Active Implantable Medical Devices Regulations, 2008
Table of Provisions (15)
Statutory Text (15 Articles)
Suppliment tal-Gazzetta tal-Gvern ta’ Malta, Nru. 18,308, 12 ta’ Settembru, 2008
L.N. 211 of 2008
PRODUCT SAFETY ACT
(CAP. 427)
Active Implantable Medical Devices Regulations, 2008
IN exercise of the powers conferred by articles 38 to 40 of the
Product Safety Act, the Minister of Finance, Economy and Investment,
on the advice of the Malta Standards Authority, has made the following
regulations>-
(1) The title of these regulations is the Active Implantable Citation and
commencement.
Medical Devices Regulations, 2008.
(2) These regulations transpose the following Directives>
a) Council Directive 90#385#EEC 1 of 20 June 1990 on the
approximation of the laws of the Member States relating to active
implantable medical devices< and
b) Directive 2007#47#EC 2 of the European Parliament and
of the Council of 5 September 2007 amending Council Directive
90#385#EEC on the approximation of the laws of the Member States
relating to active implantable medical devices, Council Directive
93#42#EEC concerning medical devices and Directive 98#8#EC
concerning the placing of biocidal products on the market.
(3) These regulations shall enter into force on the 21 st March,
2010. (4)
The following transitional measures shall remain
applicable>
OJ L.189, 20-07-1990, p.17. Directive as last amended by Regulation (EC) No
1882#2003 of the European Parliament and of the Council of 29 September 2003
adapting to Council Decision 1999#468#EC the provisions relating to committees
which assist the Commission in the exercise of its implementing powers laid down
in instruments subject to the procedure referred to in Article 251 of the EC Treaty
(OJ No. L. 284, 31-10-2003, p.1).
OJ L.247, 21-09-2007, p.21.
(a) CE marking shall not be obligatory prior to the 1 st
October 2001 in respect of stocks of active implantable medical
devices already placed on the market prior to 1 st April 2001 and
active implantable medical devices ordered before 1 st April 2001
and imported into Malta up to 30 th September 2001< and
(b) The provisions of these regulations shall not apply to
contracts made before 1 st April 2001, covering the supply of active
implantable medical devices to the Government of Malta over the
period up to and including 31 st December 2002, as well as those
contracts which had terminated before 1 st April 2001 or which,
before 1 st April 2001, were determined or dissolved by agreement
or by a judgement which had become a res judicata, or by operation
of the law< in respect of such contracts the law applicable at the
time, including the time of their termination, determination or
dissolution, shall in so far as necessary, continue to apply.
(1) These regulations shall apply to active implantable
medical devices.
(2) (a) For the purposes of these regulations, the
Regulatory Affairs Directorate will act as the Maltese national competent
authority in terms of the Committee referred to in Article 6 of Directive
93#42#EEC and with respect to the exchange of information, necessary
to enable these regulations to be applied uniformly, with other Member
States and the Commission.
(b) For the purposes of these regulations, the
Regulatory Affairs Directorate will act as Malta’s national competent
authority with respect to exchange of experience organised by the
Commission in order to coordinate the uniform application of these
regulations.
For the purposes of these regulations, the following definitions
shall apply>
“active implantable medical device” means any active medical
device which is intended to be totally or partially introduced,
surgically or medically, into the human body or by medical
intervention into a natural orifice, and which is intended to remain
after the procedure<
“active medical device” means any medical device relying
for its functioning on a source of electrical energy or any source of
power other than that directly generated by the human body or
gravity<
“authorised representative” means any natural or legal person
established in the Community or Malta who, explicitly designated
by the manufacturer, acts and may be addressed by authorities and
bodies in the Community or Malta instead of the manufacturer
with regard to the latter’s obligations under these regulations<
“CE marking” means marking as set out in Schedule IV of
these regulations<
“clinical data” means the safety and, or performance
information that is generated from the use of a device. Clinical
data are sourced from>
(a)
any clinical investigation of the device concerned<
or
(b) any clinical investigation or other studies reported
in the scientific literature, of a similar device for which
equivalence to the device in question can be demonstrated<
or
(c) published and, or unpublished reports on other
clinical experience of either the device in question or a similar
device for which equivalence to the device in question can be
demonstrated<
“the Community” means the European Community<
“custom-made device” means any device specifically made
in accordance with a duly qualified medical practitioner’s written
prescription which gives, under his responsibility, specific design
characteristics and is intended for the sole use of a particular patient.
Mass-produced devices which need to be adapted to meet the
specific requirements of the medical practitioner or any other
professional user shall not be considered to be custom-made
devices<
“device intended for clinical investigation” means any device
intended for use by a duly qualified medical practitioner when
conducting clinical investigations as referred to in Section 2(a) of
Schedule VII to these regulations in an adequate human clinical
environment. For the purpose of conducting clinical investigation,
any other person who, by virtue of his professional qualifications,
is authorised to carry out such investigation shall be accepted as
equivalent to a duly qualified medical practitioner<
“harmonised standard” means a technical specification
adopted by the European Committee for Standardisation (CEN)
or the European Committee for Electrotechnical Standardisation
(CENELEC) or the European Telecommunications Standards
Institute (ETSI), upon a remit from the European Commission in
accordance with the Notification Procedure Regulations, 2003<
“intended purpose” means the use for which the device is
intended according to the data supplied by the manufacturer on
the labelling, in the instructions and, or in promotional material<
“manufacturer””means the natural or legal person with
responsibility for the design, manufacture, packaging and labeling
of a device before it is placed on the market under his own name,
regardless of whether these operations are carried out by that person
himself or on his behalf by a third party>
(a) the obligations of these regulations to be met by
manufacturers also apply to the natural or legal person who
assembles, packages, processes, fully refurbishes and, or labels
one or more ready-made products and, or assigns to them
their intended purpose as a device with a view to their being
placed on the market under his own name<
(b) paragraph (a) hereof does not apply to the person
who, while not a manufacturer within the meaning of this
definition, assembles or adapts devices already on the market
to their intended purpose for an individual patient<
“Market Surveillance Directorate” means the Market
Surveillance Directorate of the Malta Standards Authority as
described in the Establishment of Directorates of the Malta
Standards Authority Order, 2008<
“medical device” means any instrument, apparatus, appliance,
software, material or other article, whether used alone or in
combination, together with any accessories, including the software
intended by its manufacturer to be used specifically for diagnostic
and, or therapeutic purposes and necessary for its proper
application, intended by the manufacturer to be used for human
beings for the purpose of>
(a) diagnosis, prevention, monitoring, treatment or
alleviation of disease,
(b) diagnosis, monitoring, treatment, alleviation of or
compensation for an injury or handicap,
(c) investigation, replacement or modification of the
anatomy or of a physiological process,
(d) control of conception,
and which does not achieve its principal intended action in or on
the human body by pharmacological, immunological or metabolic
means, but which may be assisted in its function by such means<
“national standard transposing the harmonised standard”
means a Standard issued by one of the members of
CEN#CENELEC#ETSI which is declared and noted to be identical
to the Harmonised Standard originally issued by
CEN#CENELEC#ETSI<
“notified body” means a body notified to the Commission
and to the Member States of the Community, assessed according
to the criteria laid down in Schedule VIII to these regulations and
appointed to carry out the procedures referred to in regulations
11(1) to (7) of these regulations. The names of such bodies are
published in the OJ<
“OJ” means the Official Journal of the European Union<
“placing on the market” means the first making available in
return for payment or free of charge of a device other than a device
intended for clinical investigation, with a view to distribution and,
or use on the Community or Maltese market, regardless of whether
it is new or fully refurbished<
“putting into service” means making available to the medical
profession for implantation.
“Regulatory Affairs Directorate” means the Regulatory
Affairs Directorate of the Malta Standards Authority as described
in the Establishment of Directorates of the Malta Standards L.N. 39 of 2008.
Authority Order, 2008.
(1) Where an active implantable medical device is intended Applicability.
to administer a substance defined as a medicinal product within the
meaning of article 2 of the Medicines Act, that device shall be governed Cap. 458.
by these regulations, without prejudice to the provisions of the Medicines
Act with regard to the medicinal product.
(2) (a) Where an active implantable medical device
incorporates, as an integral part, a substance which, if used separately,
may be considered to be a medicinal product within the meaning of
article 2 of the Medicines Act and which is liable to act upon the human
body with action that is ancillary to that of the device, that device shall
be evaluated and authorised in accordance with these regulations.
(b) Where a device incorporates, as an integral part, a
substance which, if used separately, may be considered to be a medicinal
product constituent or a medicinal product derived from human blood
or human plasma within the meaning of article 2 of the Medicines Act
and which is liable to act upon the human body with action that is
ancillary to that of the device, hereinafter referred to as a “human blood
derivative”, that device shall be assessed and authorised in accordance
with these regulations.
(3) These regulations constitute a specific Directive within
the meaning of regulation 4(5) of the Electromagnetic Compatibility
Regulations, 2007.
(4) These regulations shall not apply to>
(a) medicinal products covered by the Medicines Act. In
deciding whether a product falls under that Act or these regulations,
particular account shall be taken of the principal mode of action of
the product<
(b) human blood, blood products, plasma or blood cells of
human origin or to devices which incorporate at the time of placing
on the market such blood products, plasma or cells with the
exception of devices referred to in sub-regulation 4(2)(b) above<
(c) transplants or tissues or cells of human origin or to
products incorporating or derived from tissues or cells of human
origin, with the exception of devices referred to in paragraph 4(2)(b)
above<
(d) transplants or tissues or cells of animal origin, unless a
device is manufactured utilising animal tissue which is rendered
non-viable or non-viable products derived from animal tissue.
(1) The active implantable medical devices referred to in the
definitions of “active implantable medical device”, “custom-made
device” and”“device intended for clinical investigation”, hereinafter
referred to as “devices”, shall satisfy the essential requirements set out
in Schedule I to these regulations which apply to them, account being
taken of the intended purpose of the devices concerned.
(2) All necessary steps shall be taken to ensure that the
devices may be placed on the market and, or put into service only if
they comply with the requirements laid down in these regulations when
duly supplied, properly implanted and, or properly installed, maintained
and used in accordance with their intended purposes.
(3) Where a relevant hazard exists, devices which also
satisfy the definition of machinery in the Machinery Regulations, 2002 L.N. 369 of 2002.
shall also meet the essential health and safety requirements set out in
Schedule I to those regulations to the extent to which those essential
health and safety requirements are more specific than the essential
requirements set out in Schedule 1 to these regulations.
(4) Devices, other than those which are custom-made or
intended for clinical investigations, considered to meet the essential
requirements referred to in regulation 5(1) of these regulations must
bear the CE marking of conformity.
(5) (a) The CE marking of conformity, as shown in
Schedule IX to these regulations, must appear in a visible, legible and
indelible form on the sterile pack and, where appropriate, on the sales
packaging, if any, and on the instruction leaflet.
(b) The CE marking must be followed by the
identification number of the notified body responsible for
implementation of the procedures set out in Schedules II, IV and V to
these regulations.
(6) It is prohibited to affix markings which are likely to
deceive third parties as to the meaning and form of the CE marking.
Any other marking may be affixed to the packaging or to the instruction
leaflet accompanying the device provided that the visibility and legibility
of the CE marking is not thereby reduced.
(7) Compliance with the essential requirements referred to
in regulation 5(1) above shall be presumed in respect of devices which
are in conformity with the relevant national standards adopted pursuant
to harmonised standards the reference of which have been published in
the OJ.
(8) For the purposes of these regulations, reference to
harmonised standards also includes the monographs of the European
Pharmacopoeia notably on the interaction between medicinal products
and materials used in devices containing such medicinal products, the
references of which have been published in the OJ.
(9) (a) Where the devices are subject to other Directives
concerning other aspects and which also provide for the affixing of the
CE marking, the latter shall indicate that the devices are also presumed
to conform to the provisions of the other Directives.
(b) However, where one or more of these Directives
allow the manufacturer, during a transitional period, to choose which
arrangements to apply, the CE marking shall indicate conformity to the
provisions only of those Directives applied by the manufacturer. In this
case, particulars of the Directives applied, as published in the OJ, must
be given in the documents, notices or instructions required by the
Directives and accompanying such devices< these documents, notices
or instructions shall be accessible without it being necessary to destroy
the packaging which keeps the device sterile.
(10) Obstacles shall not be created to the placing on the
market or the putting into service within the Maltese territory of devices
complying with the provisions of these regulations and bearing the CE
marking provided for in regulations 5(4) to (6) of these regulations,
which indicates that they have been the subject of an assessment of
their conformity in accordance with regulations 6(1) to (9) of these
regulations.
(11) Obstacles shall not be created to>
(a) devices intended for clinical investigations being made
available to duly qualified medical practitioners or authorised
persons for that purpose if they satisfy the conditions laid down in
regulations 7(1) to (4) and in Schedule VI to these regulations.
(b) custom-made devices being placed on the market and
put into service if they satisfy the conditions laid down in Schedule
VI to these regulations and are accompanied by the statement,
which shall be available to the particular identified patient, referred
to in that Schedule.
Such devices shall not bear the CE marking.
(12) At trade fairs, exhibitions, demonstrations, etc., no
obstacles shall be created to the showing of devices which do not
conform to these regulations, provided that a visible sign clearly indicates
that such devices do not conform and cannot be marketed or put into
service until they have been made to comply by the manufacturer or his
authorised representative.
(13) When a device is put into service, the information
described in Sections 14, 15, 16 and 17 of Schedule I to these regulations
is required to be in, at least, Maltese and, or English.
(1) In the case of devices other than those which are custom- Conformity
Assessment
made or intended for clinical investigations, the manufacturer must, in Procedures.
order to affix the CE marking, at his own choice>
(a) follow the procedure relating to the EC declaration of
conformity set out in Schedule II to these regulations< or
(b) follow the procedure relating to the EC type-examination
set out in Schedule III to these regulations, coupled with>
(i) the procedure relating to EC verification set out in
Schedule IV to these regulations, or
(ii) the procedure relating to the EC declaration of
conformity to type set out in Schedule V to these regulations.
(2) In the case of custom-made devices, the manufacturer
must draw up the declaration provided for in Schedule VI to these
regulations before placing each device on the market.
(3) Where appropriate, the procedures provided for in
Schedule III, IV and VI to these regulations may be discharged by the
manufacturer’s authorised representative.
(4) The records and correspondence relating to the
procedures referred to in regulations 6(1), (2) and (3) above shall be in,
at least, Maltese and, or English, and, or in a language acceptable to the
notified body defined in regulations 11(1) to (7) of these regulations.
(5) During the conformity assessment procedure for a
device, the manufacturer and, or the notified body shall take account of
the results of any assessment and verification operations which, where
appropriate, have been carried out in accordance with these regulations
at an intermediate stage of manufacture.
(6) Where the conformity assessment procedure involves
the intervention of a notified body, the manufacturer, or his authorised
representative, may apply to a body of his choice within the framework
of the tasks for which the body has been notified.
(7) The notified body may require, where duly justified,
any information or data which is necessary for establishing and
maintaining the attestation of conformity in view of the chosen
procedure.
(8) Decisions taken by the notified bodies in accordance
with Schedules II, III and V to these regulations shall be valid for a
maximum of five years and may be extended on application, made at a
time agreed in the contract signed by both parties, for further periods of
a maximum length of five years.
(9) By derogation from regulations 6(1) and (2) hereof, the
Director of Regulatory Affairs, after consultation with the
Superintendent of Public Health, may authorise, on duly justified request,
the placing on the market and putting into service, within the Maltese
territory, of individual devices for which the procedures referred to in
regulations 6(1) and (2) of these regulations have not been carried out
and the use of which is in the interest of protection of health.
(10) The Regulatory Affairs Directorate shall submit a duly
substantiated request to the Commission and ask it to take the necessary
measures in the following situations>
(a) the Maltese authorities consider that the conformity of
a device or family of devices should be established, by way of
derogation from the provisions of regulations 6(1) to (9) above,
by applying solely one of the given procedures chosen from among
those referred to in regulations 6(1) to (9) of these regulations<
(b) the Maltese authorities consider that a decision is
required as to whether a particular product or product group falls
within the definition of “active implantable medical device”,
“custom-made device”, “device intended for clinical investigation”
or “medical device”.
(1) In the case of devices intended for clinical investigations,
the manufacturer or the authorised representative shall, at least 60 days
before the commencement of the investigations, submit the statement
referred to in Schedule VI to these regulations to the Regulatory Affairs
Directorate and the competent authorities in other Member States in
which the investigations are to be conducted.
(2) (a) The manufacturer may commence the relevant
clinical investigations at the end of a period of 60 days after notification,
unless the Regulatory Affairs Directorate or the Ethics Committee
established by the Clinical Trials Regulations, 2004, have notified him,
within that period, of a decision to the contrary, based on considerations
of public health or public policy.
(b) However, the Regulatory Affairs Directorate may
authorise manufacturers to start the clinical investigations in question
before the expiry of the 60-day period, provided that the Ethics
Committee mentioned above has issued a favourable opinion with
respect to the investigation programme in question, including its review
of the clinical investigation plan.
(3) If necessary, any appropriate steps to ensure public
health and public policy shall be taken according to the Public Health Cap. 465.
Act.
(4) Where a clinical investigation is refused or halted by
the Maltese authorities, the Regulatory Affairs Directorate shall
communicate such decision and the grounds thereof to all Member States
and the Commission. Where the Regulatory Affairs Directorate have
called for a significant modification or temporary interruption of a
clinical investigation, the Regulatory Affairs Directorate shall inform
the Member States concerned about such actions and the grounds for
the actions taken.
(5) The manufacturer or his authorised representative shall
notify the Regulatory Affairs Directorate and the competent authorities
of the other Member States concerned of the end of the clinical
investigation, with a justification in case of early termination. In the
case of early termination of the clinical investigation on safety grounds
this notification shall be communicated to all Member States and the
Commission. The manufacturer or his authorised representative shall
keep the report referred to in Section 2(c)(vii) of Schedule VII to these
regulations at the disposal of the Market Surveillance Directorate, the
Regulatory Affairs Directorate and the competent authorities of the other
Member States concerned.
(6) Clinical investigations shall be conducted in accordance
with the provisions of Schedule VII to these regulations.
(1) Where it is found that the devices referred to in the Withdrawal and
prohibition from
definitions of “active implantable medical device” and “custom-made market.
device”, correctly put into service and used in accordance with their
intended purpose, may compromise the health and, or safety of patients,
users or, where applicable, other persons, all appropriate measures shall
be taken by the Market Surveillance Directorate to withdraw such
devices from the market or prohibit or restrict their being placed on the
market or their being put into service.
The Regulatory Affairs Directorate shall immediately inform the
Commission of any such measure, indicating the reasons for such
decision and, in particular, whether non-compliance with these
regulations is due to>
(a) failure to meet the essential requirements referred to in
regulation 5(1), where the device does not meet in full or in part
the standards referred to in regulation 5(7)<
(b) incorrect application of those standards<
(c)
shortcomings in the standards themselves.
(2) Where a device which does not comply bears the CE
marking, the Market Surveillance Directorate shall take appropriate
action against whomsoever has affixed the mark. The Regulatory Affairs
Directorate shall inform the Commission and other Member States of
such actions.
(3) The Market Surveillance Directorate and the Regulatory
Affairs Directorate shall take the necessary steps to ensure that
information brought to their knowledge regarding the incidents
mentioned below involving a device is recorded and evaluated in a
centralised manner>
(a) any malfunction of or deterioration in the characteristics
and performances of a device, as well as any inadequacy in the
labelling or in the instructions for use which might lead to or might
have led to the death of a patient or user or to a serious deterioration
in his state of health<
(b) any technical or medical reason in relation to the
characteristics or performances of a device for the reasons referred
to in paragraph (a) above, leading to systematic recall of devices
of the same type by the manufacturer.
(4) The necessary steps shall be taken to ensure that the
manufacturer of the device concerned, or his authorised representative,
is also informed of the incident referred to in sub-regulation (3) above.
(5) (a) An assessment of the incidents mentioned in sub-
regulation (3) above will be carried out by the Regulatory Affairs
Directorate, if possible together with the manufacturer or his authorised
representative.
(b) The Regulatory Affairs Directorate shall, without
prejudice to sub-regulations (1) and (2) hereof, immediately inform the
Commission and the other Member States of the incidents referred to in
sub-regulation (3) hereof and of the measures that have been taken or
are contemplated to minimise the recurrence of such incidents, including
information on the underlying incidents.
(6) (a) Where the Market Surveillance Directorate
considers, in relation to a given product or group of products, that, in
order to ensure the protection of health and safety and, or to ensure that
public health requirements are observed, such products should be
withdrawn from the market, or their placing on the market and putting
into service should be prohibited, restricted or subjected to particular
requirements, any necessary and justified transitional measures may be
taken.
(b) The Regulatory Affairs Directorate shall then
inform the Commission and all other Member States of these transitional
measures, giving the reasons for such decisions.
(7) Any decision taken pursuant to these regulations>
(a) to refuse or restrict the placing on the market or the
putting into service of a device or the carrying out of clinical
investigations< or
(b) to withdraw devices from the market,
shall state the exact grounds on which it is based. Such decisions
shall be notified without delay to the party concerned, who shall at
the same time be informed of the remedies available to him under
the Maltese law and of the time limits to which such remedies are
subject.
(8) In the event of a decision as referred to in sub-regulation
(7) hereof, the manufacturer or his authorised representative shall have
an opportunity to put forward his viewpoint in advance, unless such
consultation is not possible because of the urgency of the measure to be
taken.
(9) Without prejudice to sub-regulations (1) and (2) hereof>
(a) where it is established that the CE marking has been
affixed unduly or is missing in violation of these regulations, the
manufacturer or his authorised representative shall be obliged to
end the infringement under conditions imposed by the Market
Surveillance Directorate.
(b) where non-compliance continues, all appropriate
measures shall be taken by the Director of Market Surveillance to
restrict or prohibit the placing on the market of the product in
question or to ensure that it is withdrawn from the market, in
accordance with the procedure as laid down in sub-regulations (1)
and (2) hereof.
(10) The provisions of sub-regulation (9) hereof shall also
apply where the CE marking has been affixed in accordance with the
procedures in these regulations, but inappropriately, on products that
are not covered by these regulations.
(1) Any manufacturer who has his registered place of
business in Malta and, under his own name, places devices on the market
in accordance with the procedure referred to in regulation 6(2) above
shall inform the Regulatory Affairs Directorate of the address of the
registered place of business and the description of the devices concerned.
(2) All data allowing for the devices to be identified, together
with the label and instructions for use, may be requested when such
devices are put into service in Malta.
(3) (a) Where a manufacturer who places a device on the
Maltese or Community market under his own name does not have a
registered place of business in a Member State, he shall designate a
single authorised representative in the European Union.
(b) When the authorised representative referred to in
sub-regulation (3)(a) has a registered place of business in Malta, he
shall inform the Regulatory Affairs Directorate of the details referred
to in sub-regulation (1) hereof.
(4) The Regulatory Affairs Directorate shall, on request,
inform the other Member States and the Commission of the details
referred to in sub-regulation (2) hereof given the manufacturer or
authorised representative.
(1) Regulatory data in accordance with these regulations
shall be stored in a European databank accessible to the competent
authorities to enable them to carry out their tasks relating to these
regulations on a well-informed basis.
(2) The databank shall contain the following>
(a) data relating to certificates issued, modified,
supplemented, suspended, withdrawn or refused according to the
procedures as laid down in Schedules II to V to these regulations<
(b) data obtained in accordance with the vigilance procedure
as defined in regulations 8(3) to (5) of these regulations<
(c) data relating to clinical investigations referred to in
regulations 7(1) to (7) of these regulations.
(3) Data shall be forwarded in a standardised format by the
Regulatory Affairs Directorate.
(1) The Regulatory Affairs Directorate shall notify the Notified bodies.
Commission and other Member States of the bodies which it has
designated for carrying out the tasks pertaining to the procedures referred
to in regulations 6(1) to (9) of these regulations together with the specific
tasks for which the bodies have been designated.
(2) The Regulatory Affairs Directorate shall apply the
criteria set out in Schedule VIII to these regulations for the designation
of bodies. Bodies that meet the criteria laid down in the national standards
which transpose the relevant harmonised standards shall be presumed
to meet the relevant criteria.
(3) The Regulatory Affairs Directorate shall withdraw the
notification if it finds that the body no longer meets the criteria referred
to in sub-regulation (2) hereof. It shall immediately inform the other
Member States and the Commission thereof.
(4) The notified body and the manufacturer, or his authorised
representative, shall lay down, by common accord, the time limits for
completion of the evaluation and verification operations referred to in
Schedules II to V to these regulations.
(5) The notified body shall inform the Regulatory Affairs
Directorate about all certificates issued, modified, supplemented,
suspended, withdrawn or refused and the other notified bodies within
the scope of these regulations about certificates suspended, withdrawn
or refused and, on request, about certificates issued. Provided that, the
notified body shall also make available, on request, all additional relevant
information.
(6) (a) Where a notified body finds that pertinent
requirements of these regulations have not been met or are no longer
met by the manufacturer or where a certificate should not have been
issued, it shall, taking account of the principle of proportionality, suspend
or withdraw the certificate issued or place any restrictions on it unless
compliance with such requirements is ensured by the implementation
of appropriate corrective measures by the manufacturer.
(b) In the case of suspension or withdrawal of the
certificate or of any restriction placed on it or in cases where an
intervention of the competent authority may become necessary, the
notified body shall inform the Regulatory Affairs Directorate thereof.
The Regulatory Affairs Directorate shall inform the other Member States
and the Commission.
(7) The notified body shall, on request, supply all relevant
information and documents including budgetary documents, required
to enable the Regulatory Affairs Directorate to verify compliance with
the criteria laid down in Schedule VIII to these regulations.
(1) Without prejudice to the existing national provisions
and practices on medical confidentiality, the Market Surveillance
Directorate shall ensure that all the parties involved in the application
of these regulations are bound to observe confidentiality with regard to
all information obtained in carrying out their tasks.
Provided that the provisions of sub-regulation (1) hereof do
not affect the obligation of the above authority with regard to mutual
information and the dissemination of warnings, nor the obligations of
the persons concerned to provide information under criminal law.
(2)
The following information shall not be treated as
confidential>
(a) information on the registration of persons responsible
for placing devices on the market in accordance with regulations
9(1) to (4) of these regulations<
(b) information to users sent out by the manufacture,
authorised representative or distributor in relation to a measure in
accordance with regulations 8(3) to (5) of these regulations<
(c) information contained in certificates issued, modified,
supplemented, suspended or withdrawn.
The penalties applicable for the infringement of any of the Penalties.
provisions above shall be those provided for in Part IV of the Product
Safety Act.
The Active Implantable Medical Devices Regulations, 2001 Revokes
L.N. 66 of 2001.
are hereby being revoked.
Schedule I
ESSENTIAL REQUIREMENTS
GENERAL REQUIREMENTS
(1)
The devices must be designed and manufactured in such a way that, when used under
the conditions and for the purposes laid down, their use does not compromise the
clinical condition or the safety of patients. They must not present any risk to the
persons implanting them or, where applicable, to other persons.
(2)
The devices must achieve the performances intended by the manufacturer, namely
that they are designed and manufactured in such a way that they are suitable for one
or more functions referred to in the definition of a “medical device” as specified by
him.
(3)
The characteristics and performances referred to in Sections 1 and 2 above must not
be adversely affected to such a degree that the clinical condition and safety of the
patients or, as appropriate, of other persons are compromised during the lifetime of
the device anticipated by the manufacturer, where the device is subjected to stresses
which may occur during normal conditions of use.
(4)
The devices must be designed, manufactured and packed in such a way that their
characteristics and performances are not adversely affected in the storage and
transport conditions laid down by the manufacturer (temperature, humidity, etc.).
(5)
Any side effects or undesirable conditions must constitute acceptable risks when
weighed against the performances intended.
(6)
Demonstration of conformity with the essential requirements must include a clinical
evaluation in accordance with Schedule VII to these regulations.
REQUIREMENTS REGARDING DESIGN AND CONSTRUCTION
(7)
The solutions adopted by the manufacturer for the design and construction of the
devices must comply with safety principles taking account of the generally
acknowledged state of the art.
(8)
Implantable devices must be designed, manufactured and packed in a non-reusable
pack according to appropriate procedures to ensure they are sterile when placed on
the market and, in the storage and transport conditions stipulated by the
manufacturer, remain so until the packaging is removed and they are implanted.
(9)
Devices must be designed and manufactured in such a way as to remove or minimise
as far as possible
(a) the risk of physical injury in connection with their physical, including
dimensional, features,
(b) risks connected with the use of energy sources with particular reference, where
electricity is used, to insulation, leakage currents and overheating of the devices,
(c) risks connected with reasonably foreseeable environmental conditions such as
magnetic fields, external electrical influences, electrostatic discharge, pressure or
variations in pressure and acceleration,
(d) risks connected with medical treatment, in particular those resulting from the use
of defibrillators or high-frequency surgical equipment,
(e) risks connected with ionising radiation from radioactive substances included in
the device, in compliance with the protection requirements laid down in the
Nuclear Safety and Radiation Protection Regulations, 2003 and the Ionising
Radiation Medical Exposure Control Regulations, 2004,
(f) risks which may arise where maintenance and calibration are impossible,
including
(i)
excessive increase of leakage currents,
(ii)
ageing of the materials used,
(iii)
excess heat generated by the device,
(iv)
decreased accuracy of any measuring or control mechanism.
(10)
The devices must be designed and manufactured in such a way as to guarantee the
characteristics and performances referred to in Sections 1 to 6 above, with particular
attention being paid to
(a) the choice of materials used, particularly as regards toxicity aspects,
(b) mutual compatibility between the materials used and biological tissues, cells and
body fluids, account being taken of the anticipated use of the device,
(c) compatibility of the devices with the substances they are intended to administer,
(d) the quality of the connections, particularly in respect of safety,
(e) the reliability of the source of energy,
(f) if appropriate, that they are leak proof,
(g) proper functioning of the programming and control system, including software.
For devices which incorporate software or which are medical software in
themselves, the software must be validated according to the state of the art
taking into account the principles of development lifecycle, risk management,
validation and verification.
(11)
Where a device incorporates, as an integral part, a substance which, if used
separately, may be considered to be a medicinal product as defined in Article 2 of the
Medicines Act, and which is liable to act upon the body with action ancillary to that
of the device, the quality, safety and usefulness of the substance must be verified by
analogy with the methods specified in Annex I to Directive 2001/83/EC.
(a) For the substances referred to in the first paragraph, the notified body shall,
having verified the usefulness of the substance as part of the medical device and
taking account of the intended purpose of the device, seek a scientific opinion
from one of the competent authorities designated by the Member States or the
European Medicines Agency (EMEA) acting particularly through its committee
in accordance with Regulation (EC) No 726/2004 3 on the quality and safety of
the substance including the clinical benefit and risk profile of the incorporation
of the substance into the device. When issuing its opinion, the competent
authority or the EMEA shall take into account the manufacturing process and the
data related to the usefulness of incorporation of the substance into the device as
determined by the notified body.
(b) Where a device incorporates, as an integral part, a human blood derivative, the
notified body shall, having verified the usefulness of the substance as part of the
device and taking account of the intended purpose of the device, seek a scientific
opinion from the EMEA, acting particularly through its committee, on the
quality and safety of the substance, including the clinical benefit and risk profile
of the incorporation of the human blood derivative into the device. When issuing
its opinion, the EMEA shall take into account the manufacturing process and the
data related to the usefulness of incorporation of the substance into the device as
determined by the notified body.
(c) Where changes are made to an ancillary substance incorporated in a device, in
particular related to its manufacturing process, the notified body shall be
informed of the changes and shall consult the relevant competent authority (i.e.
the one involved in the initial consultation), in order to confirm that the quality
and safety of the ancillary substance are maintained. The competent authority
shall take into account the data related to the usefulness of the incorporation of
the substance into the device as determined by the notified body, in order to
ensure that the changes have no negative impact on the established benefit and
risk profile of the addition of the substance in the device.
(d) When the relevant competent authority (i.e. the one involved in the initial
consultation) has obtained information on the ancillary substance, which could
have an impact on the established benefit and risk profile of the addition of the
substance to the device, it shall provide the notified body with advice, whether
this information has an impact on the established benefit and risk profile of the
addition of the substance to the device or not. The notified body shall take the
updated scientific opinion into account in reconsidering its assessment of the
conformity assessment procedure.
(12)
The devices and, if appropriate, their component parts must be identified to allow any
3 Regulation (EC) No. 726/2004 of the European Parliament and of the Council of 31 March 2004
laying down Community procedures for the authorisation and supervision of medicinal products for
human and veterinary use and establishing a European Medicines Agency (OJ L.136, 30-04-2004, p.1).
Regulation as last amended by Regulation (EC) No. 1901/2006.
necessary measure to be taken following the discovery of a potential risk in
connection with the devices and their component parts.
(13)
Devices must bear a code by which they and their manufacturer can be unequivocally
identified (particularly with regard to the type of device and year of manufacture); it
must be possible to read this code, if necessary, without the need for a surgical
operation.
(14)
When a device or its accessories bear instructions required for the operation of the
device or indicate operating or adjustment parameters, by means of a visual system,
such information must be understandable to the user and, as appropriate, the patient.
(15)
Every device must bear, legibly and indelibly, the following particular, where
appropriate in the form of generally recognised symbols
(a) On the sterile pack
(i)
the method of sterilisation,
(ii)
an indication permitting this packaging to be recognised as such,
(iii)
the name and address of the manufacturer,
(iv)
a description of the device,
(v)
if the device is intended for clinical investigations, the words
“exclusively for clinical investigations”,
(vi)
if the device is custom-made, the words “custom-made device”,
(vii)
a declaration that the implantable device is in a sterile condition,
(viii) the month and year of manufacture,
(ix)
an indication of the time limit for implanting a device safely,
(b) On the sales packaging
(i)
the name and address of the manufacturer and the name and address of
the authorised representative, where the manufacturer does not have a
registered place of business in Malta or in the Community,
(ii)
a description of the device,
(iii)
the purpose of the device,
(iv)
the relevant characteristics for its use,
(v)
if the device is intended for clinical investigations, the words
“exclusively for clinical investigations”,
(vi)
if the device is custom-made, the words: “custom-made device”,
(vii)
a declaration that the implantable device is in a sterile condition,
(viii) the month and year of manufacture,
(ix)
an indication of the time limit for implanting a device safely,
(x)
the conditions for transporting and staring the device,
(xi)
in the case of device within the meaning of regulation 4(2)(a) of these
regulations, an indication that the device contains a human blood
derivative.
(16)
When placed on the market each device must be accompanied by instructions for use
giving the following particulars
(a) the year of authorisation to affix the CE mark,
(b) the details referred to in Sections 15(a) and 15(b) above, with the exception of
those referred to in the eighth and ninth indents,
(c) the performances referred to in Section 2 of this Schedule and any undesirable
side effects,
(d) information allowing the physician to select a suitable device and the
corresponding software and accessories,
(e) information constituting the instructions for use allowing the physician and,
where appropriate, the patient to use the device, its accessories and software
correctly, as well as information on the nature, scope and times for operating
controls and trials and, where appropriate, maintenance measures,
(f) information allowing, if appropriate, certain risks in connection with
implantation of the device to be avoided,
(g) information regarding the risks of reciprocal interference 4 in connection with the
presence of the device during specific investigations or treatment,
(h) the necessary instructions in the event of the sterile pack being damaged and,
where appropriate, details of appropriate methods of re-sterilisation,
(i) an indication, if appropriate, that a device can be reused only if it is
reconditioned under the responsibility of the manufacturer to comply with the
essential requirements,
4 “Risks of reciprocal interference” means adverse effects on the device caused by instruments present
at the time of investigations or treatment, and vice versa.
(j) date of issue or the latest revision of the instructions for use.
(17)
The instruction leaflet must also include details allowing the physician to brief the
patient on the contra-indications and the precautions to be taken. These details should
cover in particular
(a) information allowing the lifetime of the energy source to be established,
(b) precautions to be taken should changes occur in the device’s performance,
(c) precautions to be taken as regards exposure, in reasonably foreseeable
environmental conditions, to magnetic fields, external electrical influences,
electrostatic discharge, pressure or variations in pressure, acceleration, etc.,
(d) adequate information regarding the medicinal products which the device in
question is designed to administer.
(18)
Confirmation that the device satisfies the requirements in respect of characteristics
and performances, as referred to in Sections 1 to 6 above, in normal conditions of
use, and the evaluation of the side effects or undesirable effects must be based on
clinical data established in accordance with Schedule VII to these regulations.
Schedule II
EC DECLARATION OF CONFORMITY
(Complete quality assurance system)
(1)
The manufacturer shall apply the quality system approved for the design,
manufacture and final inspection of the products concerned as specified in Sections 3
and 4 and shall be subject to EC surveillance as specified in Section 5 of this
Schedule.
(2)
The declaration of conformity is the procedure by means of which the manufacturer
who satisfies the obligations of Section 1 above ensures and declares that the
products concerned meet the provisions of these regulations which apply to them.
The manufacturer or his authorised representative shall affix the CE marking in
accordance with regulations 5(4) to (6) of these regulations and shall draw up a
written declaration of conformity.
This declaration shall cover one or more clearly identified devices by means of
product name, product code or other unambiguous reference and must be kept by the
manufacturer.
The CE marking shall be accompanied by the identification number of the notified
body responsible.
(3)
Quality system
(a) The manufacturer shall make an application for evaluation of his quality system
to a notified body.
The application shall include
(i)
all the appropriate items of information for the category of products
manufacture of which is envisaged,
(ii)
the quality-system documentation,
(iii)
an undertaking to fulfil the obligations arising from the quality system as
approved,
(iv)
an undertaking to maintain the approved quality system in such a way that
it remains adequate and efficacious,
(v)
an undertaking by the manufacturer to institute and keep updated a post-
market surveillance system including the provision referred to in Schedule
VII to these regulations. The undertaking shall include an obligation for
the manufacturer to notify the competent authorities of the following
incidents immediately on learning of them
(1)
any deterioration in the characteristics or performances, and any
inaccuracies in the instruction leaflet for a device which might lead
to or have led to the death of a patient or a deterioration in his state
of health;
(2)
any technical or medical reason resulting in withdrawal of a device
from the market by the manufacturer.
(b) The application of the quality system must ensure that the products confirm to
the provision of these regulations which apply to them at every stage, from
design to final controls.
All the elements, requirements and provisions adopted by the manufacturer for
his quality system shall be documented in a systematic and orderly manner in the
form of written policies and procedures. This quality-system documentation
must make possible a uniform interpretation of the quality policies and
procedures such as quality programmes, quality plans, quality manuals and
quality records. It shall include in particular the corresponding documentation,
data and records arising from the procedures referred to in point (iii) below.
It shall include in particular an adequate description of
(i)
the manufacturer’s quality objectives;
(ii)
the organisation of the business and in particular
(1)
the organisational structures, the responsibilities of the managerial
staff and their organisational authority where quality of design and
manufacture of the products is concerned,
(2)
the methods of monitoring the efficient operation of the quality
system and in particular its ability to achieve the desired quality of
the design and of the products, including control of products which
do not conform,
(3)
where the design, manufacture and, or final inspection and testing
of the products, or elements thereof, is carried out by a third party,
the methods of monitoring the efficient operation of the quality
system and in particular the type and extent of control applied to
the third party;
(iii)
the procedures for monitoring and verifying the design of the products
and in particular
(1)
the design specifications, including the standards which will be
applied and a description of the solutions adopted to fulfil the
essential requirements which apply to the products when the
standards referred to in regulation 5(7) of these regulations are not
applied in full,
(2)
the techniques of control and verification of the design, the
processes and systematic actions which will be used when the
products are being designed,
(3)
a statement indicating whether or not the device incorporates, as an
integral part, a substance or a human blood derivative referred to in
Section 11 of Schedule 1 to these regulations and the data on the
tests conducted in this connection required to assess the safety,
quality and usefulness of that substance or human blood derivative,
taking account of the intended purpose of the device,
(4)
the pre-clinical evaluation,
(5)
the clinical evaluation referred to Schedule VII to these regulations.
(iv)
the techniques of control and of quality assurance at the manufacturing
stage and in particular
(1)
the processes and procedures which will be used, particularly as
regards sterilisation, purchasing and the relevant documents,
(2)
product-identification procedures drawn up and kept up to date
from drawings, specifications or other relevant documents at every
stage of manufacture;
(v)
the appropriate tests and trials which will be effected before, during and
after production, the frequency with which they will take place, and the
test equipment used.
(c) Without prejudice to regulation 8(9) of these regulations, the notified body shall
effect an audit of the quality system to determine whether it meets the
requirements referred to in section 3(b) hereof. It shall presume conformity with
these requirements for the quality systems which use the corresponding
harmonised standards.
The team entrusted with the evaluation shall include at least one member who
has already had experience of evaluations of the technology concerned. The
evaluation procedure shall include an inspection on the manufacturer's premises
and, in duly substantiated cases, on the premises of the manufacturer's suppliers
and, or subcontractors to inspect the manufacturing processes.
The decision shall be notified to the manufacturer after the final inspection. It
shall contain the conclusions of the control and a reasoned evaluation.
(d) The manufacturer shall inform the notified body which has approved the quality
system of any plan to alter the quality system.
The notified body shall evaluate the proposed modifications and shall verify
whether the quality system so modified would meet the requirements referred to
in section 3(b) hereof; it shall notify the manufacturer of its decision. This
decision shall contain the conclusions of the control and a reasoned evaluation.
(4)
Examination of the design of the product
(a) In addition to the obligations incumbent on him under section 3, the
manufacturer shall make an application for examination of the design dossier
relating to the product which he plans to manufacture and which falls into the
category referred to in section 3(a) hereof.
(b) The application shall describe the design, manufacture, and performances of the
product in question, and it must include the documents needed to assess whether
the product conforms to the requirements of these regulations, and in particular
sections 3(b)(iii) and (iv) hereof.
It shall include inter alia
(i)
the design specifications, including the standards which have been
applied,
(ii)
the necessary proof of their appropriations, in particular where the
standards referred to in regulation 5(7) of these regulations have not been
applied in full. This proof must include the results of the appropriate tests
carried out by the manufacturer or carried out under his responsibility,
(iii)
a statement as to whether or not the device incorporates, as an integral
part, a substance as referred to in section 11 of Schedule I to these
regulations, whose action in combination with the device may result in its
bioavailability, together with data on the relevant trials conducted,
(iv)
the clinical evaluation referred to in Schedule VII to these regulations,
(v)
the draft instruction leaflet.
(c) The notified body shall examine the application and, where the product complies
with the relevant provisions of these regulations, shall issue the applicant with an
EC design examination certificate. The notified body may require the application
to be supplemented by further tests or proof so that compliance with the
requirements of the regulations may be evaluated. The certificate shall contain
the conclusions of the examination, the conditions of its validity, the data needed
for identification of the approved design and, where appropriate, a description of
the intended use of the product.
In the case of devices referred to in section 11(a) of Schedule I to these
regulations, the notified body shall, as regards the aspects referred to in that
section, consult one of the competent authorities designated by the Member
States in accordance with Directive 2001/83/EC or the EMEA before taking a
decision. The opinion of the competent national authority or the EMEA shall be
drawn up within 210 days after receipt of a valid documentation. The scientific
opinion of the competent national authority or the EMEA must be included in
the documentation concerning the device. The notified body will give due
consideration to the views expressed in this consultation when making its
decision. It will convey its final decision to the competent body concerned.
In the case of devices referred to in section 11(b) of Schedule I to these
regulations, the scientific opinion of the EMEA must be included in the
documentation concerning the device. The opinion shall be drawn up within 210
days after receipt of a valid documentation. The notified body will give due
consideration to the opinion of the EMEA when making its decision. The
notified body may not deliver the certificate if the EMEA's scientific opinion is
unfavourable. It will convey its final decision to the EMEA.
(d) The applicant shall inform the notified body which issued the EC design
examination certificate of any modification made to the approved design.
Modifications made to the approved design must obtain supplementary approval
from the notified body which issued the EC design examination certificate where
such modifications may affect conformity with the essential requirements of
these regulations or the conditions prescribed for the use of the product. This
supplementary approval shall be given in the form of an addendum to the EC
design examination certificate.
(5)
Surveillance
(a) The aim of surveillance is to ensure that the manufacturer duly fulfils the
obligations arising from the approved quality system.
(b) The manufacturer shall authorise the notified body to carry out all necessary
inspections and shall supply it with all appropriate information, in particular
(i)
the quality-system documentation,
(ii)
the data stipulated in the part of the quality system relating to design,
such as the results of analyses, calculations, tests, preclinical and clinical
evaluation, post-market clinical follow-up plan and the results of the
post-market clinical follow up, if applicable, etc.,
(iii)
the data stipulated in the part of the quality system relating to
manufacture, such as reports concerning inspections, tests,
standardisations, calibrations and the qualifications of the staff
concerned, etc.
(c) The notified body shall periodically carry out appropriate inspections and
evaluations in order to ascertain that the manufacturer is applying the approved
quality system, and shall supply the manufacturer with an evaluation report.
(d) In addition, the notified body may make unannounced visits to the manufacturer,
and shall supply him with an inspection report.
(6)
Administrative provisions
(a) For at least fifteen years from the last date of manufacture of the product, the
manufacturer or his authorised representative shall keep available for the
Regulatory Affairs Directorate and the Market Surveillance Directorate
(i)
the declaration of conformity,
(ii)
the documentation referred to in section 3(a)(ii), and in particular the
documentation, data and records referred to in the second paragraph of
section 3(b) hereof,
(iii) the amendments referred to in section 3(d) hereof,
(iv) the documentation referred to in section 4(b) hereof,
(v)
the decisions and reports of the notified body referred to in sections 3(d),
4(c), 5(c) and 5(d) of this Schedule.
(b) On request, the notified body shall make available to the other notified bodies
and Regulatory Affairs Directorate and the Market Surveillance Directorate all
relevant information on approvals of quality systems issued, refused or
withdrawn.
(7)
Application to devices referred to in regulation 4(2)(b) of these regulations
Upon completing the manufacture of each batch of devices referred to in regulation
4(2)(b) of these regulations, the manufacturer shall inform the notified body of the
release of the batch of devices and send to it the official certificate concerning the
release of the batch of human blood derivative used in the device, issued by a State
laboratory or a laboratory designated for that purpose by a Member State in
accordance with article 104A, sub-article (4) of the Medicines Act.
Schedule III
EC TYPE-EXAMINATION
(1)
EC type-examination is the procedure whereby a notified body observes and certifies
that a representative sample of the production envisaged satisfies the relevant
provisions of these regulations.
(2)
The application for EC type-examination shall be made by the manufacturer, or by
his authorised representative, to a notified body.
The application shall include
(a) the name and address of the manufacturer and the name and address of the
authorised representative if the application is made by the latter,
(b) a written declaration specifying that an application has not been made to any
other notified body,
(c) the documentation described in section 3 of this Schedule needed to allow an
evaluation to be made of the conformity of a representative sample of the
production in question, hereinafter referred to as “type”, with the requirements
of these regulations.
The applicant shall make a “type” available to the notified body. The notified body
may request other samples as necessary.
(3)
The documentation must make it possible to understand the design, the manufacture
and the performances of the product. The documentation shall contain the following
items in particular
(a) a general description of the type, including any variants planned, and its intended
use,
(b) design drawings, methods of manufacture envisaged, in particular as regards
sterilisation, and diagrams of parts, sub-assemblies, circuits, etc.,
(c) the descriptions and explanations necessary for the understanding of the above-
mentioned drawings and diagrams and of the operation of the product,
(d) a list of standards referred to in regulation 5(7) of these regulations, applied in
full or in part, and a description of the solutions adopted to satisfy the essential
requirements where the standards referred to in regulation 5(7) of these
regulations have not been applied,
(e) the results of design calculations, risk analysis, investigations and technical tests
carried out, etc.,
(f) a declaration stating whether or not the device incorporates, as an integral part, a
substance or a human blood derivative as referred to in section 11 of Schedule I
to these regulations and the data on the tests conducted in this connection
required to assess the safety, quality and usefulness of that substance or human
blood derivative, taking account of the intended purpose of the device,
(g) the pre-clinical evaluation,
(h) the clinical evaluation referred to in Schedule VII to these regulations,
(i) the draft instruction leaflet.
(4)
The notified body shall
(a) examine and evaluate the documentation, verify that the type has been
manufactured in accordance with that documentation; it shall also record the
items which have been designed in accordance with the applicable provisions of
the standards referred to in regulation 5(7) of these regulations, as well as the
items for which the design is not based on the relevant provisions of the said
standards;
(b) carry out or have carried out the appropriate inspections and the tests necessary
to verify whether the solutions adopted by the manufacturer satisfy the essential
requirements of these regulations where the standards referred to in regulation
5(7) of these regulations have not been applied;
(c) carry out or have carried out the appropriate inspections and the tests necessary
to verify whether, where the manufacturer has chosen to apply the relevant
standards, these have actually been applied;
(d) agree with the applicant on the place where the necessary inspections and tests
will be carried out.
(5)
Where the type meets the provisions of these regulations, the notified body shall
issue an EC type-examination certificate to the applicant. The certificate shall contain
the name and address of the manufacturer, the conclusions of the control, the
conditions under which the certificate is valid and the information necessary for
identification of the type approved.
The significant parts of the documentation shall be attached to the certificate and a
copy shall be kept by the notified body.
In the case of devices referred to in the second paragraph of section 11 of Schedule I
to these regulations, the notified body shall, as regards the aspects referred to in that
section, consult one of the competent authorities designated by the Member States in
accordance with Directive 2001/83/EC or the EMEA before taking a decision. The
opinion of the competent national authority or the EMEA shall be drawn up
within 210 days after receipt of a valid documentation. The scientific opinion of the
competent national authority or the EMEA must be included in the documentation
concerning the device. The notified body will give due consideration to the views
expressed in this consultation when making its decision. It will convey its final
decision to the competent body concerned.
In the case of devices referred to in the third paragraph, section 11 of Schedule I to
these regulations, the scientific opinion of the EMEA must be included in the
documentation concerning the device. The opinion shall be drawn up within 210 days
after receipt of a valid documentation. The notified body will give due consideration
to the opinion of the EMEA when making its decision. The notified body may not
deliver the certificate if the EMEA's scientific opinion is unfavourable. It will convey
its final decision to the EMEA.
(6)
The applicant shall inform the notified body which issued the EC type-examination
certificate of any modification made to the approved product.
Modifications to the approved product must receive further approval from the
notified body which issued the EC type-examination certificate where such
modifications may affect conformity with the essential requirements or with the
conditions of use specified for the product. This new approval shall be issued, where
appropriate, in the form of a supplement to the initial EC type-examination
certificate.
(7)
Administrative provisions
(a) On request, each notified body shall make available to the other notified bodies
and the competent authority, all relevant information on EC type-examination
certificates and addenda issued, refused or withdrawn.
(b) Other notified bodies may obtain a copy of the EC type-examination certificates
and, or the addenda to them. The annexes to the certificates shall be made
available to the other notified bodies when a reasoned application is made and
after the manufacturer has been informed.
(c) The manufacturer or his authorised representative shall keep with the technical
documentation a copy of the EC type-examination certificates and the
supplements to them for a period of at least 15 years from the manufacture of the
last product.
Schedule IV
EC VERIFICATION
(1)
EC verification is the procedure whereby the manufacturer or his authorised
representative ensures and declares that the products subject to the provisions of
section 3 below are in conformity with the type as described in the EC type-
examination certification and satisfy the requirements of these regulations that apply
to them.
(2)
The manufacturer or his authorised representative shall take all measures necessary in
order that the manufacturing process ensures conformity of the products to the type
as described in the EC type-examination certification and to the requirements of these
regulations that apply to them. The manufacturer or his authorised representative
shall affix the CE marking to each product and draw up a written declaration of
conformity.
(3)
The manufacturer shall, before the start of manufacture, prepare documents defining
the manufacturing processes, in particular as regards sterilisation, together with all
the routine, pre-established provisions to be implemented to ensure uniformity of
production and conformity of the products with the type as described in the EC type-
examination certificate as well as with the relevant requirements of these regulations.
(4)
The manufacturer shall undertake to institute and keep updated a post-marketing
surveillance system including the provisions referred to in Schedule VII to these
regulations. This undertaking shall include the obligation on the part of the
manufacturer to notify the Regulatory Affairs Directorate of the following events
immediately on learning of them
(a) any change in the characteristics or performances and any inaccuracies in the
instruction leaflet for a device which might lead to or have led to the death of a
patient or deterioration in his state of health;
(b) any technical or medical reason resulting in the withdrawal of a device from the
market by a manufacturer.
(5)
The notified body shall carry out the appropriate examinations and tests in order to
check the conformity of the product to the requirements of these regulations by
examination and testing of products on a statistical basis, as specified in section 6 of
this Schedule. The manufacturer must authorise the notified body to evaluate the
efficiency of the measures taken pursuant to section 3 above, by audit where
appropriate.
(6)
Statistical verification
(a) Manufacturers shall present the products manufactured in the form of uniform
batches and shall take all necessary measures in order that the manufacturing
process ensures the uniformity of each batch produced.
(b) A random sample shall be taken from each batch. Products in a sample shall be
individually examined and appropriate tests, as set out in any standard referred
to in regulation 5(7) of these regulations, or equivalent tests shall be carried out
to verify their conformity to the type as described in the EC type-examination
certificate and thereby determine whether a batch is to be accepted or rejected.
(c) Statistical control of products will be based on attributes and, or variables,
entailing sampling schemes with operational characteristics which ensure a high
level of safety and performance according to the state of the art. The sampling
schemes will be established by the harmonised standards referred to in
regulation 5(7) of these regulations, taking account of the specific nature of the
product categories in question.
(d) Where batches are accepted, the notified body shall affix, or cause to be affixed,
its identification number to each product and draw up a written certificate of
conformity relating to the tests carried out. All products in the batch may be
placed on the market except for those products from the sample which were
found not to be in conformity.
Where a batch is rejected, the notified body shall take appropriate measures to
prevent the placing on the market of that batch. In the event of frequent rejection
of batches the notified body may suspend the statistical verification.
The manufacturer may, under the responsibility of the notified body, affix the
latter's identification number during the manufacturing process.
(e) The manufacturer or his authorised representative shall ensure that he is able to
supply the notified body’s certificates of conformity on request.
(7)
Application to the devices referred to in regulation 4(2)(b) of these regulations
Upon completing the manufacture of each batch of devices referred to in regulation
4(2)(b) of these regulations, the manufacturer shall inform the notified body of the
release of the batch of devices and send to it the official certificate concerning the
release of the batch of human blood derivative used in the device, issued by a State
laboratory or a laboratory designated for that purpose by a Member State in
accordance with article 104A, sub-article (4) of the Medicines Act.
Schedule V
EC DECLARATION OF CONFORMITY TO TYPE
(Assurance of production quality)
(1)
The manufacturer shall apply the quality system approved for the manufacture and
shall conduct the final inspection of the products concerned as specified in section 3
of this Schedule; he shall be subject to the surveillance referred to in section 4 hereof.
(2)
This declaration of conformity is the procedural element whereby the manufacturer
who satisfies the obligations of section 1 hereof guarantees and declares that the
products concerned conform to the type described in the EC type-examination
certificate and meet the provisions of the regulations which apply to them.
(a) The manufacturer or his authorised representative shall affix the CE marking in
accordance with regulations 5(4) to (6) of these regulations and draw up a
written declaration of conformity. This declaration shall cover one or more
devices manufactured, clearly identified by means of product name, product
code or other unambiguous reference and must be kept by the manufacturer. The
CE marking shall be accompanied by the identification number of the notified
body responsible.
(3)
Quality system
(a) The manufacturer shall make an application for evaluation of his quality system
to a notified body.
The application shall include
(i)
all appropriate information concerning the products which it is intended to
manufacture,
(ii)
the quality-system documentation,
(iii)
an undertaking to fulfil the obligations arising from the quality system as
approved,
(iv)
an undertaking to maintain the approved quality system in such a way that
it remains adequate and efficacious,
(v)
where appropriate, the technical documentation relating to the approved
type and a copy of the EC type-examination certificate,
(vi)
an undertaking by the manufacturer to institute and keep up-dated a post-
marketing surveillance system including the provisions referred to in
Schedule VII to these regulations. The undertaking shall include an
obligation for the manufacturer to notify the Regulatory Affairs
Directorate of the following incidents immediately on learning of them
(1)
any deterioration in the characteristics or performances, and any
inaccuracies in the instruction leaflet for a device which might
lead to or have led to the death of a patient or a deterioration in
his state of health;
(2)
any technical or medical reason resulting in withdrawal of a
device from the market by the manufacturer.
(b) Application of the quality system must ensure that the products conform to the
type described in the EC type-examination certificate.
All the elements, requirements and provisions adopted by the manufacturer for
his quality system shall be documented in a systematic and orderly manner in the
form of written policies and procedures. This quality-system documentation
must make possible a uniform interpretation of the quality policies and
procedures such as quality programmes, quality plans, quality manuals and
quality records.
It shall include in particular an adequate description of
(i)
the manufacturer’s quality objectives;
(ii)
the organisation of the business and in particular
(1)
the organisational structures, the responsibilities of the managerial
staff and their organisational authority where manufacture of the
products is concerned,
(2)
the methods of monitoring the efficient operation of the quality
system and in particular its ability to achieve the desired quality of
the products, including control of products which do not conform,
(3)
where the manufacture and, or final inspection and testing of the
products, or elements thereof, is carried out by a third party, the
methods of monitoring the efficient operation of the quality system
and in particular the type and extent of control applied to the third
party.
(iii)
the techniques of control and of quality assurance at the manufacturing
stage and in particular
(1)
the processes and procedures which will be used, particularly as
regards sterilisation, purchasing and the relevant documents,
(2)
product identification procedures drawn up and kept up-to-date
from drawings, specifications or other relevant documents at every
stage of manufacture;
(iv)
the appropriate tests and trials which will be effected before, during and
after production, the frequency with which they will take place, and the
test equipment used.
(c) Without prejudice to regulation 8(9) of these regulations, the notified body shall
effect an audit of the quality system to determine whether it meets the
requirements referred to in section 3(b) hereof. It shall presume conformity with
these requirements for the quality systems which use the corresponding
harmonised standards.
The team entrusted with the evaluation shall include at least one member who
has already had experience of evaluations of the technology concerned. The
evaluation procedure shall include an inspection on the manufacturer's premises.
The decision shall be notified to the manufacturer after the final inspection. It
shall contain the conclusions of the control and a reasoned evaluation.
(d) The manufacturer shall inform the notified body which has approved the quality
system of any plan to alter that system.
The notified body shall evaluate the proposed modifications and shall verify
whether the quality system so modified would meet the requirements referred to
in section 3(b) hereof; it shall notify the manufacturer of its decision. This
decision shall contain the conclusions of the control and a reasoned evaluation.
(4)
Surveillance
(a) The aim of surveillance is to ensure that the manufacturer duly fulfils the
obligations arising from the approved quality system.
(b) The manufacturer shall authorise the notified body to carry out all necessary
inspections and shall supply it with all appropriate information, in particular
(i)
the quality-system documentation,
(ii)
the technical documentation,
(iii)
the data stipulated in the part of the quality system relating to
manufacture, such as reports concerning inspections, tests,
standardisations, calibrations and the qualifications of the staff
concerned, etc.
(c) The notified body shall periodically carry out appropriate inspections and
evaluations in order to ascertain that the manufacturer is applying the approved
quality system, and shall supply the manufacturer with an evaluation report.
(d) In addition, the notified body may make unannounced visits to the manufacturer,
and shall supply him with an inspection report.
(5)
The notified body shall communicate to the other notified bodies all relevant
information concerning approvals of quality systems issued, refused or withdrawn.
(6)
Application to the devices referred to in regulation 4(2)(b) of these regulations
Upon completing the manufacture of each batch of devices referred to in regulation
4(2)(b) of these regulations, the manufacturer shall inform the notified body of the
release of the batch of devices and send to it the official certificate concerning the
release of the batch of human blood derivative used in the device, issued by a State
laboratory or a laboratory designated for that purpose by a Member State in
accordance with article 104A, sub-article (4) of the Medicines Act.
Schedule VI
STATEMENT CONCERNING DEVICES INTENDED FOR SPECIAL PURPOSES
(1)
The manufacturer or his authorised representative shall draw up for custom-made
devices or for devices intended for clinical investigations the statement comprising
the elements stipulated in section 2 of this Schedule.
(2)
The statement must contain the following information
(a) for custom-made devices
(i)
the name and address of the manufacturer,
(ii)
the information necessary for the identification of the product in
question,
(iii)
the name of the duly qualified medical practitioner who drew up the
prescription and, if applicable, the name of the clinic concerned,
(iv)
the specific characteristics of the product revealed by the prescription,
(v)
a statement affirming that the device complies with the essential
requirements given in Schedule I to these regulations and, where
applicable, indicating which essential requirements have not been wholly
met, together with the grounds.
(b) for devices intended for the clinical investigations covered in Schedule VII to
these regulations
(i)
data allowing the devices in question to be identified,
(ii)
the clinical investigation plan,
(iii)
the investigator’s brochure,
(iv)
the confirmation of insurance of subjects,
(v)
the documents used to obtain informed consent,
(vi)
a statement indicating whether or not the device incorporates, as an
integral part, a substance or human blood derivative referred to in section
11 of Schedule I to these regulations,
(vii)
a statement indicating whether or not the device is manufactured utilising
tissues of animal origin as referred to in the Medical Devices Regulations,
2008,
(viii) the opinion of the Ethics Committee concerned and details of the aspects
covered by its opinion,
(ix)
the name of the duly qualified medical practitioner or other authorised
person and of the institution responsible for the investigations,
(x)
the place, date of commencement and duration scheduled for the
investigations,
(xi)
a statement affirming that the device in question complies with the
essential requirements apart from the aspects constituting the object of the
investigations and that, with regard to these aspects, every precaution has
been taken to protect the health and safety of the patient.
(3)
The manufacturer must also undertake to keep available for the Regulatory Affairs
Directorate and the Market Surveillance Directorate
(a) for custom-made devices, documentation, indicating manufacturing sites and
enabling the design, manufacture and performances of the product, including the
expected performances, to be understood, so as to allow conformity with the
requirements of these regulations to be assessed.
The manufacturer shall take all necessary measures to see that the manufacturing
process ensures that the products manufactured conform to the documentation
referred to in the first paragraph.
(b) for devices intended for clinical investigations, the documentation shall also
contain
(i)
a general description of the product and its intended use,
(ii)
design drawings, manufacturing methods, in particular as regards
sterilisation, and diagrams of parts, sub-assemblies, circuits, etc.,
(iii)
the descriptions and explanations necessary for the understanding of the
said drawings and diagrams and the operation of the product,
(iv)
the results of the risk analysis and a list of the standards referred to in
regulation 5(7) of these regulations, applied in full or in part, and
descriptions of the solutions adopted to satisfy the essential requirements
of these regulations where the standards referred to in regulation 5(7) of
these regulations have not been applied,
(v)
if the device incorporates, as an integral part, a substance or human blood
derivative referred to in section 11 of Schedule I to these regulations, the
data on the tests conducted in this connection which are required to
assess the safety, quality and usefulness of that substance or human blood
derivative, taking account of the intended purpose of the device,
(vi)
the results of the design calculations, checks and technical tests carried
out, etc.
The manufacturer shall take all the measures necessary to see that the
manufacturing process ensures that the products manufactured conform to the
documentation referred to in section 3(a) above and in the first paragraph of this
section.
The manufacturer may authorise the evaluation, by audit where necessary, of the
effectiveness of these measures.
(4)
The information included in the declarations covered by this Schedule shall be kept
for a period of at least 15 years from the data of manufacture of the last product.
(5)
For custom-made devices, the manufacturer must undertake to review and document
experience gained in the post-production phase, including the provisions referred to
in Schedule VII to these regulations, and to implement appropriate means to apply
any necessary corrective action. This undertaking must include an obligation for the
manufacturer to notify the Regulatory Affairs Directorate of the following incidents
immediately on learning of them and the relevant corrective actions
(a) any malfunction or deterioration in the characteristics and, or performance of a
device, as well as any inadequacy in the labelling or the instructions for use
which might lead to or might have led to the death of a patient or user or to a
serious deterioration in his state of health;
(b) any technical or medical reason connected with the characteristics or
performance of a device for the reasons referred to in paragraph (a) above
leading to systematic recall of devices of the same type by the manufacturer.
Schedule VII
CLINICAL EVALUATION
(1)
General provisions
(a) As a general rule, confirmation of conformity with the requirements concerning
the characteristics and performances referred to in sections 1 and 2 of Schedule I
to these regulations under the normal conditions of use of the device and the
evaluation of the side-effects and of the acceptability of the benefit and risk ratio
referred to in section 5 of Schedule I to these regulations, must be based on
clinical data. The evaluation of this data, hereafter referred to as clinical
evaluation, where appropriate taking account of any relevant harmonised
standards, must follow a defined and methodologically sound procedure based
on
(i)
either a critical evaluation of the relevant scientific literature currently
available relating to the safety, performance, design characteristics and
intended purpose of the device where
(1)
there is demonstration of equivalence of the device to the device
to which the data relates and,
(2)
the data adequately demonstrate compliance with the relevant
essential requirements;
(ii)
or a critical evaluation of the results of all the clinical investigations
made,
(iii)
or a critical evaluation of the combined clinical data provided in sections
1(a)(i) and (ii) heeof.
(b) Clinical investigations shall be performed unless it is duly justified to rely on
existing clinical data.
(c) The clinical evaluation and its outcome shall be documented. This
documentation shall be included and, or fully referenced in the technical
documentation of the device.
(d) The clinical evaluation and its documentation must be actively updated with data
obtained from the post-market surveillance. Where post-market clinical follow-
up as part of the post-market surveillance plan for the device is not deemed
necessary, this must be duly justified and documented.
(e) Where demonstration of conformity with essential requirements based on
clinical data is not deemed appropriate, adequate justification for any such
exclusion has to be given based on risk management output and under
consideration of the specifics of the device-body interaction, the clinical
performances intended and the claims of the manufacturer. Adequacy of
demonstration of conformity with the essential requirements by performance
evaluation, bench testing and preclinical evaluation alone has to be duly
substantiated.
(f) All data must remain confidential unless it is deemed essential that they be
divulged.
(2)
Clinical investigations
(a) Purpose
The purpose of a clinical investigation is to
(i)
verify that, under normal conditions of use, the performance of the device
comply with those indicated to in section 2 of Schedule I to these
regulations,
(ii)
determine any undesirable side-effects, under normal conditions of use,
and assess whether they constitute risks when weighed against the
intended performance of the device.
(b) Ethical consideration
Clinical investigations shall be made in accordance with the Declaration of
Helsinki approved by the 18th World Medical Assembly in Helsinki, Finland, in
1964, and amended by the 29th World Medical Assembly in Tokyo, Japan, in
1975 and the 35th World Medical Assembly in Venice, Italy, in 1983. It is
mandatory that all measures relating to the protection of human subjects are
carried out in the spirit of the Declaration of Helsinki. This includes every step
in the clinical investigation from first consideration of need and justification of
the study to publication of results.
(c) Methods
(i)
Clinical investigations shall be performed according to an appropriate
state of the art plan of investigation defined in such a way as to confirm
or refute the manufacturer’s claims for the device; the investigations shall
include an adequate number of observations to guarantee the scientific
validity of the conclusions.
(ii)
The procedures utilised to perform the investigations shall be appropriate
to the device under examination.
(iii)
Clinical investigations shall be performed in circumstances equivalent to
those which would be found in normal conditions of use of the device.
(iv)
All the appropriate features, including those involving the safety and
performances of the device, and its effect on patients, shall be examined.
(v)
All serious adverse events must be fully recorded and immediately
notified to the Market Surveillance Directorate, the Regulatory Affairs
Directorate and all competent authorities of the Member States in which
the clinical investigation is being performed.
(vi)
The investigations shall be performed under the responsibility of a duly
qualified medical practitioner or authorised person, in an appropriate
environment.
The medical practitioner or authorised person shall have access to the
technical data regarding the device.
(vii)
The written report, signed by the responsible medical practitioner or
authorised person, shall comprise a critical evaluation of all the data
collected during the clinical investigation.
Schedule VIII
MINIMUM CRITERIA TO BE MET WHEN DESIGNATING INSPECTION BODIES TO BE
NOTIFIED
(1)
Notified bodies shall be subject to the Method for Designating Conformity
Assessment Bodies Regulations, 2003, and they shall also fulfil the criteria set out
below.
(2)
The body, its director and the staff responsible for carrying out the evaluation and
verification operations shall not be the designer, manufacturer, supplier or installer of
devices which they control, nor the authorised representative of any of those parties.
They may not become directly involved in the design, construction, marketing or
maintenance of the devices, nor represent the parties engaged in these activities. This
does not preclude the possibility of exchanges of technical information between the
manufacturer and the body.
(3)
The body and its staff must carry out the evaluation and verification operations with
the highest degree of professional integrity and technical competence and must be
free from all pressures and inducements, particularly financial, which might influence
their judgement or the results of the inspection, especially from persons or groups of
persons with an interest in the results of verifications.
(4)
The body must be able to carry out all the tasks in one of Schedules II to V to these
regulations assigned to such a body and for which it has been notified, whether those
tasks are carried out by the body itself or under its responsibility. In particular, it
must have at its disposal the necessary staff and possess the necessary facilities to
enable it to perform properly the technical and administrative tasks connected with
evaluation and verification; it must also have access to the equipment necessary for
the verifications required.
(5)
The staff responsible for control operations must have
(a) sound vocational training covering all the evaluation and verification operations
for which the body has been designated,
(b) satisfactory knowledge of the requirements of the controls they carry out and
adequate experience of such operations,
(c) the ability required to draw up the certificates, records and reports to
demonstrate that the controls have been carried out.
(6)
The impartiality of inspection staff must be guaranteed. Their remuneration must not
depend on the number of controls carried out, nor on the results of such controls.
(7)
The body must take out liability insurance unless liability is assumed by the State in
accordance with Maltese law, or the Regulatory Affairs Directorate itself are directly
responsible for controls.
(8)
The staff of the body is bound to observe professional secrecy with regard to all
information gained in carrying out their tasks (except vis-à-vis the Regulatory Affairs
Directorate or other competent administrative authorities of the State in which their
activities are carried out) under these regulations or any provision of national law
giving effect to it.
Schedule IX
CE CONFORMITY MARKING
The CE conformity marking shall consist of the initials “CE” taking the following
form
For the avoidance of doubt, it is hereby declared that the grid providing the
background in the above graduated drawing is not part of the CE marking.
- If the CE marking is reduced or enlarged the proportions given in the above
graduated drawing must be respected.
- The various components of the CE marking must have substantially the same
vertical dimension, which may not be less than 5mm.
This minimum dimension may be waived for small-scale devices.
Mitbug[ fl-Istamperija tal-Gvern — Printed at the Government Printing Press
€2.24