Non-Automatic Weighing Instruments, 2016
Table of Provisions (45)
Statutory Text (45 Articles)
Suppliment tal-Gazzetta tal-Gvern ta’ Malta, Nru. 19,556, 12 ta’ April, 2016
L.N. 123 of 2016
METROLOGY ACT
(CAP. 454)
Non-Automatic Weighing Instruments, 2016
In exercise of the powers conferred by article 19 of the
Metrology Act, the Minister for Social Dialogue, Consumer Affairs
and Civil Liberties, acting on the advice of the Malta Competition and
Consumer affairs Authority, has made the following regulations:-
(1) The title of these regulations is the Non–Automatic Citation and
commencement.
Weighing Instruments, 2016.
(2) These regulations shall come into force on the 20th April,
2016.
ESSENTIAL REQUIREMENTS
The terminology used is that of the International Organisation of Legal
Metrology.
Preliminary observation
Where an instrument includes, or is connected to, more than one indicating or
printing device used for the applications listed in paragraphs (a) to (f) of
regulation 2(3), those devices which repeat the results of the weighing
operation and which cannot influence the correct functioning of the
instrument shall not be subject to the essential requirements if the weighing
results are printed or recorded correctly and indelibly by a part of the
instrument which meets the essential requirements and the results are
accessible to both parties concerned by the measurement. However, in the
case of instruments used for direct sales to the public, display and printing
devices for the vendor and the customer must fulfil the essential
requirements.
Metrological requirements
1. Units of mass
The units of mass used shall be the legal units within the meaning of Council
Directive 80/181/EEC of 20 December 1979 on the approximation of the
laws of the Member States relating to units of measurement. *
Subject to compliance with this condition, the following units are permitted
(a) SI units: kilogram, microgram, milligram, gram, tonne;
(b) imperial unit: troy ounce, if weighing precious metals;
(c) other non-SI unit: metric carat, if weighing precious stones.
For instruments that make use of the imperial unit of mass referred to above,
the relevant essential requirements specified below shall be converted to that
unit, using simple interpolation.
2. Accuracy classes
(1)
The following accuracy classes have been defined
(a) I special
* OJ L 39, 15.2.1980, p. 40.
(b) II high
(c) III medium
(d) IIII ordinary
The specifications of these classes are given in Table 1 The minimum
capacity is reduced to 5 e for instruments in classes II and III for determining
a conveying tariff.
(2)
Scale intervals
(2.1) The actual scale interval (d) and the verification scale interval (e) shall be in
the form: 1 × 10 k , 2 × 10 k , or 5 × 10 k mass units,
k being any integer or zero.
(2.2) For all instruments other than those with auxiliary indicating devices
d = e.
(2.3) For instruments with auxiliary indicating devices the following conditions
apply
e = 1 × 10 k g;
d < e ≤ 10 d.
Those conditions do not apply for instruments of class I with d < 10 –4 g, for
which e =10 –3 g.
3. Classification
Instruments with one weighing range
Instruments equipped with an auxiliary indicating device shall belong to class
I or class II. For these instruments the minimum capacity lower limits for
these two classes are obtained from Table 1 by replacement in column 3 of
the verification scale interval (e) by the actual scale interval (d).
If d < 10 –4 g, the maximum capacity of class I may be less than 50 000 e.
(2)
Instruments with multiple weighing ranges
Multiple weighing ranges are permitted, provided they are clearly indicated
on the instrument. Each individual weighing range is classified according to
point 3.1. If the weighing ranges fall into different accuracy classes the
instrument shall comply with the severest of the requirements that apply for
the accuracy classes in which the weighing ranges fall.
(3)
Multi – interval instruments
(3.1) Instruments with one weighing range may have several partial weighing
ranges (multi-interval instruments).
Multi-interval instruments shall not be equipped with an auxiliary indicating
device.
(3.2) Each partial weighing range i of multi-interval instruments is defined by
(3.3) The partial weighing ranges are classified according to Table 2. All partial
weighing ranges shall fall into the same accuracy class, that class being the
instrument’s accuracy class.
4. Accuracy
(1)
On implementation of the procedures laid down in regulation 14, the error of
indication shall not exceed the maximum permissible error of indication as
shown in Table 3. In the case of digital indication the error of indication shall
be corrected for the rounding error.
The maximum permissible errors apply to the net value and tare value for all
possible loads, excluding preset tare values.
(2)
The maximum permissible errors in service are twice the maximum
permissible errors fixed in Section 4.1.
5. Weighing results of an instrument shall be repeatable, and shall be
reproducible by the other indicating devices used and in accordance with
other methods of balancing used.
The weighing results shall be sufficiently insensitive to changes in the
position of the load on the load receptor.
6. The instrument shall react to small variations in the load.
7. Influence quantities and time
(1)
Instruments of classes II, III and IIII, liable to be used in a tilted position,
shall be sufficiently insensitive to the degree of tilting that can occur in
normal use.
(2)
The instruments shall meet the metrological requirements within the
temperature range specified by the manufacturer. The value of this range
shall be at least equal to
(a) 5°C for an instrument in class I;
(b) 15°C for an instrument in class II;
(c) 30°C for an instrument in class III or IIII.
In the absence of a manufacturer’s specification, the temperature range of –
10°C to + 40°C applies .
(3)
Instruments operated from a mains power supply shall meet the metrological
requirements under conditions of power supply within the limits of normal
fluctuation.
Instruments operated from battery power shall indicate whenever the voltage
drops below the minimum required value and shall under those circumstances
either continue to function correctly or be automatically put out of service.
(4)
Electronic instruments, except those in class I and in class II if e is less than 1
g, shall meet the metrological requirements under conditions of high relative
humidity at the upper limit of their temperature range.
(5)
Loading an instrument in class II, III or IIII for a prolonged period of time
shall have a negligible influence on the indication at load or on the zero
indication immediately after removal of the load.
(6)
Under other conditions the instruments shall either continue to function
correctly or be automatically put out of service.
Design and construction
8. General requirements
(1)
Design and construction of the instruments shall be such that the instruments
will preserve their metrological qualities when properly used and installed
and when used in an environment for which they are intended. The value of
the mass must be indicated.
(2)
When exposed to disturbances, electronic instruments shall not display the
effects of significant faults, or shall automatically detect and indicate them.
Upon automatic detection of a significant fault, electronic instruments shall
provide a visual or audible alarm that shall continue until the user takes
corrective action or the fault disappears.
(3)
The requirements of points 8.1 and 8.2 shall be met on a lasting basis during a
period of time that is normal in view of the intended use of such instruments.
Digital electronic devices shall always exercise adequate control of the
correct operation of the measuring process, of the indicating device, and of all
data storage and data transfer.
Upon automatic detection of a significant durability error, electronic
instruments shall provide a visual or audible alarm that shall continue until
the user takes corrective action or the error disappears.
(4)
When external equipment is connected to an electronic instrument through an
appropriate interface the metrological qualities of the instrument shall not be
adversely influenced.
(5)
The instruments shall have no characteristics likely to facilitate fraudulent
use, whereas possibilities for unintentional misuse shall be minimal.
Components that may not be dismantled or adjusted by the user shall be
secured against such actions.
(6)
Instruments shall be designed to permit ready execution of the statutory
controls laid down by this Regulation.
9. Indication of weighing results and other weight values
The indication of the weighing results and other weight values shall be
accurate, unambiguous and non-misleading and the indicating device shall
permit easy reading of the indication under normal conditions of use.
The names and symbols of the units referred to in point 1 of this Schedule
shall comply with the provisions of Directive 80/181/EEC with the addition
of the symbol for the metric carat which shall be the symbol ‘ct’.
Indication shall be impossible above the maximum capacity (Max), increased
by 9 e.
An auxiliary indicating device is permitted only to the right of the decimal
mark. An extended indicating device may be used only temporarily, and
printing shall be inhibited during its functioning.
Secondary indications may be shown, provided that they cannot be mistaken
for primary indications.
10. Printing of weighing results and other weight values
Printed results shall be correct, suitably identified and unambiguous. The
printing shall be clear, legible, non- erasable and durable.
11. Levelling
When appropriate, instruments shall be fitted with a levelling device and a
level indicator, sufficiently sensitive to allow proper installation.
12. Zeroing
Instruments may be equipped with zeroing devices. The operation of these
devices shall result in accurate zeroing and shall not cause incorrect
measuring results.
13. Tare devices and preset tare devices
The instruments may have one or more tare devices and a preset tare device.
The operation of the tare devices shall result in accurate zeroing and shall
ensure correct net weighing. The operation of the preset tare device shall
ensure correct determination of the calculated net value.
14. Instruments for direct sales to the public, with a maximum capacity not
greater than 100 kg: additional requirements
Instruments for direct sale to the public shall show all essential information
about the weighing operation and, in the case of price-indicating instruments,
shall clearly show the customer the price calculation of the product to be
purchased.
The price to pay, if indicated, shall be accurate.
enough for the customer to read them properly.
weighing and price computation only if all indications related to all
transactions are printed clearly and unambiguously and are conveniently
arranged on a ticket or label for the customer.
Instruments shall bear no characteristics that can cause, directly or indirectly,
indications the interpretation of which is not easy or straightforward.
Instruments shall safeguard customers against incorrect sales transactions due
to their malfunctioning.
Auxiliary indicating devices and extended indicating devices are not
permitted. Supplementary devices are permitted only if they cannot lead to
fraudulent use.
Instruments similar to those normally used for direct sales to the public which
do not satisfy the requirements of this Section must carry near to the display
the indelible marking ‘Not to be used for direct sale to the public’.
15. instruments for direct sale to the public, as far as applicable to the instrument
in question. The printing of a price label shall be impossible below a
minimum capacity.
(1)
These regulations
Scope.
(a)
implement Directive 2014/31/EU of the European
Parliament and of the Council of 26 February 2014 on the
harmonisation of the laws of the Member States relating to the
making available on the market of measuring instruments
(recast);
(b) cover non-automatic weighing instruments which
are new to the Union market when they are placed on the
market; that is to say they are either new non-automatic
weighing instruments made by a manufacturer established in the
Union or non-automatic weighing instruments, whether new or
second-hand, imported from a third country; and
(c)
apply to all non-automatic weighing instruments.
(2) For the purposes of these regulations, the following
categories of use of non-automatic weighing instruments shall be
distinguished
(a)
the
determination
of
mass
for
commercial
transactions;
(b) the determination of mass for the calculation of a
toll, tariff, tax, bonus, penalty, remuneration, indemnity or
similar type of payment;
(c)
the determination of mass for the application of laws
or regulations or for an expert opinion given in court
proceedings;
(d) the determination of mass in the practice of
medicine for weighing patients for the purposes of monitoring,
diagnosis and medical treatment;
(e)
the determination of mass for making up medicines
on prescription in a pharmacy and determination of mass in
analyses carried out in medical and pharmaceutical laboratories;
(f)
the determination of price on the basis of mass for
the purposes of direct sales to the public and the making-up of
pre-packages;
(g) all applications other than those listed in paragraphs
(a) to (f).
CONFORMITY ASSESSMENT PROCEDURES
1. Module B: EU-type examination
(1)
EU-type examination is the part of a conformity assessment procedure in
which a notified body examines the technical design of an instrument and
verifies and attests that the technical design of the instrument meets the
requirements of these regulations that apply to it.
(2)
EU-type examination may be carried out in any of the following manners
— Examination of a specimen, representative of the production envisaged,
of the complete instrument (production type);
—
assessment of the adequacy of the technical design of the instrument
through examination of the technical documentation and supporting
evidence referred to in point 1.3, plus examination of specimens,
representative of the production envisaged, of one or more critical parts
of the instrument (combination of production type and design type);
—
Assessment of the adequacy of the technical design of the instrument
through examination of the technical documentation and supporting
evidence referred to in point 1.3, without examination of a specimen
(design type).
(3)
The manufacturer shall lodge an application for EU-type examination with a
single notified body of his choice.
The application shall include
(a) The name and address of the manufacturer and, if the application is
lodged by the authorised representative, his name and address as well;
(b) A written declaration that the same application has not been lodged with
any other notified body;
(c) The technical documentation. The technical documentation shall make it
possible to assess the instrument’s conformity with the applicable
requirements of these regulations and shall include an adequate analysis
and assessment of the risk(s). The technical documentation shall specify
the applicable requirements and cover, as far as relevant for the
assessment, the design, manufacture and operation of the instrument.
The technical documentation shall contain, wherever applicable, at least
the following elements
(i)
a general description of the instrument;
(ii) conceptual design and manufacturing drawings and schemes of
components, sub-assemblies, circuits, etc.;
(iii) descriptions and explanations necessary for the understanding of
those drawings and schemes and the operation of the instrument;
(iv) a list of the harmonised standards applied in full or in part the
references of which have been published in the Official Journal of
the European Union, and, where those harmonised standards have
not been applied, descriptions of the solutions adopted to meet the
essential requirements of these regulations, including a list of other
relevant technical specifications applied. In the event of partly
applied harmonised standards, the technical documentation shall
specify the parts which have been applied;
(v) results of design calculations made, examinations carried out, etc.;
(vi) test reports;
(d) The specimens’ representative of the production envisaged. The notified
body may request further specimens if needed for carrying out the test
programme;
(e) The supporting evidence for the adequacy of the technical design
solution. This supporting evidence shall mention any documents that
have been used, in particular where the relevant harmonised standards
have not been applied in full. The supporting evidence shall include,
where necessary, the results of tests carried out in accordance with other
relevant technical specifications by the appropriate laboratory of the
manufacturer or by another testing laboratory on his behalf and under his
responsibility.
(4)
The notified body shall
For the instrument
(4.1) examine the technical documentation and supporting evidence to assess the
adequacy of the technical design of the instrument;
For the specimen(s)
(4.2) verify that the specimen(s) have been manufactured in conformity with the
technical documentation, and identify the elements which have been designed
in accordance with the applicable provisions of the relevant harmonised
standards, as well as the elements which have been designed in accordance
with other relevant technical specifications;
(4.3) carry out appropriate examinations and tests, or have them carried out, to
check whether, where the manufacturer has chosen to apply the solutions in
the relevant harmonised standards, these have been applied correctly;
(4.4) carry out appropriate examinations and tests, or have them carried out, to
check whether, where the solutions in the relevant harmonised standards have
not been applied, the solutions adopted by the manufacturer applying other
relevant technical specifications meet the corresponding essential
requirements of these regulations;
(4.5) agree with the manufacturer on a location where the examinations and tests
will be carried out.
(5)
The notified body shall draw up an evaluation report that records the
activities undertaken in accordance with point 1.4 and their outcomes.
Without prejudice to its obligations vis-à-vis the notifying authorities, the
notified body shall release the content of that report, in full or in part, only
with the agreement of the manufacturer.
(6)
Where the type meets the requirements of these regulations, that apply to the
instrument concerned, the notified body shall issue an EU-type examination
certificate to the manufacturer. That certificate shall contain the name and
address of the manufacturer, the conclusions of the examination, the
conditions (if any) for its validity and the necessary data for identification of
the approved type. The EU-type examination certificate may have one or
more Annexes attached.
The EU-type examination certificate and its Annexes shall contain all
relevant information to allow the conformity of manufactured instruments
with the examined type to be evaluated and to allow for in-service control.
The EU-type examination certificate shall have a validity period of 10 years
from the date of its issue, and may be renewed for subsequent periods of 10
years each. In the event of fundamental changes to the design of the
instrument, e.g. as a result of the application of new techniques, the validity
of EU-type examination certificate may be limited to two years and extended
by three years.
Where the type does not satisfy the applicable requirements of these
regulations, the notified body shall refuse to issue an EU-type examination
certificate and shall inform the applicant accordingly, giving detailed reasons
for its refusal.
(7)
The notified body shall keep itself appraised of any changes in the generally
acknowledged state of the art which indicate that the approved type may no
longer comply with the applicable requirements of these regulations, and
shall determine whether such changes require further investigation. If so, the
notified body shall inform the manufacturer accordingly. The manufacturer
shall inform the notified body that holds the technical documentation relating
to the EU-type examination certificate of all modifications to the approved
type that may affect the conformity of the instrument with the essential
requirements of these regulations or the conditions for validity of that
certificate. Such modifications shall require additional approval in the form
of an addition to the original EU- type examination certificate.
(8)
Each notified body shall inform its notifying authority concerning the EU-
type examination certificates and/or any additions thereto which it has issued
or withdrawn, and shall, periodically or upon request, make available to its
notifying authority the list of such certificates and/or any additions thereto
refused, suspended or otherwise restricted.
Each notified body shall inform the other notified bodies concerning the EU-
type examination certificates and/or any additions thereto which it has
refused, withdrawn, suspended or otherwise restricted, and, upon request,
concerning such certificates and/or additions thereto which it has issued.
The Commission, the Member States and the other notified bodies may, on
request, obtain a copy of the EU-type examination certificates and/or
additions thereto. On request, the Commission and the Member States may
obtain a copy of the technical documentation and the results of the
examinations carried out by the notified body. The notified body shall keep a
copy of the EU-type examination certificate, its annexes and additions, as
well as the technical file including the documentation submitted by the
manufacturer, until the expiry of the validity of that certificate.
(9)
The manufacturer shall keep a copy of the EU-type examination certificate,
its schedules and additions together with the technical documentation at the
disposal of the national authorities for
10 years after the instrument has been placed on the market.
(10) The manufacturer’s authorised representative may lodge the application
referred to in point 1.3 and fulfil the obligations set out in points 1.7 and 1.9,
provided that they are specified in the mandate.
2. Module D: Conformity to type based on quality assurance of the production
process
(1)
Conformity to type based on quality assurance of the production process is
the part of a conformity assessment procedure whereby the manufacturer
fulfils the obligations laid down in points 2.2 and 2.5, and ensures and
declares on his sole responsibility that the instruments concerned are in
conformity with the type described in the EU-type examination certificate
and satisfy the requirements of these regulations that apply to them.
(2)
Manufacturing
The manufacturer shall operate an approved quality system for production,
final product inspection and testing of the instruments concerned as specified
in point 2.3, and shall be subject to surveillance as specified in point 2.4.
(3)
Quality system
(3.1) The manufacturer shall lodge an application for assessment of his quality
system with the notified body of his choice, for the instruments concerned.
The application shall include
(a) the name and address of the manufacturer and, if the application is
lodged by the authorised representative, his name and address as well;
(b) a written declaration that the same application has not been lodged with
any other notified body;
(c) all relevant information for the instrument category envisaged;
(d) the documentation concerning the quality system; and
(e) the technical documentation of the approved type and a copy of the EU-
type examination certificate.
(3.2) The quality system shall ensure that the instruments are in conformity with
the type described in the EU-type examination certificate and comply with
the requirements of these regulations that apply to them.
All the elements, requirements and provisions adopted by the manufacturer
shall be documented in a systematic and orderly manner in the form of
written policies, procedures and instructions. The quality system
documentation shall permit a consistent interpretation of the quality
programmes, plans, manuals and records.
It shall, in particular, contain an adequate description of
(a) The quality objectives and the organisational structure, responsibilities
and powers of the management with regard to product quality;
(b) The corresponding manufacturing, quality control and quality assurance
techniques, processes and systematic actions that will be used;
(c) The examinations and tests that will be carried out before, during and
after manufacture, and the frequency with which they will be carried out;
(d) The quality records, such as inspection reports and test data, calibration
data, qualification reports on the personnel concerned, etc.;
(e) the means of monitoring the achievement of the required product quality
and the effective operation of the quality system.
(3.3) The notified body shall assess the quality system to determine whether it
satisfies the requirements referred to in point 2.3.2.
It shall presume conformity with those requirements in respect of the
elements of the quality system that comply with the corresponding
specifications of the relevant harmonised standard.
In addition to experience in quality management systems, the auditing team
shall have at least one member with experience of evaluation in the relevant
instrument field and instrument technology concerned, and knowledge of the
applicable requirements of these regulations. The audit shall include an
assessment visit to the manufacturer’s premises. The auditing team shall
review the technical documentation referred to in point 2.3.1(e) to verify the
manufacturer’s ability to identify the relevant requirements of these
regulations and to carry out the necessary examinations with a view to
ensuring compliance of the instrument with those requirements.
The decision shall be notified to the manufacturer. The notification shall
contain the conclusions of the audit and the reasoned assessment decision.
(3.4) The manufacturer shall undertake to fulfil the obligations arising out of the
quality system as approved and to maintain it so that it remains adequate and
efficient.
(3.5) The manufacturer shall keep the notified body that has approved the quality
system informed of any intended change to the quality system.
The notified body shall evaluate any proposed changes and decide whether
the modified quality system will continue to satisfy the requirements referred
to in point 2.3.2 or whether a reassessment is necessary.
It shall notify the manufacturer of its decision. The notification shall contain
the conclusions of the examination and the reasoned assessment decision.
(4)
Surveillance under the responsibility of the notified body
(4.1) The purpose of surveillance is to make sure that the manufacturer duly fulfils
the obligations arising out of the approved quality system.
(4.2) The manufacturer shall, for assessment purposes, allow the notified body
access to the manufacture, inspection, and testing and storage sites and shall
provide it with all necessary information, in particular
(a) The quality system documentation;
(b) The quality records, such as inspection reports and test data, calibration
data, qualification reports on the personnel concerned, etc.
(4.3) The notified body shall carry out periodic audits to make sure that the
manufacturer maintains and applies the quality system and shall provide the
manufacturer with an audit report.
(4.4) In addition, the notified body may pay unexpected visits to the manufacturer.
During such visits the notified body may, if necessary, carry out instrument
tests, or have them carried out, in order to verify that the quality system is
functioning correctly. The notified body shall provide the manufacturer with
a visit report and, if tests have been carried out, with a test report.
(5)
Conformity marking and EU declaration of conformity
(5.1) The manufacturer shall affix the CE marking and the supplementary
metrology marking set out in these regulations, and, under the responsibility
of the notified body referred to in point 2.3.1, the latter’s identification
number to each individual instrument that is in conformity with the type
described in the EU-type examination certificate and satisfies the applicable
requirements of these regulations.
(5.2) The manufacturer shall draw up a written EU declaration of conformity for
each instrument model and keep it at the disposal of the national authorities
for 10 years after the instrument has been placed on the market. The EU
declaration of conformity shall identify the instrument model for which it has
been drawn up.
A copy of the EU declaration of conformity shall be made available to the
relevant authorities upon request.
(6)
The manufacturer shall, for a period ending 10 years after the instrument has
been placed on the market, keep at the disposal of the national authorities
(a) the documentation referred to in point 2.3.1;
(b) the information relating to the change referred to in point 2.3.5, as
approved;
(c) the decisions and reports of the notified body referred to in points 2.3.5,
2.4.3 and 2.4.4.
(7)
Each notified body shall inform its notifying authority of quality system
approvals issued or withdrawn, and shall, periodically or upon request, make
available to its notifying authority the list of quality system approvals
refused, suspended or otherwise restricted.
(8)
Authorised representative
The manufacturer’s obligations set out in points 2.3.1, 2.3.5, 2.5 and 2.6 may
be fulfilled by his authorised representative, on his behalf and under his
responsibility, provided that they are specified in the mandate.
3. Module D1: Quality assurance of the production process
(1)
Quality assurance of the production process is the conformity assessment
procedure whereby the manufacturer fulfils the obligations laid down in
points 3.2, 3.4 and 3.7, and ensures and declares on his sole responsibility that
the instruments concerned satisfy the requirements of these regulations that
apply to them.
(2)
Technical documentation
The manufacturer shall establish the technical documentation. The
documentation shall make it possible to assess the instrument’s conformity
with the relevant requirements, and shall include an adequate analysis and
assessment of the risk(s). The technical documentation shall specify the
applicable requirements and cover, as far as relevant for the assessment, the
design, manufacture and operation of the instrument. The technical
documentation shall, wherever applicable, contain at least the following
elements
(a) a general description of the instrument;
(b) conceptual design and manufacturing drawings and schemes of
components, sub-assemblies, circuits, etc.;
(c) descriptions and explanations necessary for the understanding of those
drawings and schemes and the operation of the instrument;
(d) a list of the harmonised standards applied in full or in part the references
of which have been published in the Official Journal of the European
Union, and, where those harmonised standards have not been applied,
descriptions of the solutions adopted to meet the essential requirements
of these regulations, including a list of other relevant technical
specifications applied. In the event of partly applied harmonised
standards, the technical documentation shall specify the parts which
have been applied;
(e) results of design calculations made, examinations carried out, etc.;
(f) test reports.
(3)
The manufacturer shall keep the technical documentation at the disposal of
the relevant national authorities for 10 years after the instrument has been
placed on the market.
(4)
Manufacturing
The manufacturer shall operate an approved quality system for production,
final product inspection and testing of the instruments concerned as specified
in point 3.5, and shall be subject to surveillance as specified in point 3.6.
(5)
Quality system
(5.1) The manufacturer shall lodge an application for assessment of his quality
system with the notified body of his choice, for the instruments concerned.
The application shall include
(a) the name and address of the manufacturer and, if the application is
lodged by the authorised representative, his name and address as well;
(b) a written declaration that the same application has not been lodged with
any other notified body;
(c) all relevant information for the instrument category envisaged;
(d) the documentation concerning the quality system;
(e) the technical documentation referred to in point 3.2.
(5.2) The quality system shall ensure compliance of the instruments with the
requirements of these regulations that apply to them.
All the elements, requirements and provisions adopted by the manufacturer
shall be documented in a systematic and orderly manner in the form of
written policies, procedures and instructions. The quality system
documentation shall permit a consistent interpretation of the quality
programmes, plans, manuals and records.
It shall, in particular, contain an adequate description of
(a) the quality objectives and the organisational structure, responsibilities
and powers of the management with regard to product quality;
(b) the corresponding manufacturing, quality control and quality assurance
techniques, processes and systematic actions that will be used;
(c) the examinations and tests that will be carried out before, during and
after manufacture, and the frequency with which they will be carried out;
(d) the quality records, such as inspection reports and test data, calibration
data, qualification reports on the personnel concerned, etc.;
(e) the means of monitoring the achievement of the required product quality
and the effective operation of the quality system.
(5.3) The notified body shall assess the quality system to determine whether it
satisfies the requirements referred to in point 3.5.2.
It shall presume conformity with those requirements in respect of the
elements of the quality system that comply with the corresponding
specifications of the relevant harmonised standard.
In addition to experience in quality management systems, the auditing team
shall have at least one member with experience of evaluation in the relevant
instrument field and instrument technology concerned, and knowledge of the
applicable requirements of these regulations. The audit shall include an
assessment visit to the manufacturer’s premises. The auditing team shall
review the technical documentation referred to in point 3.2 in order to verify
the manufacturer’s ability to identify the relevant requirements of these
regulations and to carry out the necessary examinations with a view to
ensuring compliance of the instrument with those requirements.
The decision shall be notified to the manufacturer. The notification shall
contain the conclusions of the audit and the reasoned assessment decision.
(5.4) The manufacturer shall undertake to fulfil the obligations arising out of the
quality system as approved and to maintain it so that it remains adequate and
efficient.
(5.5) The manufacturer shall keep the notified body that has approved the quality
system informed of any intended change to the quality system.
The notified body shall evaluate any proposed changes and decide whether
the modified quality system will continue to satisfy the requirements referred
to in point 3.5.2 or whether reassessment is necessary.
It shall notify the manufacturer of its decision. The notification shall contain
the conclusions of the examination and the reasoned assessment decision.
(6)
Surveillance under the responsibility of the notified body
(6.1) The purpose of surveillance is to make sure that the manufacturer duly fulfils
the obligations arising out of the approved quality system .
(6.2) The manufacturer shall, for assessment purposes, allow the notified body
access to the manufacture, inspection, testing and storage sites and shall
provide it with all necessary information, in particular
(a) the quality system documentation;
(b) the technical documentation referred to in point 3.2;
(c) the quality records, such as inspection reports and test data, calibration
data, qualification reports on the personnel concerned, etc.
(6.3) The notified body shall carry out periodic audits to make sure that the
manufacturer maintains and applies the quality system and shall provide the
manufacturer with an audit report.
(6.4) In addition, the notified body may pay unexpected visits to the manufacturer.
During such visits the notified body may, if necessary, carry out product tests,
or have them carried out, in order to verify that the quality system is
functioning correctly. The notified body shall provide the manufacturer with
a visit report and, if tests have been carried out, with a test report.
(7)
Conformity marking and EU declaration of conformity
(7.1) The manufacturer shall affix the CE marking and the supplementary
metrology marking, set out in these regulations, and, under the responsibility
of the notified body referred to in point 3.5.1, the latter’s identification
number to each individual instrument that satisfies the applicable
requirements of these regulations.
(7.2) The manufacturer shall draw up a written EU declaration of conformity for
each instrument model and keep it at the disposal of the national authorities
for 10 years after the instrument has been placed on the market. The EU
declaration of conformity shall identify the instrument model for which it has
been drawn up.
A copy of the EU declaration of conformity shall be made available to the
relevant authorities upon request.
(8)
The manufacturer shall, for a period ending 10 years after the instrument has
been placed on the market, keep at the disposal of the national authorities
(a) the documentation referred to in point 3.5.1;
(b) the information relating to the change referred to in point 3.5.5, as
approved;
(c) the decisions and reports of the notified body referred to in points 3.5.5,
3.6.3 and 3.6.4.
(9)
Each notified body shall inform its notifying authority of quality system
approvals issued or withdrawn, and shall, periodically or upon request, make
available to its notifying authority the list of quality system approvals
refused, suspended or otherwise restricted.
(10) Authorised representative
The manufacturer’s obligations set out in points 3.3, 3.5.1, 3.5.5, 3.7 and 3.8
may be fulfilled by his authorised representative, on his behalf and under his
responsibility, provided that they are specified in the mandate.
4. Module F: Conformity to type based on product verification
(1)
Conformity to type based on product verification is the part of a conformity
assessment procedure whereby the manufacturer fulfils the obligations laid
down in points 4.2 and 4.5 and ensures and declares on his sole responsibility
that the instruments concerned, which have been subject to the provisions of
point 4.3, are in conformity with the type described in the EU-type
examination certificate and satisfy the requirements of these regulations that
apply to them.
(2)
Manufacturing
The manufacturer shall take all measures necessary so that the manufacturing
process and its monitoring ensure conformity of the manufactured
instruments with the approved type described in the EU-type examination
certificate and with the requirements of these regulations that apply to them.
(3)
Verification
A notified body chosen by the manufacturer shall carry out appropriate
examinations and tests in order to check the conformity of the instruments
with the approved type described in the EU-type examination certificate and
with the appropriate requirements of the Directive 2014/31.
The examinations and tests to check the conformity of the instruments with
the appropriate requirements shall be carried out by examination and testing
of every instrument as specified in point 4.4.
(4)
Verification of conformity by examination and testing of every instrument
(4.1) All instruments shall be individually examined and appropriate tests set out
in the relevant harmonised standard(s), and/or equivalent tests set out in other
relevant technical specifications, shall be carried out in order to verify
conformity with the approved type described in the EU-type examination
certificate and with the appropriate requirements of these regulations.
In the absence of such a harmonised standard, the notified body concerned
shall decide on the appropriate tests to be carried out.
(4.2) The notified body shall issue a certificate of conformity in respect of the
examinations and tests carried out, and shall affix its identification number to
each approved instrument or have it affixed under its responsibility.
The manufacturer shall keep the certificates of conformity available for
inspection by the national authorities for 10 years after the instrument has
been placed on the market.
(5)
Conformity marking and EU declaration of conformity
(5.1) The manufacturer shall affix the CE marking and the supplementary
metrology marking, set out in these regulations, and, under the responsibility
of the notified body referred to in point 4.3, the latter’s identification number
to each individual instrument that is in conformity with the approved type
described in the EU-type examination certificate and satisfies the applicable
requirements of these regulations.
(5.2) The manufacturer shall draw up a written EU declaration of conformity for
each instrument model and keep it at the disposal of the national authorities,
for 10 years after the instrument has been placed on the market. The EU
declaration of conformity shall identify the instrument model for which it has
been drawn up.
A copy of the EU declaration of conformity shall be made available to the
relevant authorities upon request.
If the notified body referred to in point 4.3 agrees and under its responsibility,
the manufacturer may also affix the notified body’s identification number to
the instruments.
(6)
If the notified body agrees and under its responsibility, the manufacturer may
affix the notified body’s identification number to the instruments during the
manufacturing process.
(7)
Authorised representative
The manufacturer’s obligations may be fulfilled by his authorised
representative, on his behalf and under his responsibility, provided that they
are specified in the mandate. An authorised representative may not fulfil the
manufacturer’s obligations set out in point 4.2.
5. Module F1: Conformity based on product verification
(1)
Conformity based on product verification is the conformity assessment
procedure whereby the manufacturer fulfils the obligations laid down in
points 5.2, 5.3 and 5.6 and ensures and declares on his sole responsibility that
the instruments concerned, which have been subject to the provisions of point
5.4, are in conformity with the requirements of these regulations that apply to
them.
(2)
Technical documentation
(2.1) The manufacturer shall establish the technical documentation. The
documentation shall make it possible to assess the instrument’s conformity
with the relevant requirements, and shall include an adequate analysis and
assessment of the risk(s). The technical documentation shall specify the
applicable requirements and cover, as far as relevant for the assessment, the
design, manufacture and operation of the instrument. The technical
documentation shall, wherever applicable, contain at least the following
elements
(a) a general description of the instrument;
(b) conceptual design and manufacturing drawings and schemes of
components, sub-assemblies, circuits, etc.;
(c) descriptions and explanations necessary for the understanding of those
drawings and schemes and the operation of the instrument;
(d) a list of the harmonised standards applied in full or in part the references
of which have been published in the Official Journal of the European
Union, and, where those harmonised standards have not been applied,
descriptions of the solutions adopted to meet the essential requirements
of these regulations, including a list of other relevant technical
specifications applied. In the event of partly applied harmonised
standards, the technical documentation shall specify the parts which
have been applied;
(e) results of design calculations made, examinations carried out, etc.;
(f) test reports.
(2.2) The manufacturer shall keep the technical documentation at the disposal of
the relevant national authorities for 10 years after the instrument has been
placed on the market.
(3)
Manufacturing
The manufacturer shall take all measures necessary so that the manufacturing
process and its monitoring ensure conformity of the manufactured
instruments with the applicable requirements of these regulations.
(4)
Verification
A notified body chosen by the manufacturer shall carry out appropriate
examinations and tests to check the conformity of the instruments with the
applicable requirements of these regulations.
The examinations and tests to check the conformity with those requirements
shall be carried out by examination and testing of every instrument as
specified in point 5.5.
(5)
Verification of conformity by examination and testing of every instrument
(5.1) All instruments shall be individually examined and appropriate tests, set out
in the relevant harmonised standards and/or equivalent tests set out in other
relevant technical specifications, shall be carried out to verify conformity
with the requirements that apply to them. In the absence of such a harmonised
standard the notified body concerned shall decide on the appropriate tests to
be carried out.
(5.2) The notified body shall issue a certificate of conformity in respect of the
examinations and tests carried out, and shall affix its identification number to
each approved instrument or have it affixed under its responsibility.
The manufacturer shall keep the certificates of conformity at the disposal of
the national authorities for 10 years after the instrument has been placed on
the market.
(6)
Conformity marking and EU declaration of conformity
(6.1) The manufacturer shall affix the CE marking and the supplementary
metrology marking, set out in these regulations, and, under the responsibility
of the notified body referred to in point 5.4, the latter’s identification number
to each individual instrument that satisfies the applicable requirements of
these regulations.
(6.2) The manufacturer shall draw up a written EU declaration of conformity for
each instrument model and keep it at the disposal of the national authorities
for 10 years after the instrument has been placed on the market. The EU
declaration of conformity shall identify the instrument model for which it has
been drawn up.
A copy of the EU declaration of conformity shall be made available to the
relevant authorities upon request.
If the notified body referred to in point 5.5 agrees and under its responsibility,
the manufacturer may also affix the notified body’s identification number to
the instruments.
(7)
If the notified body agrees and under its responsibility, the manufacturer may
affix the notified body’s identification number to the instruments during the
manufacturing process.
(8)
Authorised representative
The manufacturer’s obligations may be fulfilled by his authorised
representative, on his behalf and under his responsibility, provided that they
are specified in the mandate. An authorised representative may not fulfil the
manufacturer’s obligations set out in points 5.2.1 and 5.3.
6. Module G: Conformity based on unit verification
(1)
Conformity based on unit verification is the conformity assessment
procedure whereby the manufacturer fulfils the obligations laid down in
points 6.2, 6.3 and 6.5, and ensures and declares on his sole responsibility that
the instrument concerned, which has been subject to the provisions of point
6.4, is in conformity with the requirements of these regulations that apply to
it.
(2)
Technical documentation
(2.1) The manufacturer shall establish the technical documentation and make it
available to the notified body referred to in point 6.4. The documentation
shall make it possible to assess the instrument’s conformity with the relevant
requirements, and shall include an adequate analysis and assessment of the
risk(s). The technical documentation shall specify the applicable
requirements and cover, as far as relevant for the assessment, the design,
manufacture and operation of the instrument. The technical documentation
shall, wherever applicable, contain at least the following elements
(a) a general description of the instrument;
(b) conceptual design and manufacturing drawings and schemes of
components, sub-assemblies, circuits, etc.;
(c) descriptions and explanations necessary for the understanding of those
drawings and schemes and the operation of the instrument;
(d) a list of the harmonised standards applied in full or in part the references
of which have been published in the Official Journal of the European
Union, and, where those harmonised standards have not been applied,
descriptions of the solutions adopted to meet the essential requirements
of these regulations, including a list of other relevant technical
specifications applied. In the event of partly applied harmonised
standards, the technical documentation shall specify the parts which
have been applied;
(e) results of design calculations made, examinations carried out, etc.;
(f) test reports.
(2.2) The manufacturer shall keep the technical documentation at the disposal of
the relevant national authorities for 10 years after the instrument has been
placed on the market.
(3)
Manufacturing
The manufacturer shall take all measures necessary so that the manufacturing
process and its monitoring ensure conformity of the manufactured instrument
with the applicable requirements of these regulations.
(4)
Verification
A notified body chosen by the manufacturer shall carry out appropriate
examinations and tests, set out in the relevant harmonised standards and/or
equivalent tests set out in other relevant technical specifications, to check the
conformity of the instrument with the applicable requirements of these
Regulations, or have them carried out. In the absence of such a harmonised
standard the notified body concerned shall decide on the appropriate tests to
be carried out.
The notified body shall issue a certificate of conformity in respect of the
examinations and tests carried out and shall affix its identification number to
the approved instrument, or have it affixed under its responsibility.
The manufacturer shall keep the certificates of conformity at the disposal of
the national authorities for 10 years after the instrument has been placed on
the market.
(5)
Conformity marking and EU declaration of conformity
(5.1) The manufacturer shall affix the CE marking and the supplementary
metrology marking, set out in these Regulations, and, under the responsibility
of the notified body referred to in point 6.4, the latter’s identification number
to each instrument that satisfies the applicable requirements of these
Regulations.
(5.2) The manufacturer shall draw up a written EU declaration of conformity and
keep it at the disposal of the national authorities for 10 years after the
instrument has been placed on the market. The EU declaration of conformity
shall identify the instrument for which it has been drawn up.
A copy of the EU declaration of conformity shall be made available to the
relevant authorities upon request.
(6)
Authorised representative
The manufacturer’s obligations set out in points 6.2.2 and 6.5 may be
fulfilled by his authorised representative, on his behalf and under his
responsibility, provided that they are specified in the mandate.
7. Common provisions
(1)
The conformity assessment according to Module D, D1, F, F1 or G may be
carried out at the manufacturer’s works or any other location if transport to
the place of use does not require dismantling of the instrument, if the putting
into service at the place of use does not require assembly of the instrument or
other technical installation work likely to affect the instrument’s
performance, and if the gravity value at the place of putting into service is
taken into consideration or if the instrument’s performance is insensitive to
gravity variations. In all other cases, it shall be carried out at the place of use
of the instrument.
(2)
If the instrument’s performance is sensitive to gravity variations the
procedures referred to in point 7.1 may be carried out in two stages, with the
second stage comprising all examinations and tests of which the outcome is
gravity-dependent, and the first stage all other examinations and tests. The
second stage shall be carried out at the place of use of the instrument. If a
Member State has established gravity zones on its territory the expression ‘at
the place of use of the instrument’ may be read as ‘in the gravity zone of use
of the instrument’.
(2.1) Where a manufacturer has opted for execution in two stages of one of the
procedures mentioned in point 7.1, and where these two stages will be carried
out by different parties, an instrument which has undergone the first stage of
the procedure shall bear the identification number of the notified body
involved in that stage.
(2.2) The party which has carried out the first stage of the procedure shall issue for
each of the instruments a certificate containing the data necessary for
identification of the instrument and specifying the examinations and tests that
have been carried out.
The party which carries out the second stage of the procedure shall carry out
those examinations and tests that have not yet been carried out.
The manufacturer or his authorised representative shall ensure that he is able
to supply the notified body’s certificates of conformity on request.
(2.3) A manufacturer who has opted for Module D or D1 in the first stage may
either use this same procedure in the second stage or decide to continue in the
second stage with Module F or F1 as appropriate.
(2.4) The CE marking and the supplementary metrology marking shall be affixed
to the instrument on completion of the second stage, along with the
identification number of the notified body which took part in the second
stage.
In these regulations, unless the context otherwise requires
"accreditation" means accreditation as defined in paragraph 10
of Article 2 of Regulation (EC) No 765/2008;
"authorised representative" means any natural or legal person
established within the Union who has received a written mandate
from a manufacturer to act on his behalf in relation to specified tasks;
"Authority" means the Standards and Metrology Institute of the
Malta Competition and Consumer Affairs Authority;
"CE marking" means a marking by which the manufacturer
indicates that the instrument is in conformity with the applicable
requirements set out in Union harmonisation legislation providing for
its affixing;
"conformity assessment" means the process demonstrating
whether the essential requirements of these regulations relating to an
instrument have been fulfilled;
"conformity assessment body" means a body that performs
conformity assessment activities including calibration, testing,
certification and inspection;
"distributor" means any natural or legal person in the supply
chain, other than the manufacturer or the importer, who makes an
instrument available on the market;
" economic operators" means the manufacturer, the authorised
representative, the importer and the distributor;
"harmonised standard" means harmonised standard as defined in
paragraph (c) of paragraph 1 of Article 2 of Regulation (EU) No
1025/2012;
"importer" means any natural or legal person established within
the Union who places an instrument from a third country on the
Union market;
"making available on the market" means any supply of an
instrument for distribution or use on the Union market in the course
of a commercial activity, whether in return for payment or free of
charge;
"manufacturer" means any natural or legal person who
manufactures an instrument or has an instrument designed or
manufactured, and markets that instrument under his name or trade
mark;
"National Accreditation Board (Malta)" means the National
Accreditation Board (Malta) established in terms of regulation 3 of
the National Accreditation Board (Malta) (Establishment) S.L. 419.07.
Regulations;
"non-automatic weighing instrument" or "instrument" means a
weighing instrument requiring the intervention of an operator during
weighing;
"placing on the market" means the first making available of an
instrument on the Union market;
"recall" means any measure aimed at achieving the return of an
instrument that has already been made available to the end-user;
"technical specification" means a document that prescribes
technical requirements to be fulfilled by an instrument;
"Union harmonisation legislation" means any Union legislation
harmonising the conditions for the marketing of products;
"Weighing instrument" means a measuring instrument serving
to determine the mass of a body by using the action of gravity on that
body. A weighing instrument may also serve to determine other mass-
related magnitudes, quantities, parameters or characteristics;
"withdrawal" means any measure aimed at preventing an
instrument in the supply chain from being made available on the
market.
INSCRIPTIONS
1. Instruments intended to be used for the applications listed in paragraphs (a) to
(f) of regulation 2(3)
(1)
Those instruments shall bear visibly, legibly and indelibly the following
inscriptions
(i)
the number of the EU-type examination certificate, where
appropriate;
(ii) the manufacturer’s name, registered trade name or registered trade
mark;
(iii) the accuracy class, enclosed in an oval or in two horizontal lines
joined by two half circles;
(iv) maximum capacity, in the form Max …;
(v) minimum capacity, in the form Min …;
(vi) verification scale interval, in the form e = …;
(vii) type, batch or serial number; and when applicable
(viii) for instruments consisting of separate but associated units
identification mark on each unit;
(ix) scale interval if it is different from e, in the form d = …;
(x) maximum additive tare effect, in the form T = + …;
(xi) maximum subtractive tare effect if it is different from Max, in the
form T = – …;
(xii) tare interval if it is different from d, in the form d T = …;
(xiii) maximum safe load if it is different from Max, in the form Lim …;
(xiv) the special temperature limits, in the form … o C/… o C;
(xv) ratio between load receptor and load.
(2)
Those instruments shall have adequate facilities for the affixing of the
conformity marking and inscriptions. These shall be such that it shall be
impossible to remove the conformity marking and inscriptions without
damaging them, and that the conformity marking and inscriptions shall be
visible when the instrument is in its regular operating position.
(3)
Where a data plate is used it shall be possible to seal the plate unless it cannot
be removed without being destroyed. If the data plate is sealable it shall be
possible to apply a control mark to it.
(4)
The inscriptions Max, Min, e, and d, shall also be shown near the display of
the result if they are not already located there.
(5)
Each load measuring device which is connected or can be connected to one or
more load receptors shall bear the relevant inscriptions relating to the said
load receptors.
2. Instruments not intended to be used for the applications listed in paragraphs
(a) to (f) of regulation 2(3) shall bear visibly, legibly and indelibly
— the manufacturer’s name, registered trade name or registered trade mark;
—
maximum capacity, in the form Max ….
Those instruments shall not bear the conformity marking as set out in these
Regulations.
3. Restrictive use symbol referred to in regulation 19
The restrictive use symbol shall be constituted by a capital letter ‘M’ printed
in black on a red background at least 25 mm × 25 mm square with two
intersecting diagonals forming a cross.
(1) The Authority shall take all steps to ensure that only
instruments that meet the applicable requirements of this regulation
may be made available on the market.
(2) The Authority shall take all steps to ensure that
instruments may not be put into service for the uses referred to in
paragraphs (a) to (f) of regulation 2(3) unless they meet the
requirements of these regulations.
(3) The Authority shall take all steps to ensure that
instruments put into service for the uses referred to in paragraphs (a)
to (f) of regulation 2(3) continue to conform to the applicable
requirements of these regulations.
EU DECLARATION OF CONFORMITY (No XXXX) *
1. Instrument model/Instrument (product, type, batch or serial number)
2. Name and address of the manufacturer and, where applicable, his authorised
representative
3. This declaration of conformity is issued under the sole responsibility of the
manufacturer.
4. Object of the declaration (identification of instrument allowing traceability; it
may, where necessary for the identification of the instrument, include an
image)
5. The object of the declaration described above is in conformity with the
relevant Union harmonisation legislation
6. References to the relevant harmonised standards used or references to the
other technical specifications in relation to which conformity is declared
7. The notified body … (name, number) performed … (description of
intervention) and issued the certificate
8. Additional information
Signed for and on behalf of
(place and date of issue)
(name, function)
(signature)
* It is optional for the manufacturer to assign a number to the declaration of conformity.
€2.16
Instruments used or intended to be used for the
applications listed in paragraphs (a) to (f) of regulation 2(3) shall
satisfy the essential requirements set out in Schedule I.
Where the instrument includes, or is connected to, devices
which are not used or intended to be used for the applications listed in
paragraphs (a) to (f) of regulation 2(3), such devices shall not be
subject to those essential requirements.
(1) The Authority shall not impede the making available
on the market of instruments which meet the requirements of these
regulations.
(2) The Authority shall not impede the putting into service, for
the uses referred to in paragraphs (a) to (f) of regulation 2(3), of
instruments which meet the requirements of these regulations.
(1) When placing on the market their instruments
intended to be used for the applications listed in paragraphs (a) to (f)
of regulation 2(3), manufacturers shall ensure that they have been
designed and manufactured in accordance with the essential
requirements set out in Schedule I.
(2) For the instruments intended to be used for the applications
listed in paragraphs (a) to (f) of regulation 2(3), manufacturers shall
draw up the technical documentation referred to in Schedule II and
carry out the relevant conformity assessment procedure referred to in
regulation 14 or have it carried out .
Where compliance of an instrument intended to be used for the
applications listed in paragraphs (a) to (f) of regulation 2(3) with the
applicable requirements has been demonstrated by that conformity
assessment procedure, manufacturers shall draw up an EU declaration
of conformity and affix the CE marking and the supplementary
metrology marking.
(3) For the instruments intended to be used for the applications
listed in paragraphs (a) to (f) of regulation 2(3), manufacturers shall
keep the technical documentation and the EU declaration of
conformity for 10 years after the instrument has been placed on the
market.
(4) Manufacturers shall ensure that procedures are in place for
series production to remain in conformity with these regulations.
Changes in instrument design or characteristics and changes in the
harmonised standards or in other technical specifications by reference
to which conformity of an instrument is declared shall be adequately
taken into account.
When deemed appropriate with regard to the risks
presented by an instrument intended to be used for the applications
listed in paragraphs (a) to (f) of regulation 2(3), manufacturers shall
carry out sample testing of instruments made available on the market,
investigate, and, if necessary, keep a register of complaints, of non-
conforming instruments and instrument recalls, and shall keep
distributors informed of any such monitoring.
(5) Manufacturers shall ensure that instruments which they
have placed on the market bear a type, batch or serial number or other
element allowing their identification, as set out in Schedule III.
For the instruments intended to be used for the applications
listed in paragraphs (a) to (f) of regulation 2(3), manufacturers shall
affix the inscriptions provided for in point 1 of Schedule III.
For the instruments not intended to be used for the applications
listed in paragraphs (a) to (f) of regulation 2(3), manufacturers shall
affix the inscriptions provided for in point 2 of Schedule III.
Where an instrument which is intended to be used for any of the
applications listed in paragraphs (a) to (f) of regulation 2(3) includes,
or is connected to, devices which are not used or intended to be used
for the applications listed in paragraphs (a) to (f) of regulation 2(3),
manufacturers shall affix to each of those devices the restrictive use
symbol as provided for in regulation 19 and in point 3 of Schedule III.
(6) Manufacturers shall indicate on the instrument their name,
registered trade name or registered trade mark and the postal address
at which they can be contacted. The address shall indicate a single
paragraph at which the manufacturer can be contacted. The contact
details shall be in a language easily understood by end-users and the
Authority.
(7) Manufacturers shall ensure that the instrument intended to
be used for the applications listed in paragraphs (a) to (f) of regulation
2(3) is accompanied by instructions and information in a language
which can be easily understood by end-users, as determined by the
Authority. Such instructions and information, as well as any labelling,
shall be clear, understandable and intelligible.
(8) Manufacturers who consider or have reason to believe that
an instrument which they have placed on the market is not in
conformity with these regulations shall immediately take the
corrective measures necessary to bring that instrument into
conformity, to withdraw it or recall it, if appropriate. Furthermore,
where the instrument presents a risk, manufacturers shall immediately
inform the competent national authorities of the Member States in
which they made the instrument available on the market to that effect,
giving details, in particular, of the non-compliance and of any
corrective measures taken.
(9) Manufacturers shall, further to a reasoned request from the
Authority, provide it with all the information and documentation in
paper or electronic form necessary to demonstrate the conformity of
the instrument with these regulations, in at least English and, or
Maltese. The Authority at its request, on any action taken to eliminate
the risks posed by instruments which they have placed on the market.
(1) A manufacturer may, by a written mandate, appoint
an authorised representative.
The obligations laid down in regulation 7(1) and the obligation
to draw up technical documentation referred to in regulation 7(2)
shall not form part of the authorised representative’s mandate.
(2) An authorised representative shall perform the tasks
specified in the mandate received from the manufacturer. The
mandate shall allow the authorised representative to do at least the
following
(a)
keep the EU declaration of conformity and the
technical documentation at the disposal of the Authority for 10
years after the instrument has been placed on the market;
(b) further to a reasoned request from a competent
national authority, provide that authority with all the
information and documentation necessary to demonstrate the
conformity of an instrument;
(c)
cooperate with the competent national authorities, at
their request, on any action taken to eliminate the risks posed by
instruments covered by the authorised representative’s mandate.
(1)
Importers shall place only compliant instruments on Obligations of
importers.
the market.
(2) Before placing on the market an instrument intended to be
used for the applications listed in paragraphs (a) to (f) of regulation
2(3), importers shall ensure that the appropriate conformity
assessment procedure referred to in regulation 14 has been carried out
by the manufacturer. They shall ensure that the manufacturer has
drawn up the technical documentation, that the instrument bears the
CE marking and the supplementary metrology marking and is
accompanied by the required documents, and that the manufacturer
has complied with the requirements set out in regulation 7(5) and (6).
Where an importer considers or has reason to believe that an
instrument intended to be used for the applications listed in
paragraphs (a) to (f) of regulation 2(3) is not in conformity with the
essential requirements set out in Schedule I, he shall not place the
instrument on the market until it has been brought into conformity.
Furthermore, where the instrument presents a risk, the importer shall
inform the manufacturer and the Authority to that effect.
Before placing on the market an instrument not intended to be
used for the applications listed in paragraphs (a) to (f) of regulation
2(3) importers shall ensure that the manufacturer has complied with
the requirements set out in regulation 7(5) and (6).
(3) Importers shall indicate on the instrument their name,
registered trade name or registered trade mark and the postal address
at which they can be contacted. Where this would require the
packaging to be opened, those indications may be given on the
packaging and in a document accompanying the instrument. The
contact details shall be in at least English and, or Maltese.
(4) Importers shall ensure that the instrument intended to be
used for the applications listed in paragraphs (a) to (f) of regulation
2(3) is accompanied by instructions and information in a language
which can be easily understood by end-users, as determined by the
Authority.
(5) Importers shall ensure that, while an instrument intended
to be used for the applications listed in paragraphs (a) to (f) of
regulation 2(3) is under their responsibility, its storage or transport
conditions do not jeopardise its compliance with the essential
requirements set out in Schedule I.
(6) When deemed appropriate with regard to the risks
presented by an instrument intended to be used for the applications
listed in paragraphs (a) to (f) of regulation 2(3), importers shall carry
out sample testing of instruments made available on the market,
investigate, and, if necessary, keep a register of complaints, of non-
conforming instruments and instrument recalls, and shall keep
distributors informed of any such monitoring.
(7) Importers who consider or have reason to believe that an
instrument which they have placed on the market is not in conformity
with these regulations shall immediately take the corrective measures
necessary to bring that instrument into conformity, to withdraw it or
recall it, if appropriate. Furthermore, where the instrument presents a
risk, importers shall immediately inform the Authority, giving details,
in particular, of the non-compliance and of any corrective measures
taken.
(8) For the instruments intended to be used for the applications
listed in paragraphs (a) to (f) of regulation 2(3), importers shall, for
10 years after the instrument has been placed on the market keep a
copy of the EU declaration of conformity at the disposal of the market
surveillance authorities and ensure that the technical documentation
can be made available to those authorities, upon request.
(9) Importers shall, further to a reasoned request from the
Authority, provide it with all the information and documentation in
paper or electronic form necessary to demonstrate the conformity of
an instrument in at least English and, or Maltese. They shall
cooperate with that authority, at its request, on any action taken to
eliminate the risks posed by instruments which they have placed on
the market.
(1) When making an instrument available on the market
distributors shall act with due care in relation to the requirements of
these regulations.
(2) Before making an instrument intended to be used for the
applications listed in paragraphs (a) to (f) of regulation 2(3) available
on the market, distributors shall verify that the instrument bears the
CE marking, and the supplementary metrology marking, that it is
accompanied by the required documents and by instructions and
information in at least English and, or Maltese, and that the
manufacturer and the importer have complied with the requirements
set out in regulation 7(5) and (6) and regulation 9(3) respectively.
Where a distributor considers or has reason to believe that an
instrument intended to be used for the applications listed in
paragraphs (a) to (f) of regulation 2(3) is not in conformity with the
essential requirements set out in Schedule I, he shall not make the
instrument available on the market until it has been brought into
conformity. Furthermore, where the instrument presents a risk, the
distributor shall inform the manufacturer or the importer to that effect
as well as the Authority.
Before making an instrument not intended to be used for the
applications listed in paragraphs (a) to (f) of regulation 2(3) available
on the market, distributors shall verify that the manufacturer and the
importer have complied with the requirements set out in regulation
7(5) and (6) and regulation 9(3) respectively.
(3) Distributors shall ensure that, while an instrument intended
to be used for the applications listed in paragraphs (a) to (f) of
regulation 2(3) is under their responsibility, its storage or transport
conditions do not jeopardise its compliance with the essential
requirements set out in Schedule I.
(4) Distributors who consider or have reason to believe that an
instrument which they have made available on the market is not in
conformity with these regulations shall make sure that the corrective
measures necessary to bring that instrument into conformity, to
withdraw it or recall it, if appropriate, are taken. Furthermore, where
the instrument presents a risk, distributors shall immediately inform
the Authority to that effect, giving details, in particular, of the non-
compliance and of any corrective measures taken.
An importer or distributor shall be considered a Cases in which
manufacturer for the purposes of these regulations and he shall be obligations
manufacturers
subject to the obligations of the manufacturer under regulation 7, apply
importers and
where he places an instrument on the market under his name or trade distributors.
mark or modifies an instrument already placed on the market in such
a way that compliance with these regulations may be affected.
(1) For instruments intended to be used for the Identification of
applications listed in paragraphs (a) to (f) of regulation 2(3), economic
operators.
economic operators shall, on request, identify the following to the
Authority
(a)
any economic operator who has supplied them with
an instrument;
(b) any economic operator to whom they have supplied
an instrument.
(2) Economic operators shall be able to present the
information referred to in sub-regulation (1) for 10 years after they
have been supplied with the instrument and for 10 years after they
have supplied the instrument.
Instruments which are in conformity with harmonised
standards or parts thereof the references of which have been
published in the Official Journal of the European Union shall be
presumed to be in conformity with the essential requirements set out
in Schedule I covered by those standards or parts thereof.
(1) The conformity of instruments to the essential
requirements set out in Schedule I may be established by either of the
following conformity assessment procedures as selected by the
manufacturer
(a)
Module B as set out in point 1 of Schedule II,
followed either by Module D as set out in point 2 of Schedule II,
or by Module F as set out in point 4 of Schedule II
Provided that Module B shall not be compulsory for
instruments which do not use electronic devices and the load-
measuring device of which does not use a spring to balance the
load. For those instruments not submitted to Module B, Module
D1 as set out in point 3 of Schedule II or Module F1 as set out in
point 5 of Schedule II shall apply;
(b)
Module G as set out in point 6 of Schedule II.
(2) Any documents and correspondence relating to the
conformity assessment procedures referred to in sub-regulation (1)
shall be drawn up in at least English and, or Maltese.
(1) The EU declaration of conformity shall state that the
fulfilment of the essential requirements set out in Schedule I has been
demonstrated.
(2) The EU declaration of conformity shall have the model
structure set out in Schedule IV, shall contain the elements specified
in the relevant modules set out in Schedule II and shall be
continuously updated. It shall be in at least English and, or Maltese.
(3) Where an instrument is subject to more than one Union act
requiring an EU declaration of conformity, a single EU declaration of
conformity shall be drawn up in respect of all such Union acts. That
declaration shall contain the identification of the Union acts
concerned including their publication references.
(4) By drawing up the EU declaration of conformity, the
manufacturer shall assume responsibility for the compliance of the
instrument with the requirements laid down in these regulations.
The conformity of an instrument intended to be used for Conformity
the applications listed in paragraphs (a) to (f) of regulation 2(3) with marking.
these regulations shall be indicated by the presence, on the
instrument, of the CE marking and the supplementary metrology
marking as specified in regulation 17
(1) The CE marking shall be subject to the general General
principles of the
principles set out in Article 30 of Regulation (EC) No 765/2008
CE marking and
of the
supplementary
(2) The supplementary metrology marking shall consist of the metrology
capital letter ‘M’ and the last two digits of the year of its affixing, marking.
surrounded by a rectangle. The height of the rectangle shall be equal
to the height of the CE marking.
(3) The general principles set out in Article 30 of Regulation
(EC) No 765/2008 shall apply, mutatis mutandis, to the
supplementary metrology marking.
(1) The CE marking and the supplementary metrology Rules and
marking shall be affixed visibly, legibly and indelibly to the conditions
affixing the CE
instrument or to its data plate.
marking, the
supplementary
metrology
(2) The CE marking and the supplementary metrology marking
other markings.
marking shall be affixed before the instrument is placed on the
market.
(3) The supplementary metrology marking shall immediately
follow the CE marking.
(4) The CE marking and the supplementary metrology
marking shall be followed by the identification number(s) of the
notified body or bodies involved in the production control phase as
set out in Schedule II.
The identification number of the notified body shall be affixed
by the body itself or, under its instructions, by the manufacturer or his
authorised representative.
(5) The CE marking, the supplementary metrology marking
and the identification number(s) of the notified body or bodies may
be followed by any other mark indicating a special risk or use.
(6) The Authority shall build upon existing mechanisms to
ensure correct application of the regime governing the CE marking
and shall take appropriate action in the event of improper use of that
marking.
The symbol referred to in regulation 7(5) and specified in
point 3 of Schedule III shall be affixed to the devices in a clearly
visible and indelible form.
The Authority shall notify the Commission and the other
Member States of bodies authorised to carry out third-party
conformity assessment tasks under these regulations.
(1) The Authority shall be responsible for setting up and
carrying out the necessary procedures for the assessment and
notification of conformity assessment bodies and the monitoring of
notified bodies, including compliance with regulation 26.
(2) The assessment and monitoring referred to in sub-
regulation (1) shall be carried out by the National Accreditation
Board (Malta).
(1) The Authority shall ensure that no conflict of
interest with conformity assessment bodies occurs.
(2) The Authority shall be organised and operated so as to
safeguard the objectivity and impartiality of its activities.
(3) The Authority shall be organised in such a way that each
decision relating to notification of a conformity assessment body is
taken by competent persons different from those who carried out the
assessment.
(4) The Authority shall not offer or provide any activities that
conformity assessment bodies perform or consultancy services on a
commercial or competitive basis.
(5) The Authority shall safeguard the confidentiality of the
information it obtains.
(6) The Authority shall have a sufficient number of competent
personnel at its disposal for the proper performance of its tasks.
The Authority shall inform the Commission of their
procedures for the assessment and notification of conformity
assessment bodies and the monitoring of notified bodies, and of any
changes thereto.
(1) For the purposes of notification, a conformity
assessment body shall meet the requirements laid down in sub-
regulations (2) to (11).
(2) A conformity assessment body shall be established under
the national law of Malta and have legal personality.
(3) A conformity assessment body shall be a third-party body
independent of the organisation or the instrument it assesses.
A body belonging to a business association or professional
federation representing undertakings involved in the design,
manufacturing, provision, assembly, use or maintenance of
instruments which it assesses, may, on condition that its independence
and the absence of any conflict of interest are demonstrated, be
considered such a body
(4) A conformity assessment body, its top level management
and the personnel responsible for carrying out the conformity
assessment tasks shall not be the designer, manufacturer, supplier,
installer, purchaser, owner, user or maintainer of the instruments
which they assess, nor the representative of any of those parties. This
shall not preclude the use of assessed instruments that are necessary
for the operations of the conformity assessment body or the use of
such instruments for personal purposes.
A conformity assessment body, its top level management and
the personnel responsible for carrying out the conformity assessment
tasks shall not be directly involved in the design, manufacture or
construction, the marketing, installation, use or maintenance of those
instruments, or represent the parties engaged in those activities. They
shall not engage in any activity that may conflict with their
independence of judgement or integrity in relation to conformity
assessment activities for which they are notified. This shall in
particular apply to consultancy services.
Conformity assessment bodies shall ensure that the activities of
their subsidiaries or subcontractors do not affect the confidentiality,
objectivity or impartiality of their conformity assessment activities
(5) Conformity assessment bodies and their personnel shall
carry out the conformity assessment activities with the highest degree
of professional integrity and the requisite technical competence in the
specific field and shall be free from all pressures and inducements,
particularly financial, which might influence their judgement or the
results of their conformity assessment activities, especially as regards
persons or groups of persons with an interest in the results of those
activities.
(6) A conformity assessment body shall be capable of carrying
out all the conformity assessment tasks assigned to it by Schedule II
and in relation to which it has been notified, whether those tasks are
carried out by the conformity assessment body itself or on its behalf
and under its responsibility.
At all times and for each conformity assessment procedure and
each kind or category of instruments in relation to which it has been
notified, a conformity assessment body shall have at its disposal the
necessary
(a)
with technical knowledge and sufficient and
appropriate experience to perform the conformity assessment
tasks;
(b) descriptions of procedures in accordance with which
conformity assessment is carried out, ensuring the transparency
and the ability of reproduction of those procedures. It shall have
appropriate policies and procedures in place that distinguish
between tasks it carries out as a notified body and other
activities;
(c)
procedures for the performance of activities which
take due account of the size of an undertaking, the sector in
which it operates, its structure, the degree of complexity of the
instrument technology in question and the mass or serial nature
of the production process.
A conformity assessment body shall have the means necessary
to perform the technical and administrative tasks connected with the
conformity assessment activities in an appropriate manner and shall
have access to all necessary equipment or facilities.
(7) The personnel responsible for carrying out conformity
assessment tasks shall have the following
(a)
sound technical and vocational training covering all
the conformity assessment activities in relation to which the
conformity assessment body has been notified;
(b) satisfactory knowledge of the requirements of the
assessments they carry out and adequate authority to carry out
those assessments;
(c) appropriate knowledge and understanding of the
essential requirements set out in Schedule I, of the applicable
harmonised standards and of the relevant provisions of Union
harmonisation legislation and of national legislation;
(d)
the ability to draw up certificates, records and
reports demonstrating that assessments have been carried out.
(8) The impartiality of the conformity assessment bodies, their
top level management and of the personnel responsible for carrying
out the conformity assessment tasks shall be guaranteed.
The remuneration of the top level management and personnel
responsible for carrying out the conformity assessment tasks of a
conformity assessment body shall not depend on the number of
assessments carried out or on the results of those assessments.
(9) Conformity assessment bodies shall take out liability
insurance unless liability is assumed by the Government of Malta in
accordance with national law, or the Authority itself is directly
responsible for the conformity assessment.
(10) The personnel of a conformity assessment body shall
observe professional secrecy with regard to all information obtained
in carrying out their tasks under Schedule II or any provision of
national law giving effect to it, except in relation to the competent
authorities of Malta in which its activities are carried out. Proprietary
rights shall be protected.
(11) Conformity assessment bodies shall participate in, or
ensure that their personnel responsible for carrying the conformity
assessment tasks are informed of, the relevant standardisation
activities and the activities of the notified body coordination group
established under the relevant Union harmonisation legislation and
shall apply as general guidance the administrative decisions and
documents produced as a result of the work of that group.
Where a conformity assessment body demonstrates its Presumption of
conformity with the criteria laid down in the relevant harmonised conformity
notified bodies.
standards or parts thereof the references of which have been
published in the Official Journal of the European Union it shall be
presumed to comply with the requirements set out in regulation 24 in
so far as the applicable harmonised standards cover those
requirements .
(1) Where a notified body subcontracts specific tasks Subsidiaries of
connected with conformity assessment or has recourse to a subsidiary, and
subcontracting
it shall ensure that the subcontractor or the subsidiary meets the by
bodies.
requirements set out in regulation 24 and shall inform the Authority
accordingly.
(2) Notified bodies shall take full responsibility for the tasks
performed by subcontractors or subsidiaries wherever these are
established.
(3) Activities may be subcontracted or carried out by a
subsidiary only with the agreement of the client.
(4) Notified bodies shall keep at the disposal of the Authority
the relevant documents concerning the assessment of the
qualifications of the subcontractor or the subsidiary and the work
carried out by them under Schedule II.
(1) A conformity assessment body established in Malta
shall submit an application for notification under these regulations to
the Authority.
(2) The application for notification shall be accompanied by a
description of the conformity assessment activities, the conformity
assessment module or modules and the instrument or instruments for
which that body claims to be competent, as well as by an
accreditation certificate issued by the National Accreditation Board
(Malta) attesting that the conformity assessment body fulfils the
requirements laid down in regulation 24.
(1) The Authority may notify only conformity
assessment bodies which have satisfied the requirements laid down in
regulation 24
(2) The Authority shall notify the Commission and the other
Member States using the electronic notification tool developed and
managed by the Commission.
(3) The notification shall include full details of the conformity
assessment activities, the conformity assessment module or modules
and instrument or instruments concerned and the relevant attestation
of competence.
(4) The body concerned may perform the activities of a
notified body only where no objections are raised by the Commission
or the other Member States within two weeks of a notification where
an accreditation certificate is used.
Only such a body shall be considered a notified body for the
purposes of these regulations.
(5) The Authority shall notify the Commission and the other
Member States of any subsequent relevant changes to the notification.
(1) Where the Authority has ascertained or has been
informed that a notified body no longer meets the requirements laid
down in regulation 24, or that it is failing to fulfil its obligations, the
Authority shall restrict, suspend or withdraw notification as
appropriate, depending on the seriousness of the failure to meet those
requirements or fulfil those obligations. It shall immediately inform
the Commission and the other Member States accordingly.
(2) In the event of restriction, suspension or withdrawal of
notification, or where the notified body has ceased its activity, the
Authority shall take appropriate steps to ensure that the files of that
body are either processed by another notified body or kept available
for the Authority.
(1) Notified bodies shall carry out
conformity Operational
assessment obligations
assessments in accordance with the conformity
notified bodies.
procedures provided for in Schedule II.
(2) Conformity assessments shall be carried out in a
proportionate manner, avoiding unnecessary burdens for economic
operators. Conformity assessment bodies shall perform their activities
taking due account of the size of an undertaking, the sector in which it
operates, its structure, the degree of complexity of the instrument
technology in question and the mass or serial nature of the production
process.
In so doing they shall nevertheless respect the degree of rigour
and the level of protection required for the compliance of the
instrument with these regulations.
(3) Where a notified body finds that the essential requirements
set out in Schedule I or corresponding harmonised standards or other
technical specifications have not been met by a manufacturer, it shall
require that manufacturer to take appropriate corrective measures and
shall not issue a certificate of conformity.
(4) Where, in the course of the monitoring of conformity
following the issue of a certificate, a notified body finds that an
instrument no longer complies; it shall require the manufacturer to
take appropriate corrective measures and shall suspend or withdraw
the certificate if necessary.
(5) Where corrective measures are not taken or do not have the
required effect, the notified body shall restrict, suspend or withdraw
any certificates, as appropriate.
The Authority shall ensure that an appeal procedure Appeal against
decisions of
against decisions of the notified bodies is available.
notified bodies.
(1)
Notified bodies shall inform the Authority of the Information
obligation on
following
notified bodies.
(a)
any refusal, restriction, suspension or withdrawal of
a certificate;
(b) any circumstances affecting the scope of or
conditions for notification;
(c)
any request for information which they have
received from market surveillance authorities regarding
conformity assessment activities;
(d) on request, conformity assessment activities
performed within the scope of their notification and any other
activity performed, including cross-border activities and
subcontracting.
(2) Notified bodies shall provide the other bodies notified
under these regulations carrying out similar conformity assessment
activities covering the same instruments with relevant information on
issues relating to negative and, on request, positive conformity
assessment results.
Notified bodies shall participate, directly or by means of
designated representatives, in the appropriate coordination and
cooperation groups for notified bodies organised by the Commission.
Regulation 15(3) and Articles 16 to 29 of Regulation (EC)
No 765/2008 shall apply to instruments covered by regulation 2 of
these regulations.
(1) Where the Authority has sufficient reason to believe
that an instrument covered by these regulations presents a risk to
aspects of public interest protection covered by these regulations, it
shall carry out an evaluation in relation to the instrument concerned
covering all relevant requirements laid down in these regulations. The
relevant economic operators shall cooperate as necessary with the
Authority for that purpose.
Where, in the course of the evaluation referred to above, the
Authority finds that the instrument does not comply with the
requirements laid down in these regulations, it shall without delay
require the relevant economic operator to take all appropriate
corrective actions to bring the instrument into compliance with those
requirements, to withdraw the instrument from the market, or to recall
it within a reasonable period, commensurate with the nature of the
risk, as they may prescribe.
The Authority shall inform the relevant notified body
accordingly.
Article 21 of Regulation (EC) No 765/2008 shall apply to the
measures referred to in this regulation.
(2) Where the Authority considers that non-compliance is not
restricted to Malta, it shall inform the Commission and the other
Member States of the results of the evaluation and of the actions
which they have required the economic operator to take.
(3) The economic operator shall ensure that all appropriate
corrective action is taken in respect of all the instruments concerned
that it has made available on the market throughout the Union.
(4) Where the relevant economic operator does not take
adequate corrective action within the period referred to in sub-
regulation (1), the Authority shall take all appropriate provisional
measures to prohibit or restrict the instrument’s being made available
on their national market, to withdraw the instrument from that market
or to recall it.
The Authority shall inform the Commission and the other
Member States, without delay, of those measures.
(5) All information shall include all available details, in
particular the data necessary for the identification of the non-
compliant instrument, the origin of the instrument, the nature of the
non-compliance alleged and the risk involved, the nature and duration
of the national measures taken and the arguments put forward by the
relevant economic operator. In particular, the Authority shall indicate
whether the non- compliance is due to either of the following
(a)
failure of the instrument to meet requirements
relating to the aspects of public interest protection laid down in
these regulations; or
(b) shortcomings in the harmonised standards referred
to in regulation 13 conferring a presumption of conformity.
(6) Member States other than the Member State initiating the
procedure under this regulation shall without delay inform the
Commission and the other Member States of any measures adopted
and of any additional information at their disposal relating to the non-
compliance of the instrument concerned, and, in the event of
disagreement with the adopted national measure, of their objections.
(7) Where, within three months of receipt of the information,
no objection has been raised by either a Member State or the
Commission in respect of a provisional measure taken by the
Authority, that measure shall be deemed justified.
(8) The Authority shall ensure that appropriate restrictive
measures, such as withdrawal of the instrument from the market, are
taken in respect of the instrument concerned without delay.
If the national measure is considered justified, the
Authority shall take the necessary measures to ensure that the non-
compliant instrument is withdrawn from the market, and shall inform
the Commission accordingly. If the national measure is considered
unjustified, the Authority shall withdraw that measure.
(1) Where, having carried out an evaluation under
regulation 35, the Authority finds that although an instrument is in
compliance with these regulations, it presents a risk to aspects of
public interest protection, it shall require the relevant economic
operator to take all appropriate measures to ensure that the instrument
concerned, when placed on the market, no longer presents that risk, to
withdraw the instrument from the market or to recall it within a
reasonable period, commensurate with the nature of the risk, as it may
prescribe .
(2) The economic operator shall ensure that corrective action
is taken in respect of all the instruments concerned that he has made
available on the market throughout the Union.
(3) The Authority shall immediately inform the Commission
and the other Member States that information shall include all
available details, in particular the data necessary for the identification
of the instrument concerned, the origin and the supply chain of the
instrument, the nature of the risk involved and the nature and duration
of the national measures taken.
(1) Without prejudice to regulation 35, where the
Authority makes one of the following findings, it shall require the
relevant economic operator to put an end to the non-compliance
concerned
(a)
the CE marking or the supplementary metrology
marking has been affixed in violation of Article 30 of
Regulation (EC) No 765/2008 or of regulation 18 of these
regulations;
(b) the CE marking or the supplementary metrology
marking has not been affixed;
(c)
the inscriptions provided for in regulation 7(5) have
not been affixed or have been affixed in violation of regulation
7(5);
(d) the identification number of the notified body, where
that body is involved in the production control phase, has been
affixed in violation of regulation 17 or has not been affixed;
(e)
the EU declaration of conformity has not been
drawn up;
(f)
the EU declaration of conformity has not been
drawn up correctly;
(g) technical documentation is either not available or
not complete;
(h) the information referred to in regulation 7(6) or 9(3)
is absent, false or incomplete;
(i)
any other administrative requirement provided for in
regulation 7 or 9 is not fulfilled.
(2) Where the non-compliance referred to in regulation 38(1)
persists, the Authority shall take all appropriate measures to restrict
or prohibit the instrument being made available on the market or
ensure that it is recalled or withdrawn from the market .
Penalties applicable to infringements by economic Penalties.
operators of the provisions of national law adopted pursuant to these
regulations are stipulated in Part VI of Metrology Act.
(1) The Authority shall not impede the making available Transitory
on the market and/or the putting into service of instruments covered provisions.
by Regulation 2009/23/EC which are in conformity with that
Regulation and which were placed on the market before 20th April,
2016. (2) Certificates issued under Regulation 2009/23/EC shall be
valid under these regulations.