Regulations, 2010
Statutory Text (4 Articles)
Suppliment tal-Gazzetta tal-Gvern ta’ Malta Nru. 18,531, 8 ta’Jannar, 2010
Taqsima B
L.N. 21 of 2010
PRODUCT SAFETY ACT
(CAP. 427)
In Vitro Diagnostic Medical Devices (Amendment)
Regulations, 2010
IN exercise of the powers conferred by article 38 of the Product
Safety Act, the Minister of Finance, the Economy and Investment, on
the advice of the Malta Standards Authority, has made the following
regulations
(1) The title of these regulations is the In Vitro Citation, scope
Diagnostic Medical Devices (Amendment) Regulations, 2010, and and
commencement.
they shall be read and construed as one with the In Vitro Diagnostic S.L. 427.16.
Medical Devices Regulations, hereinafter referred to as "the principal
regulations".
(2) These regulations transpose Commission Decision
2009/ 108/EC * amending Decision 2002/364/EC ** on common
technical specifications for in vitro diagnostic medical devices.
(3) These regulations shall apply to in vitro diagnostic medical
devices falling under the scope of the principal regulations.
(4) These regulations shall enter into force on 1st
December 2009. For devices placed on the market prior to this
date, these regulations shall apply from 1st December 2010.
Manufacturers are allowed to apply the requirements set out in these
regulations before such dates.
Schedule XI to the principal regulations shall be Substitutes
Schedule XI to
substituted by the following
the principal
regulations.
* OJ L.39, 10-02-2009, p.34.
** OJ L.131, 16-05-2002, p.17.
"Schedule XI
Based on Annex I of Commission Decision 2009/108/EC of 3 February 2009 on
common technical specifications (CTS) for in vitro diagnostic medical devices
1. SCOPE
The common technical specifications set out in this Schedule shall apply
for the purposes of Schedule II, List A.
2. DEFINITIONS AND TERMS
(Diagnostic) sensitivity
The probability that the device gives a positive result in the presence of
the target marker.
True positive A specimen known to be positive for the target marker
and correctly classified by the device.
False negative A specimen known to be positive for the target marker
and misclassified by the device.
(Diagnostic) specificity The probability that the device gives a negative
result in the absence of the target marker.
False positive
A specimen known
to be negative for the target marker and
misclassified by the device.
True negative
A specimen known to be negative for the target marker and correctly
classified by the device.
Analytical sensitivity
Analytical sensitivity may be expressed as the limit of detection, i.e. the
smallest amount of the target marker that can be precisely detected.
Analytical specificity
Analytical specificity means the ability of the method to determine
solely the target marker.
Nucleic acid amplification techniques (NAT)
The term ‘NAT’ is used for tests for the detection and/or quantification
of nucleic acids by either amplification of a target sequence, by amplification
of a signal or by hybridisation.
Rapid test
‘Rapid test’ means qualitative or semi quantitative in vitro diagnostic
medical devices, used singly or in a small series, which involve non-
automated procedures and have been designed to give a fast result.
Robustness
The robustness of an analytical procedure means the capacity of an analytical
procedure to remain unaffected by small but deliberate variations in method
parameters and provides an indication of its reliability during normal usage.
Whole system failure rate
The whole system failure rate means the frequency of failures when the
entire process is performed as prescribed by the manufacturer.
Confirmation assay
Confirmation assay means an assay used for the confirmation of a reactive
result from a screening assay.
Virus typing assay
Virus typing assay means an assay used for typing with already known
positive samples, not used for primary diagnosis of infection or for screening.
Sero-conversion HIV samples
Sero-conversion HIV samples mean
-
p24 antigen and/or HIV RNA positive and;
-
recognised by all of the antibody screening tests and;
-
positive or indeterminate confirmatory assays.
Early sero-conversion HIV samples
Early sero-conversion HIV samples mean
-
p24 antigen and/or HIV RNA positive and;
-
not recognised by all of the antibody screening tests and;
-
indeterminate or negative confirmatory assays.
COMMON TECHNICAL SPECIFICATIONS
(CTS)
FOR
PRODUCTS REFERRED TO IN SCHEDULE II, LIST A.
3.1 CTS for performance evaluation of reagents and reagent products
for the detection, confirmation and quantification in human specimens of markers
of HIV infection (HIV 1 and 2), HTLV I and II, and hepatitis B, C, D
General Principles
3.1.1 Devices which detect virus infections placed on the market for use
as either screening or diagnostic tests, shall meet the requirements for sensitivity
and specificity set out in Table 1. See also principle 3.1.11 for screening assays.
3.1.2 Devices intended by the manufacturer for testing body fluids other
than serum or plasma, for example urine, saliva, etc shall meet the same CTS
requirements for sensitivity and specificity as serum or plasma tests. The
performance evaluation shall test samples from the same individuals in both the
tests to be approved and in a respective serum or plasma assay.
3.1.3 Devices intended by the manufacturer for self-test, i.e. home use,
shall meet the same CTS requirements for sensitivity and specificity as respective
devices for professional use. Relevant parts of the performance evaluation shall
be carried out (or repeated) by appropriate lay users to validate the operation of
the device and the instructions for use.
3.1.4 All performance evaluations shall be carried out in direct comparison
with an established state of the art device. The device used for comparison
shall be one bearing CE marking, if on the market at the time of the
performance evaluation.
3.1.5 If discrepant test results are identified as part of an evaluation,
these results shall be resolved as far as possible, for example
-
by evaluation of the discrepant sample in further test systems;
-
by use of an alternative method or marker;
-
by a review of the clinical status and diagnosis of the patient; and
-
by the testing of follow-up-samples.
3.1.6 Performance evaluations shall
be performed on a population
equivalent to the European population.
3.1.7 Positive specimens used in the performance evaluation shall be
selected to reflect different stages of the respective disease(s), different
antibody patterns, different genotypes, different subtypes, mutants, etc.
3.1.8 Sensitivity with true positives and sero-conversion samples shall
be evaluated as follows
3.1.8.1 Diagnostic test sensitivity during sero-conversion has to represent
the state of the art. Whether further testing of the same or additional sero-
conversion panels is conducted by the notified body or by the manufacturer
the results shall confirm the initial performance evaluation data (see Table
1). Seroconversion panels should start with a negative bleed(s) and should have
narrow bleeding intervals.
3.1.8.2 For blood screening devices (with the exception of HBsAg and
anti-HBc tests), all true positive samples shall be identified as positive by the
device to be CE marked (Table 1). For HBsAg and anti-HBc tests the new
device shall have an overall performance at least equivalent to that of the
established device (see 3.1.4).
3.1.8.3 Regarding HIV tests
-
all sero-conversion HIV samples shall be identified as positive;
and
-
at least 40 early sero-conversion HIV samples shall be tested.
Results should conform to the state of the art.
3.1.9 Performance evaluation of screening assays shall include 25 positive
(if available in the case of rare infections) "same day" fresh serum and/or
plasma samples (≤ 1 day after sampling).
3.1.10 Negative specimens used in a performance evaluation shall be
defined so as to reflect the target population for which the test is intended, for
example blood donors, hospitalised patients, pregnant women, etc.
3.1.11 For performance evaluations for screening assays (Table 1) blood
donor populations shall be investigated from at least two blood donation
centres and consist of consecutive blood donations, which have not been
selected to exclude first time donors.
3.1.12 Devices shall have a specificity of at least 99.5% on blood
donations, unless otherwise indicated in the accompanying tables. Specificity
shall be calculated using the frequency of repeatedly reactive (i.e. false
positive) results in blood donors negative for the target marker.
3.1.13 Devices shall be evaluated to establish the effect of potential
interfering substances, as part of the performance evaluation. The potential
interfering substances to be evaluated will depend to some extent on the
composition of the reagent and configuration of the assay. Potential
interfering substances shall be identified as part of the risk analysis required
by the essential requirements for each new device but may include, for example
- specimens representing "related" infections;
- specimens from multipara, i.e. women who have had more
than one pregnancy, or rheumatoid factor positive patients;
- for recombinant antigens, human antibodies to components
of the expression system, for example anti E. coli, or anti yeast.
3.1.14 For devices intended by the manufacturer to be used with serum and
plasma the performance evaluation must demonstrate serum to plasma
equivalency. This shall be demonstrated for at least 50 donations (25 positive and
25 negative).
3.1.15 For devices intended for use with plasma the performance
evaluation shall verify the performance of the device using all anticoagulants
which the manufacturer indicates for use with the device. This shall be
demonstrated for at least 50 donations (25 positive and 25 negative).
3.1.16 As part of the required risk analysis the whole system failure rate
leading to false-negative results shall be determined in repeat assays on low-
positive specimens.
3.1.17 If a new in vitro diagnostic medical device belonging to Schedule II,
List A is not specifically covered by the common technical specification, the
common technical specification for a related device should be taken into
account. Related devices may be identified on different grounds, for example
by the same or similar intended use or by similar risks.
3.2 Additional Requirements for HIV antibody/antigen combined tests
3.2.1 HIV antibody/antigen combined tests intended for anti-HIV and
p24 antigen detection which include claims for single p24 antigen detection shall
follow Table 1 and Table 5, including criteria for analytical sensitivity for p24
antigen.
3.2.2 HIV antibody/antigen combined tests intended for anti-HIV and
p24 detection which do not include claims for single p24 detection shall
follow Table 1 and Table 5, excluding criteria for analytical sensitivity for
p24.
3.3 Additional Requirements for Nucleic Acid Amplification Techniques
(NAT)
The performance evaluation criteria for NAT assays can be found in
Table 2.
3.3.1 For target sequence amplification assays, a functionality control for
each test sample (internal control) shall reflect the state of the art. This control
shall as far as possible be used throughout the whole process, i.e. extraction,
amplification/hybridisation, detection.
3.3.2 The analytical sensitivity or detection limit for NAT assays shall be
expressed by the 95% positive cut-off value. This is the analyte concentration
where 95% of test runs give positive results following serial dilutions of an
international reference material for example a WHO standard or calibrated
reference material.
3.3.3 Genotype detection shall be demonstrated by appropriate primer or
probe design validation and shall also be validated by testing characterised
genotyped samples.
3.3.4 Results of quantitative NAT assays shall be traceable to
international standards or calibrated reference materials, if available, and be
expressed in international units utilised in the specific field of application.
3.3.5 NAT assays may be used to detect virus in antibody negative
samples, i.e. pre sero conversion samples. Viruses within immune-complexes
may behave differently in comparison to free viruses, for example during a
centrifugation step. It is therefore important that during robustness studies,
antibody-negative (pre sero-conversion) samples are included.
3.3.6 For investigation of potential carry-over, at least five runs with
alternating high positive and negative specimens shall be performed during
robustness studies. The high positive samples shall comprise of samples with
naturally occurring high virus titres.
3.3.7 The whole system failure rate leading to false-negative results
shall be determined by testing low-positive specimens. Low positive specimens
shall contain a virus concentration equivalent to 3 times the 95% positive cut-off
virus concentration.
3.4 CTS for the manufacturer’s release testing of reagents and eagent
products for the detection, confirmation and quantification in human specimens of
markers of HIV infection (HIV 1 and 2), HTLV I and II, and hepatitis B, C, D
(Immunological assays only).
3.4.1 The manufacturer’s release testing criteria shall ensure that every
batch consistently identifies the relevant antigens, epitopes, and antibodies.
3.4.2. The manufacturer’s batch release testing for screening assays shall
include at least 100 specimens negative for the relevant analyte.
3.5 CTS for performance evaluation of reagents and reagent products
for determining the following blood group antigens: ABO blood group system
ABO1 (A), ABO2 (B), ABO3 (A,B); Rh blood group system RH1 (D), RH2 (C),
RH3 (E), RH4 (c), RH5 (e); Kell blood group system KEL1 (K).
Criteria for performance evaluation of reagents and reagent products for
determining the blood groups antigens: ABO blood group system ABO1 (A),
ABO2 (B), ABO3 (A,B); Rh blood group B 9 system RH1 (D), RH2 (C), RH3 (E),
RH4 (c), RH5 (e); Kell blood group system KEL1 (K) can be found in Table 9.
3.5.1 All performance evaluations shall be carried out in direct comparison
with an established state of the art device. The device used for comparison
shall be one bearing CE marking, if on the market at the time of the
performance evaluation.
3.5.2 If discrepant test results are identified as part of an evaluation,
these results shall be resolved as far as possible, for example
-
by evaluation of the discrepant sample in further test systems;
-
by use of an alternative method.
3.5.3 Performance evaluations shall be performed
on a population
equivalent to the European population.
3.5.4 Positive specimens used in the performance evaluation shall be
selected to reflect variant and weak antigen expression.
3.5.5 Devices shall be evaluated to establish the effect of potential
interfering substances, as part of the performance evaluation. The potential
interfering substances to be evaluated will depend to some extent on the
composition of the reagent and configuration of the assay. Potential
interfering substances shall be identified as part of the risk analysis required
by the essential requirements for each new device.
3.5.6 For devices intended for use with plasma the performance evaluation
shall verify the performance of the device using all anticoagulants which the
manufacturer indicates for use with the device. This shall be demonstrated for at
least 50 donations.
3.6. CTS for the manufacturers release testing of reagents and reagent
products for determining the blood group antigens: ABO blood group system
ABO1 (A), ABO2 (B), ABO3 (A,B); Rh blood group system RH1 (D), RH2 (C),
RH3 (E), RH4 (c), RH5 (e); Kell blood group system KEL1 (K).
3.6.1 The manufacturer’s release testing criteria shall ensure that every
batch consistently identifies the relevant antigens, epitopes, and antibodies.
3.6.2 Requirements for manufacturers batch release testing are outlined in
Table 10.
Table 1: "Screening" assays: anti-HIV 1 and 2, anti-HTLV I and II,
anti-HCV, HBsAg, anti-HBc
anti-HIV-1/2
anti-HTLV-
anti-HCV
HBsAg
anti-HBc
I/II
Diagnostic Positive
400 HIV-1
300 HTLV-I
400 (positive
400 Including 400
sensitivity specimens
100 HIV-2
100 HTLV-
samples)
subtype-
Including
including 40
II
consideration evaluation of
Including
non-B-
other HBV-
samples from
subtypes, all
markers
different stages of
available
infection and
HIV/1
reflecting
subtypes
different
should be
antibody patterns.
represented
by at least 3
Genotype 1-4: >
samples per
20 samples - per
subtype
genotype
(including non-a
sub-types of
genotype 4);
5: > 5 samples;
6: if available
Sero-
20 panels
To be
20 panels
20 panels
To be defined
conversion
10 further
defined
10 further panels
10 further
when
panels
panels (at
when
(at Notified Body
panels (at
available
Notified Body
available
or manufacturer)
Notified Body
or
or
manufacturer)
manufacturer)
Analytical
Standards
0.130 IU/ml
sensitivity
(Second
International
Standard for
HBsAg,
subtype adw2,
genotype A,
NIBSC code
00/588)
Specificity
Unselected
donors
(including1 st
time donors)
Hospitalized
patients
Potentially
cross-
reacting
blood-
specimens
(RF+, related
viruses,
pregnant
women, etc)
Table 2: NAT assays for HIV1, HCV, HBV, HTLV I/II (qualitative and
quantitative; not molecular typing)
HIV1
HCV
HBV
HTLV I/II
Acceptanc
e criteria
NAT
qualitative
quantitative
qualitative
quantitative
quantitative
qualitative
quantitative
As for HIV
As for HIV
As for HIV
quantitative
quantitative
quantitative
Sensitivity
According
Detection
According
According
According to
Detection
to EP
limit: as for
to EP
to EP
EP
limit
validation
qualitative
validation
validation
validation
Detection
guideline
tests;
guideline
guideline
guideline
( 1 ): several
( 1 ): several
( 1 ): several
of
(1)
Quantific-
analytical
several
ation limit
dilution
dilution
dilution
sensitivity
dilution
dilutions
series into
series into
series into
(IU/ml;
series into
(half-log10
borderline
borderline
borderline
defined on
borderline
or less) of
concentra-
concentr-
concentr-
WHO
concentra-
calibrated
tion;
ation;
ation;
standards or
tion;
reference
statistical
statistical
statistical
calibrated
statistical
prepara-
analysis
analysis
analysis
reference
analysis
tions,
(e.g. Probit
(e.g. Probit
(e.g. Probit
materials)
(e.g.
definition of
analysis)
analysis) on
analysis) on
Probit
lower, upper
on the
the basis of
the basis of
analysis)
quantifi-
basis of at
at least 24
at least 24
on the
cation limit,
least 24
replicates;
replicates;
basis of at
precision,
replicates;
calculation
calculation
least 24
accuracy,
calculation
of 95% cut-
of 95% cut-
replicates;
"linear"mea
of 95%
off value
off value
calculatio
suring
cut-off
n of 95%
range,
value
cut-off
"dynamic
value
range".
Reproduci-
bility at
different
concentra-
tion levels
to be shown
Genotype/
At least 10
Dilution
At least 10
As far as
As far as
subtype
samples per
series of all
samples
calibrated
calibrated
detection/
subtype (as
relevant
per
genotype
genotype
quantific-
far as
genotypes/
genotype
reference
reference
ation
available)
subtypes,
(as far as
materials
materials
efficiency
preferably
available)
are
are
of
available
available
reference
materials,
as far as
available
Cell culture
Transcripts
supernatant
or plasmids
s (could
quantified
substitute
by
for rare
appropriate
HIV-1
meethods
subtypes)
may be
According
used
to EP
According
According
According
validation
to EP
to EP
to EP
guideline
validation
validation
validation
( 1 ) as far as
guideline ( 1 )
guideline
guideline
calibrated
( 1 ) as far as
( 1 ) as far as
as far as
subtype
calibrated
calibrated
calibrated
reference
subtype
subtype
subtype
materials
reference
reference
reference
are
materials
materials
materials
available;
are
are
are
in vitro
available;
available; in
available; in
transcripts
in vitro
vitro
vitro
could be an
transcripts
transcripts
transcripts
option
could be an
could be an
could be an
option
option
option
Diagnostic
500 blood
100 blood
500 blood
500 blood
500 blood
specificity
donors
donors
donors
donors
donors
negative
samples
HIV1
HCV
HBV
HTLV I/II
Acceptanc
e criteria
qualitative
quantitative
qualitative
quantitative
quantitative
qualitative
quantitative
As for HIV
As for HIV
As for HIV
quantitative
quantitative
quantitative
By
As for
By assays
By assays
By assays
suitable
qualitative
design and/
design and/
design and/
assay
tests
or testing
or testing of
or testing of
design
of at least
at least 10
at least 10
evidence
10 human
other DNA-
human
(e.g.
flavivirus
virus
retrivirus
sequence
(e.g. HGV,
positive
(e.g. HIV-)
compari-
YFV)
samples
positive
son) and/
positive
samples
or testing
samples
of at least
10 human
retrovirus
(e.g.
HTLV)-
positive
samples
As for
qualitative
tests
At least 5
At least 5
At least 5
At least 5
runs using
runs using
runs using
runs using
alternating
alternating
alternating
alternating
high
high
high
high
positive
positive
positive
positive
(known to
(known to
(known to
(known to
occur
occur
occur
occur
naturally)
naturally)
naturally)
naturally)
and
and
and
and
negative
negative
negative
negative
samples
samples
samples
samples
Internal
Internal
Internal
Internal
control
control
control
control
preferably
preferably
preferably
preferably
to go
to go
to go
to go
through the
through the
through the
through the
whole
whole NAT
whole NAT
whole NAT
NAT
procedure
procedure
procedure
procedure
At least
At least
At least
At least
99 / 100
assays
samples
samples
samples
samples
positive
virus-
virus-
virus-
virus-
spiked with
spiked
spiked with
spiked with
3 x the 95%
with 3 x
3 x the 95
3 x the 95
pos cut-off
the 95%
% pos cut-
% pos cut-
concentra-
pos cut-off
off
off
tion
concentra-
concentra-
concentr-
tion
tion
ation
( 1 ) European Pharmacopoeia guideline
Acceptance criteria for ‘whole system failure rate leading to false-neg results’ is 99/100 assays
positive.
For quantitative NATs a study shall be performed on at least 100 positive specimens reflecting
the routine conditions of users (e.g. no pre-selection of specimens). Comparative results with
another NAT test system shall be generated in parallel.
For qualitative NATs a study on diagnostic sensitivity shall be performed using at least 10 sero-
conversion panels.
Comparative results with another NAT test system shall be generated in parallel.
Table 3: Rapid tests: anti-HIV 1 and 2, anti-HCV, HBsAg, anti-HBc,
anti-HTLV I and II
anti-HIV 1/2
anti-HCV
HBsAg
anti-HBc
anti-HTLV I/II
Acceptance
criteria
Diagnostic Positive
Same criteria
Same criteria
Same criteria
Same criteria
Same criteria
Same criteria as
sensitivity specimens
as for
as for
as for
as for
as for
for screening
screening
screening
screening
screening
screening
assays
assays
assays
assays
assays
assays
Sero-
Same criteria
Same criteria
Same criteria
Same criteria
Same criteria
Same criteria as
conversion
as for
as for
as for
as for
as for
for screening
panels
screening
screening
screening
screening
screening
assays
assays
assays
assays
assays
assays
> 99 % (anti-
Negative
1000 blood
1000 blood
1000 blood
1000 blood
1000 blood
HBc: > 96 %)
specimens
donations 200
donations
donations
donations
donations
clinical
200 clinical
200 clinical
200 clinical
200 clinical
specimens
specimens 200
specimens
specimens
specimens 200
200 samples
samples from
200 samples
samples from
from pregnant
pregnant
from pregnant
pregnant
women
women
women
women
100 potentially 100
100 potentially 100 potentially
interfering
potentially
potentially
interfering
interfering
samples
interfering
interfering
samples
samples
samples
samples
Table 4: Confirmatory/supplementary assays for anti-HIV 1 and 2,
anti-HTLV I and II, anti-HCV, HBsAg
anti-HIV
anti-HTLV
HCV
HBsAg
Acceptance
Confirmatory Confirmatory Supplementary Confirmatory
criteria
Assay
Assay
Assay
Assay
Diagnostic
Positive
200 HIV-1
200 HTLV-I
300 HCV
300 HBsAg
Correct
sensitivity
specimens
and 100 HIV- and 100
(positive
identificationas
HTLV-II
samples)
positive (or
indeterminate),
not negative
Including
Including
Including
samples from
samples from
samples from
different
different stages
different stages
stages of
of infection and
of infection 20
infection and
reflecting
“high pos”
reflecting
different
samples (> 26
different
antibody
IU/ml); 20
antibody
patterns.
samples in the
patterns
Genotypes 1 -
cut-off range
4: > 20 samples
(including non-
a sub-types of
genotype 4);
5: > 5 samples;
6: if available
Sero-con-
15 sero-
version
conversion
seroconversion
seroconversion
panels
panels/ low
panels/ low
panels/ low
titre panels
titre panels
titre panels
anti-HIV
anti-HTLV
HCV
HBsAg
Acceptance
Confirmatory
Confirmatory
Supplementary
Confirmatory
criteria
Assay
Assay
Assay
Assay
Analytical
Standards
Second
sensitivity
International
Standard for
HBsAg,
subtype adw2,
genotype A,
NIBSC code
00/588
Diagnostic
Negative
200 blood
200 blood
200 blood
10 false
No false-
specificity
specimens
donations
donations
donations
positives as
positive
available from
results/ (1) no
the
neutralisation
performance
evaluation of
the screening
assay. (1)
200 clinical
200 clinical
200 clinical
samples
samples
samples
including
including
including
pregnant
pregnant
pregnant
women
women
women
50 potentially
50 potentially
50 potentially
potentially
interfering
interfering
interfering
interfering
samples,
samples,
samples
samples,
including
including
including
samples with
samples with
samples with
indeterminate
indeterminate
indetermin-
results in
results in other
ate results in
other
confirmatory
other
confirmatory
assays
confirmatory
assays
assays
Table 5: HIV 1 Antigen
HIV-1 Antigen Assay
Acceptance criteria
Diagnostic
Positive specimens
50 HIV-1 Ag-positive
Correct identification (after
sensitivity
50 cell culture supernatants
neutralisation)
including different HIV-1
subtypes and HIV-2
Sero-conversion panels
20 sero-conversion panels/
low titre panels
Analytical
Standards
HIV-1 p24 Antigen, 1st
≤ 2 IU/ml
sensitivity
International Reference
Reagent, NIBSC code: 90/
Diagnostic
200 blood donations 200
>99.5% after neutralisation
specificity
clinical samples 50
potentially interfering
samples
Table 6: Serotyping and Genotyping Assay: HCV
HCV Serotyping and Genotyping
Acceptance criteria
Assay
Diagnostic
Positive specimens
200 (positive samples) Including
> 95% agreement between
sensitivity
samples from different stages of
serotyping and genotyping
infection and reflecting different
> 95% agreement between
antibody patterns. Genotypes 1 -
genotyping and sequencing
4: > 20 samples (including non-a
sub-types of genotype 4);
5: > 5 samples;
6: if available
Diagnostic
Negative specimens
specificity
Table 7: HBV Markers: anti-HBs, anti HBc IgM, anti-HBe, HBeAg
anti-HBs
anti-HBc IgM
anti-HBe
HBeAg
Acceptance
criteria
Diagnostic
Positive
100 vaccinees
200 Including
200 Including
200 Including
> 98%
sensitivity
specimens 100 naturally
samples from
samples from
samples from
infected
different stages
different stages
different
persons
of
infection
of
infection
stages
of
(acute/chronic
(acute/chronic
infection
etc.)
etc.)
(acute/chronic
etc.)
The
acceptance
criteria should
only
be
applied
on
samples from
acute infection
stage
Sero-con-
10 follow-ups When availa-
version
or
anti-HBs
ble
panels
sero-
conversions
Analytical
Standards
WHO
1st
HBe
- Anti-HBs
sensitivity
International
Referenzan-
<10 mIU/ml
Reference
tigen 82; PEI
Preparation
Germany
1977; NIBSC,
United
Kingdom
Diagnostic No
500 Including 200
blood
blood 200
blood
>98%
specificity negative
clinical
donations
donations
donations
specimens
samples
clinical
clinical
200 clinical
samples
samples
samples
50 potentially 50 potentially
50 potentially
interfering
interfering
interfering
samples
samples
samples
Table 8: HDV markers: anti-HDV, anti-HDV IgM, Delta Antigen
anti-HDV
anti-HDV IgM
Delta Antigen
Acceptance
criteria
Diagnostic
Positive
>98%
sensitivity
specimens
Specifying
HBV- Specifying HBV-
markers
markers
Diagnostic
Negative
>98%
specificity
specimens
Including
clinical Including clinical Including
clinical
samples
samples
samples
potentially 50
potentially 50
potentially
interfering samples interfering samples interfering samples
Table 9: Blood group antigens in the ABO, Rh and Kell blood group systems
Specificity
Number of tests per
Total number of samples
Total number of samples
recommended method
to be tested for a launch
to be tested for a new
product
formulation, or use of
well-characterized
reagents
Anti-ABO1(anti-A),anti-
Anti-ABO2 (anti-B), Anti-
ABO3 (anti-A,B)
Anti-RH1 (anti-D)
Anti-RH2
(anti-C
),
anti-RH4
(anti-c),
anti-RH3 (anti-E)
Anti-RH5 (anti-e)
Anti-KEL1 (anti-K)
Acceptance criteria
All of the above reagents shall show comparable test results with
established reagents with acceptable performance with regard to claimed
reactivity of the device. For established reagents, where the application or use has
been changed or extended, further testing should be carried out in accordance with
the requirements outlined in column 1 (above).
Performance evaluation of anti-D-reagents shall include tests against a range
of weak RH1 (D) and partial RH1 (D) samples, depending on the intended use of
the product.
Qualifications
Clinical samples
10% of the test population
Neonatal specimens: > 2% of the test population
ABO samples
> 40% A, B positives
‘weak D’
> 2% of RH1 (D) positives
Table 10: Batch release criteria for reagents and reagent products to
determine blood group antigens in the ABO, Rh and Kell blood group systems
Specificity Testing Requirements on each reagent
1. Test reagents
Blood Group Reagents
Minimum number of control cells to be tested
Positive reactions
Negative reactions
A2B
Ax
B
Anti-ABO1 (anti-A)
2*
B
A1B
Anti-ABO2 (anti-B)
Ax
B
Anti-ABO3 (anti-A,B)
R1r
R2r
Weak
r’r
r’’r
rr
D
Anti-RH1 (anti-D)
2*
R1R2
R1r
r’r
R2R2 r’’r
rr
Anti-RH2 (anti-C)
R1R2
R1r
r’r
R1R1
Anti-RH4 (anti-c)
R1R2
R2r
r”r
R1R1 r’r
rr
Anti-RH 3 (anti-E)
R1R2
R2r
r”r
R2R2
Anti-RH5 (anti-e)
Kk
kk
Anti-KEL1 (anti-K)
(*) Only by recommended techniques where reactivity against these antigens is claimed
Note: Polyclonal reagents must be tested against a wider panel of cells to confirm specificity
and exclude presence of unwanted contaminating antibodies.
Acceptance Criteria
Each batch of reagent must exhibit unequivocal positive or negative results by all recommended
techniquesin accordance with the results obtained from the performance evaluation data.
2. Control Materials (red Cells)
The phenotype of red cells used in the control of blood typing reagents listed
above should be confirmed using established device.".