

- to deternune the safety and

compatibility

with

potential

recipients, or

- to monitor therapeutic measures,

and includes a specimen receptacle but

not a product for general laboratory use,

unless that product, in view of its

characteristics, is specifically intended

by its manufacturer to de used for in

vitro diagnostic exaxnination;

3.1.3.3.

Products for general laboratory use are

not in vitro diagnostic medical devices

unless such products, in view of their

characteristics, are specifically intended

by their manufacturer to be used for in

vitro diagnostic examination.

3.1.4.

"accessory" means an article intended specifically

by its manufacturer to be used together with an in

vitro diagnostic medical device to enable that

device to be used in accordance with its intended

purpose, which is not -

3.1.4.1.

itself an in vitro diagnostic medical

device;

3.1.4.2.

an invasive sampling medical device; or

3.1.4.3.

a medical device which is directly

applied to the human body for the

purpose of obtaining a specimen within

the meaning of Directive 93142/EEC;

3.1.5.

"calibration and control material" means any

substance, material or article intended by its

manufacturer either to establish measurement

relationships or to verify the performance

characteristics of a relevant device in conjunction

with the intended use of that device;

3.1.6.

"device for self-testing" means any device intended

by the manufacturer to be able to be used by lay

persons in a home environment;

3.1.7.

"device for performance evaluation" means any

device intended by the manufacturer to be subject

to one or more performance evaluation studies in

laboratories for medical analyses or in other

appropriate environments outside his own

premises;

3.1.8.

"manufacturer" means the natural or legal person

with responsibility for the design, manufacture,

packaging and labelling of a device before it is

placed on the market under his own name,

regardless of whether these operations are carried

out by that person himself or on his behalf by a

third party;

3.1.8.1.

The obligations of the regulations to be

met by manufacturers also apply to the

natural or legal person who assembles,

packages, processes, fully refurbishes

and, or labels one or more ready-made

products and, or assigns to them their

intended purpose as devices with a view

to their being placed on the market under

his own name.

3.1.8.2.

Regulation 3.1.8.1 does not apply to the

person who, while not a manufacturer

within the meaning of regulation 3.1.7,

assembles or adapts devices already on

the market to their intended purpose for

an individual patient;

3.1.9. "Medicines Authority" means the Authority as

established under article 4 of the Medicines Act;

3.1.10. "intended purpose" means the use for which the

device is intended according to the data supplied

by the manufacturer on the labelling, in the

instructions for use and, or in promotional

materials;

3.1.11. "placing on the market" means the first making

available in return for payment or free of charge of

a device other than a device intended for

performance evaluation with a view to distribution

IN VITRO DIAGNOSTIC MEDICAL DEVICES

and, or use on the market, regardless of whether it

is new or fully refurbished;

3.1.12. "putting into service" means the stage at which a

device has been made available to the final user as

being ready for use on the market for the first time

for its intended purpose;

3.1.13. "specimen receptacles" are those devices, whether

vacuum-type or not, specifically intended by their

manufacturers for the primary containment and

preservation of specimens derived from the human

body for the purpose of in vitro diagnostic

examination;

3.1.14. for the purposes of these regulations, the removal,

collection and use of tissues, cells and substances

of human origin shall be governed, in relation to

ethics, by the principles laid down in the

Convention of the Council of Europe for the

protection of human rights and dignity of the

human being with regard to the application of

biology and medicine and by any Member States

regulations for that matter;

3.1.15. in these regulations, unless the context otherwise

requires, a reference -

3.1.15.1. to a numbered regulation is to the

regulation of these regulations bearing

that number;

3.1.15.2. in a regulation to a numbered or lettered

paragraph is to the paragraph of that

regulation bearing that number or letter;

3.1.15.3. in a paragraph to a numbered or lettered

sub-paragraph is to the sub-paragraph in

that paragraph bearing that number or

letter; and

3.1.15.4. to a numbered article or Annex is to the

article or Annex of the Directive bearing

that number.

3.1.16. "harmonised standard" means a technical

specification adopted by the European Committee

for Standardisation (CEN) or the European

Committee for Electrotechnical Standardisation

(CENELEC) or the European Telecommunications

Standards Institute (ETSI), upon a remit from the

European Commission in accordance with

Directive 98/34/EC of the European Parliament and

of the Council laying down a procedure for the

provision of information in the field of technical

standards and regulations as amended by Directive

98/48/EC of the European Parliament and of the

Council;

3.1.16.1. Where, in these regulations, reference is

made to harmonised standards, this is

also meant to refer to the common

technical specifications;

3.1.17. "national standard transposing the harmonised

standard" means a standard issued by one of the

members of CEN/CENELEC/ETSI which is

declared and noted to be technically equivalent to

the harmonised standard originally issued by CEN/

CENELEC/ETSI;

3.1.18. "OJ" means the Official Joumal of the European

Communities;

3.1.19. "Community" means the European Community;

3.1.20. "essential requirements" means the requirements in

Annex I of the Directive which is set out in

Schedule I;

3.1.21. "notified body" means a body responsible for

carrying out functions relating to the conformity

assessment procedures set out in Article 9 of the

Directive and described in regulation 7 of these

regulations which has been notified by the Member

State concerned to the Commission and the other

Member States pursuant to Article 15(1) of the

Directive;

3.1.22. "CE marking" means marking as set out in Annex

10 of the Directive, which is set out in Schedule X;

3.1.23. "local autliorised representative" means any person

(or organisation) established in Malta who has been

appointed by the manufacturer to act on his behalf

in carrying out certain tasks required by these

regulations;

3.1.24. "importer" means any person who places medical

devices from a foreign country on to the Maltese

market so that he supplies such devices on the first

occasion that it is supplied in Malta;

3.1.25. "person who supplies medical devices" means

wholesalers, distributors, retailers, etc.;

3.1.26. "responsible person" in relation to medical devices

means

3.1.26.1. the manufacturer thereof;

3.1.26.2. the manufacturer’s local authorised

representative; or

3.1.26.3. where

the

manufacturer

is

not

established in Malta and he has not

appointed

a

local

authorised

representative, the person who places the

devices on the local market;

3.1.27. "common technical specification" means a

technical specification for a relevant device

referred to in a list in Annex II of the Directive, set

out in Schedule II, which has been adopted in

accordance with the procedure set out in Article

7(2) of the Directive and published in the OJ, as set

out in Schedule XI.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

Based on Annex III of the Directive

EC DECLARATION OF CONFORMITY

1. The EC declaration of conformity is the procedure whereby the

manufacturer or his authorised representative who fulfils the obligations imposed by

sections 2 to 5 of this Schedule and additionally, in the case of devices for self-

testing, the obligations imposed by section 6 of this Schedule, ensures and declares

that the products concerned meet the provisions of these regulations which apply to

them. The manufacturer must affix the CE marking in accordance with regulation 12

of these regulations.

2. The manufacturer must prepare the technical documentation described in

section 3 of this Schedule and ensure that the manufacturing process follows the

principles of quality assurance as set out in section 4 of this Schedule.

3. The technical documentation must allow assessment of the conformity of the

product with the requirements of these regulations. It must include in particular

a general description of the product, including any variants planned,

the documentation of the quality system,

design information, including the determination of the characteristics of

the basic materials, characteristics and limitation of the performance of

the devices, methods of manufacture and, in the case of instruments,

design drawings, diagrams of components, sub-assemblies, circuits,

etc.,

in the case of devices containing tissues of human origin or substances

derived from such tissue, information on the origin of such material and

on the conditions in which it was collected,

the descriptions and explanations necessary to understand the

abovementioned characteristics, drawings and diagrams and the

operation of the product,

the results of the risk analysis and, where appropriate, a list of the

standards referred to in Article 5 of the Directive, applied in full or in

part, and descriptions of the solutions adopted to meet the essential

requirements of the Directive if the standards referred to in Article 5 of

the Directive have not been applied in full,

in the case of sterile products or products with a special microbiological

state or state of cleanliness, a description of the procedures used,

the results of the design calculations and of the inspections carried out,

etc.,

if the device is to be combined with other medical device(s) in order to

operate as intended, proof must be provided that it conforms to the

essential requirements when combined with any such device(s) having

the characteristics specified by the manufacturer,

the test reports,

adequate performance evaluation data showing the performances

claimed by the manufacturer and supported by a reference measurement

system (when available), with information on the reference methods,

the reference materials, the known reference values, the accuracy and

measurement units used; such data should originate from studies in a

clinical or other appropriate environment or result from relevant

biographical references,

the labels and instructions for use,

the results of stability studies.

4. The manufacturer shall take necessary measures to ensure that the

manufacturing process follows the principles of quality assurance as appropriate for

the products manufactured.

The system shall address

the organisational structure and responsibilities,

the manufacturing processes and systematic quality control of

production,

the means to monitor the performance of the quality system.

5. The manufacturer shall institute and keep up to date a systematic procedure

to review experience gained from devices in the post-production phase and to

implement appropriate means to apply any necessary corrective actions, taking

account of the nature and risks in relation to the product. He shall notify the

competent authorities of the following incidents immediately on learning of them

(i)

any malfunction, failure or deterioration in the characteristics and, or

performance of a device, as well as any inadequacy in the labelling or

the instructions for use which, directly or indirectly, might lead to, or

might have led to, the death of a patient or user or other persons or to a

serious deterioration in his or their state of health;

(ii)

any technical or medical reason connected with the characteristics or

the performance of a device for the reasons referred to in paragraph (i)

leading to systematic recall of devices of the same type by the

manufacturer.

6. For devices for self-testing the manufacturer shall lodge an application or

examination of the medical design with a notified body.

6.1. The application shall enable the design of the device to be understood and

shall enable conformity with the design-related requirements of the regulations to be

assessed.

It shall include

-

test reports including, where appropriate, results of studies carried out

with lay persons,

-

data showing the handling suitability of the device in view of its

intended purpose for self-testing,

-

the information to be provided with the device on its label and its

instructions for use.

6.2. The notified body shall examine the application and, if the design conforms

to the relevant provisions of these regulations, shall issue the applicant with an EC

design-examination certificate. The notified body may require the application to be

completed by further tests or proof to allow assessment of conformity with the

design-related requirements of the regulations. The certificate shall contain the

conclusions of the examination, the conditions of validity, the data needed for

identification of the approved design and, where appropriate, a description of the

intended purpose of the product.

6.3. The applicant shall inform the notified body which issued the EC design-

examination certificate of any significant change made to the approved design.

Changes to the approved design must receive further approval from the notified body

which issued the EC design-examination certificate wherever the changes could

affect conformity with the essential requirements of the regulations or with the

conditions prescribed for use of the product. This additional approval shall take the

form of a supplement to the EC design-examination certificate.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. The provisions of these regulations are applicable to all

imported in vitro diagnostic medical devices as well as those

manufactured for the local market.

4.1.1.

No person may manufacture, import, place on the

market or put into service any in vitro diagnostic

medical devices which do not comply with the

provisions of these regulations.

4.2. Subject to regulation 4.3, the provisions of these

regulations shall not apply to in vitro diagnostic medical devices

intended exclusively for export.

4.3. Subject to regulation 11, regulation 4.2 shall not apply if

the CE marking or any inscription liable to be confused therewith is

affixed to the in vitro diagnostic medical devices.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

Based on Annex IV of the Directive

EC DECLARATION OF CONFORMITY

(FULL QUALITY ASSURANCE SYSTEM)

1. The manufacturer must ensure application of the quality system approved

for the design, manufacture and final inspection of the devices concerned, as

specified in section 3 of this Schedule, and is subject to audit as laid down in point

3.3 and to the surveillance as specified in section 5 of this Schedule. In addition, the

manufacturer must follow, for devices covered by Annex II, List A, of the Directive,

which is set out in Schedule II, the procedures laid down in sections 4 and 6 of this

Schedule.

2. The declaration of conformity is the procedure whereby the manufacturer

who fulfils the obligations imposed by section 1 ensures and declares that the

devices concerned meet the provisions of these regulations which apply to them.

The manufacturer shall affix the CE marking in accordance with regulation

12 of these regulations, and shall draw up a declaration of conformity covering the

devices concerned.

3. Quality system

3.1. The manufacturer must lodge an application for assessment of his quality

system with a notified body.

The application must include

the name and address of the manufacturer and any additional

manufacturing site covered by the quality system,

adequate information on the device or device category covered by the

procedure,

a written declaration that no such application has been lodged with any

other notified body for the same device-related quality system,

the documentation on the quality system,

an undertaking by the manufacturer to fulfil the obligations imposed by

the quality system approved,

an undertaking by the manufacturer to keep the approved quality system

adequate and efficacious,

an undertaking by the manufacturer to institute and keep up to date a

systematic procedure to review experience gained from devices in the

post-production phase and to implement appropriate means to apply any

necessary corrective action and notification as referred to in Annex III,

section 5 of the Directive, which is set out in Schedule III.

3.2. Application of the quality system must ensure that the devices conform to

the provisions of these regulations which apply to them at every stage, from design

to final inspection. All the elements, requirements and provisions adopted by the

manufacturer for his quality system must be documented in a systematic and orderly

manner in the form of written policies and procedures, such as quality programmes,

quality plans, quality manuals and quality records.

It shall include in particular an adequate description of

(a) the manufacturer’s quality objectives;

(b) the organisation of the business and in particular

-

the organisational structures, the responsibilities of the

managerial staff and their organisational authority where quality

of design and manufacture of the devices is concerned,

-

the methods of monitoring the efficient operation of the quality

system and in particular its ability to achieve the desired quality

of design and of product, including control of devices which fail

to conform;

(c) the procedures for monitoring and verifying the design of the devices

and in particular

-

a general description of the device, including any variants

planned,

-

all documentation referred to in Annex III, section 3, indents 3 to

13 of the Directive, which is set out in Schedule III,

-

in the case of devices for self-testing, the information referred to

in Annex III, section 6.1 of theþirective, which is set out in

Schedule III,

-

the techniques used to control and verify the design and the

processes and systematic measures which will be used when the

devices are being designed;

(d) the inspection and quality assurance techniques at the manufacturing

stage and in particular

-

the processes and procedures which will be used, particularly as

regards sterilisation,

-

the procedures in relation to purchasing,

-

the product identification procedures drawn up and kept up to

date from drawings, specifications or other relevant documents at

every stage of manufacture;

(e) the appropriate tests and trials which will be carried out before, during

and after manufacture, the frequency with which they will take place,

and the test equipment used; it must be possible to trace back the

calibration.

The manufacturer shall carry out the required controls and tests according to

the latest state of the art. The controls and tests shall cover the manufacturing

process including the characterisation of the raw material and the individual devices

or each batch of devices manufactured.

In testing the devices covered by Annex II, List A of the Directive, which is

set out in Schedule II, the manufacturer shall take into account the most recent

available information, in particular as regards the biological complexity and

variability of the specimens to be tested with the in vitro device concerned.

3.3. The notified body must audit the quality system to determine whether it

meets the requirements referred to in point 3.2 of this Schedule. It must presume that

quality systems which implement the relevant harmonised standards conform to the

requirements.

The assessment team must have experience of assessments of the technology

concerned. The assessment procedure must include an inspection on the

manufacturer’s premises and, in duly substantiated cases, on the premises of the

manufacturer ’s suppliers and, or subcontractors to inspect the manufacturing

processes.

The decision shall be notified to the manufacturer. It must contain the

conclusions of the inspection and a reasoned assessment.

3.4. The manufacturer must inform the notified body which approved the quality

system of any plan for substantial changes to the quality system or the product-range

covered.

The notified body must assess the changes proposed and verify whether after

these changes the quality system still meets the requirements referred to in point 3.2

of this Schedule. It must notify the manufacturer of its decision. This decision must

contain the conclusions of the inspection and a reasoned assessment.

4. Examination of the design of the product

4.1. For devices covered by Annex II, List A of the Directive, which is set out in

Schedule II, in addition to the obligations imposed by section 3 of this Schedule, the

manufacturer must lodge with the notified body an application for examination of

the design dossier relating to the device which he plans to manufacture and which

falls into the category referred to in point 3.1 of this Schedule.

4.2. The application must describe the design, manufacture and performances of

the device in question. It must include the documents needed to assess whether the

device conforms to the requirements of these regulations, as referred to in point

IN VITRO DIAGNOSTIC MEDICAL DEVICES

3.2(c) of this Schedule.

4.3. The notified body must examine the application and, if the device conforms

to the relevant provisions of the regulations, issue the application with an EC design-

examination certificate. The notified body may require the application to be

completed by further tests or proof to allow assessment of conformity with the

requirements of the regulations. The certificate must contain the conclusions of the

examination, the conditions of validity, the data needed for the identification of the

approved design and, where appropriate, a description of the intended purpose of the

device.

4.4. Changes to the approved design must receive further approval from the

notified body which issued the EC design-examination certificate wherever the

changes could affect conformity with the essential requirements of the regulations or

with the conditions prescribed for use of the device. The applicant shall inform the

notified body which issued the EC design-examination certificate of any such

changes made to the approved design. The additional approval must take the form of

a supplement to the EC design-examination certificate.

4.5. The manufacturer shall inform the notified body without delay if it has

obtained information about changes to the pathogen and markers of infections to be

tested, in particular as a consequence of biological complexity and variability. In this

connection, the manufacturer shall inform the notified body whether any such

change is likely to affect the performance of the in vitro diagnostic device

concerned.

5. Surveillance

5.1. The aim of surveillance is to ensure that the manufacturer duly fulfils the

obligations imposed by the approved quality system.

5.2. The manufacturer must authorise the notified body to carry out all the

necessary inspections and supply it with all relevant information, in particular

-

the documentation on the quality system,

-

the data stipulated in the part of the quality system relating to design,

such as the results of analyses, calculation, tests, etc.,

-

the data stipulated in the part of the quality system relating to

manufacture, such as inspection reports and test data, calibration data,

qualification reports of the personnel concerned, etc.

5.3. The notified body must periodically carry out appropriate inspections and

assessments to make sure that the manufacturer applies the approved quality system

and must supply the manufacturer with an assessment report.

5.4. In addition, the notified body may pay unannounced visits to the

manufacturer. At the time of such visits, the notified body may, where necessary,

carry out or ask for tests in order to check that the quality system is working

properly. It must provide the manufacturer with an inspection report and, if a test has

been carried out, with a test report.

6. Verification of manufactured products covered by Schedule II, List A

6.1. In the case of devices covered by Annex II, List A of the Directive, which is

set out in Schedule II, the manufacturer shall forward to the notified body without

delay after the conclusion of the controls and tests the relevant reports on the tests

carried out on the manufactured devices or each batch of devices.

Furthermore, the manufacturer shall make the samples of manufactured

devices or batches of devices available to the notified body in accordance with pre-

agreed conditions and modalities.

6.2. The manufacturer may place the devices on the market, unless the notified

body communicates to the manufacturer within the agreed time-frame, but not later

than thirty days after reception of the samples, any other decision, including in

particular any condition of validity of delivered certificates.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. All necessary steps shall be taken to ensure that devices are

placed on the market and, or put into service only if they comply

with the requirements laid down in these regulations when duly

supplied and properly installed, maintained and used in accordance

with their intended purpose.

5.1.1.

This regulation applies also to devices made

available for performance evaluation.

5.2. Devices must meet the essential requirements set out in

Annex I of the Directive, which is set out in Schedule I, which

apply to them, taking account of the intended purpose of the

devices concerned.

5.3. A device shall be treated as complying with an essential

requirement if as respects that requirement, it complies with the

relevant national standard, unless there are reasonable indications

that the device does not comply with that requirement.

5.4. Without prejudice to regulation 5.3, a device specified in

List A or B of Annex II of the Directive, set out in Schedule II,

shall be treated as complying with the relevant essential

requirements if the device is designed and manufactured in

conformity with the common technical specifications drawn up for

the device, unless there are reasonable indications that the device

does not comply with those requirements.

5.4.1.

A device to which regulation 5.4 relates shall be

treated as having been manufactured in accordance

with the common technical specifications drawn up

for the device if, for justifiable reasons, it is not so

manufactured but is manufactured in accordance

with technical specifications that are at least

equivalent

to

such

common

technical

specifications, unless there are reasonable

indications that the device does not comply with

those requirements.

5.4.1.1.

The technical specifications set out in

Schedule XI are adopted as common

technical specifications for devices in

list A of Schedule II.

5.5. No person shall place on the market any device to which

these regulations apply unless the requirements of regulation 5.2

have been complied with in relation to it.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

Based on Annex V of the Directive

EC TYPE-EXAMINATION

1. EC type-examination is the part of the procedure whereby a notified body

ascertains and certifies that a representative sample of the production envisaged

fulfils the relevant provisions of these regulations.

2. The application for EC type-examination shall be lodged by the

manufacturer or by his authorised representative with a notified body.

The application shall include

the name and the address of the manufacturer and the name and address

of the authorised representative if the application is lodged by the

representative,

the documentation described in section 3 of this Schedule needed to

assess the conformity of the representative sample of the production in

question, hereinafter referred to as the "type", with the requirements of

these regulations. The applicant shall make a "type" available to the

notified body. The notified body may request other samples as

necessary,

a written declaration that no application has been lodged with any other

notified body for the same type.

3. The documentation must allow an understanding of the design, the

manufacture and the performances of the device. The documentation shall contain

the following items in particular

a general description of the type, including any variants planned,

all documentation referred to in Annex III, section 3, indents 3 to 13 of

the Directive, which is set out in Schedule III,

in the case of devices for self-testing, the information referred to in

Annex III, section 6.1 of the Directive, which is set out it Schedule III.

4. The notified body shall

4.1. examine and assess the documentation and verify that the type has been

manufactured in conformity with that documentation; it shall also

record the items designed in conformity with the applicable provisions

of the standards referred to in Article 5 of the Directive, as well as the

items not designed on the basis of the relevant provisions of the

abovementioned standards;

4.2. perform or have performed appropriate examinations and the tests

necessary to verify whether the solutions adopted by the manufacturer

meet the essential requirements of these regulations if the standards

IN VITRO DIAGNOSTIC MEDICAL DEVICES

referred to in Article 5 of the Directive have not been applied; if the

device is to be combined with other device(s) in order to operate as

intended, proof must be provided that it conforms to the essential

requirements when combined with any such device(s) having the

characteristics specified by the manufactucer;

4.3. carry out or ask for the appropriate examinations and the tests necessary

to verify whether, if the manufacturer has chosen to apply the relevant

standards, these have actually been applied;

4.4. agree with the applicant on the place where the necessary examinations

and tests will be carried out.

5. If the type conforms to the provisions of these regulations, the notified body

shall issue the applicant with an EC type-examination certificate. The certificate

shall contain the name and address of the manufacturer, the conclusions of the

examination, the conditions of validity and the data needed for identification of the

type approved. The relevant parts of the documentation shall be annexed to the

certificate and a copy shall be kept by the notified body.

6. The manufacturer shall inform the notified body without delay if it has

obtained information about changes to the pathogen and markers of infections to be

tested, in particular as a consequence of biological complexity and variability. In this

connection, the manufacturer shall inform the notified body whether any such

change is likely to affect the performance of the in vitro device concerned.

6.1. Changes to the approved device must receive further approval from the

notified body which issued the EC type-examination certificate wherever the

changes may affect conformity with the essential requirements of the regulations or

with the conditions prescribed for use of the device. The applicant shall inform the

notified body which issued the EC type-examination certificate of any such change

made to the approved device. This new approval shall take the form of a supplement

to the initial EC type-exanunation certificate.

7. Administrative provisions

7.1. Other notified bodies may obtain a copy of the EC type-examination

certificates and, or the supplements thereto. The annexes to the certificates must be

available to the other notified bodies on reasoned application, after the manufacturer

has been informed.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. The placing on the market or putting into service under the

conditions specified by the manufacturer of devices referred to in

regulation 3, which comply with these regulations and bear the CE

marking provided for in regulation 11, indicating that they have

undergone conformity assessment in accordance with regulation 7

shall not be prohibited, restricted or impeded.

6.2. No obstacle shall be created to devices intended for

performance evaluation being made available for that purpose to

the laboratories or other institutions listed in the statement referred

to in Annex VIII of the Directive, which is set out in Schedule VIII,

if they meet the conditions laid down in regulation 7.4 and Annex

VIII of the Directive, which is set out in Schedule VIII.

6.3. At trade fairs, exhibitions, demonstrations, scientific or

technical gatherings, etc., no obstacle shall be created to the

showing of devices which do not conform to these regulations,

provided that such devices are not used on specimens taken from

the participants and that a visible sign clearly indicates that such

devices cannot be marketed or put into service until they have been

made to comply.

6.4. The information to be supplied pursuant to Annex I, part B,

section 8 of the Directive, which is set out in Schedule I, is required

to be in either the Maltese or English language, when the device

reaches the final user.

6.4.1.

Provided that safe and correct use of the device is

ensured, the information referred to in regulation

6.4 may be authorised to be in one or more other

official Community language(s).

6.4.1.1.

In the application of this provision,

account shall be taken of the principle of

proportionality and, in particular

- whether the information can be

supplied by harmonised symbols or

recognised codes or other measures;

- the type of user anticipated for the

device.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

Based on Annex VI of the Directive

EC VERIFICATION

1. EC verification is the procedure whereby the manufacturer or his authorized

representative ensures and declares that the products which have been subject to the

procedure set out in section 4 of this Schedule conform to the type described in the

EC type-examination certificate and meet the requirements of these regulations

which apply to them.

2.1. The manufacturer must take all the measures necessary to ensure that the

manufacturing process produces products which conform to the type described in the

EC type-examination certificate and the requirements of the regulations which apply

to them. Before the start of manufacture, the manufacturer must prepare documents

defining the manufacturing process, in particular as regards sterilisation and the

suitability of starting materials, where necessary, and define the necessary testing

procedures according to the state of the art. All the routine, pre-established

provisions must be implemented to ensure homogeneous production and conformity

of the products with the type described in the EC type-examination certificate and

with the requirements of these regulations which apply to them.

2.2. To the extent that for certain aspects the final testing according to point 6.3

of this Schedule is not appropriate, adequate process testing, monitoring and control

methods shall be established by the manufacturer with the approval of the notified

body. The provisions of Annex IV, section 5 of the Directive, which is set out in

Schedule IV, shall apply accordingly in relation to the abovementioned approved

procedures.

3. The manufacturer must undertake to institute and keep up to date a

systematic procedure to review experience gained from devices in the post-

production phase and to implement appropriate means to apply any necessary

corrective and notification action as referred to in Annex III, section 5 of the

Directive, which is set out in Schedule III.

4. The notified body must carry out the appropriate examinations and tests

taking account of point 2.2 of this Schedule in order to verify the conformity of the

product with the requirements of the regulations either by examining and testing

every product as specified in section 5 of this Schedule or by examining and testing

products on a statistical basis as specified in section 6 of this Schedule, as the

manufacturer decides. When carrying out statistical verification according to section

6, the notified body has to decide when statistical procedures for lot-by-lot

inspection or isolated lot inspection have to be applied. Such decision must be taken

in consultation with the manufacturer.

In as far as the conduct of examinations and tests on a statistical basis is not

appropriate, examinations and tests may be carried out on a random basis provided

that such procedure in conjunction with the measures taken in accordance with point

2.2 of this Schedule ensures an equivalent level of conformity.

5. Verification by examination and testing of every product

5.1. Every product is examined individually and the appropriate tests defined in

the relevant standard(s) referred to in Article 5 of the Directive or equivalent tests

must be carried out in order to verify the conformity of the products with the EC type

described in the type-examination certificate and with the requirements of the

regulations which apply to them.

5.2. The notified body must affix, or have affixed, its identification number to

each approved product and must draw up a written certificate of conformity relating

to the tests carried out.

6. Statistical verification

6.1. The manufacturer must present the manufactured products in the form of

homogeneous batches.

6.2. One or more random samples, as necessary, are taken from each batch. The

products which make up the sample are examined and the appropriate tests defined

in the relevant standard(s) referred to in Article 5 of the Directive or equivalent tests

must be carried out to verify, where appropriate, the conformity of the products with

the type described in the EC type-examination certificate and with the requirements

of the regulations which apply to them in order to determine whether to accept or

reject the batch.

IN VITRO DIAGNOSTIC MEDICAL DEVICES

6.3. Statistical control of products will be based on attributes and, or variables,

entailing sampling schemes with operational characteristics which ensure a high

level of safety and performance according to the state of the art. The sampling

scheme will be established by the harmonised standards referred to in Artic1e 5 of

the Directive taking account of the specific nature of the product categories in

question.

6.4. If the batch is accepted, the notified body affixes, or has affixed its

identification number to each product and draws up a written certificate of

conformity relating to the tests carried out. All products in the batch may be put on

the market except any in the sample which failed to conform.

If the batch is rejected the competent notified body must take appropriate

measures to prevent the batch from being placed on the market. In the event of

frequent rejection of batches, the notified body may suspend the statistical

verification.

The manufacturer may, on the responsibility of the notified body, affix the

notified body’s identification number during the manufacturing process.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. For all devices other than those covered by Annex II of the

Directive, which i s s et out in S c hed u le I I , a nd d ev i ces f or

performance evaluation, the manufacturer shall, in order to affix

the CE marking, follow the procedure referred to in Annex III of

the Directive, which is set out in Schedule III, and draw up the EC

declaration of conformity required before placing the devices on

the market.

7.1.1.

For all devices for self-testing other than those

covered by Annex II of the Directive, which is set

out in Schedule II, and medical devices for

performance evaluation, the manufacturer shall,

prior to the drawing up of the aforementioned

declaration of conformity, fulfil the supplementary

requirements set out in Annex III, point 6 of the

Directive, which is set out in Schedule III. Instead

of applying this procedure, the manufacturer may

follow the procedure referred to in regulations 7.2

and 7.3.

7.2. For all devices referred to in List A in Annex II of the

Directive, which is set out in Schedule II, other than those intended

for performance evaluation, the manufacturer shall, in order to

affix the CE marking either

7.2.1.

follow the procedure relating to the EC declaration

of conformity set out in Annex IV (full quality

assurance) of the Directive, which is set out in

Schedule IV, or

7.2.2.

follow the procedure relating to EC type-

examination set out in Annex V of the Directive,

which is set out in Schedule V, coupled with the

procedure relating to the EC declaration of

conformity set out in Annex VII (production

quality assurance) of the Directive, which is set out

in Schedule VII.

7.3. For all devices referred to in List B in Annex II of the

Directive, which is set out in Schedule II, other than those intended

for performance evaluation, the manufacturer shall for the purposes

of affixing the CE marking, follow either

7.3.1.

the procedure relating to the EC declaration of

conformity set out in Annex IV (full quality

assurance) of the Directive, which is set out in

Schedule IV, or

7.3.2.

the procedure relating to EC type-examination set

out in Annex V of the Directive, which is set out in

Schedule V, coupled with

7.3.2.1.

the procedure relating to EC verification

set out in Annex VI, which is set out in

Schedule VI, or

7.3.2.2.

the procedure relating to the EC

declaration of conformity set out in

Annex

VII

(production

quality

assurance), which is set out in Schedule

VII.

7.4. In the case of devices for performance evaluation, the

manufacturer shall follow the procedure referred to in Annex VIII

of the Directive, which is set out in Schedule VIII, and draw up the

statement set out in that Schedule before such devices are made

available.

7.4.1.

This provision does not affect national regulations

relating to the ethical aspects of carrying out

performance evaluation studies using tissues or

substances of human origin.

7.5. During the conformity assessment procedure for a device,

the manufacturer and, if involved, the notified body shall take

account of the results of any assessment and verification operations

which, where appropriate, have been carried out in accordance with

these regulations at an intermediate state of manufacture.

7.6. The manufacturer may instruct his authorised

representative to initiate the procedures provided for in Annexes

III, V, VI and VIII of the Directive, which are set out in Schedules

III, V, VI and VIII.

7.7. The manufacturer must keep the declaration of conformity,

the technical documentation referred to in Annexes III to VIII of

the Directive, which are set out in Schedules III and VIII, as well as

the decisions, reports and certificates, established by notified

bodies, and make it available to the national authorities for

inspection purposes for a period ending five years after the last

product has been manufactured. Where the manufacturer is not

established in Malta, the obligation to make the aforementioned

documentation available on request applies to his authorised

representative.

7.8. Where the conformity assessment procedure involves

intervention of a notified body, the manufacturer, or his authorised

representative, may apply to a body of his choice within the

framework of tasks for which the body has been notified.

7.9. The notified body may require, where duly justified, any

information or data, which is necessary for establishing and

maintaining the attestation of conformity in view of the chosen

procedure.

7.10. Decisions taken by the notified bodies in accordance with

Annexes III, IV, and V of the Directive, which are set out in

Schedules III, IV, and V, shall be valid for a maximum of five years

and may be extended on application, made at a time agreed in the

contract signed by both parties, for further periods of up to five

years.

7.11. The records and correspondence relating to the procedures

referred to in regulations 7.1 to 7.4 shall be in an official language.

7.12. By way of derogation from regulations 7.1 to 7.4, the

competent authorities may authorise, on duly justified request, the

placing on the market and putting into service, of individual

medical devices for which the procedures referred to in regulations

7.1 to 7.4 have not been carried out and the use of which is in the

interest of protection of health.

7.13. The provisions of regulation 7 shall apply accordingly to

any natural or legal person who manufactures devices covered by

these regulations and, without placing them on the market, puts

them into service and uses them in the context of his professional

activity.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

Based on Annex VII of the Directive

EC DECLARATION OF CONFORMITY

(PRODUCTION QUALITY ASSURANCE)

1. The manufacturer must ensure application of the quality system approved

for the manufacture of the devices concerned and carry out the final inspection, as

specified in section 3 of this Schedule, and is subject to the surveillance referred to

in section 4 of this Schedule.

2. The declaration of conformity is the part of the procedure whereby the

manufacturer who fulfils the obligations imposed by section 1 ensures and declares

that the products concerned conform to the type described in the EC type-

exanunation certificate and meet the provisions of these regulations which apply to

them.

The manufacturer must affix the CE marking in accordance with regulation

12 of these regulations and draw up a declaration of conformity covering the devices

concerned.

3. Quality system

3.1. The manufacturer must lodge an application for assessment of his quality

system with a notified body.

The application must include the technical documentation on the types

approved and a copy of the EC type-examination certificates.

3.2. Application of the quality system must ensure that the devices conform to

the type described in the EC type-examination certificate.

All the elements, requirements and provisions adopted by the manufacturer

for his quality system must be documented in a systematic and orderly manner in the

form of written policy statements and procedures. This quality system

documentation must permit uniform interpretation of the quality policy and

procedures such as quality progranunes, plans, manuals and records.

It must include in particular an adequate description of

(a) the manufacturer’s quality objectives;

(b) the organisation of the business and in particular

-

the organisational structures, the responsibilities of the

managerial staff and their organisational authority where quality

of manufacture of the devices is concerned,

-

the methods of monitoring the efficient operation of the quality

system and in particular its ability to achieve the desired quality

of product, including control of devices which fail to conform;

(c) the inspection and quality assurance techniques at the manufacturing

stage and in particular

-

the processes and procedures which will be used, particularly as

regards sterilisation,

-

the procedures in relation to purchasing,

-

the product identification procedures drawn up and kept up to

date from drawings, specifications or other relevant documents at

every stage of manufacture;

(d) the appropriate tests and trials to be carried out before, during and after

manufacture, the frequency with which they will take place, and the test

equipment used; it must be possible to trace back the calibration.

3.3. The notified body must audit the quality system to determine whether it

meets the requirements referred to in section 3.2 of this Schedule. It must presume

that quality systems which implement the relevant harmonised standards conform to

these requirements.

The assessment team must have past experience of assessments of the

technology concerned. The assessment procedure must include an inspection on the

manufacturer’s premises and, in duly substantiated cases, on the premises of the

manufacturer ’s suppliers and, or subcontractors to inspect the manufacturing

processes.

The decision must be notified to the manufacturer. It must contain the

conclusions of the inspection and a reasoned assessment.

3.4. The manufacturer shall inform the notified body which approved the quality

system of any plan for substantial changes to the quality system. The notified body

must assess the changes proposed and verify whether after these changes the quality

system still meets the requirements referred to in section 3.2 of this Schedule. It

must notify the manufacturer of its decision. This decision must contain the

conclusions of the inspection and a reasoned assessment.

4. Surveillance

The provisions of Annex IV, section 5 of the Directive, which is set out in

Schedule IV, shall apply.

5. Verification of manufactured products covered by Schedule II, List A.

5.1. In the case of devices covered by Annex II, List A of the Directive, which is

set out in Schedule II, the manufacturer shall forward to the notified body without

delay after the conclusion of the controls and tests the relevant reports on the tests

carried out on the manufactured devices or each batch of devices.

Furthermore, the manufacturer shall make the samples of manufactured

devices or batches of devices available to the notified body in accordance with pre-

IN VITRO DIAGNOSTIC MEDICAL DEVICES

agreed conditions and modalities.

5.2. The manufacturer may place the devices on the market, unless the notified

body communicates to the manufacturer within the agreed time-frame, but not later

than thirty days after reception of the samples, any other decision, including in

particular any condition of validity of delivered certificates.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. Where it is ascertained that the devices referred to in

regulation 6.1, when correctly installed, maintained and used in

accordance with their intended purpose, may compromise the

health and, or safety of patients, users or, where applicable other

persons, or the safety of property, the Medicines Authority shall take

all appropriate interim measures to withdraw such devices from the

market, or prohibit or restrict their being placed on the market or

put into service.

8.1.1.

The Medicines Authority shall immediately inform

Commission of any such measures, indicating the

reasons for the decision and in particular, whether

non-compliance is due to

8.1.1.1.

failure

to

meet

the

essential

requirements referred to in regulation 5;

8.1.1.2.

incorrect application of the standards

referred to in Article 5 of the Directive,

insofar as it is claimed that the standards

have been applied;

8.1.1.3.

shortconungs

in

the

standards

themselves.

8.2. Where a non-complying medical device referred to in

regulation 3 bears the CE marking, the Medicines Authority shall take

appropriate action against whomsoever has affixed the marking or

drawn the declaration and shall so inform the Commission and the

other Member States thereof.

8.3. Any decision taken pursuant to these regulations which

restricts the placing on the market and putting into service of

devices shall state the exact grounds on which it is based.

8.3.1.

Such a decision shall be notified as soon as

possible to the party concerned, who shall at the

same time be informed of the legal remedies

available to him under the laws in force in Malta

and of the time limits to which such remedies are

subject.

8.4. Where the Medicines Authority considers, in relation to a

given product or group of products, that, in order to ensure

protection of health and safety and, or to ensure that public health

requirements are observed pursuant to Article 36 of the Treaty, the

availability of such products should be prohibited, restricted or

made subject to particular requirements, necessary and justified

transitional measures may be taken.

8.5. The Medicines Authority shall immediately inform the

Commission of any measures taken in pursuance of regulations

8.1.1, 8.2 and 8.4.

8.6. In the event of a decision as referred to in regulation 8.2,

the manufacturer or his authorised representative shall have an

opportunity to put forward his point of view in advance, unless

such consultation is not possible because of the urgency of the

measure to be taken as justified in particular by public health

requirements.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

Based on Annex VIII of the Directive

STATEMENT AND PROCEDURES CONCERNING DEVICES

FOR PERFORMANCE EVALUATION

1. For devices for performance evaluation the manufacturer or his authorized

representative shall draw up the statement containing the information stipulated in

section 2 of this Schedule and ensure that the relevant provisions of these regulations

are met.

2. The statement shall contain the following information

-

data allowing identification of the device in question,

-

an evaluation plan stating in particular the purpose, scientific, technical

or medical grounds, scope of the evaluation and number of devices

concerned,

-

the list of laboratories or other institutions taking part in the evaluation

study,

-

the starting date and scheduled duration for the evaluations and, in the

case of devices for self-testing, the location and number of lay persons

involved,

-

a statement that the device in question conforms to the requirements of

the regulations, apart from the aspects covered by the evaluation and

apart from those specifically itenused in the statement, and that every

precaution has been taken to protect the health and safety of the patient,

user and other persons.

3. The manufacturer shall also undertake to keep available for the competent

national authorities the documentation allowing an understanding of the design,

manufacture and performances of the product, including the expected performances,

so as to allow assessment of conformity with the requirements of these regulations.

This documentation must be kept for a period ending at least five years after the end

of the performance evaluation.

The manufacturer shall take all the measures necessary for the

manufacturing process to ensure that the products manufactured conform to the

documentation mentioned in the first paragraph.

4. The provisions of regulations 9.1, 9.3 and 9.5 of these regulations shall

apply to devices intended for performance evaluation.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. Any manufacturer who places devices on the Maltese

market under his own name shall notify the Medicines Authority

9.1.1.

of the address of the registered place of business,

9.1.2.

in the case of an authorised representative,

sufficient evidence that he is the authorised

representative of the manufacturer;

9.1.3.

of information relating to the reagents, reagent

products and calibration and control materials in

terms of common technological characteristics and,

or analytes and of any significant change thereto

including discontinuation of placing on the market;

for other devices, the appropriate indications;

9.1.4.

in the case of devices covered by Annex II of the

Directive, which is set out in Schedule II, and of

devices for self-testing, of all data allowing for

identification of such devices, the analytical and,

where appropriate, diagnostic parameters as

referred to in Annex I, part A, section 3 of the

Directive, which is set out in Schedule I, the

outcome of performance evaluation pursuant to

Annex VIII of the Directive, which is set out in

Schedule VIII, certificates and any significant

change thereto, including discontinuation of

placing on the market.

9.2. For devices covered by Annex II of the Directive, which is

set out in Sched ule I I, and for devices for self-t esting, the

Medicines Authority may request to be informed of the data

allowing identification together with the label and the instructions

for use when such medical devices are placed on the market and, or

put into service within their territory.

9.2.1.

The measures given in regulation 9.2 cannot

constitute a precondition for the placing on the

market and, or putting into service of medical

devices which are in conformity with these

regulations.

9.3. Where a manufacturer who places medical devices on the

market under his own name does not have a registered place of

business in Malta, he shall designate an authorised representative.

The authorized representative shall notify the Medicines Authority

of all paiticulars as referred to in regulation 9.1.

9.4. The notification referred to in regulation 9.1 shall also

include any new device. In addition, where, in the context of such

notification, a device notified, bearing the CE marking, is a "new

product", the manufacturer shall indicate this fact on his

notification.

For the purposes of regulation 9, a medical device is "new"

if

9.4.1.

there has been no such device continuously

available on the Cornmunity market during the

previous three years for the relevant analyte or

other parameter;

9.4.2.

the procedure involves analytical technology not

continuously used in connection with a given

analyte or other parameter on the Community

market during the previous three years.

9.5. The notifications referred to in regulations 9.1 and 9.3 are

to be registered immediately in the databank described in Article 12

of the Directive. The procedures for implementing this regulation

and in particular those referring to the notification and the concept

of significant change shall be adopted in accordance with the

procedure referred to in Article 7 of the Directive.

9.6. Transitionally, pending the establishment of a European

Databank accessible to the competent authorities of the Member

States and containing the data relating to all medical devices

available on the territory of the Comrnunity, the manufacturer shall

give such notification to the competent authorities of each Member

State concerned by the placing on the market.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

Based on Annex IX of the Directive

CRITERIA FOR THE DESIGNATION OF NOTIFIED BODIES

1. The notified body, its director and the assessment and verification staff shall

not be the designer, manufacturer, supplier, installer or user of the devices which

they inspect, nor the authorised representative of any of these persons. They may not

be directly involved in the design, construction, marketing or maintenance of the

devices, nor represent the parties engaged in these activities. This in no way

precludes the possibility of exchanges of technical information between the

manufacturer and the body.

2. The notified body and its staff must carry out the assessment and

verification operations with the highest degree of professional integrity and the

requisite competence in the field of medical devices and must be free from all

pressures and inducements, particularly financial, which might influence their

judgment or the results of the inspection, especially from persons or groups of

persons with an interest in the results of the verifications.

Should the notified body subcontract specific tasks connected with the

establishment and verification of the facts, it must first ensure that the subcontractor

meets the provisions of the regulations. The notified body shall keep at the disposal

of the national authorities the relevant documents assessing the subcontractor’s

qualifications and the work carried out by the subcontractor under these regulations.

3. The notified body must be able to carry out all the tasks assigned to such

bodies by one of Annexes III to VII of the Directive, which are set out in Schedules

III to VII, and for which it has been notified, whether these tasks are carried out by

the body itself or on its responsibility. In particular, it must have the necessary staff

and possess the facilities needed to perform properly the technical and

administrative tasks entailed in assessment and verification. This includes the

availability of sufficient scientific staff within the organisation who possess

adequate experience and knowledge necessary to assess the biological and medical

functionality and performance of devices for which it has been notified, in relation to

the requirements of this Directive and, in particular, with the requirements of Annex

I of the Directive, which is set out in Schedule I. The notified body must also have

access to the equipment necessary for the verifications required.

4. The inspection staff must have

-

sound vocational training covering all the assessment and verification

operations for which the body has been designated,

-

satisfactory knowledge of the rules on the inspections which they carry

out and adequate experience of such inspections,

-

the ability required to draw up the certificates, records and reports to

demonstrate that the inspections have been carried out.

5. The impartiality of the inspection staff must be guaranteed. Their

remuneration must not depend on the number of inspections carried out, nor on the

results of the inspections.

6. The body must take out civil liability insurance, unless liability is assumed

by the State under domestic legislation or the Member State itself carries out the

inspections directly.

7. The staff of the inspection body are bound to observe professional secrecy

with regard to all information gained in the course of their duties (except vis-a-vis

the competent administrative authorities of the State in which their activities are

carried out) under these regulations or any provision of national law putting it into

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. The Medicines Authority shall take the necessary steps to

ensure that any information brought to his knowledge, in

accordance with the provisions of these regulations, regarding the

incidents mentioned below involving devices bearing the CE

marking is recorded and evaluated centrally

10.1.1. any malfunction, failure or deterioration in the

characteristics and, or performance of a device, as

well as any inadequacy in the labelling or the

instructions for use which, directly or indirectly,

might lead to or might have led to the death of a

patient, or user or of other persons or to a serious

deterioration in their state of health;

10.1.2. any technical or medical reason in relation to the

characteristics or performance of a device for the

reasons referred to in regulation 10.1.1, leading to

systematic recall of medical devices of the same

type by the manufacturer.

10.2. Where medical practitioners, medical institutions or the

organizers of external quality assessment schemes are required to

inform the competent authorities of any incidents referred to in

regulation 10.1, necessary steps shall be taken to ensure that the

manufacturer of the device concerned, or his authorized

representative, is also informed on the incident.

10.3. After carrying out an assessment, if possible together with the

manufacturer, the Medicines Authority shall, without prejudice to

Article 8, immediately inform the Commission and the other Member

States of the incidents referred to in regulation 10.1 for which

appropriate measures, including possible withdrawal, have been taken

or are contemplated.

10.4. Where, in the context of notification referred to in

regulation 9, a device notified, bearing the CE marking, is a "new"

product, the manufacturer shall indicate this fact on his

notification. The Medicines Authority so notified may at any time

within the following two years and on justified grounds, require the

manufacturer to submit a report relating to the experience gained

with the device subsequent to its being placed on the market.

10.5. The Medicines Authority of the Malta Standards Authority

shall immediately inform the Commission of any measures taken in

pursuance of regulation 10.3 of these regulations.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

Based on Annex X of the Directive

CE MARKING OF CONFORMITY

The CE conformity marking shall consist of the initials "CE" taking the

following form

For the avoidance of doubt, it is hereby decdared that the grid providing the

background in the above graduated drawing is not part of the CE marking.

if the marking is reduced or enlarged the proportions given in the above

graduated drawing must be respected,

the various components of the CE marking must have substantially the

same vertical dimension, which may not be less than 5mm. This

minimum dimension may be waived for small-scale devices.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. The CE marking consists of the initials "CE" in accordance

with the model in Annex X of the Directive, which is set out in

Schedule X.

11.2. Devices, other than medical devices for performance

evaluation, considered to meet the essential requirements referred

to in regulation 5, must bear the CE marking of conformity when

they are placed on the market.

11.3. The CE marking of conformity, as shown in Annex X of

the Directive, which is set out in Schedule X, must appear in a

visible, legible and indelible form on the device, where practicable

and appropriate, and on the instructions for use. The CE marking of

conformity must also appear on the sales packaging.

11.3.1. The CE marking shall be accompanied by the

identification number of the notified body

responsible for implementation of the procedures

set out in Annexes III, IV, VI and VII of the

Directive, which are set out in Schedules III, IV,

VI, and VII.

11.4. No person shall affix on any device, to which these

regulations apply, any markings or inscriptions which are likely to

mislead third parties with regard to the meaning or the graphics of

the CE marking.

11.5. Without prejudice to regulation 11.4, nothing in these

regulations shall prohibit the affixing on device, to the packaging

or to the instruction leaflet accompanying the device, provided that

the visibility and legibility of the CE marking are not thereby

reduced.

11.6. Without prejudice to regulation 8

11.6.1. where it is established that the CE marking has

been wrongly affixed, the manufacturer or his

authorized representative shall be obliged to end

the infringement under conditions laid down by the

Medicines Authority;

11.6.2. where non-compliance continues, the Medicines

Authority shall take all appropriate measures to

restrict or prohibit the placing on the market of the

product in question or to ensure that it is

withdrawn from the market, in accordance with the

procedure laid down in regulation 8.

11.7. The provisions stated in regulation 11.6 shall also apply

where the CE marking has been affixed in accordance with the

procedures in these regulations, but inappropriately, on products

that are not covered by these regulations.

11.8. Subject to regulation 11.9, where the devices are subject to

other directives, concerning other aspects which also provide for

the affixing of the CE marking, the latter shall indicate that the

devices also fulfil the provisions of the other directives.

11.9. Where one or more of the other directives referred to in

regulation 11.8 allow the manufacturer, during a transitional

period, to choose which arrangements to apply, the CE marking

shall indicate that the devices fulfil the provisions only of those

directives applied by the manufacturer. In case, the particulars of

these directives, as published in the Official Journal of the

European Communities, must be given in the documents, notices or

instructions required by the directives and accompanying such

devices.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

L.N. 30 of 2004.

Substituted by

L.N. 21 of 2010.

Based on Annex I of Commission Decision 2009/108/EC of 3 February 2009 on common

technical specifications (CTS) for in vitro diagnostic medical devices

1. SCOPE

The common technical specifications set out in this Schedule shall apply for

the purposes of Schedule II, List A.

2. DEFINITIONS AND TERMS

(Diagnostic) sensitivity

The probability that the device gives a positive result in the presence of the

target marker.

True positive A specimen known to be positive for the target marker and

correctly classified by the device.

False negative A specimen known to be positive for the target marker and

misclassified by the device.

(Diagnostic) specificity The probability that the device gives a negative result

in the absence of the target marker.

False positive

A specimen known to be negative for the target marker and misclassified by

the device.

True negative

A specimen known to be negative for the target marker and correctly

classified by the device.

Analytical sensitivity

Analytical sensitivity may be expressed as the limit of detection, i.e. the smallest

amount of the target marker that can be precisely detected.

Analytical specificity

Analytical specificity means the ability of the method to determine solely the

target marker.

Nucleic acid amplification techniques (NAT)

The term ‘NAT’ is used for tests for the detection and/or quantification of

nucleic acids by either amplification of a target sequence, by amplification of a

signal or by hybridisation.

Rapid test

‘Rapid test’ means qualitative or semi quantitative in vitro diagnostic

medical devices, used singly or in a small series, which involve non-automated

procedures and have been designed to give a fast result.

Robustness

The robustness of an analytical procedure means the capacity of an analytical

procedure to remain unaffected by small but deliberate variations in method

parameters and provides an indication of its reliability during normal usage.

Whole system failure rate

The whole system failure rate means the frequency of failures when

the

entire process is performed as prescribed by the manufacturer.

Confirmation assay

Confirmation assay means an assay used for the confirmation of a reactive result

from a screening assay.

Virus typing assay

Virus typing assay means an assay used for typing with already known positive

samples, not used for primary diagnosis of infection or for screening.

Sero-conversion HIV samples

IN VITRO DIAGNOSTIC MEDICAL DEVICES

Sero-conversion HIV samples mean

-

p24 antigen and/or HIV RNA positive and;

-

recognised by all of the antibody screening tests and;

-

positive or indeterminate confirmatory assays.

Early sero-conversion HIV samples

Early sero-conversion HIV samples mean

-

p24 antigen and/or HIV RNA positive and;

-

not recognised by all of the antibody screening tests and;

-

indeterminate or negative confirmatory assays.

3. COMMON TECHNICAL SPECIFICATIONS (CTS) FOR PRODUCTS

REFERRED TO IN SCHEDULE II, LIST A.

3.1 CTS for performance evaluation of reagents and reagent products for the

detection, confirmation and quantification in human specimens of markers of HIV

infection (HIV 1 and 2), HTLV I and II, and hepatitis B, C, D

General Principles

3.1.1 Devices which detect virus infections placed on the market for use as

either screening or diagnostic tests, shall meet the requirements for sensitivity and

specificity set out in Table 1. See also principle 3.1.11 for screening assays.

3.1.2 Devices intended by the manufacturer for testing body fluids other than

serum or plasma, for example urine, saliva, etc shall meet the same CTS

requirements for sensitivity and specificity as serum or plasma tests. The

performance evaluation shall test samples from the same individuals in both the tests

to be approved and in a respective serum or plasma assay.

3.1.3 Devices intended by the manufacturer for self-test, i.e. home use, shall

meet the same CTS requirements for sensitivity and specificity as respective devices

for professional use. Relevant parts of the performance evaluation shall be

carried out (or repeated) by appropriate lay users to validate the operation of the

device and the instructions for use.

3.1.4 All performance evaluations shall be carried out in direct comparison with

an established state of the art device. The device used for comparison shall be

one bearing CE marking, if on the market at the time of the performance

evaluation.

3.1.5 If discrepant test results are identified as part of an evaluation, these

results shall be resolved as far as possible, for example

-

by evaluation of the discrepant sample in further test systems;

-

by use of an alternative method or marker;

-

by a review of the clinical status and diagnosis of the patient; and

-

by the testing of follow-up-samples.

3.1.6 Performance evaluations shall

be

performed

on

a population

equivalent to the European population.

3.1.7 Positive specimens used in the performance evaluation shall be

selected to reflect different stages of the respective disease(s), different

antibody patterns, different genotypes, different subtypes, mutants, etc.

3.1.8 Sensitivity with true positives and sero-conversion samples shall be

evaluated as follows

3.1.8.1 Diagnostic test sensitivity during sero-conversion has to represent the

state of the art. Whether further testing of the same or additional sero-conversion

panels is conducted by the notified body or by the manufacturer the results

shall confirm the initial performance evaluation data (see Table 1).

Seroconversion panels should start with a negative bleed(s) and should have narrow

bleeding intervals.

3.1.8.2 For blood screening devices (with the exception of HBsAg and anti-

HBc tests), all true positive samples shall be identified as positive by the device

to be CE marked (Table 1). For HBsAg and anti-HBc tests the new device shall

have an overall performance at least equivalent to that of the established device

(see 3.1.4).

3.1.8.3 Regarding HIV tests

-

all sero-conversion HIV samples shall be identified as positive; and

-

at least 40 early sero-conversion HIV samples shall be tested. Results

should conform to the state of the art.

3.1.9 Performance evaluation of screening assays shall include 25 positive (if

available in the case of rare infections) "same day" fresh serum and/or plasma

samples (≤ 1 day after sampling).

3.1.10 Negative specimens used in a performance evaluation shall be defined

so as to reflect the target population for which the test is intended, for example blood

donors, hospitalised patients, pregnant women, etc.

3.1.11 For performance evaluations for screening assays (Table 1) blood donor

populations shall be investigated from at least two blood donation centres and

consist of consecutive blood donations, which have not been selected to exclude

first time donors.

3.1.12 Devices shall have a specificity of at least 99.5% on blood donations,

unless otherwise indicated in the accompanying tables. Specificity shall be

calculated using the frequency of repeatedly reactive (i.e. false positive)

results in blood donors negative for the target marker.

3.1.13 Devices shall be evaluated to establish the effect of potential

interfering substances, as part of the performance evaluation. The potential

interfering substances to be evaluated will depend to some extent on the

composition of the reagent and configuration of the assay. Potential interfering

substances shall be identified as part of the risk analysis required by the

essential requirements for each new device but may include, for example

- specimens representing "related" infections;

- specimens from multipara, i.e. women who have had more than

one pregnancy, or rheumatoid factor positive patients;

- for recombinant antigens, human antibodies to components of the

expression system, for example anti E. coli, or anti yeast.

3.1.14 For devices intended by the manufacturer to be used with serum and

plasma the performance evaluation must demonstrate serum to plasma

equivalency. This shall be demonstrated for at least 50 donations (25 positive and 25

negative).

3.1.15 For devices intended for use with plasma the performance

evaluation shall verify the performance of the device using all anticoagulants

which the manufacturer indicates for use with the device. This shall be demonstrated

IN VITRO DIAGNOSTIC MEDICAL DEVICES

for at least 50 donations (25 positive and 25 negative).

3.1.16 As part of the required risk analysis the whole system failure rate

leading to false-negative results shall be determined in repeat assays on low-

positive specimens.

3.1.17 If a new in vitro diagnostic medical device belonging to Schedule II, List

A is not specifically covered by the common technical specification, the

common technical specification for a related device should be taken into account.

Related devices may be identified on different grounds, for example by the

same or similar intended use or by similar risks.

3.2

Additional Requirements for HIV antibody/antigen combined tests

3.2.1 HIV antibody/antigen combined tests intended for anti-HIV and p24

antigen detection which include claims for single p24 antigen detection shall follow

Table 1 and Table 5, including criteria for analytical sensitivity for p24 antigen.

3.2.2 HIV antibody/antigen combined tests intended for anti-HIV and p24

detection which do not include claims for single p24 detection shall follow

Table 1 and Table 5, excluding criteria for analytical sensitivity for p24.

3.3 Additional Requirements for Nucleic Acid Amplification Techniques

(NAT)

The performance evaluation criteria for NAT assays can be found in Table 2.

3.3.1 For target sequence amplification assays, a functionality control for

each test sample (internal control) shall reflect the state of the art. This control

shall as far as possible be used throughout the whole process, i.e. extraction,

amplification/hybridisation, detection.

3.3.2 The analytical sensitivity or detection limit for NAT assays shall be

expressed by the 95% positive cut-off value. This is the analyte concentration where

95% of test runs give positive results following serial dilutions of an international

reference material for example a WHO standard or calibrated reference material.

3.3.3 Genotype detection shall be demonstrated by appropriate primer or probe

design validation and shall also be validated by testing characterised genotyped

samples.

3.3.4 Results of quantitative NAT assays shall be traceable to international

standards or calibrated reference materials, if available, and be expressed in

international units utilised in the specific field of application.

3.3.5 NAT assays may be used to detect virus in antibody negative samples,

i.e. pre sero conversion samples. Viruses within immune-complexes may

behave differently in comparison to free viruses, for example during a

centrifugation step. It is therefore important that during robustness studies,

antibody-negative (pre sero-conversion) samples are included.

3.3.6 For investigation of potential carry-over, at least five runs with alternating

high positive and negative specimens shall be performed during robustness

studies. The high positive samples shall comprise of samples with naturally

occurring high virus titres.

3.3.7 The whole system failure rate leading to false-negative results shall be

determined by testing low-positive specimens. Low positive specimens shall

contain a virus concentration equivalent to 3 times the 95% positive cut-off virus

concentration.

3.4

CTS for the manufacturer’s release testing of reagents and eagent products

for the detection, confirmation and quantification in human specimens of markers of

HIV infection (HIV 1 and 2), HTLV I and II, and hepatitis B, C, D (Immunological

assays only).

3.4.1 The manufacturer’s release testing criteria shall ensure that every batch

consistently identifies the relevant antigens, epitopes, and antibodies.

3.4.2. The manufacturer’s batch release testing for screening assays shall

include at least 100 specimens negative for the relevant analyte.

3.5 CTS for performance evaluation of reagents and reagent products for

determining the following blood group antigens: ABO blood group system ABO1

(A), ABO2 (B), ABO3 (A,B); Rh blood group system RH1 (D), RH2 (C), RH3 (E),

RH4 (c), RH5 (e); Kell blood group system KEL1 (K).

Criteria for performance evaluation of reagents and reagent products for

determining the blood groups antigens: ABO blood group system ABO1 (A),

ABO2 (B), ABO3 (A,B); Rh blood group B 9 system RH1 (D), RH2 (C), RH3 (E),

RH4 (c), RH5 (e); Kell blood group system KEL1 (K) can be found in Table 9.

3.5.1 All performance evaluations shall be carried out in direct comparison with

an established state of the art device. The device used for comparison shall be

one bearing CE marking, if on the market at the time of the performance

evaluation.

3.5.2 If discrepant test results are identified as part of an evaluation, these

results shall be resolved as far as possible, for example

-

by evaluation of the discrepant sample in further test systems;

-

by use of an alternative method.

3.5.3 Performance evaluations shall

be

performed

on

a population

equivalent to the European population.

3.5.4 Positive specimens used in the performance evaluation shall be selected

to reflect variant and weak antigen expression.

3.5.5 Devices shall be evaluated to establish the effect of potential

interfering substances, as part of the performance evaluation. The potential

interfering substances to be evaluated will depend to some extent on the

composition of the reagent and configuration of the assay. Potential interfering

substances shall be identified as part of the risk analysis required by the

essential requirements for each new device.

3.5.6 For devices intended for use with plasma the performance evaluation shall

verify the performance of the device using all anticoagulants which the

manufacturer indicates for use with the device. This shall be demonstrated for at

least 50 donations.

3.6. CTS for the manufacturers release testing of reagents and reagent products

for determining the blood group antigens: ABO blood group system ABO1 (A),

ABO2 (B), ABO3 (A,B); Rh blood group system RH1 (D), RH2 (C), RH3 (E), RH4

(c), RH5 (e); Kell blood group system KEL1 (K).

3.6.1 The manufacturer’s release testing criteria shall ensure that every batch

consistently identifies the relevant antigens, epitopes, and antibodies.

3.6.2 Requirements for manufacturers batch release testing are outlined in Table

10. Table 1: "Screening" assays: anti-HIV 1 and 2, anti-HTLV I and II,

IN VITRO DIAGNOSTIC MEDICAL DEVICES

anti-HCV, HBsAg, anti-HBc

anti-HIV-1/2

anti-HTLV-

anti-HCV

HBsAg

anti-HBc

I/II

Diagnostic

Positive

400 HIV-1 100

300 HTLV-I 400 (positive

400 Including

sensitivity

specimens

HIV-2

100 HTLV-II samples)

subtype-

Including

including 40

consideration

evaluation

Including

non-B-

of other

samples from

subtypes, all

HBV-

different stages

available HIV/

markers

of infection and

1 subtypes

reflecting

should be

different

represented by

antibody

at least 3

patterns.

samples per

subtype

Genotype 1-4: >

20 samples - per

genotype

(including non-a

sub-types of

genotype 4);

5: > 5 samples;

6: if available

Sero-

20 panels

To be

20 panels

20 panels

To be

conversion

10 further

defined

10 further panels

10 further

defined

panels

panels (at

when

(at Notified

panels (at

when

Notified Body

available

Body or

Notified Body

available

or

manufacturer)

or

manufacturer)

manufacturer)

Analytical

Standards

0.130 IU/ml

sensitivity

(Second

International

Standard for

HBsAg,

subtype adw2,

genotype A,

NIBSC code

00/588)

Specificity

Unselected

donors

(including1

st

time

donors)

Hospitalize

d patients

Potentially

cross-

reacting

blood-

specimens

(RF+,

related

viruses,

pregnant

women,

etc)

Table 2: NAT assays for HIV1, HCV, HBV, HTLV I/II (qualitative and

quantitative; not molecular typing)

HIV1

HCV

HBV

HTLV I/II

Accep-

tance

NAT

qualitative

quantitative

qualitative

quantitative

quantitative

qualitative

quantitative

criteria

As for HIV

As for HIV

As for HIV

quantitative

quantitative

quantitative

Sensitivity

According

Detection

According

According

According

Detection

to EP

limit: as

to EP

to EP

to EP

limit

validation

for

validation

validation

validation

Detection

guideline

qualitative

guideline

guideline

guideline

( 1 )

( 1 ): several

( 1 ): several

of

(1)

tests;

analytical

several

Quantific-

several

dilution

dilution

sensitivity

dilution

ation limit

dilution

series into

series into

(IU/ml;

series into

dilutions

series into

borderline

borderline

defined on

borderline

(half-

borderline

concentr-

concentr-

WHO

concentra-

log10 or

concentra-

ation;

ation;

standards

tion;

less) of

tion;

statistical

statistical

or

statistical

calibrated

statistical

analysis

analysis

calibrated

analysis

reference

analysis

(e.g.

(e.g. Probit

reference

(e.g.

prepara-

(e.g.

Probit

analysis)

materials)

Probit

tions,

Probit

analysis)

on the basis

analysis)

definition

analysis)

on the

of at least

on the

of lower,

on the

basis of at

basis of at

upper

basis of at

least 24

replicates;

least 24

quantifi-

least 24

replicates;

calculation

replicates;

cation

replicates;

calculation

of 95% cut-

calculatio

limit,

calculatio

of 95%

off value

n of 95%

precision,

n of 95%

cut-off

cut-off

accuracy,

cut-off

value

value

"linear"

value

measuring

range,

"dynamic

range".

Reproduci

-bility at

different

concentra-

tion levels

to be

shown

Genotype/

At least 10

Dilution

At least 10

As far as

As far as

subtype

samples per

series of

samples

calibrated

calibrated

detection/

subtype (as

all

per

genotype

genotype

quantific-

far as

relevant

genotype

reference

reference

ation

available)

genotypes

(as far as

materials

materials

efficiency

/subtypes,

available)

are

are

preferabl

available

available

y of

reference

materials,

as far as

available

IN VITRO DIAGNOSTIC MEDICAL DEVICES

HIV1

HCV

HBV

HTLV I/II

Accep-

tance

NAT

qualitative

quantitative

qualitative

quantitative

quantitative

qualitative

quantitative

criteria

As for HIV

As for HIV

As for HIV

quantitative

quantitative

quantitative

Cell culture

Transcrip

supernatant

ts or

s (could

plasmids

substitute

quantified

for rare

by

HIV-1

appropria

subtypes)

te

According

meethods

to EP

may be

validation

used

guideline

According

According

According

( 1 ) as far as

to EP

to EP

to EP

calibrated

validation

validation

validation

subtype

guideline

guideline

guideline

reference

( 1 ) as far as

( 1 ) as far as

( 1 ) as far

materials

calibrated

calibrated

as

are

subtype

subtype

calibrated

available;

reference

reference

subtype

in vitro

materials

materials

reference

transcripts

are

are

materials

could be an

available;

available;

are

option

available;

in vitro

in vitro

transcripts

transcripts

in vitro

could be an

could be an

transcripts

option

option

could be

an option

Diagnosti

500 blood

500 blood

500 blood

500 blood

c

donors

blood

donors

donors

donors

specificit

donors

y

negative

samples

Potential

By

As for

By assays

By assays

By assays

cross

suitable

qualitative design and/

design and/

design

reactive

assay

tests

or testing

or testing of

and/or

markers

design

of at least

at least 10

testing of

evidence

10 human

other DNA-

at least 10

(e.g.

flavivirus

virus

human

sequence

(e.g. HGV,

positive

retrivirus

compari-

YFV)

samples

(e.g. HIV-)

son) and/

positive

positive

or testing

samples

samples

of at least

10 human

retrovirus

(e.g.

HTLV)-

positive

samples

Robustness

As for

qualitativ

e tests

Cross-

At least 5

At least 5

At least 5

At least 5

contami-

runs using

runs using

runs using

runs using

nation

alternatin

alternating

alternating

alternatin

g high

high

high

g high

positive

positive

positive

positive

(known to

(known to

(known to

(known to

occur

occur

occur

occur

naturally)

naturally)

naturally)

naturally)

and

and

and

and

negative

negative

negative

negative

samples

samples

samples

samples

Inhibition

Internal

Internal

Internal

Internal

control

control

control

control

preferably

preferably

preferably

preferably

to go

to go

to go

to go

through

through the

through the

through

the whole

whole NAT

whole NAT

the whole

NAT

procedure

procedure

NAT

procedure

procedure

HIV1

HCV

HBV

HTLV I/II

Accep-

tance

qualitative

quantitative

qualitative

quantitative

quantitative

qualitative

quantitative

criteria

As for HIV

As for HIV

As for HIV

quantitative

quantitative

quantitative

At least

At least

At least

At least

99 / 100

assays

samples

samples

samples

samples

positive

virus-

virus-

virus-

virus-

spiked with

spiked

spiked

spiked

3 x the

with 3 x

with 3 x

with 3 x

95% pos

the 95%

the 95 %

the 95 %

cut-off

pos cut-off

pos cut-off

pos cut-off

concentra-

concentra-

concentra-

concentr-

tion

tion

tion

ation

( 1 ) European Pharmacopoeia guideline

Acceptance criteria for ‘whole system failure rate leading to false-neg results’ is 99/100

assays positive.

For quantitative NATs a study shall be performed on at least 100 positive specimens

reflecting the routine conditions of users (e.g. no pre-selection of specimens). Comparative

results with another NAT test system shall be generated in parallel.

For qualitative NATs a study on diagnostic sensitivity shall be performed using at least 10

sero-conversion panels.

Comparative results with another NAT test system shall be generated in parallel.

Table 3: Rapid tests: anti-HIV 1 and 2, anti-HCV, HBsAg, anti-HBc, anti-

HTLV I and II

anti-HIV 1/2

anti-HCV

HBsAg

anti-HBc

anti-HTLV I/

Acceptance

II

criteria

Positive

Same criteria

Same criteria

Same criteria

Same criteria

Same criteria

Same criteria as

specimens

as for

as for

as for

as for

as for

for screening

screening

screening

screening

screening

screening

assays

assays

assays

assays

assays

assays

Sero-

Same criteria

Same criteria

Same criteria

Same criteria

Same criteria

Same criteria as

conversion

as for

as for

as for

as for

as for

for screening

panels

screening

screening

screening

screening

screening

assays

assays

assays

assays

assays

assays

99 % (anti-

Negative

1000 blood

1000 blood

1000 blood

1000 blood

1000 blood

HBc: 96 %)

specimens

donations 200

donations

donations

donations

donations

clinical

200 clinical

200 clinical

200 clinical

200 clinical

specimens

specimens 200

specimens 200 specimens

specimens

200 samples

samples from

samples from

200 samples

from pregnant

pregnant

pregnant

from pregnant

women

women

women

women

100 potentially 100 potentially 100

interfering

interfering

potentially

potentially

potentially

samples

samples

interfering

interfering

interfering

samples

samples

samples

IN VITRO DIAGNOSTIC MEDICAL DEVICES

Table 4: Confirmatory/supplementary assays for anti-HIV 1 and 2, anti-HTLV

I and II, anti-HCV, HBsAg

anti-HIV

anti-HTLV

HCV

HBsAg

Acceptance

Confirmatory

Confirmatory Supplementary

Confirmatory

criteria

Assay

Assay

Assay

Assay

Diagnostic Positive

200 HIV-1 and 200 HTLV-I

300 HCV

300 HBsAg

Correct

sensitivity specimens

100 HIV-2

and 100

(positive

identification

HTLV-II

samples)

as positive (or

indeterminate),

not negative

Including

Including

Including

samples from

samples from

samples from

different

different

different stages

stages of

stages of

of infection 20

infection and

infection and

“high pos”

reflecting

reflecting

samples (> 26

different

different

IU/ml); 20

antibody

antibody

samples in the

patterns

patterns.

cut-off range

Genotypes 1 -

4: > 20

samples

(including

non-a sub-

types of

genotype 4);

5: > 5 samples;

6: if available

Sero-con-

15 sero-

version

conversion

seroconversion

seroconversion

panels

panels/ low

panels/ low

panels/ low

titre panels

titre panels

titre panels

Analytical

Standards

Second

sensitivity

International

Standard for

HBsAg,

subtype adw2,

genotype A,

NIBSC code

00/588

anti-HTLV

HCV

HBsAg

Acceptance

anti-HIV

Confirmatory

Confirmatory Supplementary

Confirmatory

criteria

Assay

Assay

Assay

Assay

Diagnostic

Negative

200 blood

200 blood

200 blood

10 false

No false-

specificity

specimens

donations

donations

donations

positives as

positive

available from

results/ (1) no

the

neutralisation

performance

evaluation of

the screening

assay. (1)

200 clinical

200 clinical

200 clinical

samples

samples

samples

including

including

including

pregnant

pregnant

pregnant

women

women

women

50 potentially

50 potentially

potentially

potentially

interfering

interfering

interfering

interfering

samples,

samples

samples,

samples,

including

including

including

samples with

samples with

samples with

indeterminate

indetermin-

indeterminat

results in

ate results in

e results in

other

other

other

confirmatory

confirmatory

confirmatory

assays

assays

assays

Table 5: HIV 1 Antigen

HIV-1 Antigen Assay

Acceptance criteria

Diagnostic

Positive specimens

50 HIV-1 Ag-positive

Correct identification (after

sensitivity

50 cell culture supernatants

neutralisation)

including different HIV-1

subtypes and HIV-2

Sero-conversion panels

20 sero-conversion panels/low

titre panels

Analytical

Standards

HIV-1 p24 Antigen, 1st

≤ 2 IU/ml

sensitivity

International Reference

Reagent, NIBSC code: 90/636

Diagnostic

200 blood donations 200

>99.5% after neutralisation

specificity

clinical samples 50 potentially

interfering samples

Table 6: Serotyping and Genotyping Assay: HCV

HCV Serotyping and

Acceptance criteria

Genotyping Assay

Diagnostic

Positive specimens

(positive

samples)

> 95% agreement between

sensitivity

Including

samples

from

serotyping and genotyping

different stages of infection and

> 95% agreement between

reflecting different antibody

genotyping and sequencing

patterns. Genotypes 1 - 4: > 20

samples (including non-a sub-

types of genotype 4);

5: > 5 samples;

6: if available

IN VITRO DIAGNOSTIC MEDICAL DEVICES

HCV Serotyping and

Acceptance criteria

Genotyping Assay

Diagnostic

Negative specimens

specificity

Table 7: HBV Markers: anti-HBs, anti HBc IgM, anti-HBe, HBeAg

anti-HBs

anti-HBc IgM

anti-HBe

HBeAg

Diagnostic

Positive

100 vaccinees 200 Including 200 Including

200 Including

sensitivity

specimens 100 naturally samples from samples from

samples from

infected

different stages different stages

different

persons

of

infection of

infection

stages

(acute/chronic (acute/chronic

infection

etc.)

etc.)

(acute/chronic

etc.)

The

acceptance

criteria should

only

be

applied

on

samples from

acute infection

stage

Sero-con-

follow- When available

version

ups or anti-

panels

HBs

sero-

conversions

Analytical

Standards

WHO

1st

HBe

sensitivity

International

Referenzan-

Reference

tigen 82; PEI

Preparation

Germany

1977;

NIBSC,

United

Kingdom

Diagnostic

No

blood

blood

specificity

negative

Including

donations

donations

donations

specimens

clinical

samples

clinical

clinical

200 clinical

samples

samples

samples

50 potentially

50 potentially

50 potentially

interfering

interfering

interfering

samples

samples

samples

Table 8: HDV markers: anti-HDV, anti-HDV IgM, Delta Antigen

anti-HDV

anti-HDV IgM

Delta Antigen

Diagnostic

Positive

sensitivity

specimens

Specifying

HBV- Specifying HBV-

markers

markers

Diagnostic

Negative

specificity

specimens

Including

clinical Including clinical Including

clinical

samples

samples

samples

potentially 50

potentially 50

potentially

interfering samples interfering samples interfering samples

Table 9: Blood group antigens in the ABO, Rh and Kell blood group systems

Specificity

Number of tests per

Total number of samples

Total number of samples

recommended

to be tested for a launch

to be tested for a new

method

product

formulation, or use of

well-characterized

reagents

Anti-ABO1(anti-A),anti-

Anti-ABO2 (anti-B), Anti-

ABO3 (anti-A,B)

Anti-RH1 (anti-D)

Anti-RH2 (anti-C ), anti-

RH4

(anti-c),

anti-RH3 (anti-E)

Anti-RH5 (anti-e)

Anti-KEL1 (anti-K)

Acceptance criteria

All of the above reagents shall show comparable test results with

established reagents with acceptable performance with regard to claimed reactivity

of the device. For established reagents, where the application or use has been

changed or extended, further testing should be carried out in accordance with the

requirements outlined in column 1 (above).

Performance evaluation of anti-D-reagents shall include tests against a range of

weak RH1 (D) and partial RH1 (D) samples, depending on the intended use of the

product.

Qualifications

Clinical samples

10% of the test population

Neonatal specimens

> 2% of the test population

ABO samples

> 40% A, B positives

‘weak D’

> 2% of RH1 (D) positives

Table 10: Batch release criteria for reagents and reagent products to

determine blood group antigens in the ABO, Rh and Kell blood group systems

Specificity Testing Requirements on each reagent

1. Test reagents

Blood Group Reagents

Minimum number of control cells to be tested

Positive reactions

Negative reactions

A2B

Ax

B

Anti-ABO1 (anti-A)

2*

B

A1B

Anti-ABO2 (anti-B)

Ax

B

Anti-ABO3 (anti-A,B)

R1r

R2r

Weak D

r’r

r’’r

rr

Anti-RH1 (anti-D)

2*

R1R2

R1r

r’r

R2R2

r’’r

rr

Anti-RH2 (anti-C)

R1R2

R1r

r’r

R1R1

IN VITRO DIAGNOSTIC MEDICAL DEVICES

Blood Group Reagents

Minimum number of control cells to be tested

Positive reactions

Negative reactions

Anti-RH4 (anti-c)

R1R2

R2r

r”r

R1R1

r’r

rr

Anti-RH 3 (anti-E)

R1R2

R2r

r”r

R2R2

Anti-RH5 (anti-e)

Kk

kk

Anti-KEL1 (anti-K)

(*) Only by recommended techniques where reactivity against these antigens is claimed

Note: Polyclonal reagents must be tested against a wider panel of cells to confirm specificity

and exclude presence of unwanted contaminating antibodies.

Acceptance Criteria

Each batch of reagent must exhibit unequivocal positive or negative results by all

recommended techniquesin accordance with the results obtained from the performance

evaluation data.

2. Control Materials (red Cells)

The phenotype of red cells used in the control of blood typing reagents listed

above should be confirmed using established device.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

1. Without prejudice to national law and practice on medical

secrecy, it shall be ensured that all the parties involved in the

application of these regulations are bound to observe

confidentiality with regard to information obtained in carrying out

their tasks. This does not affect the obligations of the persons

concerned to provide information under criminal law.

Laws of Malta — Subsidiary Legislation 427.16 — IN VITRO DIAGNOSTIC MEDICAL DEVICES REGULATIONS

12.97, p. 19).

†Council Directive 88/379/EEC of 7 June 1988 on the approximation of the laws, regulations and

administrative provisions of the Member States relating to the classification, packaging and labelling of

dangerous preparations (OJ No. L 187, 16.07.88, P. 14). Directive as last amended by Commission Directive

96/65/EC (OJ No. L 265, 18.10.96 p. 15).

The provisions of the aforementioned Directives on the safety data sheet

shall apply, unless all relevant information as appropriate is already made available

by the instructions for use.

8.4. The label must bear the following particulars which may take the form of

symbols as appropriate

(a) the name or trade name and address of the manufacturer. For devices

imported into the Community with a view to their distribution in the

Community, the label, the outer packaging, or the instructions for use

shall contain in addition the name and address of the authorised

representative of the manufacturer;

(b) the details strictly necessary for the user to uniquely identify the device

and the contents of the packaging;

(c) where appropriate, the word "STERILE" or a statement indicating any

special microbiological state or state of cleanliness;

(d) the batch code, preceded by the word "LOT", or the serial number;

(e) if necessary, an indication of the date by which the device or part of it

should be used, in safety, without degradation of performance,

expressed as the year, the month and, where relevant, the day, in that

order;

(f)

in case of devices for performance evaluation, the words "for

performance evaluation only";

(g) where appropriate, a statement indicating the in vitro use of the device;

(h) any particular storage and, or handling conditions;

(i)

where applicable, any particular operating instructions;

(j)

appropriate warnings and, or precautions to take;

(k) if the device is intended for self-testing, that fact must be clearly stated.

8.5. If the intended purpose of the device is not obvious to the user, the

manufacturer must clearly state the intended purpose in the instructions for use and,

if appropriate, on the label.

8.6. Wherever reasonable and practicable, the devices and separate components

must be identified, where appropriate in terms of batches, to allow all appropriate

action to detect any potential risk posed by the devices and detachable components.

8.7. Where appropriate, the instructions for use must contain the following

particulars

(a) the details referred to in point 8.4 of this section with the exception of

points (d) and (e);

(b) composition of the reagent product by nature and amount or

concentration of the active ingredient(s) of the reagent(s) or kit as well

as a statement, where appropriate, that the device contains other

ingredients which might influence the measurement;

(c) the storage conditions and shelf life following the first opening of the

primary container, together with the storage conditions and stability of

working reagents;

(d) the performances referred to in point 3 of part A of this Schedule;

(e) an indication of any special equipment required including information

necessary for the identification of that special equipment for proper use;

(f)

the type of specimen to be used, any special conditions of collection,

pre-treatment and, if necessary, storage conditions and instructions for

the preparation of the patient;

(g) a detailed description of the procedure to be followed in using the

device;

(h) the measurement procedure to be followed with the device including as

appropriate

-

the principle of the method,

-

the specific analytical performance characteristics (e.g.

sensitivity, specificity, accuracy, repeatability, reproducibility,

limits of detection and measurement range, including information

needed for the control of known relevant interferences),

limitations of the method and information about the use of

available reference measurement procedures and materials by the

user,

-

the details of any further procedure or handling needed before the

device can be used (for example, reconstitution, incubation,

dilution, instrument checks, etc.),

-

the indication whether any particular training is required;

(i)

the mathematical approach upon which the calculation of the analytical

result is made;

(j)

measures to be taken in the event of changes in the analytical

performance of the device;

(k) information appropriate to users on

-

internal quality control including specific validation procedures,

-

the traceability of the calibration of the device;

(l)

the reference intervals for the quantities being determined, including a

description of the appropriate reference population;

(m) if the device must be used in combination with or installed with or

connected to other devices or equipment in order to operate as required

for its intended purpose, sufficient details of its characteristics to

identify the correct devices or equipment to use in order to obtain a safe

and proper combination;

(n) all the information needed to verify whether the device is properly

installed and can operate correctly and safely, plus details of the nature

and frequency of the maintenance and calibration needed to ensure that

the device operates properly and safely; information about safe waste

disposal;

(o) details of any further treatment or handling needed before the device

can be used (for example, sterilisation, final assembly, etc.);

(p) the necessary instructions in the event of damage to the protective

packaging and details of appropriate methods of resterilisation or

decontamination;

(q) if the device is reusable, information on the appropriate processes to

allow reuse, including cleaning, disinfection, packaging and

resterilisation or decontamination, and any restriction on the number of

reuses;

(r) precautions to be taken as regards exposure, in reasonably foreseeable

environmental conditions, to magnetic fields, external electrical

influences, electrostatic discharge, pressure or variations in pressure,

acceleration, thermal ignition sources, etc.;

(s) precautions to be taken against any special, unusual risks related to the

use or disposal of the device including special protective measures;

where the device includes substances of human or animal origin,

attention must be drawn to their potential infectious nature;

(t)

specifications for devices for self-testing

-

the results need to be expressed and presented in a way that is

readily understood by a lay person; information needs to be

provided with a device to the user on action to be taken (in case of

positive, negative or indeterminate result) and on the possibility

of false positive or false negative result,

-

specific particulars may be omitted provided that the other

information supplied by the manufacturer is sufficient to enable

the user to use the device and to understand the result(s) produced

by the device,

-

the information provided must include a statement clearly

directing that the user should not take any decision of medical

relevance without first consulting his or her medical practitioner,

-

the information must also specify that when the device for self-

testing is used for the monitoring of an existing disease, the

patient should only adapt the treatment if he has received the

appropriate training to do so;

(u) date of issue or latest revision of the instructions for use.