Medical Devices (Amendment) Regulations, 2004
Statutory Text (4 Articles)
Suppliment tal-Gazzetta tal-Gvern ta’ Malta, Nru. 17,532, 23 ta’ Jannar, 2004
L.N. 27 of 2004
PRODUCT SAFETY ACT
(ACT NO. V OF 2001)
Medical Devices (Amendment) Regulations, 2004
IN exercise of the powers conferred by articles 38 to 40 of the
Product Safety Act, the Minister of Finance and Economic Affairs, on
the advice of the Malta Standards Authority, has made the following
regulations:-
(1) The title of these regulations is Medical Devices Citation
commencement.
(Amendment) Regulations, 2004, and they shall be read and construed
as one with the Medical Devices Regulations, 2003, hereinafter referred L.N. 47 of 2003.
to as “the principal regulations”.
(2) These regulations shall come into force on the date of
their publication
Provided that the provisions of the following regulations shall
enter into force on 1 st April 2004
– item 1.2.2 in regulation 2 of these regulations;
– items 3.1.23 to 3.1.34 in regulation 4 of these regulations; and
– regulations 5, 6, 7 and 8 of these regulations.
Immediately after regulation 1.2.1 of the principal regulations, Amends regulation
1 of the principal
there shall be inserted the following regulations
regulations.
“1.2.2. Holders of EC design-examination certificates or
EC type-examination certificates issued before 1 st April 2004 for
medical devices referred to in Article 1(1) of Directive 2003/32
shall apply for a complementary EC design-examination certificate
or EC type-examination certificate attesting to compliance with
the specifications laid down in Schedule XIII to these regulations.
1.2.3. Breast implants placed on the market before 1 st
September, 2003 pursuant to regulation 8.3(a) or regulation
3(b)(iii) of these regulations shall be subject to a conformity
reassessment procedure as Class III medical devices before the 1 st
March, 2004.
1.2.4. Decisions taken by notified bodies before the 1 st
September 2003 under regulation 8.3(a) of these regulations may
not be extended.”.
3. For regulation 2.1.3 and 2.1.3.1 of the principal regulations,
there shall be substituted the following
“2.1.3. Devices shall be divided into Classes I, IIa, IIb and
III. Classification shall be carried out in accordance with Annex
IX of the Directive, which is set out in Schedule IX to these
regulations, read with Directive 2003/12.”.
4. Immediately after regulation 3.1.21 of the principal regulations,
there shall be inserted the following
“3.1.22. “Directive 2003/12” means Commission Directive
2003/12/EC of 3 rd February 2003 on the reclassification of breast
implants in the framework of Directive 93/42/EEC concerning
medical devices;
3.1.23. “animal” means an animal from a bovine, ovine or
caprine species, as well as deer, elk, mink and cats;
3.1.24. “cell” means the smallest organised unit of any living
form which is capable of independent existence and of replacement
of its own substance in a suitable environment;
3.1.25. “derivative” means a material obtained from an animal
tissue by a manufacturing process such as collagen, gelatine,
monoclonal antibodies;
3.1.26. “Directive 2003/32” means Commission Directive
2003/32/EC of 23 rd April 2003 introducing detailed specifications
as regards the requirements laid down in Council Directive 93/42/
EEC with respect to medical devices manufactured utilising tissues
of animal origin;
3.1.27. “inactivation” means a process by which the ability
to cause infection or pathogenic reaction by transmissible agents
is reduced;
3.1.28. “non-viable” means having no potential for
metabolism or multiplication;
3.1.29. “reduction, elimination or removal” means a process
by which the number of transmissible agents is reduced, eliminated
or removed in order to prevent infection or pathogenic reaction;
3.1.30. “source country” means the country in which the
animal was born, has been reared and, or has been slaughtered;
3.1.31. “starting materials” means raw materials or any other
product of animal origin out of which, or with the help of which,
the devices referred to in regulation 2.2 of these regulations are
produced;
3.1.32. “tissue” means an organisation of cells and/or extra-
cellular constituents;
3.1.33. “transmissible agents” means unclassified pathogenic
entities, prions and such entities as bovine spongiform
encephalophathies agents and scrapie agents;
3.1.34. “TSE certificate” means TSE certificate of suitability
issued by the European Directorate for the Quality of Medicines.”.
5. Regulations 4.6 and 4.7 of the principal regulations shall Amends regulation
be renumbered as regulations 4.7 and 4.8 respectively and immediately 4 regulations.
after regulation 4.5.1 of the principal regulations, there shall be inserted
the following
“4.6. Subject to regulation 4.6.2 of these regulations, no
person shall put into service or supply a relevant device (if that
supply is also a placing on the market or if that supply is of a
relevant device that has been placed on the market) if that device
is manufactured utilising either animal tissue which is rendered
non-viable or non-viable products derived from animal tissue
unless–
(a) there is in respect of that device a risk analysis and
risk management scheme which is in accordance with the
requirements of Directive 2003/32; and
(b) the manufacturer of the device has fulfilled his
obligations under the procedures within that scheme.
4.6.1. If collagen, gelatine or tallow is of animal origin and
has been used in the manufacture of a relevant device, then subject
to regulation 4.6.2 of these regulations, no person shall put into
service or supply that device (if that supply is also a placing on the
market or if that supply is of a device that has been placed on the
market), unless that collagen, gelatine or tallow was fit for human
consumption.
4.6.2. Regulations 4.6 and 4.6.1 of these regulations shall
not apply to a relevant device which is not intended to come into
contact with the human body or which is intended to come into
contact with intact skin only.”.
6. Immediately after regulation 5.2 of the principal regulations,
there shall be inserted the following
“5.2.1. All necessary steps shall be taken to ensure that devices
referred to in Article 1(1) of Directive 2003/32 are placed on the
market and put into service only if they comply with the provisions
of these regulations and the specifications laid down in Schedule
XIII to these regulations.”.
7. (1) Immediately after regulation 8.1 of the principal
regulations, there shall be inserted the following
“8.1.1. Before lodging an application for a conformity
assessment pursuant to regulation 8.1 of these regulations, the
manufacturer of medical devices referred to Article 1(1) of
Directive 2003/32, shall carry out the risk analysis and the risk
management scheme set out in Schedule XIII to these regulations.”.
(2) Immediately after regulation 8.12 of the principal
regulations, there shall be added the following
“8.13. Conformity assessment procedures for medical devices
referred to in Article 1(1) of Directive 2003/32, shall include the
evaluation of their compliance with the essential requirements of
the Directive and the specifications laid down in the Annex to
Directive 2003/32, which is set out in Schedule XIII to these
regulations.
8.14. Notified bodies shall evaluate the manufacturer’s risk
analysis and risk management strategy, and in particular
(a) the information provided by the manufacturer;
(b) the justification for the use of animal tissues or
derivatives;
(c) the results of elimination and, or inactivation studies
or of literature search;
(d) the manufacturer’s control of the sources of raw
materials, finished products and subcontractors;
(e) the need to audit matters related to sourcing,
including third party supplies.
8.15. Notified bodies shall, during the evaluation of the risk
analysis and risk management in the framework of the conformity
assessment procedure, take account of the TSE certificate, for
starting materials, where available.
8.16. Except for medical devices using starting materials for
which a TSE certificate has been issued as referred to in regulation
8.15, national bodies shall, through the CIGD, seek the opinion of
the competent authorities of the other Member States on their
evaluation of and conclusions on the risk analysis and risk
management of the tissues or the derivatives intended to be
incorporated in the medical device as established by the
manufacturer.
8.16.1. Before issuing an EC design-examination certificate
or an EC type-examination certificate, the notified bodies shall
give due consideration to any comments received within 12 weeks
from the date on which the opinion of the national competent
authorities were sought.”.
8. Immediately after Schedule XII of the principal regulations, Adds new Schedule
XIII to the principal
there shall be added the following Schedule
regulations.
“Schedule XIII
Based on the Annex of Directive 2003/32
1. RISK ANALYSIS AND RISK MANAGEMENT
1.1. Justification for the use of animal tissues or derivatives
The manufacturer must justify, on the basis of his overall risk analysis and risk
management strategy for a specific medical device, the decision to use animal tissues
or derivatives, referred to in Article 1(1) of Directive 2003/32, (specifying animal species
and tissues) taking into account the expected clinical benefit, potential residual risk
and suitable alternatives.
1.2. Assessment procedure
In order to ensure a high level of protection for patients or users, the manufacturer
of devices utilising animal tissues or derivatives referred to in point 1.1 must implement
an appropriate and well documented risk analysis and risk management strategy, to
address all relevant relating aspects to TSE. He must identify the hazards associated
with those tissues or derivatives, establish documentation on measures taken to minimise
the risk of transmission and demonstrate the acceptability of the residual risk associated
with the device utilising such tissues or derivatives, taking into account the intended
use and the benefit of the device.
The safety of a device, in terms of its potential for passing on a transmissible
agent, is dependent on all the factors described in points 1.2.1 to 1.2.7, which must be
analysed, evaluated and managed. These measures in combination determine the device
safety.There are two key steps that must be considered.
These are
—selecting starting materials (tissues or derivatives) considered appropriate
regarding their potential contamination with transmissible agents (see 1.2.1, 1.2.2
and 1.2.3) taking into account further processing,
— applying a production process to remove or inactivate transmissible agents
on controlled sourced tissues or derivatives (see 1.2.4).
Furthermore, the characteristics of the device and its intended use must be taken
into account (see 1.2.5, 1.2.6 and 1.2.7).
In performing the risk analysis and risk management strategy, due consideration
must be given to opinions adopted by the relevant scientific committees, and where
appropriate to the opinions of the Committee for Proprietary Medicinal Products CPMP,
the references of which have been published in the Official Journal of the European
Union.
B
1.2.1. Animals as a source of material
The TSE risk is related to the source species, strains and nature of the starting
tissue. As the accumulation of TSE infectivity occurs over an incubation period of
several years, sourcing from young healthy animals is considered to be a factor reducing
the risk. Risk animals such as fallen stock, emergency slaughtered and TSE suspected
animals must be excluded.
1.2.2. Geographical sourcing
Pending the classification of countries according to the BSE status in Regulation
(EC) No 999/2001 of theEuropean Parliament and of the Council of 22 May 2001
laying down rules for the prevention, control and eradication of certain transmissible
spongiform encephalopathies 12 , the Geographical BSE risk (GBR) is used when
assessing the risk of the source country. The GBR is a qualitative indicator of the
likelihood of the presence of one or more cattle being infected with BSE, pre-clinically
as well as clinically, at a given point in time, in a country. Where presence is confirmed,
the GBR gives an indication of the level of infection as specified in the table below.
GBR level
Presence of one or more cattle clinically
or pre-clinically infected with the BSE agent
in a geographical region/country
I
Highly unlikely
II
Unlikely but not excluded
III
Likely but not confirmed or confirmed, at a lower level
IV
Confirmed, at a higher level
Certain factors influence the geographical risk of BSE infection associated with
the use of raw tissues or derivatives from individual countries. These factors are defined
in Article 2.3.13.2, point 1) of the International Animal Health Code of the OIE (Office
International des Épizooties), which is available at web-site www.oie.int/eng/normes/
Mcode/A_00067.htm
The Scientific Steering Committee has made an assessment of Geographical BSE
Risk (GBR) of several third countries and Member States, and will continue to do so
for all the countries, which applied for BSE status categorisation, taking the main OIE
factors into account.
1.2.3. Nature of starting tissue
The manufacturer must take into account the classification of the hazards relating
to different type of starting tissue. Sourcing of animal tissue must be subject to control
and individual inspection by a veterinarian and the animal carcass must be certified as
fit for human consumption.
OJ No. L 147, 31-05-2001, P.1.
The manufacturer must ensure that no risk of cross-contamination occurs at the
time of slaughtering.
The manufacturer must not source animal tissue or derivatives classified as
potentially high TSE infectivity, unless sourcing of these materials is necessary in
exceptional circumstances, taking into account the important benefit for the patient
and the absence of an alternative starting tissue.
In addition, the provisions in Regulation (EC) No 1774/2002 of the European
Parliament and of the Council of 3 October 2002 laying down health rules concerning
animal by-products not intended for human consumption must be applied.
1.2.3.1. Sheep and goats
A classification of infectivity in tissues for sheep and goats has been established
based on actual knowledge on the basis of the titres of transmissible agents in tissues
and body fluids from naturally infected sheep and goats with clinical scrapie. A table
was presented in the Scientific Steering Committee (SSC) opinion of 22-23 July 1999
on ‘The policy of breeding and genotyping of sheep’, (as an annex) 13 and further updated
in the opinion of the SSC — TSE infectivity distributed in ruminant tissues state of
knowledge December 2001 — adopted 10-11 January 2002 14 .
The classification may be reviewed in the light of new scientific evidence (for
example using relevant opinions from the Scientific Committees, the Committee for
Proprietary Medicinal Products CPMP and Commission Measures regulating the use
of material presenting risks as regards TSE). A review of the references to relevant
documents/opinions will be published in the Official Journal of the European Union
and will be listed after a Commission Decision has been taken.
1.2.3.2. Cattle
The list of specified risk material (SRM) laid down in Regulation (EC) No 999/
2001 shall be considered as potentially high TSE infective.
1.2.4. Inactivation or removal of transmissible agents
1.2.4.1. For devices which cannot withstand an inactivation/elimination process
without undergoing unacceptable degradation, the manufacturer must rely principally
on the control of sourcing.
1.2.4.2. For other devices, if claims are made by the manufacturer for the ability
of manufacturing processes to remove or inactivate transmissible agents, these will
have to be substantiated by appropriate documentation.
Available on the website of the Commission
http://europa.eu.int/comm/food/fs/sc/ssc/outcome_en.html.
Available on the website of the Commission
http://europa.eu.int/comm/food/fs/sc/ssc/outcome_en.html.
B
Relevant information from an appropriate scientific literature search and analysis
can be used to support inactivation/elimination factors, where the specific processes
referred to the literature are comparable to those used for the device. This search and
analysis should also cover the available scientific opinions that may have been adopted
by a EU Scientific Committee. These opinions shall serve as a reference, in cases
where there are conflicting opinions.
When the literature search failed to substantiate the claims, the manufacturer must
set up a specific inactivation and/or elimination study on a scientific basis and the
following need to be considered
— the identified hazard associated with the tissue,
— identification of the relevant model agents,
— rationale for the choice of the particular combinations of model agents,
— identification of stage chosen to eliminate and/or inactivate the
transmissible agents,
— calculation of the reduction factors.
A final report must identify manufacturing parameters and limits that are critical
to the effectiveness of the inactivation or elimination process.
Appropriate documented procedures must be applied to ensure that the validated
processing parameters are applied during routine manufacture.
1.2.5. Quantities of animal starting tissues or derivatives required to produce one
unit of the medical device
The manufacturer must evaluate the quantity of raw tissues or derivatives of animal
origin required to produce a single unit of the medical device. Where a purification
process is involved, the manufacturer must assess whether it may have the potential to
concentrate levels of transmissible agents present in the animal starting tissues or
derivatives.
1.2.6. Tissues or derivatives of animal origin coming into contact with the patients
and users
The manufacturer must consider
(i) the quantity of animal tissues or derivatives;
(ii) the contact area: its surface, type (e.g. skin, mucous tissue, brain) and
condition (e.g. healthy or damaged);
(iii) the type of the tissues or derivatives coming into contact with the patients
and/or users; and
(iv) how long the device is intended to remain in contact with the body
(including bioresorption effect).
The number of medical devices that could be used in a given procedure shall be
taken into account.
1.2.7. Route of administration
The manufacturer must take into account the route of administration recommended
in the product information, from the highest risk down to the lowest.
1.3. Review of the assessment
The manufacturer must establish and maintain a systematic procedure to review
information gained about their medical device or similar devices in the post-production
phase. The information must be evaluated for possible relevance to safety, especially
(a) if previously unrecognised hazards are detected;
(b) if the estimated risk arising from a hazard is no longer acceptable;
(c) if the original assessment is otherwise invalidated.
If any of the above apply, the results of the evaluation shall be fed back as an
input to the risk managementprocess.
In the light of this new information, a review of the appropriate risk management
measures for the device must be considered (including rationale for choosing an animal
tissue or derivative). If there is a potential that the residual risk or its acceptability has
changed, the impact on previously implemented risk control measures must be re-
evaluated and justified.
The results of this evaluation must be documented.
2. EVALUATION OF CLASS III MEDICAL DEVICES BY NOTIFIED BODIES
For devices falling into Class III under rule 17 15 of Annex IX to the Directive, set
out in Schedule IX to these regulations, manufacturers must provide to the notified
bodies referred to in Article 4 of Directive 2003/32, all relevant information to allow
evaluation of their current risk analysis and risk management strategy.
All medical devices utilising animal tissues or derivatives rendered non-viable except devices intended to
come into contact with intact skin only.
B
Any new information on TSE risk, collected by the manufacturer and relevant for
their devices must be sent to the notified body for information.
Any change in relation to processes of sourcing, collection, handling and
inactivation/elimination and that could modify the result of the manufacturer’s risk
management dossier must be transmitted to the notified body for the purpose of an
additional approval prior to its implementation.”.