ACTIVE IMPLANTABLE MEDICAL DEVICES REGULATIONS
Table of Provisions (14)
Statutory Text (14 Articles)
SUBSIDIARY LEGISLATION 427.10
ACTIVE IMPLANTABLE MEDICAL DEVICES
REGULATIONS
21st March, 2010
LEGAL NOTICE 211 of 2008, as amended by Legal Notice 319 of 2020.
(1) The title of these regulations is Active Implantable
Medical Devices Regulations.
(2)
These regulations transpose the following Directives
(a) Council Directive 90/385/EEC * of 20 June 1990 on the
approximation of the laws of the Member States
relating to active implantable medical devices; and
(b) Directive 2007/47/EC † of the European Parliament
and of the Council of 5 September 2007 amending
Council Directive 90/385/EEC on the approximation
of the laws of the Member States relating to active
implantable medical devices, Council Directive 93/42/
EEC concerning medical devices and Directive 98/8/
EC concerning the placing of biocidal products on the
market.
(3) The
following
transitional
measures
shall
remain
applicable
(a) CE marking shall not be obligatory prior to the 1st
October 2001 in respect of stocks of active
implantable medical devices already placed on the
market prior to 1st April 2001 and active implantable
medical devices ordered before 1st April 2001 and
imported into Malta up to 30th September 2001; and
(b) The provisions of these regulations shall not apply to
contracts made before 1st April 2001, covering the
supply of active implantable medical devices to the
Government of Malta over the period up to and
including 31st December 2002, as well as those
contracts which had terminated before 1st April 2001
or which, before 1st April 2001, were determined or
dissolved by agreement or by a judgement which had
become a res judicata, or by operation of the law; in
respect of such contracts the law applicable at the
time, including the time of their termination,
determination or dissolution, shall in so far as
necessary, continue to apply.
*OJ L.189, 20-07-1990, p.17. Directive as last amended by Regulation (EC) No 1882/
2003 of the European Parliament and of the Council of 29 September 2003 adapting to
Council Decision 1999/468/EC the provisions relating to committees which assist the
Commission in the exercise of its implementing powers laid down in instruments
subject to the procedure referred to in Article 251 of the EC Treaty (OJ No. L. 284,
31-10-2003, p.1).
†OJ L.247, 21-09-2007, p.21.
(1) These regulations shall apply to active implantable
medical devices.
(2)
(a) For the purposes of these regulations, the Medicines
Authority will act as the Maltese national competent
authority in terms of the Committee referred to in
Article 6 of Directive 93/42/EEC and with respect to
the exchange of information, necessary to enable these
r e g u l a t i o n s t o b e a p p l i e d u n i f o r m l y, w i t h o t h e r
Member States and the Commission.
(b) For the purposes of these regulations, the Medicines
Authority will act as Malta’s national competent
authority with respect to exchange of experience
organised by the Commission in order to coordinate
the uniform application of these regulations.
For the purposes of these regulations, the following
definitions shall apply
"active implantable medical device" means any active medical
device which is intended to be totally or partially introduced,
surgically or medically, into the human body or by medical
intervention into a natural orifice, and which is intended to remain
after the procedure;
"active medical device" means any medical device relying for its
functioning on a source of electrical energy or any source of power
other than that directly generated by the human body or gravity;
"authorised representative" means any natural or legal person
established in the Community or Malta who, explicitly designated
by the manufacturer, acts and may be addressed by authorities and
bodies in the Community or Malta instead of the manufacturer with
regard to the latter’s obligations under these regulations;
"CE marking" means marking as set out in Schedule IV;
"clinical data" means the safety and, or performance information
that is generated from the use of a device. Clinical data are sourced
from
(a) any clinical investigation of the device concerned; or
(b) any clinical investigation or other studies reported in
the scientific literature, of a similar device for which
equivalence to the device in question can be
demonstrated; or
(c) published and, or unpublished reports on other clinical
experience of either the device in question or a similar
device for which equivalence to the device in question
can be demonstrated;
"the Community" means the European Community;
"custom-made device" means any device specifically made in
accordance with a duly qualified medical practitioner’s written
prescription which gives, under his responsibility, specific design
characteristics and is intended for the sole use of a particular
patient. Mass-produced devices which need to be adapted to meet
the specific requirements of the medical practitioner or any other
professional user shall not be considered to be custom-made
devices;
"device intended for clinical investigation" means any device
intended for use by a duly qualified medical practitioner when
conducting clinical investigations as referred to in section 2(a) of
Schedule VII in an adequate human clinical environment. For the
purpose of conducting clinical investigation, any other person who,
by virtue of his professional qualifications, is authorised to carry
out such investigation shall be accepted as equivalent to a duly
qualified medical practitioner;
"harmonised standard" means a technical specification adopted
by the European Committee for Standardisation (CEN) or the
E u r o p e a n C o m m i t t e e f o r E l e c t r o t e c h n i c a l St a n d a r d i s a t i o n
(CENELEC) or the European Telecommunications Standards
Institute (ETSI), upon a remit from the European Commission in
accordance with the Notification Procedure Regulations;
"intended purpose" means the use for which the device is
intended according to the data supplied by the manufacturer on the
labelling, in the instructions and, or in promotional material;
"manufacturer" means the natural or legal person with
responsibility for the design, manufacture, packaging and labeling
of a device before it is placed on the market under his own name,
regardless of whether these operations are carried out by that
person himself or on his behalf by a third party
(a) the obligations of these regulations to be met by
manufacturers also apply to the natural or legal person
who assembles, packages, processes, fully refurbishes
and, or labels one or more ready-made products and,
or assigns to them their intended purpose as a device
with a view to their being placed on the market under
his own name;
(b) paragraph (a) does not apply to the person who, while
not a manufacturer within the meaning of this
definition, assembles or adapts devices already on the
market to their intended purpose for an individual
patient;
"medical device" means any instrument, apparatus, appliance,
software, material or other article, whether used alone or in
combination, together with any accessories, including the software
intended by its manufacturer to be used specifically for diagnostic
and, or therapeutic purposes and necessary for its proper
application, intended by the manufacturer to be used for human
beings for the purpose of -
(a) diagnosis, prevention,
monitoring,
treatment
or
alleviation of disease,
(b) diagnosis, monitoring, treatment, alleviation of or
compensation for an injury or handicap,
(c) investigation, replacement or modification of the
anatomy or of a physiological process,
(d) control of conception,
and which does not achieve its principal intended action in or on
the human body by pharmacological, immunological or metabolic
means, but which may be assisted in its function by such means;
"Medicines Authority" means the Authority as established under
article 4 of the Medicines Act;
"national standard transposing the harmonised standard" means a
Standard issued by one of the members of CEN/CENELEC/ETSI
which is declared and noted to be identical to the Harmonised
Standard originally issued by CEN/CENELEC/ETSI;
"notified body" means a body notified to the Commission and to
the Member States of the Community, assessed according to the
criteria laid down in Schedule VIII and appointed to carry out the
procedures referred to in regulation 11(1) to (7). The names of such
bodies are published in the OJ;
"OJ" means the Official Journal of the European Union;
"placing on the market" means the first making available in
return for payment or free of charge of a device other than a device
intended for clinical investigation, with a view to distribution and,
or use on the Community or Maltese market, regardless of whether
it is new or fully refurbished;
"putting into service" means making available to the medical
profession for implantation;
(1) Where an active implantable medical device is
intended to administer a substance defined as a medicinal product
within the meaning of article 2 of the Medicines Act, that device
shall be governed by these regulations, without prejudice to the
provisions of the Medicines Act with regard to the medicinal
product.
(2)
(a) Where an active implantable medical device
incorporates, as an integral part, a substance which, if
used separately, may be considered to be a medicinal
product within the meaning of article 2 of the
Medicines Act and which is liable to act upon the
human body with action that is ancillary to that of the
device, that device shall be evaluated and authorised in
accordance with these regulations.
(b) Where a device incorporates, as an integral part, a
substance which, if used separately, may be considered
to be a medicinal product constituent or a medicinal
product derived from human blood or human plasma
within the meaning of article 2 of the Medicines Act
and which is liable to act upon the human body with
action that is ancillary to that of the device, hereinafter
referred to as a "human blood derivative", that device
shall be assessed and authorised in accordance with
these regulations.
(3) These regulations constitute a specific Directive within the
meaning of regulation 4(5) of the Electromagnetic Compatibility
Regulations.
(4)
These regulations shall not apply to
(a) medicinal products covered by the Medicines Act. In
deciding whether a product falls under that Act or
these regulations, particular account shall be taken of
the principal mode of action of the product;
(b) human blood, blood products, plasma or blood cells of
human origin or to devices which incorporate at the
time of placing on the market such blood products,
plasma or cells with the exception of devices referred
to in sub-regulation (2)(b);
(c) transplants or tissues or cells of human origin or to
products incorporating or derived from tissues or cells
of human origin, with the exception of devices
referred to in subregulation (2)(b);
(d) transplants or tissues or cells of animal origin, unless a
device is manufactured utilising animal tissue which is
rendered non-viable or non-viable products derived
from animal tissue.
(1) The active implantable medical devices referred to in
the definitions of "active implantable medical device", "custom-
made device" and "device intended for clinical investigation",
hereinafter referred to as "devices", shall satisfy the essential
requirements set out in Schedule I which apply to them, account
being taken of the intended purpose of the devices concerned.
(2) All necessary steps shall be taken to ensure that the devices
may be placed on the market and, or put into service only if they
comply with the requirements laid down in these regulations when
duly supplied, properly implanted and, or properly installed,
maintained and used in accordance with their intended purposes.
(3) Where a relevant hazard exists, devices which also satisfy
the definition of machinery in the Machinery Regulations shall also
meet the essential health and safety requirements set out in
Schedule I to the extent to which those essential health and safety
requirements are more specific than the essential requirements set
out in Schedule 1.
(4) Devices, other than those which are custom-made or
intended for clinical investigations, considered to meet the
essential requirements referred to in regulation 5(1), must bear the
CE marking of conformity.
(5)
(a) The CE marking of conformity, as shown in Schedule
IX, must appear in a visible, legible and indelible form
on the sterile pack and, where appropriate, on the sales
packaging, if any, and on the instruction leaflet.
(b) The CE marking must be followed by the
identification number of the notified body responsible
for implementation of the procedures set out in
Schedules II, IV and V.
(6)
It is prohibited to affix markings which are likely to
deceive third parties as to the meaning and form of the CE marking.
Any other marking may be affixed to the packaging or to the
instruction leaflet accompanying the device provided that the
visibility and legibility of the CE marking is not thereby reduced.
(7) Compliance with the essential requirements referred to in
sub-regulation (1) shall be presumed in respect of devices which
are in conformity with the relevant national standards adopted
pursuant to harmonised standards the reference of which have been
published in the OJ.
(8) For the purposes of these regulations, reference to
h armo ni sed sta nd ard s al so in cl ud es t he m on og rap hs of t he
European Pharmacopoeia notably on the interaction between
medicinal products and materials used in devices containing such
medicinal products, the references of which have been published in
the OJ.
(9)
(a) Where the devices are subject to other Directives
concerning other aspects and which also provide for
the affixing of the CE marking, the latter shall indicate
that the devices are also presumed to conform to the
provisions of the other Directives.
(b) However, where one or more of these Directives allow
the manufacturer, during a transitional period, to
choose which arrangements to apply, the CE marking
shall indicate conformity to the provisions only of
those Directives applied by the manufacturer. In this
case, particulars of the Directives applied, as
published in the OJ, must be given in the documents,
notices or instructions required by the Directives and
accompanying such devices; these documents, notices
or instructions shall be accessible without it being
necessary to destroy the packaging which keeps the
device sterile.
(10) Obstacles shall not be created to the placing on the market
or the putting into service within the Maltese territory of devices
complying with the provisions of these regulations and bearing the
CE marking provided for in sub-regulations (4) to (6), which
indicates that they have been the subject of an assessment of their
conformity in accordance with regulation 6(1) to (9).
(11) Obstacles shall not be created to
(a) devices intended for clinical investigations being made
available to duly qualified medical practitioners or
authorised persons for that purpose if they satisfy the
conditions laid down in regulation 7(1) to (4) and in
Schedule VI;
(b) custom-made devices being placed on the market and
put into service if they satisfy the conditions laid down
in Schedule VI and are accompanied by the statement,
which shall be available to the particular identified
patient, referred to in that Schedule.
Such devices shall not bear the CE marking.
(12) At trade fairs, exhibitions, demonstrations, etc., no
obstacles shall be created to the showing of devices which do not
conform to these regulations, provided that a visible sign clearly
indicates that such devices do not conform and cannot be marketed
or put into service until they have been made to comply by the
manufacturer or his authorised representative.
(13) When a device is put into service, the information
described in sections 14, 15, 16 and 17 of Schedule I is required to
be in, at least, Maltese and, or English.
(1) In the case of devices other than those which are
custom-made or intended for clinical investigations, the
manufacturer must, in order to affix the CE marking, at his own
choice
(a) follow the procedure relating to the EC declaration of
conformity set out in Schedule II; or
(b) follow the procedure relating to the EC type-
examination set out in Schedule III, coupled with
(i)
the procedure relating to EC verification set out
in Schedule IV, or
(ii) the procedure relating to the EC declaration of
conformity to type set out in Schedule V.
(2) In the case of custom-made devices, the manufacturer must
draw up the declaration provided for in Schedule VI before placing
each device on the market.
(3) Where appropriate, the procedures provided for in Schedule
III, IV and VI may be discharged by the manufacturer’s authorised
representative.
(4) The records and correspondence relating to the procedures
referred to in sub-regulations (1), (2) and (3) shall be in, at least,
Maltese and, or English, and, or in a language acceptable to the
notified body defined in regulation 11(1) to (7).
(5) During the conformity assessment procedure for a device,
the manufacturer and, or the notified body shall take account of the
results of any assessment and verification operations which, where
appropriate, have been carried out in accordance with these
regulations at an intermediate stage of manufacture.
(6) Where the conformity assessment procedure involves the
intervention of a notified body, the manufacturer, or his authorised
representative, may apply to a body of his choice within the
framework of the tasks for which the body has been notified.
(7) The notified body may require, where duly justified, any
information or data which is necessary for establishing and
maintaining the attestation of conformity in view of the chosen
procedure.
(8) Decisions taken by the notified bodies in accordance with
Schedules II, III and V shall be valid for a maximum of five years
and may be extended on application, made at a time agreed in the
contract signed by both parties, for further periods of a maximum
length of five years.
(9) By derogation from sub-regulations (1) and (2), the
Medicines Authority, after consultation with the Superintendent of
Public Health, may authorise, on duly justified request, the placing
on the market and putting into service, within the Maltese territory,
of individual devices for which the procedures referred to in sub-
regulations (1) and (2) have not been carried out and the use of
which is in the interest of protection of health.
(10) The Medicines Authority shall submit a duly substantiated
request to the Commission and ask it to take the necessary
measures in the following situations
(a) the Maltese authorities consider that the conformity of
a device or family of devices should be established, by
way of derogation from the provisions of sub-
regulations (1) to (9), by applying solely one of the
given procedures chosen from among those referred to
in sub-regulations (1) to (9);
(b) the Maltese authorities consider that a decision is
required as to whether a particular product or product
group falls within the definition of "active implantable
medical device", "custom-made device", "device
intended for clinical investigation" or "medical
device".
(1) In the case of devices intended for clinical
investigations, the manufacturer or the authorised representative
shall, at least sixty days before the commencement of the
investigations, submit the statement referred to in Schedule VI to
the Medicines Authority and the competent authorities in other
Member States in which the investigations are to be conducted.
(2)
(a) The manufacturer may commence the relevant clinical
investigations at the end of a period of sixty days after
notification, unless the Medicines Authority or the
Ethics Committee established by the Clinical Trials
Regulations have notified him, within that period, of a
decision to the contrary, based on considerations of
public health or public policy.
(b) However, the Medicines Authority may authorise
manufacturers to start the clinical investigations in
question before the expiry of the sixty-day period,
provided that the Ethics Committee mentioned above
has issued a favourable opinion with respect to the
investigation programme in question, including its
review of the clinical investigation plan.
(3) If necessary, any appropriate steps to ensure public health
and public policy shall be taken according to the Public Health Act.
(4) Where a clinical investigation is refused or halted by the
Maltese authorities, the Medicines Authority shall communicate
such decision and the grounds thereof to all Member States and the
Commission. Where the Medicines Authority have called for a
significant modification or temporary interruption of a clinical
investigation, the Medicines Authority shall inform the Member
States concerned about such actions and the grounds for the actions
taken.
(5) The manufacturer or his authorised representative shall notify
the Medicines Authority and the competent authorities of the Member
States concerned of the end of the clinical investigation, with a
justification in case of early termination. In the case of early
termination of the clinical investigation on safety grounds this
notification shall be communicated to all Member States and the
Commission. The manufacturer or his authorised representative shall
keep the report referred to in section 2(c)(vii) of Schedule VII at the
disposal of the Medicines Authority and the competent authorities of
the other Member States concerned.
(6) Clinical investigations shall be conducted in accordance
with the provisions of Schedule VII.
(1) Where it is found that the devices referred to in the
definitions of "active implantable medical device" and "cusom
device", correctly put into service and used in accordance with
their intended purpose, may compromise the health and, or safety
o f p a t i e n t s , u s e r s o r, w h e r e a p p l i c a b l e , o t h e r p e r s o n s , a l l
appropriate measures shall be taken by the Medicines Authority to
withdraw such devices from the market or prohibit or restrict their
being placed on the market or their being put into service.
The Medicines Authority shall immediately inform the
Commission of any such measure, indicating the reasons for such
decisions and, in particular, whether non-compliance with these
regulations is due to
(a) failure to meet the essential requirements referred to in
regulation 5(1), where the device does not meet in full
or in part the standards referred to in regulation 5(7);
(b) incorrect application of those standards;
(c) shortcomings in the standards themselves.
(2) Where a device which does not comply bears the CE marking,
the Medicines Authority shall take appropriate action against
whomsoever has affixed the mark. The Medicines Authority shall
inform the Commission and other Member States of such actions.
(3) The Medicines Authority shall take the necessary steps to
ensure that information brought to their knowledge regarding the
incidents mentioned below involving a device is recorded and
evaluated in a centralised manner
(a) any malfunction of or deterioration in the
characteristics and performances of a device, as well
as any inadequacy in the labelling or in the
instructions for use which might lead to or might have
led to the death of a patient or user or to a serious
deterioration in his state of health;
(b) any technical or medical reason in relation to the
characteristics or performances of a device for the
reasons referred to in paragraph (a), leading to
systematic recall of devices of the same type by the
manufacturer.
(4) The necessary steps shall be taken to ensure that the
manufacturer of the device concerned, or his authorised
representative, is also informed of the incident referred to in sub-
regulation (3).
(5)
(a) An assessment of the incidents mentioned in
subregulation (3) will be carried out by the Medicines
Authority, if possible together with the manufacturer or
his authorised representative.
(b) The Medicines Authority shall, without prejudice to
sub-regulations (1) and (2), immediately inform the
Commission and the other Member States of the
incidents referred to in sub-regulation (3) and of the
measures that have been taken or are contemplated to
minimise the recurrence of such incidents, including
information on the underlying incidents.
(6)
(a) Where the Medicines Authority considers, in relation to
a given product or group of products, that, in order to
ensure the protection of health and safety and, or to
ensure that public health requirements are observed,
such products should be withdrawn from the market, or
their placing on the market and putting into service
should be prohibited, restricted or subjected to
particular requirements, any necessary and justified
transitional measures may be taken.
(b) The Medicines Authority shall then inform the
Commission and all other Member States of these
transitional measures, giving the reasons for such
decisions.
(7)
Any decision taken pursuant to these regulations -
(a) to refuse or restrict the placing on the market or the
putting into service of a device or the carrying out of
clinical investigations; or
(b) to withdraw devices from the market,
shall state the exact grounds on which it is based. Such decisions
shall be notified without delay to the party concerned, who shall at
the same time be informed of the remedies available to him under
the Maltese law and of the time limits to which such remedies are
subject.
(8) In the event of a decision as referred to in sub-regulation
(7), the manufacturer or his authorised representative shall have an
opportunity to put forward his viewpoint in advance, unless such
consultation is not possible because of the urgency of the measure
to be taken.
(9)
Without prejudice to sub-regulations (1) and (2)
(a) where it is established that the CE marking has been
affixed unduly or is missing in violation of these
regulations, the manufacturer or his authorised
representative shall be obliged to end the infringement
under conditions imposed by the Medicines Authority;
(b) where non-compliance continues, all appropriate
measures shall be taken by the Medicines Authority to
restrict or prohibit the placing on the market of the
product in question or to ensure that it is withdrawn
from the market, in accordance with the procedure as
laid down in sub-regulations (1) and (2).
(10) The provisions of sub-regulation (9) shall also apply where
the CE marking has been affixed in accordance with the procedures
in these regulations, but inappropriately, on products that are not
covered by these regulations.
(1) Any manufacturer who has his registered place of
business in Malta and, under his own name, places devices on the
market in accordance with the procedure referred to in regulation
6(2) shall inform the Medicines Authority of the address of the
registered place of business and the description of the devices
concerned.
(2) All data allowing for the devices to be identified, together
with the label and instructions for use, may be requested when such
devices are put into service in Malta.
(3)
(a) Where a manufacturer who places a device on the
Maltese or Community market under his own name
does not have a registered place of business in a
Member State, he shall designate a single authorised
representative in the European Union.
(b) When the authorised representative referred to in
paragraph (a) has a registered place of business in
Malta, he shall inform the Medicines Authority of the
details referred to in sub-regulation (1).
(4) The Medicines Authority shall, on request, inform the other
Member States and the Commission of the details referred to in
sub-regulation (2) given the manufacturer or authorised
representative.
(1) Regulatory data in accordance with these regulations
shall be stored in a European databank accessible to the competent
authorities to enable them to carry out their tasks relating to these
regulations on a well-informed basis.
(2)
The databank shall contain the following
(a) data relating to certificates issued, modified,
supplemented, suspended, withdrawn or refused
according to the procedures as laid down in Schedules
II to V;
(b) data obtained in accordance with the vigilance
procedure as defined in regulation 8(3) to (5);
(c) data relating to clinical investigations referred to in
regulation 7(1) to (7).
(3)
Data shall be forwarded in a standardised format by the
Medicines Authority.
(1) The Medicines Authority shall notify the Commission
and other Member States of the bodies which it has designated for
carrying out the tasks pertaining to the procedures referred to in
regulation 6(1) to (9) together with the specific tasks for which the
bodies have been designated.
(2) The Medicines Authority shall apply the criteria set out in
Schedule VIII for the designation of bodies. Bodies that meet the
criteria laid down in the national standards which transpose the
relevant harmonised standards shall be presumed to meet the
relevant criteria.
(3) The Medicines Authority shall withdraw the notification if it
finds that the body no longer meets the criteria referred to in sub-
regulation (2). It shall immediately inform the other Member States
and the Commission thereof.
(4) The notified body and the manufacturer, or his authorised
representative, shall lay down, by common accord, the time limits
for completion of the evaluation and verification operations
referred to in Schedules II to V.
(5) The notified body shall inform the Medicines Authority
about all certificates issued, modified, supplemented, suspended,
withdrawn or refused and the other notified bodies within the scope
of these regulations about certificates suspended, withdrawn or
refused and, on request, about certificates issued
Provided that, the notified body shall also make available,
on request, all additional relevant information.
(6)
(a) Where a notified body finds that pertinent
requirements of these regulations have not been met or
are no longer met by the manufacturer or where a
certificate should not have been issued, it shall, taking
account of the principle of proportionality, suspend or
withdraw the certificate issued or place any restrictions
on it unless compliance with such requirements is
ensured by the implementation of appropriate
corrective measures by the manufacturer.
(b) In the case of suspension or withdrawal of the
certificate or of any restriction placed on it or in cases
where an intervention of the competent authority may
become necessary, the notified body shall inform the
Medicines Authority thereof. The Medicines Authority
shall inform the other Member States and the
Commission.
(7) The notified body shall, on request, supply all relevant
information and documents including budgetary documents,
required to enable the Medicines Authority to verify compliance
with the criteria laid down in Schedule VIII.
(1) Without prejudice to the existing national provisions
and practices on medical confidentiality, the Medicines Authority
shall ensure that all the parties involved in the application of these
regulations are bound to observe confidentiality with regard to all
information obtained in carrying out their tasks
Provided that the provisions of this sub-regulation do not
affect the obligation of the above authority with regard to mutual
information and the dissemination of warnings, nor the obligations
of the persons concerned to provide information under criminal
law.
(2) The following information shall not be treated as
confidential
(a) information on the registration of persons responsible
for placing devices on the market in accordance with
regulation 9(1) to (4);
(b) information to users sent out by the manufacture,
authorised representative or distributor in relation to a
measure in accordance with regulation 8(3) to (5);
(c) information contained in certificates issued, modified,
supplemented, suspended or withdrawn.
Deleted by Legal Notice 319 of 2020.
Schedule I
ESSENTIAL REQUIREMENTS
I
GENERAL REQUIREMENTS
(1) The devices must be designed and manufactured in such a way that, when
used under the conditions and for the purposes laid down, their use does not
compromise the clinical condition or the safety of patients. They must not present
any risk to the persons implanting them or, where applicable, to other persons.
(2) The devices must achieve the performances intended by the manufacturer,
namely that they are designed and manufactured in such a way that they are suitable
for one or more functions referred to in the definition of a “medical device” as
specified by him.
(3) The characteristics and performances referred to in sections 1 and 2 must
not be adversely affected to such a degree that the clinical condition and safety of
the patients or, as appropriate, of other persons are compromised during the lifetime
of the device anticipated by the manufacturer, where the device is subjected to
stresses which may occur during normal conditions of use.
(4) The devices must be designed, manufactured and packed in such a way that
their characteristics and performances are not adversely affected in the storage and
transport conditions laid down by the manufacturer (temperature, humidity, etc.).
(5) Any side effects or undesirable conditions must constitute acceptable risks
when weighed against the performances intended.
(6) Demonstration of conformity with the essential requirements must include a
clinical evaluation in accordance with Schedule VII to these regulations.
II
REQUIREMENTS REGARDING DESIGN AND CONSTRUCTION
(7) The solutions adopted by the manufacturer for the design and construction
of the devices must comply with safety principles taking account of the generally
acknowledged state of the art.
(8) Implantable devices must be designed, manufactured and packed in a non-
reusable pack according to appropriate procedures to ensure they are sterile when
placed on the market and, in the storage and transport conditions stipulated by the
manufacturer, remain so until the packaging is removed and they are implanted.
(9) Devices must be designed and manufactured in such a way as to remove or
minimise as far as possible
(a) the risk of physical injury in connection with their physical, including
dimensional, features,
(b) risks connected with the use of energy sources with particular reference,
where electricity is used, to insulation, leakage currents and overheating
of the devices,
(c) risks connected with reasonably foreseeable environmental conditions
such as magnetic fields, external electrical influences, electrostatic
discharge, pressure or variations in pressure and acceleration,
(d) risks connected with medical treatment, in particular those resulting
from the use of defibrillators or high-frequency surgical equipment,
(e) risks connected with ionising radiation from radioactive substances
included in the device, in compliance with the protection requirements
laid down in the Nuclear Safety and Radiation Protection Regulations,
and the Ionising Radiation Medical Exposure Control Regulations,
(f)
risks which may arise where maintenance and calibration are
impossible, including
(i)
excessive increase of leakage currents,
(ii)
ageing of the materials used,
(iii)
excess heat generated by the device,
(iv)
decreased accuracy of any measuring or control mechanism.
(10) The devices must be designed and manufactured in such a way as to
guarantee the characteristics and performances referred to in sections 1 to 6, with
particular attention being paid to
(a) the choice of materials used, particularly as regards toxicity aspects,
(b) mutual compatibility between the materials used and biological tissues,
cells and body fluids, account being taken of the anticipated use of the
device,
(c) compatibility of the devices with the substances they are intended to
administer,
(d) the quality of the connections, particularly in respect of safety,
(e) the reliability of the source of energy,
(f)
if appropriate, that they are leak proof,
(g) proper functioning of the programming and control system, including
software. For devices which incorporate software or which are medical
software in themselves, the software must be validated according to the
state of the art taking into account the principles of development
lifecycle, risk management, validation and verification.
(11) Where a device incorporates, as an integral part, a substance which, if used
separately, may be considered to be a medicinal product as defined in article 2 of the
Medicines Act, and which is liable to act upon the body with action ancillary to that
of the device, the quality, safety and usefulness of the substance must be verified by
analogy with the methods specified in Annex I to Directive 2001/83/EC.
(a) For the substances referred to in the first paragraph, the notified body
shall, having verified the usefulness of the substance as part of the
medical device and taking account of the intended purpose of the
device, seek a scientific opinion from one of the competent authorities
designated by the Member States or the European Medicines Agency
(EMEA) acting particularly through its committee in accordance with
Regulation (EC) No 726/2004 * on the quality and safety of the
substance including the clinical benefit and risk profile of the
incorporation of the substance into the device. When issuing its opinion,
the competent authority or the EMEA shall take into account the
manufacturing process and the data related to the usefulness of
incorporation of the substance into the device as determined by the
notified body.
(b) Where a device incorporates, as an integral part, a human blood
derivative, the notified body shall, having verified the usefulness of the
substance as part of the device and taking account of the intended
*Regulation (EC) No. 726/2004 of the European Parliament and of the Council of 31 March 2004 laying
down Community procedures for the authorisation and supervision of medicinal products for human and
veterinary use and establishing a European Medicines Agency (OJ L.136, 30-04-2004, p.1). Regulation as
last amended by Regulation (EC) No. 1901/2006.
purpose of the device, seek a scientific opinion from the EMEA, acting
particularly through its committee, on the quality and safety of the
substance, including the clinical benefit and risk profile of the
incorporation of the human blood derivative into the device. When
issuing its opinion, the EMEA shall take into account the manufacturing
process and the data related to the usefulness of incorporation of the
substance into the device as determined by the notified body.
(c) Where changes are made to an ancillary substance incorporated in a
device, in particular related to its manufacturing process, the notified
body shall be informed of the changes and shall consult the relevant
competent authority (i.e. the one involved in the initial consultation), in
order to confirm that the quality and safety of the ancillary substance
are maintained. The competent authority shall take into account the data
related to the usefulness of the incorporation of the substance into the
device as determined by the notified body, in order to ensure that the
changes have no negative impact on the established benefit and risk
profile of the addition of the substance in the device.
(d) When the relevant competent authority (i.e. the one involved in the
initial consultation) has obtained information on the ancillary
substance, which could have an impact on the established benefit and
risk profile of the addition of the substance to the device, it shall
provide the notified body with advice, whether this information has an
impact on the established benefit and risk profile of the addition of the
substance to the device or not. The notified body shall take the updated
scientific opinion into account in reconsidering its assessment of the
conformity assessment procedure.
(12) The devices and, if appropriate, their component parts must be identified to
allow any necessary measure to be taken following the discovery of a potential risk
in connection with the devices and their component parts.
(13) Devices must bear a code by which they and their manufacturer can be
unequivocally identified (particularly with regard to the type of device and year of
manufacture); it must be possible to read this code, if necessary, without the need for
a surgical operation.
(14) When a device or its accessories bear instructions required for the operation
of the device or indicate operating or adjustment parameters, by means of a visual
system, such information must be understandable to the user and, as appropriate, the
(15) Every device must bear, legibly and indelibly, the following particular,
where appropriate in the form of generally recognised symbols
(a) On the sterile pack
(i)
the method of sterilisation,
(ii)
an indication permitting this packaging to be recognised as such,
(iii)
the name and address of the manufacturer,
(iv)
a description of the device,
(v)
if the device is intended for clinical investigations, the words
"exclusively for clinical investigations",
(vi)
if the device is custom-made, the words "custom-made device",
(vii) a declaration that the implantable device is in a sterile condition,
(viii) the month and year of manufacture,
(ix)
an indication of the time limit for implanting a device safely,
(b) On the sales packaging
(i)
the name and address of the manufacturer and the name and
address of the authorised representative, where the manufacturer
does not have a registered place of business in Malta or in the
Community,
(ii) a description of the device,
(iii) the purpose of the device,
(iv) the relevant characteristics for its use,
(v) if the device is intended for clinical investigations, the words
"exclusively for clinical investigations",
(vi) if the device is custom-made, the words: "custom-made device",
(vii)
a declaration that the implantable device is in a sterile condition,
(viii)
the month and year of manufacture,
(ix)
an indication of the time limit for implanting a device safely,
(x)
the conditions for transporting and staring the device,
(xi)
in the case of device within the meaning of regulation 4(2)(a), an
indication that the device contains a human blood derivative.
(16) When placed on the market each device must be accompanied by
instructions for use giving the following particulars
(a) the year of authorisation to affix the CE mark,
(b) the details referred to in sections 15(a) and (b), with the exception of
those referred to in the eighth and ninth indents,
(c) the performances referred to in section 2 and any undesirable side
effects,
(d) information allowing the physician to select a suitable device and the
corresponding software and accessories,
(e) information constituting the instructions for use allowing the physician
and, where appropriate, the patient to use the device, its accessories and
software correctly, as well as information on the nature, scope and times
for operating controls and trials and, where appropriate, maintenance
measures,
(f)
information allowing, if appropriate, certain risks in connection with
implantation of the device to be avoided,
(g) information regarding the risks of reciprocal interference * in connection
with the presence of the device during specific investigations or
treatment,
(h) the necessary instructions in the event of the sterile pack being damaged
and, where appropriate, details of appropriate methods of re-
sterilisation,
(i)
an indication, if appropriate, that a device can be reused only if it is
reconditioned under the responsibility of the manufacturer to comply
with the essential requirements,
(j)
date of issue or the latest revision of the instructions for use.
*"Risks of reciprocal interference" means adverse effects on the device caused by instruments present at the
time of investigations or treatment, and vice versa.
(17) The instruction leaflet must also include details allowing the physician to
brief the patient on the contra-indications and the precautions to be taken. These
details should cover in particular
(a) information allowing the lifetime of the energy source to be established,
(b) precautions to be taken should changes occur in the device’s
performance,
(c) precautions to be taken as regards exposure, in reasonably foreseeable
environmental conditions, to magnetic fields, external electrical
influences, electrostatic discharge, pressure or variations in pressure,
acceleration, etc.,
(d) adequate information regarding the medicinal products which the device
in question is designed to administer.
(18) Confirmation that the device satisfies the requirements in respect of
characteristics and performances, as referred to in sections 1 to 6, in normal
conditions of use, and the evaluation of the side effects or undesirable effects must
be based on clinical data established in accordance with Schedule VII.
Schedule II
EC DECLARATION OF CONFORMITY
(Complete quality assurance system)
(1) The manufacturer shall apply the quality system approved for the design,
manufacture and final inspection of the products concerned as specified in sections 3
and 4 and shall be subject to EC surveillance as specified in section 5.
(2) The declaration of conformity is the procedure by means of which the
manufacturer who satisfies the obligations of Section 1 above ensures and declares
that the products concerned meet the provisions of these regulations which apply to
them.
The manufacturer or his authorised representative shall affix the CE marking
in accordance with regulation 5(4) to (6) and shall draw up a written declaration of
conformity.
This declaration shall cover one or more clearly identified devices by means
of product name, product code or other unambiguous reference and must be kept by
the manufacturer.
The CE marking shall be accompanied by the identification number of the
notified body responsible.
(3)
Quality system
(a) The manufacturer shall make an application for evaluation of his quality
system to a notified body.
The application shall include
(i)
all the appropriate items of information for the category of
products manufacture of which is envisaged,
(ii) the quality-system documentation,
(iii) an undertaking to fulfil the obligations arising from the quality
system as approved,
(iv)
an undertaking to maintain the approved quality system in such a
way that it remains adequate and efficacious,
(v)
an undertaking by the manufacturer to institute and keep updated
a postmarket surveillance system including the provision referred
to in Schedule VII. The undertaking shall include an obligation
for the manufacturer to notify the competent authorities of the
following incidents immediately on learning of them
1. any deterioration in the characteristics or performances, and
any inaccuracies in the instruction leaflet for a device which
might lead to or have led to the death of a patient or a
deterioration in his state of health;
2. any technical or medical reason resulting in withdrawal of a
device from the market by the manufacturer.
(b) The application of the quality system must ensure that the products
confirm to the provision of these regulations which apply to them at
every stage, from design to final controls.
All the elements, requirements and provisions adopted by the
manufacturer for his quality system shall be documented in a systematic
and orderly manner in the form of written policies and procedures. This
quality-system documentation must make possible a uniform
interpretation of the quality policies and procedures such as quality
programmes, quality plans, quality manuals and quality records. It shall
include in particular the corresponding documentation, data and records
arising from the procedures referred to in point (iii) hereunder.
It shall include in particular an adequate description of
(i) the manufacturer’s quality objectives;
(ii) the organisation of the business and in particular
1. the organisational structures, the responsibilities of the
managerial staff and their organisational authority where
quality of design and manufacture of the products is
concerned,
2. the methods of monitoring the efficient operation of the
quality system and in particular its ability to achieve the
desired quality of the design and of the products, including
control of products which do not conform,
3. where the design, manufacture and, or final inspection and
testing of the products, or elements thereof, is carried out by
a third party, the methods of monitoring the efficient
operation of the quality system and in particular the type and
extent of control applied to the third party;
(iii) the procedures for monitoring and verifying the design of the
products and in particular
1. the design specifications, including the standards which will
be applied and a description of the solutions adopted to fulfil
the essential requirements which apply to the products when
the standards referred to in regulation 5(7) are not applied in
full,
2. the techniques of control and verification of the design, the
processes and systematic actions which will be used when
the products are being designed,
3. a statement indicating whether or not the device
incorporates, as an integral part, a substance or a human
blood derivative referred to in section 11 of Schedule 1 and
the data on the tests conducted in this connection required to
assess the safety, quality and usefulness of that substance or
human blood derivative, taking account of the intended
purpose of the device,
4. the pre-clinical evaluation,
5. the clinical evaluation referred to Schedule VII;
(iv)
the techniques of control and of quality assurance at the
manufacturing stage and in particular
1. the processes and procedures which will be used, particularly
as regards sterilisation, purchasing and the relevant
documents,
2. product-identification procedures drawn up and kept up to
date from drawings, specifications or other relevant
documents at every stage of manufacture;
(v)
the appropriate tests and trials which will be effected before,
during and after production, the frequency with which they will
take place, and the test equipment used.
(c) Without prejudice to regulation 8(9), the notified body shall effect an
audit of the quality system to determine whether it meets the
requirements referred to in section 3(b). It shall presume conformity
with these requirements for the quality systems which use the
corresponding harmonised standards.
The team entrusted with the evaluation shall include at least one
member who has already had experience of evaluations of the
technology concerned. The evaluation procedure shall include an
inspection on the manufacturer's premises and, in duly substantiated
cases, on the premises of the manufacturer's suppliers and, or
subcontractors to inspect the manufacturing processes.
The decision shall be notified to the manufacturer after the final
inspection. It shall contain the conclusions of the control and a reasoned
evaluation.
(d) The manufacturer shall inform the notified body which has approved the
quality system of any plan to alter the quality system.
The notified body shall evaluate the proposed modifications and shall
verify whether the quality system so modified would meet the
requirements referred to in section 3(b); it shall notify the manufacturer
of its decision. This decision shall contain the conclusions of the control
and a reasoned evaluation.
(4)
Examination of the design of the product
(a) In addition to the obligations incumbent on him under section 3, the
manufacturer shall make an application for examination of the design
dossier relating to the product which he plans to manufacture and which
falls into the category referred to in section 3(a).
(b) The application shall describe the design, manufacture, and
performances of the product in question, and it must include the
documents needed to assess whether the product conforms to the
requirements of these regulations, and in particular section 3(b)(iii) and
(iv).
It shall include inter alia
(i)
the design specifications, including the standards which have
been applied,
(ii) the necessary proof of their appropriations, in particular where
the standards referred to in regulation 5(7) have not been applied
in full. This proof must include the results of the appropriate tests
carried out by the manufacturer or carried out under his
responsibility,
(iii) a statement as to whether or not the device incorporates, as an
integral part, a substance as referred to in section 11 of Schedule
I, whose action in combination with the device may result in its
bioavailability, together with data on the relevant trials
conducted,
(iv) the clinical evaluation referred to in Schedule VII,
(v) the draft instruction leaflet.
(c) The notified body shall examine the application and, where the product
complies with the relevant provisions of these regulations, shall issue
the applicant with an EC design examination certificate. The notified
body may require the application to be supplemented by further tests or
proof so that compliance with the requirements of the regulations may
be evaluated. The certificate shall contain the conclusions of the
examination, the conditions of its validity, the data needed for
identification of the approved design and, where appropriate, a
description of the intended use of the product.
In the case of devices referred to in section 11(a) of Schedule I, the
notified body shall, as regards the aspects referred to in that section,
consult one of the competent authorities designated by the Member
States in accordance with Directive 2001/83/EC or the EMEA before
taking a decision. The opinion of the competent national authority or
the EMEA shall be drawn up within 210 days after receipt of a valid
documentation. The scientific opinion of the competent national
authority or the EMEA must be included in the documentation
concerning the device. The notified body will give due consideration to
the views expressed in this consultation when making its decision. It
will convey its final decision to the competent body concerned.
In the case of devices referred to in section 11(b) of Schedule I, the
scientific opinion of the EMEA must be included in the documentation
concerning the device. The opinion shall be drawn up within 210 days
after receipt of a valid documentation. The notified body will give due
consideration to the opinion of the EMEA when making its decision.
The notified body may not deliver the certificate if the EMEA's
scientific opinion is unfavourable. It will convey its final decision to the
EMEA.
(d) The applicant shall inform the notified body which issued the EC design
examination certificate of any modification made to the approved
design. Modifications made to the approved design must obtain
supplementary approval from the notified body which issued the EC
design examination certificate where such modifications may affect
conformity with the essential requirements of these regulations or the
conditions prescribed for the use of the product. This supplementary
approval shall be given in the form of an addendum to the EC design
examination certificate.
(5)
Surveillance
(a) The aim of surveillance is to ensure that the manufacturer duly fulfils
the obligations arising from the approved quality system.
(b) The manufacturer shall authorise the notified body to carry out all
necessary inspections and shall supply it with all appropriate
information, in particular
(i) the quality-system documentation,
(ii) the data stipulated in the part of the quality system relating to
design, such as the results of analyses, calculations, tests,
preclinical and clinical evaluation, post-market clinical follow-up
plan and the results of the post-market clinical follow up, if
applicable, etc.,
(iii) the data stipulated in the part of the quality system relating to
manufacture, such as reports concerning inspections, tests,
standardisations, calibrations and the qualifications of the staff
concerned, etc.
(c) The notified body shall periodically carry out appropriate inspections
and evaluations in order to ascertain that the manufacturer is applying
the approved quality system, and shall supply the manufacturer with an
evaluation report.
(d) In addition, the notified body may make unannounced visits to the
manufacturer, and shall supply him with an inspection report.
(6)
Administrative provisions
(a) For at least fifteen years from the last date of manufacture of the
product, the manufacturer or his authorised representative shall keep
available for the Regulatory Affairs Directorate and the Market
Surveillance Directorate
(i) the declaration of conformity,
(ii) the documentation referred to in section 3(a)(ii), and in particular
the documentation, data and records referred to in the second
paragraph of section 3(b),
(iii) the amendments referred to in section 3(d),
(iv) the documentation referred to in section 4(b),
(v) the decisions and reports of the notified body referred to in
sections 3(d), 4(c), 5(c) and 5(d).
(b) On request, the notified body shall make available to the other notified
bodies and Regulatory Affairs Directorate and the Market Surveillance
Directorate all relevant information on approvals of quality systems
issued, refused or withdrawn.
(7)
Application to devices referred to in regulation 4(2)(b)
Upon completing the manufacture of each batch of devices referred to in
regulation 4(2)(b), the manufacturer shall inform the notified body of the release of
the batch of devices and send to it the official certificate concerning the release of
the batch of human blood derivative used in the device, issued by a State laboratory
or a laboratory designated for that purpose by a Member State in accordance with
article 104A(4) of the Medicines Act.
Schedule III
EC TYPE-EXAMINATION
(1) EC type-examination is the procedure whereby a notified body observes and
certifies that a representative sample of the production envisaged satisfies the
relevant provisions of these regulations.
(2) The application for EC type-examination shall be made by the manufacturer,
or by his authorised representative, to a notified body.
The application shall include
(a) the name and address of the manufacturer and the name and address of
the authorised representative if the application is made by the latter,
(b) a written declaration specifying that an application has not been made to
any other notified body,
(c) the documentation described in section 3 needed to allow an evaluation
to be made of the conformity of a representative sample of the
production in question, hereinafter referred to as "type", with the
requirements of these regulations.
The applicant shall make a "type" available to the notified body. The
notified body may request other samples as necessary.
(3) The documentation must make it possible to understand the design, the
manufacture and the performances of the product. The documentation shall contain
the following items in particular
(a) a general description of the type, including any variants planned, and its
intended use,
(b) design drawings, methods of manufacture envisaged, in particular as
regards sterilisation, and diagrams of parts, sub-assemblies, circuits,
etc.,
(c) the descriptions and explanations necessary for the understanding of the
abovementioned drawings and diagrams and of the operation of the
product,
(d) a list of standards referred to in regulation 5(7), applied in full or in
part, and a description of the solutions adopted to satisfy the essential
requirements where the standards referred to in regulation 5(7) have not
been applied,
(e) the results of design calculations, risk analysis, investigations and
technical tests carried out, etc.,
(f)
a declaration stating whether or not the device incorporates, as an
integral part, a substance or a human blood derivative as referred to in
section 11 of Schedule I and the data on the tests conducted in this
connection required to assess the safety, quality and usefulness of that
substance or human blood derivative, taking account of the intended
purpose of the device,
(g) the pre-clinical evaluation,
(h) the clinical evaluation referred to in Schedule VII,
(i)
the draft instruction leaflet.
(4)
The notified body shall
(a) examine and evaluate the documentation, verify that the type has been
manufactured in accordance with that documentation; it shall also
record the items which have been designed in accordance with the
applicable provisions of the standards referred to in regulation 5(7), as
well as the items for which the design is not based on the relevant
provisions of the said standards;
(b) carry out or have carried out the appropriate inspections and the tests
necessary to verify whether the solutions adopted by the manufacturer
satisfy the essential requirements of these regulations where the
standards referred to in regulation 5(7) have not been applied;
(c) carry out or have carried out the appropriate inspections and the tests
necessary to verify whether, where the manufacturer has chosen to
apply the relevant standards, these have actually been applied;
(d) agree with the applicant on the place where the necessary inspections
and tests will be carried out.
(5) Where the type meets the provisions of these regulations, the notified body
shall issue an EC type-examination certificate to the applicant. The certificate shall
contain the name and address of the manufacturer, the conclusions of the control, the
conditions under which the certificate is valid and the information necessary for
identification of the type approved.
The significant parts of the documentation shall be attached to the certificate
and a copy shall be kept by the notified body.
In the case of devices referred to in the second paragraph of section 11 of
Schedule I, the notified body shall, as regards the aspects referred to in that section,
consult one of the competent authorities designated by the Member States in
accordance with Directive 2001/83/EC or the EMEA before taking a decision. The
opinion of the competent national authority or the EMEA shall be drawn up within
210 days after receipt of a valid documentation. The scientific opinion of the
competent national authority or the EMEA must be included in the documentation
concerning the device. The notified body will give due consideration to the views
expressed in this consultation when making its decision. It will convey its final
decision to the competent body concerned.
In the case of devices referred to in the third paragraph, section 11 of
Schedule I, the scientific opinion of the EMEA must be included in the
documentation concerning the device. The opinion shall be drawn up within 210
days after receipt of a valid documentation. The notified body will give due
consideration to the opinion of the EMEA when making its decision. The notified
body may not deliver the certificate if the EMEA's scientific opinion is
unfavourable. It will convey its final decision to the EMEA.
(6) The applicant shall inform the notified body which issued the EC type-
examination certificate of any modification made to the approved product.
Modifications to the approved product must receive further approval from
the notified body which issued the EC type-examination certificate where such
modifications may affect conformity with the essential requirements or with the
conditions of use specified for the product. This new approval shall be issued, where
appropriate, in the form of a supplement to the initial EC type-examination
certificate.
(7)
Administrative provisions
(a) On request, each notified body shall make available to the other notified
bodies and the competent authority, all relevant information on EC
type-examination
certificates
and
addenda
issued,
refused
or
withdrawn.
(b) Other notified bodies may obtain a copy of the EC type-examination
certificates and, or the addenda to them. The annexes to the certificates
shall be made available to the other notified bodies when a reasoned
application is made and after the manufacturer has been informed.
(c) The manufacturer or his authorised representative shall keep with the
technical documentation a copy of the EC type-examination certificates
and the supplements to them for a period of at least 15 years from the
manufacture of the last product.
Schedule IV
Amended by
EC VERIFICATION
L.N. 319 of 2020.
(1) EC verification is the procedure whereby the manufacturer or his authorised
representative ensures and declares that the products subject to the provisions of
section 3 below are in conformity with the type as described in the EC type-
examination certification and satisfy the requirements of these regulations that apply
to them.
(2) The manufacturer or his authorised representative shall take all measures
necessary in order that the manufacturing process ensures conformity of the products
to the type as described in the EC type-examination certification and to the
requirements of these regulations that apply to them. The manufacturer or his
authorised representative shall affix the CE marking to each product and draw up a
written declaration of conformity.
(3) The manufacturer shall, before the start of manufacture, prepare documents
defining the manufacturing processes, in particular as regards sterilisation, together
with all the routine, pre-established provisions to be implemented to ensure
uniformity of production and conformity of the products with the type as described
in the EC type-examination certificate as well as with the relevant requirements of
these regulations.
(4) The manufacturer shall undertake to institute and keep updated a post-
marketing surveillance system including the provisions referred to in Schedule VII.
This undertaking shall include the obligation on the part of the manufacturer to
notify the Medicines Authority of the following events immediately on learning of
them
(a) any change in the characteristics or performances and any inaccuracies
in the instruction leaflet for a device which might lead to or have led to
the death of a patient or deterioration in his state of health;
(b) any technical or medical reason resulting in the withdrawal of a device
from the market by a manufacturer.
(5) The notified body shall carry out the appropriate examinations and tests in
order to check the conformity of the product to the requirements of these regulations
by examination and testing of products on a statistical basis, as specified in section
6. The manufacturer must authorise the notified body to evaluate the efficiency of
the measures taken pursuant to section 3, by audit where appropriate.
(6)
Statistical verification
(a) Manufacturers shall present the products manufactured in the form of
uniform batches and shall take all necessary measures in order that the
manufacturing process ensures the uniformity of each batch produced.
(b) A random sample shall be taken from each batch. Products in a sample
shall be individually examined and appropriate tests, as set out in any
standard referred to in regulation 5(7), or equivalent tests shall be
carried out to verify their conformity to the type as described in the EC
type-examination certificate and thereby determine whether a batch is
to be accepted or rejected.
(c) Statistical control of products will be based on attributes and, or
variables, entailing sampling schemes with operational characteristics
which ensure a high level of safety and performance according to the
state of the art. The sampling schemes will be established by the
harmonised standards referred to in regulation 5(7), taking account of
the specific nature of the product categories in question.
(d) Where batches are accepted, the notified body shall affix, or cause to be
affixed, its identification number to each product and draw up a written
certificate of conformity relating to the tests carried out. All products in
the batch may be placed on the market except for those products from
the sample which were found not to be in conformity.
Where a batch is rejected, the notified body shall take appropriate
measures to prevent the placing on the market of that batch. In the event
of frequent rejection of batches the notified body may suspend the
statistical verification.
The manufacturer may, under the responsibility of the notified body,
affix the latter's identification number during the manufacturing
process.
(e) The manufacturer or his authorised representative shall ensure that he is
able to supply the notified body’s certificates of conformity on request.
(7)
Application to the devices referred to in regulation 4(2)(b)
Upon completing the manufacture of each batch of devices referred to in
regulation 4(2)(b), the manufacturer shall inform the notified body of the release of
the batch of devices and send to it the official certificate concerning the release of
the batch of human blood derivative used in the device, issued by a State laboratory
or a laboratory designated for that purpose by a Member State in accordance with
article 104A(4) of the Medicines Act.
Schedule V
Amended by
L.N. 319 of 2020.
EC DECLARATION OF CONFORMITY TO TYPE
(Assurance of production quality)
(1) The manufacturer shall apply the quality system approved for the
manufacture and shall conduct the final inspection of the products concerned as
specified in section 3; he shall be subject to the surveillance referred to in section 4.
(2) This declaration of conformity is the procedural element whereby the
manufacturer who satisfies the obligations of section 1 guarantees and declares that
the products concerned conform to the type described in the EC type-examination
certificate and meet the provisions of the regulations which apply to them.
(a) The manufacturer or his authorised representative shall affix the CE
marking in accordance with regulation 5(4) to (6) and draw up a written
declaration of conformity. This declaration shall cover one or more
devices manufactured, clearly identified by means of product name,
product code or other unambiguous reference and must be kept by the
manufacturer. The CE marking shall be accompanied by the
identification number of the notified body responsible.
(3)
Quality system
(a) The manufacturer shall make an application for evaluation of his quality
system to a notified body.
The application shall include
(i)
all appropriate information concerning the products which it is
intended to manufacture,
(ii) the quality-system documentation,
(iii) an undertaking to fulfil the obligations arising from the quality
system as approved,
(iv) an undertaking to maintain the approved quality system in such a
way that it remains adequate and efficacious,
(v) where appropriate, the technical documentation relating to the
approved type and a copy of the EC type-examination certificate,
(vi) an undertaking by the manufacturer to institute and keep up-dated
a post-marketing surveillance system including the provisions
referred to in Schedule VII. The undertaking shall include an
obligation for the manufacturer to notify the Medicines Authority
of the following incidents immediately on learning of them
1. any deterioration in the characteristics or performances, and
any inaccuracies in the instruction leaflet for a device which
might lead to or have led to the death of a patient or a
deterioration in his state of health;
2. any technical or medical reason resulting in withdrawal of a
device from the market by the manufacturer.
(b) Application of the quality system must ensure that the products conform
to the type described in the EC type-examination certificate.
All the elements, requirements and provisions adopted by the
manufacturer for his quality system shall be documented in a systematic
and orderly manner in the form of written policies and procedures. This
quality-system documentation must make possible a uniform
interpretation of the quality policies and procedures such as quality
programmes, quality plans, quality manuals and quality records.
It shall include in particular an adequate description of
(i) the manufacturer’s quality objectives;
(ii) the organisation of the business and in particular
1. the organisational structures, the responsibilities of the
managerial staff and their organisational authority where
manufacture of the products is concerned,
2. the methods of monitoring the efficient operation of the
quality system and in particular its ability to achieve the
desired quality of the products, including control of products
which do not conform,
3. where the manufacture and, or final inspection and testing of
the products, or elements thereof, is carried out by a third
party, the methods of monitoring the efficient operation of
the quality system and in particular the type and extent of
control applied to the third party;
(iii) the techniques of control and of quality assurance at the
manufacturing stage and in particular
1. the processes and procedures which will be used, particularly
as regards sterilisation, purchasing and the relevant
documents,
2. product identification procedures drawn up and kept up-to-
date from drawings, specifications or other relevant
documents at every stage of manufacture;
(iv)
the appropriate tests and trials which will be effected before,
during and after production, the frequency with which they will
take place, and the test equipment used.
(c) Without prejudice to regulation 8(9), the notified body shall effect an
audit of the quality system to determine whether it meets the
requirements referred to in section 3(b). It shall presume conformity
with these requirements for the quality systems which use the
corresponding harmonised standards.
The team entrusted with the evaluation shall include at least one
member who has already had experience of evaluations of the
technology concerned. The evaluation procedure shall include an
inspection on the manufacturer's premises.
The decision shall be notified to the manufacturer after the final
inspection. It shall contain the conclusions of the control and a reasoned
evaluation.
(d) The manufacturer shall inform the notified body which has approved the
quality system of any plan to alter that system.
The notified body shall evaluate the proposed modifications and shall
verify whether the quality system so modified would meet the
requirements referred to in section 3(b); it shall notify the manufacturer
of its decision. This decision shall contain the conclusions of the control
and a reasoned evaluation.
(4)
Surveillance
(a) The aim of surveillance is to ensure that the manufacturer duly fulfils
the obligations arising from the approved quality system.
(b) The manufacturer shall authorise the notified body to carry out all
necessary inspections and shall supply it with all appropriate
information, in particular
(i) the quality-system documentation,
(ii) the technical documentation,
(iii) the data stipulated in the part of the quality system relating to
manufacture, such as reports concerning inspections, tests,
standardisations, calibrations and the qualifications of the staff
concerned, etc.
(c) The notified body shall periodically carry out appropriate inspections
and evaluations in order to ascertain that the manufacturer is applying
the approved quality system, and shall supply the manufacturer with an
evaluation report.
(d) In addition, the notified body may make unannounced visits to the
manufacturer, and shall supply him with an inspection report.
(5) The notified body shall communicate to the other notified bodies all relevant
information concerning approvals of quality systems issued, refused or withdrawn.
(6)
Application to the devices referred to in regulation 4(2)(b)
Upon completing the manufacture of each batch of devices referred to in
regulation 4(2)(b), the manufacturer shall inform the notified body of the release of
the batch of devices and send to it the official certificate concerning the release of
the batch of human blood derivative used in the device, issued by a State laboratory
or a laboratory designated for that purpose by a Member State in accordance with
article 104A(4) of the Medicines Act.
Schedule VI
Amended by
STATEMENT CONCERNING DEVICES INTENDED FOR
L.N. 319 of 2020.
SPECIAL PURPOSES
(1) The manufacturer or his authorised representative shall draw up for custom-
made devices or for devices intended for clinical investigations the statement
comprising the elements stipulated in section 2.
(2)
The statement must contain the following information
(a) for custom-made devices
(i) the name and address of the manufacturer,
(ii) the information necessary for the identification of the product in
question,
(iii) the name of the duly qualified medical practitioner who drew up
the prescription and, if applicable, the name of the clinic
concerned,
(iv)
the specific characteristics of the product revealed by the
prescription,
(v)
a statement affirming that the device complies with the essential
requirements given in Schedule I and, where applicable,
indicating which essential requirements have not been wholly
met, together with the grounds;
(b) for devices intended for the clinical investigations covered in Schedule
(i)
data allowing the devices in question to be identified,
(ii)
the clinical investigation plan,
(iii)
the investigator’s brochure,
(iv)
the confirmation of insurance of subjects,
(v)
the documents used to obtain informed consent,
(vi)
a statement indicating whether or not the device incorporates, as
an integral part, a substance or human blood derivative referred to
in section 11 of Schedule I,
(vii)
a statement indicating whether or not the device is manufactured
utilising tissues of animal origin as referred to in the Medical
Devices Regulations,
(viii) the opinion of the Ethics Committee concerned and details of the
aspects covered by its opinion,
(ix) the name of the duly qualified medical practitioner or other
authorised person and of the institution responsible for the
investigations,
(x) the place, date of commencement and duration scheduled for the
investigations,
(xi) a statement affirming that the device in question complies with
the essential requirements apart from the aspects constituting the
object of the investigations and that, with regard to these aspects,
every precaution has been taken to protect the health and safety of
the patient.
(3) The manufacturer must also undertake to keep available for the Medicines
(a) for custom-made devices, documentation, indicating manufacturing
sites and enabling the design, manufacture and performances of the
product, including the expected performances, to be understood, so as to
allow conformity with the requirements of these regulations to be
assessed.
The manufacturer shall take all necessary measures to see that the
manufacturing process ensures that the products manufactured conform
to the documentation referred to in the first paragraph;
(b) for devices intended for clinical investigations, the documentation shall
also contain
(i) a general description of the product and its intended use,
(ii) design drawings, manufacturing methods, in particular as regards
sterilisation, and diagrams of parts, sub-assemblies, circuits, etc.,
(iii) the descriptions and explanations necessary for the understanding
of the said drawings and diagrams and the operation of the
product,
(iv) the results of the risk analysis and a list of the standards referred
to in regulation 5(7), applied in full or in part, and descriptions of
the solutions adopted to satisfy the essential requirements of these
regulations where the standards referred to in regulation 5(7) have
not been applied,
(v) if the device incorporates, as an integral part, a substance or
human blood derivative referred to in section 11 of Schedule I,
the data on the tests conducted in this connection which are
required to assess the safety, quality and usefulness of that
substance or human blood derivative, taking account of the
intended purpose of the device,
(vi) the results of the design calculations, checks and technical tests
carried out, etc.
The manufacturer shall take all the measures necessary to see that the
manufacturing process ensures that the products manufactured conform
to the documentation referred to in section 3(a) and in the first
paragraph of this section.
The manufacturer may authorise the evaluation, by audit where
necessary, of the effectiveness of these measures.
(4) The information included in the declarations covered by this Schedule shall
be kept for a period of at least 15 years from the data of manufacture of the last
(5) For custom-made devices, the manufacturer must undertake to review and
document experience gained in the post-production phase, including the provisions
referred to in Schedule VII, and to implement appropriate means to apply any
necessary corrective action. This undertaking must include an obligation for the
manufacturer to notify the Medicines Authority of the following incidents
immediately on learning of them and the relevant corrective actions
(a) any malfunction or deterioration in the characteristics and, or
performance of a device, as well as any inadequacy in the labelling or
the instructions for use which might lead to or might have led to the
death of a patient or user or to a serious deterioration in his state of
health;
(b) any technical or medical reason connected with the characteristics or
performance of a device for the reasons referred to in paragraph (a)
leading to systematic recall of devices of the same type by the
manufacturer.
Schedule VII
Amended by
L.N. 319 of 2020.
CLINICAL EVALUATION
(1)
General provisions
(a) As a general rule, confirmation of conformity with the requirements
concerning the characteristics and performances referred to in sections
1 and 2 of Schedule I under the normal conditions of use of the device
and the evaluation of the side-effects and of the acceptability of the
benefit and risk ratio referred to in section 5 of Schedule I, must be
based on clinical data. The evaluation of this data, hereafter referred to
as clinical evaluation, where appropriate taking account of any relevant
harmonised standards, must follow a defined and methodologically
sound procedure based on
(i)
either a critical evaluation of the relevant scientific literature
currently available relating to the safety, performance, design
characteristics and intended purpose of the device where
1. there is demonstration of equivalence of the device to the
device to which the data relates and,
2. the data adequately demonstrate compliance with the
relevant essential requirements;
(ii) or a critical evaluation of the results of all the clinical
investigations made,
(iii) or a critical evaluation of the combined clinical data provided in
section 1(a)(i) and (ii).
(b) Clinical investigations shall be performed unless it is duly justified to
rely on existing clinical data.
(c) The clinical evaluation and its outcome shall be documented. This
documentation shall be included and, or fully referenced in the
technical documentation of the device.
(d) The clinical evaluation and its documentation must be actively updated
with data obtained from the post-market surveillance. Where post-
market clinical follow-up as part of the post-market surveillance plan
for the device is not deemed necessary, this must be duly justified and
documented.
(e) Where demonstration of conformity with essential requirements based
on clinical data is not deemed appropriate, adequate justification for any
such exclusion has to be given based on risk management output and
under consideration of the specifics of the device-body interaction, the
clinical performances intended and the claims of the manufacturer.
Adequacy of demonstration of conformity with the essential
requirements by performance evaluation, bench testing and preclinical
evaluation alone has to be duly substantiated.
(f)
All data must remain confidential unless it is deemed essential that they
be divulged.
(2)
Clinical investigations
(a) Purpose
The purpose of a clinical investigation is to
(i)
verify that, under normal conditions of use, the performance of
the device comply with those indicated to in section 2 of Schedule
I,
(ii) determine any undesirable side-effects, under normal conditions
of use, and assess whether they constitute risks when weighed
against the intended performance of the device.
(b) Ethical consideration
Clinical investigations shall be made in accordance with the Declaration
of Helsinki approved by the 18th World Medical Assembly in Helsinki,
Finland, in 1964, and amended by the 29th World Medical Assembly in
Tokyo, Japan, in 1975 and the 35th World Medical Assembly in Venice,
Italy, in 1983. It is mandatory that all measures relating to the
protection of human subjects are carried out in the spirit of the
Declaration of Helsinki. This includes every step in the clinical
investigation from first consideration of need and justification of the
study to publication of results.
(c) Methods
(i)
Clinical investigations shall be performed according to an
appropriate state of the art plan of investigation defined in such a
way as to confirm or refute the manufacturer’s claims for the
device; the investigations shall include an adequate number of
observations to guarantee the scientific validity of the
conclusions.
(ii) The procedures utilised to perform the investigations shall be
appropriate to the device under examination.
(iii) Clinical investigations shall be performed in circumstances
equivalent to those which would be found in normal conditions of
use of the device.
(iv) All the appropriate features, including those involving the safety
and performances of the device, and its effect on patients, shall be
examined.
(v) All serious adverse events must be fully recorded and
immediately notified to the Medicines Authority and all competent
authorities of the Member States in which the clinical
investigation is being performed.
(vi) The investigations shall be performed under the responsibility of
a duly qualified medical practitioner or authorised person, in an
appropriate environment.
The medical practitioner or authorised person shall have access to
the technical data regarding the device.
(vii)
The written report, signed by the responsible medical practitioner
or authorised person, shall comprise a critical evaluation of all the
data collected during the clinical investigation.
Schedule VIII
Amended by
L.N. 319 of 2020.
MINIMUM CRITERIA TO BE MET WHEN DESIGNATING
INSPECTION BODIES TO BE NOTIFIED
(1) Notified bodies shall be subject to the Method for Designating Conformity
Assessment Bodies Regulations, and they shall also fulfil the criteria set out below.
(2) The body, its director and the staff responsible for carrying out the
evaluation and verification operations shall not be the designer, manufacturer,
supplier or installer of devices which they control, nor the authorised representative
of any of those parties. They may not become directly involved in the design,
construction, marketing or maintenance of the devices, nor represent the parties
engaged in these activities. This does not preclude the possibility of exchanges of
technical information between the manufacturer and the body.
(3) The body and its staff must carry out the evaluation and verification
operations with the highest degree of professional integrity and technical
competence and must be free from all pressures and inducements, particularly
financial, which might influence their judgement or the results of the inspection,
especially from persons or groups of persons with an interest in the results of
verifications.
(4) The body must be able to carry out all the tasks in one of Schedules II to V
assigned to such a body and for which it has been notified, whether those tasks are
carried out by the body itself or under its responsibility. In particular, it must have at
its disposal the necessary staff and possess the necessary facilities to enable it to
perform properly the technical and administrative tasks connected with evaluation
and verification; it must also have access to the equipment necessary for the
verifications required.
(5)
The staff responsible for control operations must have
(a) sound vocational training covering all the evaluation and verification
operations for which the body has been designated,
(b) satisfactory knowledge of the requirements of the controls they carry
out and adequate experience of such operations,
(c) the ability required to draw up the certificates, records and reports to
demonstrate that the controls have been carried out.
(6) The impartiality of inspection staff must be guaranteed. Their remuneration
must not depend on the number of controls carried out, nor on the results of such
controls.
(7) The body must take out liability insurance unless liability is assumed by the
State in accordance with Maltese law, or the Medicines Authority itself are directly
responsible for controls.
(8) The staff of the body is bound to observe professional secrecy with regard to
all information gained in carrying out their tasks (except vis-à-vis the Medicines
Authority or other competent administrative authorities of the State in which their
activities are carried out) under these regulations or any provision of national law
giving effect to it.
Schedule IX
CE CONFORMITY MARKING
The CE conformity marking shall consist of the initials “CE” taking the following
For the avoidance of doubt, it is hereby declared that the grid providing the
background in the above graduated drawing is not part of the CE marking.
If the CE marking is reduced or enlarged the proportions given in the
above graduated drawing must be respected.
The various components of the CE marking must have substantially the
same vertical dimension, which may not be less than 5mm.
This minimum dimension may be waived for small-scale devices.