Medical Devices Regulations, 2003
Table of Provisions (13)
Statutory Text (13 Articles)
Suppliment tal-Gazzetta tal-Gvern ta’ Malta, Nru. 17,349, 31 ta’ Jannar, 2003
L.N. 47 of 2003
PRODUCT SAFETY ACT
(ACT NO. V OF 2001)
Medical Devices Regulations, 2003
IN exercise of the powers conferred by articles 38 to 40 of the
Product Safety Act, 2001, the Minister for Economic Services, on the
advice of the Malta Standards Authority, has made the following
regulations:-
1. The title of these regulations is the Medical Devices Citation and
commencement.
Regulations, 2003.
1.2. These regulations shall come into force on the date of
publication in the Gazette.
1.2.1. In the case of medical devices which have been
subjected to EEC pattern approval in accordance with Directive
76/764/EEC 1 , their being placed on the market and put into service
shall be accepted during the period up to 30 th June, 2004.
1.3. The requirements to provide information to the European
Commission referred to in regulation 7.9 of these regulations shall come
into force on such date as the Minister responsible for consumer affairs,
after consultation with the Malta Standards Authority may prescribe.
1. These regulations shall apply to medical devices and their Scope and
Classification.
accessories.
2.1.1. For the purposes of these regulations, accessories shall
be treated as medical devices in their own right.
2.1.2. Both medical devices and accessories shall hereinafter
be termed ‘devices’.
_____
OJ No.L262, 27-09-76, P.139. Directive as last amended by Directive 84/414/EEC
(OJ No.L228, 25-08-84, P.25)
2.1.3. Devices shall be divided into Classes I, IIa, IIb and
III. Classification shall be carried out in accordance with Annex
IX of the Directive, which is set out in Schedule IX to these
regulations.
2.1.3.1. The classification rules set out in Annex IX of
the Directive, which is set out in Schedule IX to these
regulations may be adapted in accordance with the procedure
referred to in Article 7 (2) of the Directive in the light of
technical progress and any information which becomes
available under the information system provided for in
regulation 7.4 of these regulations and Article 10 of the
Directive.
1. For the purposes of these regulations, the following
definitions shall apply
3.1.1. “medical device” means any instrument, apparatus,
appliance, material or other article, whether used alone or in
combination, including the software necessary for its proper
application intended by the manufacturer to be used for human
beings for the purpose of
- diagnosis, prevention, monitoring, treatment or
alleviation of disease,
- diagnosis, monitoring, treatment, alleviation of or
compensation for an injury or handicap,
- investigation, replacement or modification of the
anatomy or of a physiological process,
- control of conception,
and which does not achieve its principal intended action in or on
the human body by pharmacological, immunological or metabolic
means, but which may be assisted in its function by such means;
3.1.2. “accessory” means an article which whilst not being
a device is intended specifically by its manufacturer to be used
together with a device to enable it to be used in accordance with
the use of the device intended by the manufacturer of the device;
3.1.3. “in vitro diagnostic medical device” means any
medical device which is a reagent, reagent product, calibrator,
control material, kit, instrument, apparatus, equipment or system,
whether used alone or in combination, intended by the manufacturer
to be used in vitro for the examination of specimens, including
blood and tissue donations, derived from the human body, solely
or principally for the purpose of providing information
- concerning a physiological or pathological state, or
- concerning a congenital abnormality, or
- to determine the safety and compatibility with potential
recipients, or
- to monitor therapeutic measures
Provided that -
Specimen receptacles are considered to be in vitro diagnostic
medical devices; and
“Specimen receptacles” are those devices, whether vacuum-
type or not, specifically intended by their manufacturers for the
primary containment and preservation of specimens derived from
the human body for the purpose of in vitro diagnostic examination
Products for general laboratory use are not in vitro diagnostic
medical devices unless such products, in view of their
characteristics, are specifically intended by their manufacturer to
be used for in vitro diagnostic examination;
3.1.4. “custom-made device” means any device specifically
made in accordance with a duly qualified medical practitioner’s
written prescription which gives, under his responsibility, specific
design characteristics and is intended for the sole use of a particular
patient.
3.1.4.1. The abovementioned prescription may also be
made out by any other person authorised by virtue of his
professional qualifications to do so; and
3.1.4.2. Mass-produced devices which need to be
adapted to meet the specific requirements of the medical
practitioner or any other professional user are not considered
to be custom-made devices;
3.1.5. “device intended for clinical investigation” means any
device intended for use by a duly qualified medical practitioner
when conducting investigations, as referred to in section 2.1 of
Annex X of the Directive, which is set out in Schedule X to these
regulations, in an adequate human clinical environment
3.1.5.1. For the purpose of conducting clinical
investigation, any other person who, by virtue of his
professional qualifications, is authorised to carry out such
investigation shall be accepted as equivalent to a duly qualified
medical practitioner;
3.1.6. “manufacturer” means the natural or legal person with
responsibility for the design, manufacture, packaging and labelling
of a device before it is placed on the market under his own name,
regardless of whether these operations are carried out by that person
himself or on his behalf by a third party
3.1.6.1. The obligations of these regulations to be met
by manufacturers also apply to the natural or legal person
who assembles, packages, processes, fully refurbishes and/
or labels one or more ready-made products and/or assigns to
them their intended purpose as a device with a view to their
being placed on the market under his own name; and
3.1.6.2. Regulation 3.1.6.1 does not apply to the person
who, while not a manufacturer within the meaning of
regulation 3.1.6, assembles or adapts devices already on the
market to their intended purpose for an individual patient;
3.1.7. “intended purpose” means the use for which the
device is intended according to the data supplied by the
manufacturer on the labelling, in the instructions and/or in
promotional materials;
3.1.8. “placing on the market” means the first making
available in return for payment or free of charge of a device other
than a device intended for clinical investigation, with a view to
distribution and/or use on the Community or Maltese market,
regardless of whether it is new or fully refurbished
3.1.8.1. Making available can be either for payment
or free of charge; and
3.1.8.2. Putting into service takes place at the moment
of first use within the Community or Malta by the end user.
3.1.9. “putting into service” means the stage at which a
device has been made available to the final user as being ready for
use on the Community or Maltese market for the first time for its
intended purpose;
3.1.10. “the Directive” means Council Directive 93/42/EEC 2
on the approximation of the laws of the Member States concerning
medical devices;
3.1.11. “notified body” means a body notified to the
Commission and to the Member States of the Community, assessed
according to the criteria laid down in Annex XI of the Directive,
which is set out in Schedule XI to these regulations and appointed
to carry out the procedures referred to in regulations 8.1 to 8.12 of
these regulations and Article 11 of the Directive. The names of
such bodies are published in the Official Journal of the European
Communities;
3.1.12. “Harmonized Standard” means a technical
specification adopted by the European Committee for
Standardization (CEN) or the European Committee for
Electrotechnical Standardization (CENELEC) or the European
Telecommunications Standards Institute (ETSI), upon a remit from
the European Commission in accordance with Directive 98/34/
EC 3 of the European Parliament and of the Council laying down a
procedure for the provision of information in the field of technical
standards and regulations as amended by Directive 98/48/EC 4 of
the European Parliament and of the Council;
3.1.13. “National Standard transposing the Harmonized
Standard” means a Standard issued by one of the members of CEN/
CENELEC/ETSI which is declared and noted to be identical to
the Harmonized Standard originally issued by CEN/CENELEC/
ETSI.
3.1.14. “OJ” means the Official Journal of the European
Communities;
3.1.15.
“Community” means the European Community;
3.1.16. “CE marking” means marking as set out in Annex
XII of the Directive, which is set out in Schedule XII of these
regulations;
3.1.17. “authorised representative” means any person (or
organisation) established in the Community or Malta who has been
appointed by the manufacturer to act on his behalf in carrying out
certain tasks required by these regulations;
3.1.18. “importer” means any person who places devices
from a foreign country on to the Maltese market so that he supplies
such devices on the first occasion that they are supplied in Malta;
_____
OJ No. L169, 12-07-93, P.1
OJ No. L204, 21-07-98, P.37
OJ No. L217, 05-08-98, P.18
3.1.19. “person who supplies devices” means wholesalers,
distributors, retailers etc. ;
3.1.20.
“responsible person” in relation to devices means
(a)
the manufacturer thereof;
(b)
the manufacturer’s authorised representative; or
(c) where the manufacturer is not established in Malta
and he has not appointed an authorised representative, the person
who places the devices on the Community or Maltese market;
3.1.21. “authorised representative” means any person (or
organisation) established in the Community or Malta who has been
appointed by the manufacturer to act on his behalf in carrying out
certain tasks required by these regulations.
1. The provisions of these regulations are applicable to all imported
medical devices as well as to those manufactured for the local market.
4.1.1. No person may manufacture, import, place on the
market or put into service any medical devices which do not comply
with the provisions of these regulations.
4.1.2. Regulation 4.1 above shall not be construed to prohibit
any person from possessing for supply at any time before it is
supplied, offered for supply or exposed for supply in Malta for the
first time, or before it is agreed for the first time to supply it in
Malta, any device referred to in regulation 2.1 of these regulations.
4.2. Subject to regulation 4.3 below, the provisions of these
regulations shall not apply to medical devices intended exclusively for
export.
4.3. Subject to regulation 12 below, regulation 4.2 of these
regulations shall not apply if the CE marking or any inscription liable
to be confused therewith is affixed to the medical devices.
4.4. Where a device is intended to administer a medicinal product
within the meaning of Article 1 of Directive 65/65/EEC 5 on the
approximation of provisions laid down by law, regulation or
administrative action relating to proprietary medicinal products, that
device shall be governed by these regulations, without prejudice to the
provisions
of Directive 65/65/EEC with regard to the medicinal product.
_____
OJ No.L22, 09-06-65, P. 369/65. Directive as last amended by Directive 93/39/EEC
(OJ No.L214, 24-08-93, P. 22)
4.4.1. If, however, such a device is placed on the market in
such a way that the device and the medicinal product form a single
integral product which is intended exclusively for use in the given
combination and which is not reusable, that single product shall
be governed by Directive 65/65/EEC.
4.4.2. The relevant essential requirements of Annex I of the
Directive, which is set out in Schedule I to these regulations shall
apply as far as safety and performance related device features are
concerned.
4.5. Where a device incorporates, as an integral part, a substance
which, if used separately, may be considered to be a medicinal product
within the meaning of Article 1 of Directive 65/65/EEC and which is liable
to act upon the body with action ancillary to that of the device, that device
must be assessed and authorised in accordance with these regulations.
4.5.1. Where a device incorporates, as an integral part, a substance
which, if used separately, may be considered to be a medicinal product
constituent or a medicinal product derived from human blood or human
plasma within the meaning of Article 1 of Directive 89/381/EEC 6 and
which is liable to act upon the human body with action ancillary to that
of the device, hereinafter referred to as a ‘human blood derivative’, that
device must be assessed and authorised in accordance with these
regulations.
4.6.
These regulations do not apply to
(a) in vitro diagnostic devices;
(b) active implantable devices covered by Directive 90/385/
EEC, as implemented by L.N. 66 of 2001: Active Implantable
Medical Devices Regulations, 2001;
(c) medicinal products covered by Directive 65/65/EEC 7 ,
including medicinal products derived from blood as covered by
Directive 89/381/EEC;
(d) cosmetic products covered by Directive 76/768/EEC 8 ;
(e) human blood, blood products, plasma or blood cells of
human origin or to devices which incorporate at the time of placing
on the market such blood products, plasma or cells, with the
_____
OJ No. L 181, 28-06-89, P. 44-46.
OJ No. P 022, 09-02-65, P. 369-373.
OJ No.L262, 27-09-76, P.169. Directive as last amended by Commission Directive 97/45/EC
(OJ No. L196, 24-07-97, P.17)
exception of devices referred to in regulation 4.5.1 of these
regulations;
(f) transplants or tissues or cells of human origin nor to
products incorporating or derived from tissues or cells of human
origin;
(g) transplants or tissues or cells of animal origin, unless a
device is manufactured utilising animal tissue which is rendered
non-viable or non-viable products derived from animal tissue.
4.7. These regulations do not apply to personal protective
equipment covered by Directive 89/686/EEC, as implemented by L.N.
122 of 1998: Personal Protective Equipment Order (M34 of 1998). In
deciding whether a product falls under those regulations or the present
regulations, particular account shall be taken of the principal intended
purpose of the product.
1. The devices must meet the essential requirements set out in
Annex I of the Directive, which is set out in Schedule I of these
regulations which apply to them, taking account of the intended purpose
of the devices concerned.
5.2. All necessary steps shall be taken to ensure that devices may
be placed on the market and/or put into service only if they comply
with the requirements laid down in these regulations when duly supplied
and properly installed, maintained and used in accordance with their
intended purpose.
5.3. No obstacle shall be created to the placing on the market or
the putting into service of devices bearing the CE marking provided for
in regulations 12.1, 12.2 and 12.3 of these regulations which indicate
that they have been the subject of an assessment of their conformity in
accordance with the provisions of regulations 8.1 to 8.12 of these
regulations.
5.4.
No obstacle shall be created to
- devices intended for clinical investigation being made
available to medical practitioners or authorised persons for that
purpose if they meet the conditions laid down in regulation 11.1
and in Annex VIII of the Directive, which is set out in Schedule
VIII to these regulations,
- custom-made devices being placed on the market and put
into service if they meet the conditions laid down in regulations
1 to 8.12 in combination with Annex VIII of the Directive, which
is set out in Schedule VIII to these regulations; Class IIa, IIb and
III devices shall be accompanied by the statement referred to in
Schedule VIII to these regulations.
5.4.1.
These devices shall not bear the CE marking.
5.5. At trade fairs, exhibitions, demonstrations, etc. no obstacle
shall be created to the showing of devices which do not conform to
these regulations, provided that a visible sign clearly indicates that such
devices cannot be marketed or put into service until they have been
made to comply.
5.6. When a device reaches the final user, regardless of whether
it is for professional or other use, the information which must be made
available to the user and the patient in accordance with section 13 of
Schedule I to these regulations, must be provided in Maltese or in any
other language of the Community acceptable to the Maltese authorities.
6.1. Compliance with the essential requirements referred to in CE marking and
regulation 5.1 of these regulations shall be presumed in respect of devices certificates
conformity.
which are in conformity with the relevant national standards adopted
pursuant to the harmonized standards the references of which have been
published in the Official Journal of the European Communities.
6.1.1. For the purposes of these regulations, reference to
harmonized standards also includes the monographs of the
European Pharmacopoeia notably on surgical sutures and on
interaction between medicinal products and materials used in
devices containing such medicinal products, the references of which
have been published in the Official Journal of the European
Communities.
6.2. Where the devices are subject to other Directives concerning
other aspects and which also provide for the affixing of the CE marking,
the latter shall indicate that the devices also fulfil the provisions of the
other Directives.
6.2.1 However, should one or more of these Directives allow
the manufacturer, during a transitional period, to choose which
arrangements to apply, the CE marking shall indicate that the
devices fulfil the provisions only of those Directives applied by
the manufacturer. In this case, the particulars of these Directives,
as published in the Official Journal of the European Communities,
must be given in the documents, notices or instructions required
by the Directives and accompanying such devices.
7.1. Where it is ascertained that the devices referred to in
regulation 5.3 and in the second indent of regulation 5.4 of these
regulations, when correctly installed, maintained and used for their
intended purpose, may compromise the health and/or safety of patients,
users or, where applicable, other persons, all appropriate interim
measures shall be taken by the Director of Consumer Affairs to withdraw
such devices from the market or prohibit or restrict their being placed
on the market or put into service.
7.2. If a device which is not in conformity with the relevant
requirements bears the CE marking all appropriate measures shall be
taken with regard to those responsible for affixing the mark.
7.3. The Director of Consumer Affairs shall take the necessary
steps to ensure that any information brought to his knowledge, in
accordance with the provisions of these regulations, regarding the
incidents mentioned below involving a Class I, IIa, IIb or III device is
recorded and evaluated centrally
(a) any malfunction or deterioration in the characteristics and/
or performance of a device, as well as any inadequacy in the
labelling or the instructions for use which might lead to or might
have led to the death of a patient or user or to a serious deterioration
in his state of health;
(b) any technical or medical reason in relation to the
characteristics or performance of a device for the reasons referred
to in subparagraph (a), leading to systematic recall of devices of
the same type by the manufacturer.
7.4. The Consumer and Industrial Goods Directorate 9 of the Malta
Standards Authority shall be immediately informed of any such measures
taken under regulations 7.1, 7.2 and 7.3 above, indicating the reasons
for such decision and, in particular, whether non-compliance is due to
(a) failure to satisfy the essential requirements referred to in
regulation 5.1 of these regulations;
(b) incorrect application of the standards referred to in
regulations 6.1 and 6.1.1 of these regulations, in so far as it is
claimed that the standards have been applied;
(c) shortcomings in the standards themselves.
7.4.1. An assessment of the incidents mentioned in regulations
7.3 above will be carried out by the Malta Standards Authority, if possible
together with the manufacturer.
L.N. 213 of 2000 – Malta Standards Authority (Establishment of Directorates) Order, 2000, published in
the Government Gazette No. 17,002 of 20 th October, 2000.
7.5.
Any decision taken pursuant to these regulations
(a) to refuse or restrict the placing on the market or the putting
into service of a device or the carrying out of clinical investigations;
or
(b) to withdraw devices from the market, shall state the exact
grounds on which it is based.
7.5.1. Such decisions shall be notified without delay to the
party concerned, who shall at the same time be informed of the
remedies available to him under the Maltese law and of the time
limits to which such remedies are subject.
7.6. In the event of a decision as referred to in regulation 7.5 of
these regulations, the responsible person shall have an opportunity to
put forward his viewpoint in advance, unless such consultation is not
possible because of the urgency of the measure to be taken.
7.7. Without prejudice to the existing national provisions and
practices on medical secrets, the Director of Consumer Affairs shall
ensure that all the parties involved in the application of these regulations
are bound to observe confidentiality with regard to all information
obtained in carrying out their tasks. This does not affect the obligation
of the above authority with regard to the dissemination of warnings,
nor the obligations of the persons concerned to provide information
under criminal law.
7.8. Where the Director of Consumer Affairs considers, in relation
to a product or group of products, that, in order to ensure protection of
health and safety and/or to ensure that public health requirements are
observed pursuant to Article 36 of the Treaty, the availability of such
products should be prohibited, restricted or subjected to particular
requirements, any necessary and justified transitional measures may be
taken.
7.9. The Consumer and Industrial Goods Directorate of the Malta
Standards Authority shall immediately inform the Commission of any
measures taken in pursuance of regulations 7.1, 7.2, 7.3 of these
regulations.
8.1. In the case of devices falling within Class III, other than Conformity
devices which are custom-made or intended for clinical investigations, Assessment
Procedures.
the responsible person shall, in order to affix the CE marking, either
(a) follow the procedure relating to the EC declaration of
conformity set out in Annex II of the Directive, which is set out in
Schedule II (full quality assurance system) to these regulations; or
(b) follow the procedure relating to the EC type-examination
set out in Annex III of the Directive, which is set out in Schedule
III to these regulations, coupled with
(i) the procedure relating to the EC verification set out
in Annex IV of the Directive, which is set out in Schedule IV
to these regulations; or
(ii) the procedure relating to the EC declaration of
conformity set out in Annex V of the Directive, which is set
out in Schedule V (production quality assurance) to these
regulations.
8.2. In the case of devices falling within Class IIa, other than
devices which are custom-made or intended for clinical investigations,
the responsible person shall, in order to affix the CE marking, follow
the procedure relating to the EC declaration of conformity set out in
Annex VII of the Directive, which is set out in Schedule VII to these
regulations, coupled with either
(a) the procedure relating to the EC verification set out in
Annex IV of the Directive, which is set out in Schedule IV to these
regulations; or
(b) the procedure relating to the EC declaration of
conformity set out in Annex V of the Directive, which is set out in
Schedule V (production quality assurance) to these regulations;
or
(c) the procedure relating to the EC declaration of conformity
set out in Annex VI of the Directive, which is set out in Schedule
VI (product quality assurance) to these regulations.
8.2.1. Instead of applying these procedures, the
manufacturer may also follow the procedure referred to in
regulation 8.3(a) below.
8.3. In the case of devices falling within Class IIb, other than
devices which are custom-made or intended for clinical investigations,
the responsible person shall, in order to affix the CE marking, either
(a) follow the procedure relating to the EC declaration of
conformity set out in Annex II of the Directive, which is set out in
Schedule II (full quality assurance) to these regulations; in this
case, point 4 of Annex II of the Directive, which is set out in
Schedule II to these regulations is not applicable; or
(b) follow the procedure relating to the EC type-examination
set out in Annex III of the Directive, which is set out in ScheduleIII
of these regulations, coupled with
(i) the procedure relating to the EC verification set out
in Annex IV of the Directive, which is set out in Schedule IV
of these regulations; or
(ii) the procedure relating to the EC declaration of
conformity set out in Annex V of the Directive, which is set
out in Schedule V (production quality assurance) to these
regulations; or
(iii) the procedure relating to the EC declaration of
conformity set out in Annex VI of the Directive, which is set
out in ScheduleVI (product quality assurance) to these
regulations.
8.4. In the case of devices falling within Class I, other than devices
which are custom-made or intended for clinical investigations, the
responsible person shall, in order to affix the CE marking, follow the
procedure referred to in Annex VII of the Directive, which is set out in
Schedule VII to these regulations and draw up the EC declaration of
conformity required before placing the device on the market.
8.5. In the case of custom-made devices, the responsible person
shall follow the procedure referred to in Annex VIII of the Directive,
which is set out in Schedule VIII to these regulations and draw up the
statement set out in that Schedule before placing each device on the
market.
8.5.1. The responsible person shall submit to the Director
General Health and the Malta Standards Authority a list of such
devices which have been put into service in Malta.
8.6. During the conformity assessment procedure for a device,
the responsible person and/or the notified body shall take account of
the results of any assessment and verification operations which, where
appropriate, have been carried out in accordance with these regulations
at an intermediate stage of manufacture.
8.7. The manufacturer may instruct his authorized representative
or the importer to initiate the procedures provided for in Annexes III,
IV, VII and VIII of the Directive, which are set out in Schedules III, IV,
VII and VIII to these regulations.
8.8. Where the conformity assessment procedure involves the
intervention of a notified body, the responsible person may apply to a
body of his choice within the framework of the tasks for which the
body has been notified.
8.9. The notified body may require, where duly justified, any
information or data, which is necessary for establishing and maintaining
the attestation of conformity in view of the chosen procedure.
8.10. Decisions taken by the notified bodies in accordance with
Annexes II and III of the Directive, which are set out in Schedules II
and III to these regulations, shall be valid for a maximum of five years
and may be extended on application, made at a time agreed in the contract
signed by both parties, for further periods of five years.
8.11. The records and correspondence relating to the procedures
referred to in regulations 8.1 to 8.5 of these regulations shall be in
Maltese or in another Community language acceptable to the notified
body.
8.12. By derogation from regulations 8.1 to 8.5 of these
regulations, the Director of Consumer Affairs, after consulting the
Director General Health may authorise, on duly justified request, the
placing on the market and putting into service, within the Maltese
territory, of individual devices for which the procedures referred to in
regulations 8.1 to 8.5 have not been carried out and the use of which is
in the interest of protection of health.
Particular procedure
1. By way of derogation from regulations 8.1 to 8.12 of these
regulations, the following subparagraphs shall apply to systems and
procedure packs
9.1.1. Any natural or legal person who puts devices bearing
the CE marking together within their intended purpose and within
the limits of use specified by their manufacturers, in order to place
them on the market as a system or procedure pack, shall draw up a
declaration by which he states that
(a) he has verified the mutual compatibility of the devices
in accordance with the manufacturers’ instructions and has
carried out his operations in accordance with these
instructions; and
(b) he has packaged the system or procedure pack and
supplied relevant information to users incorporating relevant
instructions from the manufacturers; and
(c) the whole activity is subjected to appropriate methods
of internal control and inspection.
9.1.1.1. Where the conditions above are not met, as in cases
where the system or procedure pack incorporate devices which do
not bear a CE marking or where the chosen combination of devices
is not compatible in view of their original intended use, the system
or procedure pack shall be treated as a device in its own right and
as such be subjected to the relevant procedure pursuant to
regulations 8.1 to 8.12 of these regulations.
9.1.2. Any natural or legal person who sterilised, for the
purpose of placing on the market, systems or procedure packs
referred to in regulation 9.1.1 of these regulations or other CE-
marked medical devices designed by their manufacturers to be
sterilised before use, shall, at his choice, follow one of the
procedures referred to in Annexes IV, V or VI of the Directive,
which are set out in Schedules IV, V or VI to these regulations.
The application of the abovementioned Annexes and the
intervention of the notified body are limited to the aspects of the
procedure relating to the obtaining of sterility. The person shall
draw up a declaration stating that sterilisation has been carried out
in accordance with the manufacturer’s instructions.
9.1.3. The products referred to in regulations 9.1.1 and 9.1.2
of these regulations shall not themselves bear an additional CE
marking. They shall be accompanied by the information referred
to in point 13 of Annex I of the Directive, which is set out in
Schedule I to these regulations which includes, where appropriate,
the information supplied by the manufacturers of the devices which
have been put together. The declaration referred to in regulations
9.1.1 and 9.1.2 of these regulations shall be kept at the disposal of
the Director of Consumer Affairs for a period of five years.
1. Any manufacturer who, under his own name, places devices Registration of
on the market in accordance with the procedures referred to in regulations persons
for placing devices
8.4 and 8.5 of these regulations and any other natural or legal person on the market.
engaged in the activities referred to in regulation 9.1 of these regulations
shall inform the Malta Standards Authority of his registered place of
business and the description of the devices concerned.
10.1.1. For all medical devices of classes IIb and III, all
data allowing for identification of such devices together with the
label and the instructions for use, may be requested when such
devices are put into service within Malta.
10.2. Where a manufacturer who places devices, referred to in
regulation 10.1, on the market under his own name does not have a
registered place of business in a Member State, he shall designate the
person(s) responsible for marketing them who is (are) established in
the Community.
10.2.1 The persons referred to in regulation 10.2 above, who
have their registered place of business in Malta, shall inform the
Malta Standards Authority of the address of the registered place of
business and the category of devices concerned.
1. In the case of devices intended for clinical investigations,
the responsible person shall follow the procedure referred to in Annex
VIII of the Directive, which is set out in Schedule VIII to these
regulations and notify the Director General Health about the
investigations that are to be conducted.
11.1.1. In the case of devices falling within Class III and
implantable and long-term invasive devices falling within Class
IIa or IIb, the manufacturer may commence the relevant clinical
investigation at the end of a period of 60 days after notification,
unless the Director General Health has notified him within that
period of a decision to the contrary based on considerations of
public health or public policy.
11.1.1.1. However, manufacturers may be authorised
to commence the relevant clinical investigations before the
expiry of the period of 60 days, in so far as the Director
General Health, acting on the advice of the Ethics Committee,
has issued a favourable opinion on the programme of
investigation in question.
11.1.2. In the case of devices other than those referred to in
regulation 11.1.1 above, manufacturers may be authorised to
commence clinical investigations, immediately after the date of
notification, provided that the Director General Health, acting on
the advice of the Ethics Committee, has delivered a favourable
opinion with regard to the investigational plan.
11.1.3. The clinical investigations must be conducted in
accordance with the provisions of Annex X of the Directive, which
is set out in Schedule X to these regulations.
11.1.4. The Director General Health shall, if necessary, take
the appropriate steps to ensure public health and public policy.
11.1.5. The responsible person shall keep the report referred
to in point 2.3.7 of Annex X of the Directive, which is set out in
Schedule X to these regulations at the disposal of the Director
General Health and the Director of Consumer Affairs.
11.1.6. The provisions of regulations 11.1 and 11.1.1 of these
regulations do not apply where the clinical investigations are
conducted using devices which are authorised in accordance with
regulations 8.1 to 8.12 of these regulations to bear the CE marking
unless the aim of these investigations is to use the devices for a
purpose other than that referred to in the relevant conformity
assessment procedure. The relevant provisions of Annex X of the
Directive, which is set out in Schedule X to these regulations remain
applicable.
1. Devices, other than devices which are custom-made or CE Marking.
intended for clinical investigations, considered to meet the essential
requirements referred to in regulation 5.1 of these regulations must bear
the CE marking of conformity when they are placed on the market.
12.2. The CE marking of conformity, as shown in Annex XII of
the Directive, which is set out in Schedule XII to these regulations,
must appear in a visible, legible and indelible form on the device or its
sterile pack, where practicable and appropriate, and on the instructions
for use. Where applicable, the CE marking must also appear on the
sales packaging.
12.2.1. The CE marking shall be accompanied by the
identification number of the notified body responsible for
implementation of the procedures set out in Annexes II, IV, V and
VI of the Directive, which are set out in Schedules II, IV, V and VI
to these regulations.
12.3. It is prohibited to affix marks or inscriptions which are
likely to mislead third parties with regard to the meaning or the graphics
of the CE marking. Any other mark may be affixed to the device, to the
packaging or to the instruction leaflet accompanying the device provided
that the visibility and legibility of the CE marking is not thereby reduced.
12.4.
Without prejudice to regulation 7 of these regulations
12.4.1. where it is established that the CE marking has been
affixed unduly, the responsible person shall be obliged to end the
infringement, under the conditions imposed by the Director of
Consumer Affairs.
12.4.2. where non-compliance continues, all appropriate
measures shall be taken by the Director of Consumer Affairs to
restrict or prohibit the placing on the market of the product in
question or to ensure that it is withdrawn from the market in
accordance with the procedure as laid down in regulation 7 of these
regulations.
12.5. Those provisions shall also apply where the CE marking
has been affixed in accordance with the procedures in these regulations,
but inappropriately, on products that are not covered by these regulations.
1.
The Medical Devices Regulations, 2001 are hereby being
revoked.
B
Schedule I
Based on Annex I of the Directive
ESSENTIAL REQUIREMENTS
I.
GENERAL REQUIREMENTS
1. The devices must be designed and manufactured in such a way that, when
used under the conditions and for the purposes intended, they will not compromise the
clinical condition or the safety of patients, or the safety and health of users or, where
applicable, other persons, provided that any risks which may be associated with their
use constitute acceptable risks when weighed against the benefits to the patient and are
compatible with a high level of protection of health and safety.
2. The solutions adopted by the manufacturer for the design and construction of
the devices must conform to safety principles, taking account of the generally
acknowledged state of the art.
In selecting the most appropriate solutions, the manufacturer must apply the
following principles in the following order
- eliminate or reduce risks as far as possible (inherently safe design and
construction),
- where appropriate take adequate protection measures including alarms if
necessary, in relation to risks that cannot be eliminated,
- inform users of the residual risks due to any shortcomings of the protection
measures adopted.
3. The devices must achieve the performances intended by the manufacturer
and be designed, manufactured and packaged in such a way that they are suitable for
one or more of the functions referred to in regulation 3.1.1 of these regulations, as
specified by the manufacturer.
4. The characteristics and performances referred to in sections 1, 2 and 3 of this
Schedule must not be adversely affected to such a degree that the clinical conditions
and safety of the patients and, where applicable, of other persons are compromised
during the lifetime of the device as indicated by the manufacturer, when the device is
subjected to the stresses which can occur during normal conditions of use.
5. The devices must be designed, manufactured and packed in such a way that
their characteristics and performances during their intended use will not be adversely
affected during transport and storage taking account of the instructions and information
provided by the manufacturer.
6. Any undesirable side-effect must constitute an acceptable risk when weighed
against the performances intended.
II. REQUIREMENTS REGARDING DESIGN AND CONSTRUCTION
7. Chemical, physical and biological properties
7.1 The devices must be designed and manufactured in such a way as to guarantee
the characteristics and performances referred to in section I of this Schedule relating to
the ‘General requirements’. Particular attention must be paid to
- the choice of materials used, particularly as regards toxicity and, where
appropriate, flammability,
- the compatibility between the materials used and biological tissues, cells
and body fluids, taking account of the intended purpose of the device.
7.2 The devices must be designed, manufactured and packed in such a way as
to minimise the risk posed by contaminants and residues to the persons involved in the
transport, storage and use of the devices and to the patients taking account of the intended
purpose of the product. Particular attention must be paid to the tissues exposed and to
the duration and frequency of exposure.
7.3 The devices must be designed and manufactured in such a way that they can
be used safely with the materials, substances and gases with which they enter into
contact during their normal use or during routine procedures; if the devices are intended
to administer medicinal products they must be designed and manufactured in such a
way as to be compatible with the medicinal products concerned according to the
provisions and restrictions governing these products and that their performance is
maintained in accordance with the intended use.
7.4 Where a device incorporates, as an integral part, a substance which, if used
separately, may be considered to be a medicinal product as defined in Article 1 of
Directive 65/65/EEC and which is liable to act upon the body with action ancillary to
that of the device, the safety, quality and usefulness of the substance must be verified,
taking account of the intended purpose of the device, by analogy with the appropriate
methods specified in Directive 75/318/EEC.
7.4.1. Where a device incorporates, as an integral part, a human blood
derivative, the notified body shall seek a scientific opinion from the European
Agency for the Evaluation of Medicinal Products (EMEA) on the quality and
safety of the derivative, taking account of the appropriate Community provisions
and, in particular, by analogy with the provisions of Directives 75/318/EEC 10 and
89/381/EEC. The usefulness of the derivative as a part of the medical device shall
be verified, taking account of the intended purpose of the device.
_____
OJ No. L 147, 09-06-75, P. 01-12.
B
7.4.2. In accordance with Article 4(3) of Directive 89/381/EEC, a sample
from each batch of bulk and/or finished product of the human blood derivative
shall be tested by a State laboratory or a laboratory designated for that purpose by
a Member State.
7.5 The devices must be designed and manufactured in such a way as to reduce
to a minimum the risks posed by substances leaking from the device.
7.6 Devices must be designed and manufactured in such a way as to reduce, as
much as possible, risks posed by the unintentional ingress of substances into the device
taking into account the device and the nature of the environment in which it is intended
to be used.
8. Infection and microbial contamination
8.1 The devices and manufacturing processes must be designed in such a way
as to eliminate or reduce as far as possible the risk of infection to the patient, user and
third parties. The design must allow easy handling and, where necessary, minimise
contamination of the device by the patient or vice versa during use.
8.2 Tissues of animal origin must originate from animals that have been subjected
to veterinary controls and surveillance adapted to the intended use of the tissues.
Notified bodies shall retain information on the geographical origin of the animals.
Processing, preservation, testing and handling of tissues, cells and substances of
animal origin must be carried out so as to provide optimal security. In particular safety
with regard to viruses and other transferable agents must be addressed by implementation
of validated methods of elimination or viral inactivation in the course of the
manufacturing process.
8.3 Devices delivered in a sterile state must be designed, manufactured and
packed in a non-reusable pack and/or according to appropriate procedures to ensure
that they are sterile when placed on the market and remain sterile, under the storage
and transport conditions laid down, until the protective packaging is damaged or opened.
8.4 Devices delivered in a sterile state must have been manufactured and sterilised
by an appropriate, validated method.
8.5 Devices intended to be sterilised must be manufactured in appropriately
controlled (e.g. environmental) conditions.
8.6 Packaging systems for non-sterile devices must keep the product without
deterioration at the level of cleanliness stipulated and, if the devices are to be sterilised
prior to use, minimize the risk of microbial contamination; the packaging system must
be suitable taking account of the method of sterilisation indicated by the manufacturer.
8.7 The packaging and/or label of the device must distinguish between identical
or similar products sold in both sterile and non-sterile condition.
9. Construction and environmental properties
9.1 If the device is intended for use in combination with other devices or
equipment, the whole combination, including the connection system must be safe and
must not impair the specified performances of the devices. Any restrictions on use
must be indicated on the label or in the instructions for use.
9.2 Devices must be designed and manufactured in such a way as to remove or
minimise as far is possible
- the risk of injury, in connection with their physical features, including
the volume/pressure ratio, dimensional and where appropriate ergonomic features,
- risks connected with reasonably foreseeable environmental conditions,
such as magnetic fields, external electrical influences, electrostatic discharge,
pressure, temperature or variations in pressure and acceleration,
- the risks of reciprocal interference with other devices normally used in
the investigations or for the treatment given,
- risks arising where maintenance or calibration are not possible (as with
implants), from ageing of materials used or loss of accuracy of any measuring or
control mechanism.
9.3 Devices must be designed and manufactured in such a way as to minimise
the risks of fire or explosion during normal use and in single fault condition. Particular
attention must be paid to devices whose intended use includes exposure to flammable
substances or to substances which could cause combustion.
10. Devices with a measuring function
10.1 Devices with a measuring function must be designed and manufactured in
such a way as to provide sufficient accuracy and stability within appropriate limits of
accuracy and taking account of the intended purpose of the device. The limits of accuracy
must be indicated by the manufacturer.
10.2 The measurement, monitoring and display scale must be designed in line
with ergonomic principles taking account of the intended purpose of the device.
10.3 The measurements made by the devices with a measuring function must
be expressed in legal units conforming to the provisions of Council Directive 80/181/
EEC 11 .
_____
OJ No. L39, 15-02-80, P. 40, Directive as last amended by Directive 89/617/EEC
(OJ No. L357, 07-12-89, P28).
B
11. Protection against radiation
11.1
General
11.1.1 Devices shall be designed and manufactured in such a way that exposure
of patients, users and other persons to radiation shall be reduced as far as possible
compatible with the intended purpose, whilst not restricting the application of
appropriate specified levels for therapeutic and diagnostic purposes.
11.2
Intended radiation
11.2.1 Where devices are designed to emit hazardous levels of radiation necessary
for a specific medical purpose the benefit of which is considered to outweigh the risks
inherent in the emission, it must be possible for the user to control the emissions. Such
devices shall be designed and manufactured to ensure reproducibility and tolerance of
relevant variable parameters.
11.2.2 Where devices are intended to emit potentially hazardous, visible and/or
invisible radiation, they must be fitted, where practicable, with visual displays and/
or audible warnings of such emissions.
11.3
Unintended radiation
11.3.1 Devices shall be designed and manufactured in such a way that exposure
of patients, users and other persons to the emission of unintended, stray or scattered
radiation is reduced as far as possible.
11.4
Instructions
11.4.1 The operating instructions for devices emitting radiation must give detailed
information as to the nature of the emitted radiation, means of protecting the patient
and the user and on ways of avoiding misuse and of eliminating the risks inherent in
installation.
11.5
Ionising radiation
11.5.1 Devices intended to emit ionising radiation must be designed and
manufactured in such a way as to ensure that, where practibile, the quantity, geometry
and quality of radiation emitted can be varied and controlled taking into account the
intended use.
11.5.2 Devices emitting ionising radiation intended for diagnostic radiology shall
be designed and manufactured in such a away as to achieve appropriate image and/or
output quality for the intended medical purpose whilst minimising radiation exposure
of the patient and user.
11.5.3 Devices emitting ionising radiation, intended for therapeutic radiology
shall be designed and manufactured in such a way as to enable reliable monitoring and
control of the delivered dose, the beam type and energy and where appropriate the
quality of radiation.
12 Requirements for medical devices connected to or equipped with an
energy source
12.1 Devices incorporating electronic programmable systems must be designed
to ensure the repeatability, reliability and performance of these systems according to
the intended use. In the event of a single fault condition (in the system) appropriate
means should be adopted to eliminate or reduce as far as possible consequent risks.
12.2 Devices where the safety of the patients depends on an internal power
supply must be equipped with a means of determining the state of the power supply.
12.3 Devices where the safety of the patients depends on an external power
supply must include an alarm system to signal any power failure.
12.4 Devices intended to monitor one or more clinical parameters of a patient
must be equipped with appropriate alarm systems to alert the user of situations which
could lead to death or severe deterioration of the patient’s state of health.
12.5 Devices must be designed and manufactured in such a way as to minimise
the risks of creating electromagnetic fields which could impair the operation of other
devices or equipment in the usual environment.
12.6
Protection against electrical risks
Devices must be designed and manufactured in such a way as to avoid, as far as
possible, the risk of accidental electric shocks during normal use and in single fault
condition, provided the devices are installed correctly.
12.7
Protection against mechanical and thermal risks
12.7.1 Devices must be designed and manufactured in such a way as to protect
the patient and user against mechanical risks connected with, for example resistance,
stability and moving parts.
2.2.2 Devices must be designed and manufactured in such a way as to reduce to
the lowest possible level the risks arising from vibration generated by the devices,
taking account of technical progress and of the means available for limiting vibrations,
particularly at source, unless the vibrations are part of the specified performance.
12.7.3 Devices must be designed and manufactured in such a way as to reduce
to the lowest possible level the risks arising from the noise emitted, taking account of
B
technical progress and of the means available to reduce noise, particularly at source,
unless the noise emitted is part of the specified performance.
12.7.4 Terminals and connectors to the electricity, gas or hydraulic and pneumatic
energy supplies which the user has to handle must be designed and constructed in such
a way as to minimise all possible risks.
12.7.5 Accessible parts of the devices (excluding the parts or areas intended to
supply heat or reach given temperatures) and their surroundings must not attain
potentially dangerous temperatures under normal use.
12.8 Protection against the risks posed to the patient by energy supplies or
substances
12.8.1 Devices for supplying the patient with energy or substances must be
designed and constructed in such a way that the flow-rate can be set and maintained
accurately enough to guarantee the safety of the patient and of the user.
12.8.2 Devices must be fitted with the means of preventing and/or indicating
any inadequacies in the flow-rate which could pose a danger.
Devices must incorporate suitable means to prevent, as far as possible, the
accidental release of dangerous levels of energy from an energy and/or substance source.
12.9
The function of the controls and indicators must be clearly specified on the
devices.
Where a device bears instructions required for its operation or indicates operating
or adjustment parameters by means of a visual system, such information must be
understandable to the user and, as appropriate, the patient.
13. Information supplied by the manufacturer
13.1 Each device must be accompanied by the information needed to use it
safely and to identify the manufacturer, taking account of the training and knowledge
of the potential users.
This information comprises the details on the labels and the data in the instructions
for use.
As far as practicable and appropriate, the information needed to use the device
safely must be set out on the device itself and/or on the packaging for each unit, where
appropriate, on the sales packaging. If individual packaging of each unit is not
practicable, the information must be set out in the leaflet supplied with one or more
devices.
Instructions for use must be included in the packaging for every device. By way
of exception, no such instructions for use are needed for devices in Class I or IIa if they
can be used safely without any such instructions.
13.2 Where appropriate, this information should take the form of symbols. Any
symbol or identification colour used must conform to the harmonized standards. In
areas for which no standards exist, the symbols and colours must be described in the
documentation supplied with the device.
13.3
The label must bear the following particulars
(a) the name or trade name and address of the manufacturer. For devices
imported in Malta, in view of their distribution locally, the label, or the outer
packaging, or instructions for use, shall contain in addition the name and address
of either the person responsible referred to in regulation 8.15 of these regulations
or of the authorised representative of the manufacturer or of the importer established
in Malta, as appropriate;
(b) the details strictly necessary for the user to identify the device and the
contents of the packaging;
(c)
where appropriate, the word ‘STERILE’;
(d) where appropriate, the batch code, preceded by the word ‘LOT’, or the
serial number;
(e) where appropriate, an indication of the date by which the device should
be used, in safety, expressed as the year and month;
(f)
where appropriate, an indication that the device is for single use;
(g) if the device is custom-made, the words ‘custom-made device’;
(h) if the device is intended for clinical investigations, the words
‘exclusively for clinical investigations’;
(i)
any special storage and/or handling conditions;
(j)
any special operating instructions;
(k) any warnings and/or precautions to take;
(l) year of manufacture for active devices other than those covered by (e).
This indication may be included in the batch or serial number;
(m)
where applicable, method of sterilisation.
B
(n) in the case of a device within the meaning of regulation 4.5 of these
regulations, an indication that the device contains a human blood derivative.
13.4 If the intended purpose of the device is not obvious to the user, the
manufacturer must clearly state it on the label and in the instructions for use.
13.5 Wherever reasonable and practicable, the devices and detachable
components must be identified, where appropriate in terms of batches, to allow all
appropriate action to detect any potential risk posed by the devices and detachable
components.
13.6 Where appropriate, the instructions for use must contain the following
particulars
(a)
the details referred to in section 13.3, with the exception of (d) and (e);
(b) the performances referred to in section 3 and any undesirable side-
effects;
(c) if the device must be installed with or connected to other medical devices
or equipment in order to operate as required for its intended purpose, sufficient
details of its characteristics to identify the correct devices or equipment to use in
order to obtain a safe combination;
(d) all the information needed to verify whether the device is properly
installed and can operate correctly and safely, plus details of the nature and
frequency of the maintenance and calibration needed to ensure that the devices
operate properly and safely at all times;
(e) where appropriate, information to avoid certain risks in connection with
implantation of the device;
(f) information regarding the risks of reciprocal interference posed by the
presence of the device during specific investigations or treatment;
(g) the necessary instructions in the event of damage to the sterile packaging
and, where appropriate, details of appropriate methods of resterilisation;
(h) if the device is reusable, information on the appropriate processes to
allow reuse, including cleaning, disinfection, packaging and, where appropriate,
the method of sterilisation of the device to be resterilised, and any restriction on
the number of reuses;
Where devices are supplied with the intention that they be sterilised before use,
the instructions for cleaning and sterilisation must be such that, if correctly followed,
the device will still comply with the requirements in section I;
(i) details of any further treatment or handling needed before the device
can be used (for example, sterilisation, final assembly etc.);
(j) in the case of devices emitting radiation for medical purposes, details
of the nature, type, intensity and distribution of this radiation;
The instructions for use must also include details allowing the medical staff to
brief the patient on any contra-indications and any precautions to be taken. These
details should cover in particular
(k) precautions to be taken in the event of changes in the performance of
the device;
(l) precautions to be taken as regards exposure, in reasonably foreseeable
environmental conditions, to magnetic fields, external electrical influences,
electrostatic discharge, pressure or variations in pressure, acceleration, thermal
ignition sources, etc;
(m) adequate information regarding the medicinal product or products
which the device in question is designed to administer, including any limitations
in the choice of substances to be delivered;
(n) precautions to be taken against any special, unusual risks related to the
disposal of the device;
(o) medicinal substances incorporated into the device as an integral part in
accordance with section 7.4;
(p) degree of accuracy claimed for devices with a measuring function.
Where conformity with the essential requirements must be based on clinical
data, as in section I (6), such data must be established in accordance with Annex X of
the Directive, which is set out in Schedule X to these regulations.
B
Schedule II
Based on Annex II of The Directive
EC DECLARATION OF CONFORMITY (Full quality assurance system)
1. The manufacturer must ensure application of the quality system approved for
the design, manufacture and final inspection of the products concerned, as specified in
section 3 and is subject to audit as laid down in sections 3.3 and 4 and to Community
surveillance as specified in section 5.
2. The declaration of conformity is the procedure whereby the manufacturer who
fulfils the obligations imposed by section 1 ensures and declares that the products
concerned meet the provisions of these regulations which apply to them.
The manufacturer must affix the CE marking in accordance with regulation 9 of
these regulations and draw up a written declaration of conformity. This declaration
must cover a given number of the products manufactured and be kept by the
manufacturer.
3. Quality system
3.1. The manufacturer must lodge an application for assessment of his quality
system with a notified body.
The application must include
- the name and address of the manufacturer and any additional manufacturing
site covered by the quality system,
- all the relevant information on the product or product category covered by
the procedure,
- a written declaration that no application has been lodged with any other
notified body for the same product-related quality system,
- the documentation on the quality system,
- an undertaking by the manufacturer to fulfil the obligations imposed by the
quality system approved,
- an undertaking by the manufacturer to keep the approved quality system
adequate and efficacious,
- an undertaking by the manufacturer to institute and keep up to date a
systematic procedure to review experience gained from devices in the post-
production phase and to implement appropriate means to apply any necessary
corrective action. This undertaking must include an obligation for the manufacturer
to notify the competent authorities of the following incidents immediately on
learning of them
(i) any malfunction or deterioration in the characteristics and/or
performance of a device, as well as any inadequacy in the instructions for
use which might lead to or might have led to the death of a patient or user or
to a serious deterioration in his state of health;
(ii) any technical or medical reason connected with the characteristics
or performance of a device leading for the reasons referred to in subparagraph
(i) to systematic recall of devices of the same type by the manufacturer.
3.2. Application of the quality system must ensure that the products conform to
the provisions of these regulations which apply to them at every stage, from design to
final inspection. All the elements, requirements and provisions adopted by the
manufacturer for his quality system must be documented in a systematic and orderly
manner in the form of written policies and procedures such as quality programmes,
quality plans, quality manuals and quality records.
It shall include in particular an adequate description of
(a) the manufacturer’s quality objectives;
(b) the organization of the business and in particular
- the organizational structures, the responsibilities of the managerial
staff and their organizational authority where quality of design and
manufacture of the products is concerned,
- the methods of monitoring the efficient operation of the quality system
and in particular its ability to achieve the desired quality of design and of
product, including control of products which fail to conform;
(c) the procedures for monitoring and verifying the design of the products
and in particular
- a general description of the product, including any variants planned,
- the design specifications, including the standards which will be applied
and the results of the risk analysis, and also a description of the solutions
adopted to fulfil the essential requirements which apply to the products if
the standards referred to in Article 5 are not applied in full,
B
- the techniques used to control and verify the design and the processes
and systematic measures which will be used when the products are being
designed,
- if the device is to be connected to other device(s) in order to operate as
intended, proof must be provided that it conforms to the essential requirements
when connected to any such device(s) having the characteristics specified
by the manufacturer,
- a statement indicating whether or not the device incorporates, as an
integral part, a substance or a human blood derivative referred to in section
7.4 of Annex I of the Directive, which is set out in Schedule I hereto and the
data on the tests conducted in this connection required to assess the safety,
quality and usefulness of that substance or human blood derivative, taking
account of the intended purpose of the device,
- the clinical data referred to in Annex X of the Directive, which is set
out in Schedule X to these regulations,
- the draft label and, where appropriate, instructions for use;
(d) the inspection and quality assurance techniques at the manufacturing
stage and in particular
- the processes and procedures which will be used, particularly as regards
sterilization, purchasing and the relevant documents,
- the product identification procedures drawn up and kept up to date
from drawings, specifications or other relevant documents at every stage of
manufacture;
(e) the appropriate tests and trials which will be carried out before, during
and after manufacture, the frequency with which they will take place, and the test
equipment used; it must be possible to trace back the calibration of the test
equipment adequately.
3.3. The notified body must audit the quality system to determine whether it meets
the requirements referred to in section 3.2. It must presume that quality systems which
implement the relevant harmonized standards conform to these requirements.
The assessment team must include at least one number with past experience of
assessments of the technology concerned. The assessment procedure must include an
inspection on the manufacturer’s premises and, in duly substantiated cases, on the
premises of the manufacturer’s suppliers and/or subcontractors to inspect the
manufacturing processes.
The decision is notified to the manufacturer. It must contain the conclusions of
the inspection and a reasoned assessment.
3.4. The manufacturer must inform the notified body which approved the quality
system of any plan for substantial changes to the quality system or the product-range
covered. The notified body must assess the changes proposed and verify whether after
these changes the quality system still meets the requirements referred to in section 3.2.
It must notify the manufacturer of its decision. This decision must contain the
conclusions of the inspection and a reasoned assessment.
4. Examination of the design of the product
4.1. In addition to the obligations imposed by section 3, the manufacturer must
lodge with the notified body an application for examination of the design dossier relating
to the product which he plans to manufacture and which falls into the category referred
to in section 3.1.
4.2. The application must describe the design, manufacture and performances of
the product in question. It must include the documents needed to assess whether the
product conforms to the requirements of these regulations, as referred to in section
3.2(c).
4.3. The notified body must examine the application and, if the product conforms
to the relevant provisions of these regulations, issue the application with an EC design-
examination certificate. The notified body may require the application to be completed
by further tests or proof to allow assessment of conformity with the requirements of
the regulations. The certificate must contain the conclusions of the examination, the
conditions of validity, the data needed for identification of the approved design, where
appropriate, a description of the intended purpose of the product.
In the case of devices referred to in Annex I, section 7.4, first subparagraph, of the
Directive, which is set out in Schedule I hereto the notified body shall, as regards the
aspects referred to in that section, consult one of the competent bodies designated by
the Member States in accordance with Directive 65/65/EEC before taking a decision.
The notified body will give due consideration to the views expressed in this consultation
when making its decision. It will convey its final decision to the competent body
concerned.
In the case of devices referred to in Annex I, section 7.4, second subparagraph of
the Directive, which is set out in Schedule I hereto, the scientific opinion of the EMEA
must be included in the documentation concerning the device. The notified body will
give due consideration to the opinion of the EMEA when making its decision. The
notified body may not deliver the certificate if the EMEA’s scientific opinion is
unfavourable. It will convey its final decision to the EMEA.
4.4. Changes to the approved design must receive further approval from the notified
body which issued the EC design-examination certificate wherever the changes could
B
affect conformity with the essential requirements of the regulations or with the conditions
prescribed for use of the product. The applicant shall inform the notified body which
issued the EC design-examination certificate of any such changes made to the approved
design. This additional approval must take the form of a supplement to the EC design-
examination certificate.
5. Surveillance
5.1. The aim of surveillance is to ensure that the manufacturer duly fulfils the
obligations imposed by the approved quality system.
5.2. The manufacturer must authorize the notified body to carry out all the necessary
inspections and supply it with all relevant information, in particular
- the documentation on the quality system,
- the data stipulated in the part of the quality system relating to design, such
as the results of analyses, calculation tests, etc.,
- the data stipulated in the part of the quality system relating to manufacture,
such as inspection reports and test data, calibration data, qualification reports of
the personnel concerned, etc.
5.3. The notified body must periodically carry out appropriate inspections and
assessments to make sure that the manufacturer applies the approved quality system
and must supply the manufacturer with an assessment report.
5.4. In addition, the notified body may pay unannounced visits to the manufacturer.
At the time of such visits, the notified body may, where necessary, carry out or ask for
tests in order to check that the quality system is working properly. It must provide the
manufacturer with an inspection report and, if a test has been carried out, with a test
report.
6. Administrative provisions
6.1. The manufacturer must, for a period ending at least five years after the last
product has been manufactured, keep at the disposal of the national authorities
- the declaration of conformity,
- the documentation referred to in the fourth indent of section 3.1,
- the changes referred to in section 3.4,
- the documentation referred to in section 4.2, and
- the decisions and reports from the notified body as referred to in sections
3.3, 4.3, 4.4, 5.3 and 5.4.
6.2. In respect of devices subject to the procedure in section 4, when neither the
manufacturer nor his authorized representative is established in the Community or
Malta, the obligation to keep available the technical documentation shall fall to the
person responsible for placing the device on the Community market or the importer
referred to in Annex I, section 13.3(a) of the Directive, which is set out in Schedule I
hereto.
7. Application to devices in Classes IIa and IIb
In line with regulations 8.2 and 8.3 of these regulations, this Schedule may apply
to products in Classes IIa and IIb. Section 4, however, does not apply.
8. Application to the devices referred to regulation 4.5 of these regulations
Upon completing the manufacture of each batch of devices referred to in regulation
4.5 of these regulations, the manufacturer shall inform the notified body of the release
of the batch of devices and send to it the official certificate concerning the release of
the batch of human blood derivative used in the device, issued by a State laboratory or
a laboratory designated for that purpose by a Member State in accordance with Article
4(3) of Directive 89/381/EEC.
B
Schedule III
Based on Annex III of the Directive
EC TYPE-EXAMINATION
1. EC type-examination is the procedure whereby a notified body ascertains and
certifies that a representative sample of the production covered fulfils the relevant
provisions of these regulations.
2. The application includes
- the name and address of the manufacturer and the name and address of the
authorized representative if the application is lodged by the representative,
- the documentation described in section 3 needed to assess the conformity
of the representative sample of the production in question, hereinafter referred to
as the ‘type’, with the requirements of these regulations. The applicant must make
a ‘type’ available to the notified body. The notified body may request other samples
as necessary,
- a written declaration that no application has been lodged with any other
notified body for the same type.
3. The documentation must allow an understanding of the design, the manufacture
and the performances of the product and must contain the following items in particular
- a general description of the type, including any variants planned,
- design drawings, methods of manufacture envisaged, in particular as regards
sterilization, and diagrams of components, sub-assemblies, circuits, etc.,
- the descriptions and explanations necessary to understand the
abovementioned drawings and diagrams and the operation of the product,
- a list of the standards referred to in regulation 6 of these regulations, applied
in full or in part, and descriptions of the solutions adopted to meet the essential
requirements if the standards referred to in regulation 6 of these regulations have
not been applied in full,
- the results of the design calculations, risk analysis, investigations, technical
tests, etc. carried out,
- a declaration stating whether or not the device incorporates, as an integral
part, a substance or human blood derivative, referred to in section 7.4 of Annex I
of the Directive, which is set out in Schedule I hereto, and the data on the tests
conducted in this connection which are required to assess the safety, quality and
usefulness of that substance or human blood derivative, taking account of the
intended purpose of the device,
- the clinical data referred to in Annex X of the Directive, which is set out in
Schedule X hereto,
- the draft label and, where appropriate, instructions for use.
4. The notified body must
4.1. examine and assess the documentation and verify that the type has been
manufactured in conformity with that documentation; it must also record the items
designed in conformity with the applicable provisions of the standards referred to in
regulation 6 of these regulations, as well as the items not designed on the basis of the
relevant provisions of the abovementioned standards;
4.2. carry out or arrange for the appropriate inspections and the tests necessary to
verify whether the solutions adopted by the manufacturer meet the essential requirements
of these regulations if the standards referred to in regulation 6 of these regulations
have not been applied; if the device is to be connected to other device(s) in order to
operate as intended, proof must be provided that it conforms to the essential requirements
when connected to any such device(s) having the characteristics specified by the
manufacturer;
4.3. carry out or arrange for the appropriate inspections and the tests necessary to
verify whether, if the manufacturer has chosen to apply the relevant standards, these
have actually been applied;
4.4. agree with the applicant on the place where the necessary inspections and
tests will be carried out.
5. If the type conforms to the provisions of these regulations, the notified body
issues the applicant with an EC type-examination certificate. The certificate must contain
the name and address of the manufacturer, the conclusions of the inspection, the
conditions of validity and the data needed for identification of the type approved. The
relevant parts of the documentation must be annexed to the certificate and a copy kept
by the notified body.
In the case of devices referred to in Annex I, section 7.4, first subparagraph of the
Directive, which is set out in Schedule I hereto, the notified body shall, as regards the
aspects referred to in that section, consult one of the competent bodies designated by
the Member States in accordance with Directive 65/65/EEC before taking a decision.
The notified body will give due consideration to the views expressed in this consultation
when making its decision. It will convey its final decision to the competent body
concerned.
B
In the case of devices referred to in Annex I, section 7.4, second subparagraph of
the Directive, which is set out in Schedule I hereto, the scientific opinion of the EMEA
must be included in the documentation concerning the device. The notified body will
give due consideration to the opinion of the EMEA when making its decision. The
notified body may not deliver the certificate if the EMEA’s scientific opinion is
unfavourable. It will convey its final decision to the EMEA.
6. The applicant must inform the notified body which issued the EC type-
examination certificate of any significant change made to the approved product.
Changes to the approved product must receive further approval from the notified
body which issued the EC type-examination certificate wherever the changes may
affect conformity with the essential requirements or with the conditions prescribed for
use of the product. This new approval must, where appropriate, take the form of a
supplement to the initial EC type-examination certificate.
7. Administrative provisions
7.1. The notified body must make available to the other notified bodies on request,
all relevant information on EC type-examination certificates and supplements issued,
refused or withdrawn.
7.2. Other notified bodies may obtain a copy of the EC type-examination
certificates and/or the supplements thereto. The Annexes to the certificates must be
made available to other notified bodies on reasoned application, after the manufacturer
has been informed.
7.3. The manufacturer or his authorized representative must keep with the technical
documentation copies of EC type-examination certificates and their additions for a
period ending at least five years after the last device has been manufactured.
7.4. When neither the manufacturer nor his authorized representative is established
in the Community or Malta, the obligation to keep available the technical documentation
shall fall to the person responsible for placing the device on the Community or Maltese
market or the importer referred to in Annex I of the Directive, which is set out in
Schedule I, section 13.3 (a) hereto.
Schedule IV
Based on Annex IV of the Directive
EC VERIFICATION
1. EC verification is the procedure whereby the manufacturer or his authorized
representative ensures and declares that the products which have been subject to the
procedure set out in section 4 conform to the type described in the EC type-examination
certificate and meet the requirements of these regulations which apply to them.
2. The manufacturer must take all the measures necessary to ensure that the
manufacturing process produces products which conform to the type described in the
EC type-examination certificate and to the requirements of the regulations which apply
to them. Before the start of manufacture, the manufacturer must prepare documents
defining the manufacturing process, in particular as regards sterilization where necessary,
together with all the routine, pre-established provisions to be implemented to ensure
homogeneous production and, where appropriate, conformity of the products with the
type described in the EC type-examination certificate and with the requirements of
these regulations which apply to them. The manufacturer must affix the CE marking in
accordance with regulation 9 of these regulations and draw up a declaration of
conformity.
In addition, for products placed on the market in sterile condition, and only for
those aspects of the manufacturing process designed to secure and maintain sterility,
the manufacturer must apply the provisions of Annex V of the Directive, which is set
out in Schedule V, sections 3 and 4 hereto.
3. The manufacturer must undertake to institute and keep up to date a systematic
procedure to review experience gained from devices in the post-production phase and
to implement appropriate means to apply any necessary corrective action. This
undertaking must include an obligation for the manufacturer to notify the competent
authorities of the following incidents immediately on learning of them
(i) any malfunction or deterioration in the characteristics and/or performance
of a device, as well as any inadequacy in the labelling or the instructions for use
which might lead to or might have led to the death of a patient or user or to a
serious deterioration in his state of health;
(ii) any technical or medical reason connected with the characteristics or
performance of a device for the reasons referred to in subparagraph (i) leading to
systematic recall of devices of the same type by the manufacturer.
4. The notified body must carry out the appropriate examinations and tests in
order to verify the conformity of the product with the requirements of the regulations
either by examining and testing every product as specified in section 5 or by examining
B
and testing products on a statistical basis as specified in section 6, as the manufacturer
decides. The aforementioned checks do not apply to those aspects of the manufacturing
process designed to secure sterility.
5. Verification by examination and testing of every product
5.1. Every product is examined individually and the appropriate tests defined in
the relevant standard(s) referred to in regulation 6 of these regulations or equivalent
tests must be carried out in order to verify, where appropriate, the conformity of the
products with the EC type described in the type-examination certificate and with the
requirements of the regulations which apply to them.
5.2. The notified body must affix, or have affixed its identification number to
each approved product and must draw up a written certificate of conformity relating to
the tests carried out.
6. Statistical verification
6.1. The manufacturer must present the manufactured products in the form of
homogeneous batches.
6.2. A random sample is taken from each batch. The products which make up the
sample are examined individually and the appropriate tests defined in the relevant
standard(s) referred to in regulation 6 of these regulations or equivalent tests must be
carried out to verify, where appropriate, the conformity of the products with the type
described in the EC type-examination certificate and with the requirements of the
regulations which apply to them in order to determine whether to accept or reject the
batch.
6.3. Statistical control of products will be based on attributes, entailing a sampling
system ensuring a limit quality corresponding to a probability of acceptance of 5 %,
with a non-conformity percentage of between 3 and 7 %. The sampling method will be
established by the harmonized standards referred to in regulation 6 of these regulations,
taking account of the specific nature of the product categories in question.
6.4. If the batch is accepted, the notified body affixes or has affixed its identification
number to each product and draws up a written certificate of conformity relating to the
tests carried out. All products in the batch may be put on the market except any in the
sample which failed to conform.
If a batch is rejected, the competent notified body must take appropriate measures
to prevent the batch from being placed on the market. In the event of frequent rejection
of batches, the notified body may suspend the statistical verification.
The manufacturer may, on the responsibility of the notified body, affix the notified
body’s identification number during the manufacturing process.
7. Administrative provisions
The manufacturer or his authorized representative must, for a period ending at
least five years after the last product has been manufactured, make available to the
national authorities
- the declaration of conformity,
- the documentation referred to in section 2,
- the certificates referred to in sections 5.2 and 6.4,
- where appropriate, the type-examination certificate referred to in Annex III
of the Directive, which is set out in Schedule III hereto.
8. Application to devices in Class IIa
In line with regulation 8.2 of these regulations, this Schedule may apply to products
in Class IIa, subject to the following exemptions
8.1. in derogation from sections 1 and 2, by virtue of the declaration of conformity
the manufacturer ensures and declares that the products in Class IIa are manufactured
in conformity with the technical documentation referred to in section 3 of Annex VII
of the Directive, which is set out in Schedule VII hereto and meet the requirements of
these regulations which apply to them;
8.2. in derogation from sections 1, 2, 5 and 6, the verifications conducted by the
notified body are intended to confirm the conformity of the products in Class IIa with
the technical documentation referred to in section 3 of Annex VII of the Directive,
which is set out in Schedule VII hereto.
9. Application to devices referred to in regulation 4.5 of these regulations
In the case of section 5, upon completing the manufacture of each batch of devices
referred to in regulation 4.5 of these regulations, and in the case of verification under
section 6, the manufacturer shall inform the notified body of the release of this batch of
devices and send to it the official certificate concerning the release of the batch of
human blood derivative used in the device issued by a State laboratory or a laboratory
designated for that purpose by a Member State in accordance with Article 4(3) of
Directive 89/381/EEC.
B
Schedule V
Based on Annex V of the Directive
EC DECLARATION OF CONFORMITY (Production quality assurance)
1. The manufacturer must ensure application of the quality system approved for
the manufacture of the products concerned and carry out the final inspection, as specified
in section 3, and is subject to the Community surveillance referred to in section 4.
2. The declaration of conformity is the part of the procedure whereby the
manufacturer who fulfils the obligations imposed by section 1 ensures and declares
that the products concerned conform to the type described in the EC type-examination
certificate and meets the provisions of these regulations which apply to them.
The manufacturer must affix the CE marking in accordance with regulation 9 of
these regulations and draw up a written declaration of conformity. This declaration
must cover a given number of identified specimens of the products manufactured and
must be kept by the manufacturer.
3. Quality system
3.1. The manufacturer must lodge an application for assessment of his quality
system with a notified body.
The application must include
- the name and address of the manufacturer,
- all the relevant information on the product or product category covered by
the procedure,
- a written declaration that no application has been lodged with any other
notified body for the same products,
- the documentation on the quality system,
- an undertaking to fulfil the obligations imposed by the quality system is
approved,
- an undertaking to maintain the practicability and effectiveness of the
approved quality system,
- where appropriate, the technical documentation on the types approved and
a copy of the EC type-examination certificates,
- an undertaking by the manufacturer to institute and keep up to date a
systematic procedure to review experience gained from devices in the post-
production phase and to implement appropriate means to apply any necessary
corrective action. This undertaking must include an obligation for the manufacturer
to notify the competent authorities of the following incidents immediately on
learning of them
(i) any malfunction or deterioration in the characteristics and/or
performance of a device, as well as any inadequacy in the labelling or the
instructions for use which might lead to or might have led to the death of a
patient or user or to a serious deterioration in his state of health;
(ii) any technical or medical reason connected with the characteristics
or performance of a device for the reasons referred to in subparagraph (i)
above leading to a systematic recall of devices of the same type by the
manufacturer.
3.2. Application of the quality system must ensure that the products conform to
the type described in the EC type-examination certificate.
All the elements, requirements and provisions adopted by the manufacturer for
his quality system must be documented in a systematic and orderly manner in the form
of written policy statements and procedures. This quality system documentation must
permit uniform interpretation of the quality policy and procedures such as quality
programmes, plans, manuals and records.
It must include in particular an adequate description of
(a) the manufacturer’s quality objectives;
(b) the organization of the business and in particular
- the organizational structures, the responsibilities of the managerial
staff and their organizational authority where manufacture of the products is
concerned,
- the methods of monitoring the efficient operation of the quality system
and in particular its ability to achieve the desired quality of product, including
control of products which fail to conform;
(c) the inspection and quality assurance techniques at the manufacturing stage
and in particular
- the processes and procedures which will be used, particularly as regards
sterilization, purchasing and the relevant documents,
B
- the product identification procedures drawn up and kept up to date
from drawings, specifications or other relevant documents at every stage of
manufacture;
(d) the appropriate tests and trials to be carried out before, during and after
manufacture, the frequency with which they will take place, and the test equipment
used; it must be possible adequately to trace back the calibration of the test
equipment.
3.3. The notified body must audit the quality system to determine whether it meets
the requirements referred to in section 3.2. It must presume that quality systems which
implement the relevant harmonized standards conform to these requirements.
The assessment team must include at least one member with past experience of
assessments of the technology concerned. The assessment procedure must include an
inspection on the manufacturer’s premises and, in duly substantiated cases, on the
premises of the manufacturer’s suppliers to inspect the manufacturing processes.
The decision must be notified to the manufacturer after the final inspection and
contain the conclusions of the inspection and a reasoned assessment.
3.4. The manufacturer must inform the notified body which approved the quality
system of any plan for substantial changes to the quality system.
The notified body must assess the changes proposed and verify whether after
these changes the quality system still meets the requirements referred to in section 3.2.
After the abovementioned information has been received the decision is notified to the
manufacturer. It must contain the conclusions of the inspection and a reasoned
assessment.
4. Surveillance
4.1. The aim of surveillance is to ensure that the manufacturer duly fulfils the
obligations imposed by the approved quality system.
4.2. The manufacturer authorizes the notified body to carry out all the necessary
inspections and must supply it with all relevant information, in particular
- the documentation on the quality system,
- the data stipulated in the part of the quality system relating to manufacture,
such as inspection reports and test data, calibration data, qualification reports of
the personnel concerned, etc.
4.3. The notified body must periodically carry out appropriate inspections and
assessments to make sure that the manufacturer applies the approved quality system
and supply the manufacturer with an assessment report.
4.4. In addition, the notified body may pay unannounced visits to the manufacturer.
At the time of such visits, the notified body may, where necessary, carry out or ask for
tests in order to check that the quality system is working properly. It must provide the
manufacturer with an inspection report and, if a test has been carried out, with a test
report.
5. Administrative provisions
5.1. The manufacturer must, for a period ending at least five years after the last
product has been manufactured, make available to the national authorities
- the declaration of conformity,
- the documentation referred to in the fourth indent of section 3.1,
- the changes referred to in section 3.4,
- the documentation referred to in the seventh indent of section 3.1,
- the decisions and reports from the notified body as referred to in sections
4.3 and 4.4,
- where appropriate, the type-examination certificate referred to in Annex III
of the Directive, which is set out in Schedule III hereto.
6. Application to devices in Class IIa
In line with regulation 8.2 of these regulations, this Schedule may apply to products
in Class IIa, subject to the following exemption
6.1. in derogation from sections 2, 3.1 and 3.2, by virtue of the declaration of
conformity the manufacturer ensures and declares that the products in Class IIa are
manufactured in conformity with the technical documentation referred to in section 3
of Annex VII of the Directive, which is set out in Schedule VII and meet the requirements
of these regulations which apply to them.
7. Application to devices referred to in regulation 4.5 of these regulations
Upon completing the manufacture of each batch of devices referred to in regulation
4.5 of these regulations, the manufacturer shall inform the notified body of the release
of the batch of devices and send to it the official certificate concerning the release of
the batch of human blood derivative used in the device issued by a State laboratory or
a laboratory designated for that purpose by a Member State in accordance with Article
4(3) of Directive 89/381/EEC.
B
Schedule VI
Based on Annex VI of the Directive
EC DECLARATION OF CONFORMITY (Product quality assurance)
1. The manufacturer must ensure application of the quality system approved for
the final inspection and testing of the product, as specified in section 3 and must be
subject to the surveillance referred to in section 4.
In addition, for products placed on the market in sterile condition, and only for
those aspects of the manufacturing process designed to secure and maintain sterility,
the manufacturer must apply the provisions of Annex V of the Directive, which is set
out in Schedule V, sections 3 and 4 hereto.
2. The declaration of conformity is the part of the procedure whereby the
manufacturer who fulfils the obligations imposed by section 1 ensures and declares
that the products concerned conform to the type described in the EC type-examination
certificate and meet the provisions of these regulations which apply to them.
The manufacturer affixes the CE marking in accordance with regulation 9 of these
regulations and draws up a written declaration of conformity. This declaration must
cover a given number of identified specimens of the products manufactured and be
kept by the manufacturer. The CE marking must be accompanied by the identification
number of the notified body which performs the tasks referred to in this Schedule.
3. Quality system
3.1. The manufacturer lodges an application for assessment of his quality system
with a notified body.
The application must include
- the name and address of the manufacturer,
- all the relevant information on the product or product category covered by
the procedure,
- a written declaration specifying that no application has been lodged with
any other notified body for the same products,
- the documentation on the quality system,
- an undertaking by the manufacturer to fulfil the obligations imposed by the
quality system approved,
- an undertaking by the manufacturer to keep the approved quality system
adequate and efficacious,
- where appropriate, the technical documentation on the types approved and
a copy of the EC type-examination certificates,
- an undertaking by the manufacturer to institute and keep up to date a
systematic procedure to review experience gained from devices in the post-
production phase and to implement appropriate means to apply any necessary
corrective action. This undertaking must include an obligation for the manufacturer
to notify the competent authorities of the following incidents immediately on
learning of them
(i) any malfunction or deterioration in the characteristics and/or
performance of a device, as well as any inadequacy in the labelling or the
instructions for use which might lead to or might have led to the death of a
patient or user or to a serious deterioration in his state of health;
(ii) any technical or medical reason connected with the characteristics
or the performance of a device for the reasons referred to in subparagraph (i)
leading to a systematic recall of devices of the same type by the manufacturer.
3.2. Under the quality system, each product or a representative sample of each
batch is examined and the appropriate tests defined in the relevant standard(s) referred
to in regulation 6 of these regulations or equivalent tests are carried out to ensure that
the products conform to the type described in the EC type-examination certificate and
fulfil the provisions of these regulations which apply to them. All the elements,
requirements and provisions adopted by the manufacturer must be documented in a
systematic and orderly manner in the form of written measures, procedures and
instructions. This quality system documentation must permit uniform interpretation of
the quality programmes, quality plans, quality manuals and quality records.
It must include in particular an adequate description of
- the quality objectives and the organizational structure, responsibilities and
powers of the managerial staff with regard to product quality,
- the examinations and tests that will be carried out after manufacture; it
must be possible to trace back the calibration of the test equipment adequately,
- the methods of monitoring the efficient operation of the quality system
- the quality records, such as reports concerning inspections, tests, calibration
and the qualifications of the staff concerned, etc.
The aforementioned checks do not apply to those aspects of the manufacturing
process designed to secure sterility.
B
3.3. The notified body audits the quality system to determine whether it meets the
requirements referred to in section 3.2. It must presume that quality systems which
implement the relevant harmonized standards conform to these requirements. The
assessment team must include at least one member with past experience of assessments
of the technology concerned. The assessment procedure must include an inspection on
the manufacturer’s premises and, in duly substantiated cases, on the premises of the
manufacturer’s suppliers to inspect the manufacturing processes. The decision must
be notified to the manufacturer. It must contain the conclusions of the inspection and a
reasoned assessment.
3.4. The manufacturer must inform the notified body which approved the quality
system of any plan for substantial changes to the quality system.
The notified body must assess the changes proposed and verify whether after
these changes the quality system will still meet the requirements referred to in section
3.2. After receiving the abovementioned information it must notify the manufacturer
of its decision. This decision must contain the conclusions of the inspection and a
reasoned assessment.
4. Surveillance
4.1. The aim of surveillance is to ensure that the manufacturer duly fulfils the
obligations imposed by the approved quality system.
4.2. The manufacturer must allow the notified body access for inspection purposes
to the inspection, testing and storage locations and supply it with all relevant information,
in particular
- the documentation on the quality system,
- the technical documentation,
- the quality records, such as inspection reports, test data, calibration data,
qualification reports of the staff concerned, etc.
4.3. The notified body must periodically carry out appropriate inspections and
assessments to make sure that the manufacturer applies the quality system and must
supply the manufacturer with an assessment report.
4.4. In addition, the notified body may pay unannounced visits to the manufacturer.
At the time of such visits, the notified body may, where necessary, carry out or ask for
tests in order to check that the quality system is working properly and that the production
conforms to the requirements of the regulations which apply to it. To this end, an
adequate sample of the final products, taken on site by the notified body, must be
examined and the appropriate tests defined in the relevant standard(s) referred to in
regulation 6 of these regulations or equivalent tests must be carried out. Where one or
more of the samples fails to conform, the notified body must take the appropriate
measures.
It must provide the manufacturer with an inspection report and, if a test has been
carried out, with a test report.
5. Administrative provisions
5.1. The manufacturer must, for a period ending at least five years after the last
product has been manufactured, make available to the national authorities
- the declaration of conformity,
- the documentation referred to in the seventh indent of section 3.1,
- the changes referred to in section 3.4,
- the decisions and reports from the notified body as referred to in the final
indent of section 3.4 and in sections 4.3 and 4.4,
- where appropriate, the certificate of conformity referred to in Annex III of
the Directive, which is set out in Schedule III hereto.
6. Application to devices in Class IIa
In line with regulation 8.2 of these regulations, this Schedule may apply to products
in Class IIa, subject to this derogation
6.1. by derogation from sections 2, 3.1 and 3.2 by virtue of the declaration of
conformity the manufacturer ensures and declares that the products in Class IIa are
manufactured in conformity with the technical documentation referred to in section 3
of Annex VII of the Directive, which is set out in Schedule VII and meet the requirements
of these regulations which apply to them.
B
Schedule VII
Based on Annex VII of the Directive
EC DECLARATION OF CONFORMITY
1. The EC declaration of conformity is the procedure whereby the manufacturer
or his authorized representative who fulfils the obligations imposed by section 2 and,
in the case of products placed on the market in a sterile condition and devices with a
measuring function, the obligations imposed by section 5 ensures and declares that the
products concerned meet the provisions of these regulations which apply to them.
2. The manufacturer must prepare the technical documentation described in section
3. The manufacturer or his authorized representative must make this documentation,
including the declaration of conformity, available to the national authorities for
inspection purposes for a period ending at least five years after the last product has
been manufactured.
Where neither the manufacturer nor his authorized representative are established
in the Community or Malta, this obligation to keep the technical documentation available
must fall to the person(s) who place(s) the product on the Community or Maltese
market.
3. The technical documentation must allow assessment of the conformity of the
product with the requirements of the regulations. It must include in particular
- a general description of the product, including any variants planned,
- design drawings, methods of manufacture envisaged and diagrams of
components, sub-assemblies, circuits, etc.,
- the descriptions and explanations necessary to understand the
abovementioned drawings and diagrams and the operations of the product,
- the results of the risk analysis and a list of the standards referred to in
regulation 6 of these regulations, applied in full or in part, and descriptions of the
solutions adopted to meet the essential requirements of the regulations if the
standards referred to in regulation 6 of these regulations have not been applied in
full,
- in the case of products placed on the market in a sterile condition, description
of the methods used,
- the results of the design calculations and of the inspections carried out,
etc.; if the device is to be connected to other device(s) in order to operate as
intended, proof must be provided that it conforms to the essential requirements
when connected to any such device(s) having the characteristics specified by the
manufacturer,
- the test reports and, where appropriate, clinical data in accordance with
Annex X of the Directive, which is set out in Schedule X hereto,
- the label and instructions for use.
4. The manufacturer shall institute and keep up to date a systematic procedure to
review experience gained from devices in the post-production phase and to implement
appropriate means to apply any necessary corrective actions, taking account of the
nature and risks in relation to the product. He shall notify the competent authorities of
the following incidents immediately on learning of them
(i) any malfunction or deterioration in the characteristics and/or performance
of a device, as well as any inadequacy in the labelling or the instructions for use
which might lead to or might have led to the death of a patient or user or to a
serious deterioration in his state of health;
(ii) any technical or medical reason connected with the characteristics on the
performance of a device for the reasons referred to in subparagraph (i) leading to
systematic recall of devices of the same type by the manufacturer.
5. With products placed on the market in sterile condition and Class I devices
with a measuring function, the manufacturer must observe not only the provisions laid
down in this Schedule but also one of the procedures referred to in Annexes IV, V or VI
of the Directive, which are set out in Schedules IV, V or VI hereto.
Application of the abovementioned Schedules and the intervention by the notified
body is limited to
- in the case of products placed on the market in sterile condition, only the
aspects of manufacture concerned with securing and maintaining sterile conditions,
- in the case of devices with a measuring function, only the aspects of
manufacture concerned with the conformity of the products with the metrological
requirements.
Section 6.1. of this Schedule is applicable.
6. Application to devices in Class IIa
In line with regulation 8.2 of these regulations, this Schedule may apply to products
in Class IIa, subject to the following derogation
6.1. where this Schedule is applied in conjunction with the procedure referred to
in Annexes IV, V or VII of the Directive, which are set out in Schedules IV, V or VI
hereto, the declaration of conformity referred to in the abovementioned Schedules
forms a single declaration. As regards the declaration based on this Schedule, the
manufacturer must ensure and declare that the product design meets the provisions of
these regulations which apply to it.
B
Schedule VIII
Based on Annex VIII of the Directive
STATEMENT CONCERNING DEVICES FOR SPECIAL PURPOSES
1. For custom-made devices or for devices intended for clinical investigations the
manufacturer or his authorized representative must draw up the statement containing
the information stipulated in section 2.
2. The statement must contain the following information
2.1. for custom-made devices
- data allowing identification of the device in question,
- a statement that the device is intended for exclusive use by a particular
patient, together with the name of the patient,
- the name of the medical practitioner or other authorized person who made
out the prescription and, where applicable, the name of the clinic concerned,
- the particular features of the device as specified in the relevant medical
prescription,
- a statement that the device in question conforms to the essential requirements
set out in Schedule I and, where applicable, indicating which essential requirements
have not been fully met, together with the grounds;
2.2. for devices intended for the clinical investigations covered by Annex X of the
Directive, which is set out in Schedule X hereto
- data allowing identification of the device in question,
- an investigation plan stating in particular the purpose, scientific, technical
or medical grounds, scope and number of devices concerned,
- the opinion of the ethics committee concerned and details of the aspects
covered by its opinion,
- the name of the medical practitioner or other authorized person and of the
institution responsible for the investigations,
- the place, starting date and scheduled duration for the investigations,
- a statement that the device in question conforms to the essential requirements
apart from the aspects covered by the investigations and that, with regard to these
aspects, every precaution has been taken to protect the health and safety of the
patient.
3. The manufacturer must also undertake to keep available for the competent
national authorities
3.1. for custom-made devices, documentation allowing an understanding of the
design, manufacture and performances of the product, including the expected
performances, so as to allow assessment of conformity with the requirements of these
regulations.
The manufacturer must take all the measures necessary to ensure that the
manufacturing process produces products which are manufactured in accordance with
the documentation mentioned in the first paragraph;
3.2. for devices intended for clinical investigations, the documentation must
contain
- a general description of the product,
- design drawings, methods of manufacture envisaged, in particular as regards
sterilization, and diagrams of components, sub-assemblies, circuits, etc.,
- the descriptions and explanations necessary to understand the
abovementioned drawings and diagrams and the operation of the product,
- the results of the risk analysis and a list of the standards referred to in
regulation 6 of these regulations, applied in full or in part, and descriptions of the
solutions adopted to meet the essential requirements of these regulations if the
standards referred to in regulation 6 of these regulations have not been applied,
- the results of the design calculations, and of the inspections and technical
tests carried out, etc.
The manufacturer must take all the measures necessary to ensure that the
manufacturing process produces products which are manufactured in accordance with
the documentation referred to in the first paragraph of this section.
The manufacturer must authorize the assessment, or audit where necessary, of the
effectiveness of these measures.
4. The information contained in the declarations concerned by this Schedule should
be kept for a period of time of at least five years.
B
Schedule IX
Based on Annex IX of the Directive
CLASSIFICATION CRITERIA
I.
DEFINITIONS
1. Definitions for the classification rules
1.1
Duration
Transient
Normally intended for continuous use for less than 60 minutes.
Short term
Normally intended for continuous use for not more than 30 days.
Long term
Normally intended for continuous use for more than 30 days.
1.2
Invasive devices
Invasive device
A device which, in whole or in part, penetrates inside the body, either through a
body orifice or through the surface of the body.
Body orifice
Any natural opening in the body, as well as the external surface of the eyeball, or
any permanent artificial opening, such as a stoma.
Surgically invasive device
An invasive device which penetrates inside the body through the surface of the
body, with the aid or in the context of a surgical operation.
For the purposes of these regulations devices other than those referred to in the
previous subparagraph and which produce penetration other than through an established
body orifice, shall be treated as surgically invasive devices.
Implantable device
Any device which is intended
-
to be totally introduced into the human body or,
-
to replace an epithelial surface or the surface of the eye,
by surgical intervention which is intended to remain in place after the procedure.
Any device intended to be partially introduced into the human body through
surgical intervention and intended to remain in place after the procedure for at least 30
days is also considered an implantable device.
1.3
Reusable surgical instrument
Instrument intended for surgical use by cutting, drilling, sawing, scratching,
scraping, clamping, retracting, clipping or similar procedures, without connection to
any active medical device and which can be reused after appropriate procedures have
been carried out.
1.4
Active medical device
Any medical device operation of which depends on a source of electrical energy
or any source of power other than that directly generated by the human body or gravity
and which acts by converting this energy. Medical devices intended to transmit energy,
substances or other elements between an active medical device and the patient, without
any significant change, are not considered to be active medical devices.
1.5.
Active therapeutical device
Any active medical device, whether used alone or in combination with other
medical devices, to support, modify, replace or restore biological functions or structures
with a view to treatment or alleviation of an illness, injury or handicap.
1.6
Active device for diagnosis
Any active medical device, whether used alone or in combination with other
medical devices, to supply information for detecting, diagnosing, monitoring or treating
physiological conditions, states of health, illnesses or congenital deformances.
1.7
Central circulatory system
For the purposes of these regulations, ‘central circulatory system’ means the
following vessels
arteriae pulmonales, aorta ascendens, arteriae coronariae, arteria carotis
communis, arteria carotis externa, arteria carotis interna, arteriae cerebrales,
B
truncus brachicephalicus, venae cordis, venae pulmonales, vena cava superior,
vena cava inferior,
1.8
Central nervous system
For the purposes of these regulations, ‘central nervous system’ means brain,
meninges and spinal cord.
II.
IMPLEMENTING RULES
2. Implementing rules
2.1 Application of the classification rules shall be governed by the intended
purpose of the devices.
2.2 If the device is intended to be used in combination with another device, the
classification rules shall apply separately to each of the devices. Accessories are
classified in their own right separately from the device with which they are used.
2.3 Software, which drives a device or influences the use of a device, falls
automatically in the same class.
2.4 If the device is not intended to be used solely or principally in a specific part
of the body, it must be considered and classified on the basis of the most critical specified
use.
2.5 If several rules apply to the same device, based on the performance specified
for the device by the manufacturer, the strictest rules resulting in the higher classification
shall apply.
III. CLASSIFICATION
1. Non-invasive devices
1.1
Rule 1
All non-invasive devices are in Class I, unless one of the rules set out hereinafter
applies.
1.2
Rule 2
All non-invasive devices intended for channelling or storing blood, body liquids
or tissues, liquids or gases for the purpose of eventual infusion, administration or
introduction into the body are in Class IIa
- if they may be connected to an active medical device in Class IIa or a
higher class,
- if they are intended for use for storing or channelling blood or other body
liquids or for storing organs, parts of organs or body tissues,
in all other cases they are in Class I.
1.3
Rule 3
All non-invasive devices intended for modifying the biological or chemical
composition of blood, other body liquids or other liquids intended for infusion into the
body are in Class IIb, unless the treatment consists of filtration, centrifugation or
exchanges of gas, heat, in which case they are in Class IIa.
1.4
Rule 4
All non-invasive devices which come into contact with injured skin
- are in Class I if they are intended to be used as a mechanical barrier, for
compression or for absorption of exudates,
- are in Class IIb if they are intended to be used principally with wounds which
have breached the dermis and can only heal by secondary intent,
- are in Class IIa in all other cases, including devices principally intended to
manage the micro-environment of a wound.
2. Invasive devices
2.1
Rule 5
All invasive devices with respect to body orifices, other than surgically invasive
devices and which are not intended for connection to an active medical device
-
are in Class I if they are intended for transient use,
- are in Class IIa if they are intended for short-term use, except if they are
used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or
in a nasal cavity, in which case they are in Class I,
- are in Class IIb if they are intended for long-term use, except if they are
used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or
in a nasal cavity and are not liable to be absorbed by the mucous membrane, in
which case they are in Class IIa.
All invasive devices with respect to body orifices, other than surgically invasive
devices, intended for connection to an active medical device in Class IIa or a higher
class, are in Class IIa.
B
Rule 6
All surgically invasive devices intended for transient use are in Class IIa unless
they are
- intended specifically to diagnose, monitor or correct a defect of the heart
or of the central circulatory system through direct contact with these parts of the
body, in which case they are in Class III,
-
reusable surgical instruments, in which case they are in Class I,
- intended to supply energy in the form of ionising radiation in which case
they in Class IIb,
- intended to have a biological effect or to be wholly or mainly absorbed in
which case they are in Class IIb,
- intended to administer medicines by means of a delivery system, if this is
done in a manner that is potentially hazardous taking account of the mode of
application, in which they are in Class IIb.
Rule 7
All surgically invasive devices intended for short-term use are in Class IIa unless
they are intended
- either specifically to diagnose, monitor or correct a defect of the heart or
of the central circulatory system through direct contact with these parts of the
body, in which case they are in Class III,
- or specifically for use in direct contact with the central nervous system,
in which case they are in Class III,
- or to supply energy in the form of ionising radiation in which case they
are in Class IIb,
- or to have a biological effect or to be wholly or mainly absorbed in which
case they are in Class III,
- or to undergo chemical change in the body, except if the devices are placed
in the teeth, or to administer medicines, in which case they are in Class IIb.
Rule 8
All implantable devices and long-term surgically invasive devices are in Class
IIb unless they are intended
-
to be placed in the teeth, in which case they are in Class IIa,
- to be used in direct contact with the heart, the central circulatory system
or the central nervous system, in which case they are in Class III,
- to have a biological effect or to be wholly or mainly absorbed, in which
case they are in Class III,
- or to undergo chemical change in the body, except if the devices are placed
in the teeth, or to administer medicines, in which case they are in Class III.
3. Additional rules applicable to active devices
3.1
Rule 9
All active therapeutic devices intended to administer or exchange energy are in
Class IIa unless their characteristics are such that they may administer or exchange
energy to or from the human body in a potentially hazardous way, taking account of
the nature, the density and site of application of the energy, in which case they are in
Class IIb.
All active devices intended to control or monitor the performance of active
therapeutic devices in Class IIb, or intended directly to influence the performance of
such devices are in Class IIb.
3.2
Rule 10
Active devices intended for diagnosis are in Class IIa
- if they are intended to supply energy which will be absorbed by the human
body, except for devices used to illuminate the patient’s body, in the visible
spectrum,
-
if they are intended to image in vivo distribution of radiopharmaceuticals,
- if they are intended to allow direct diagnosis or monitoring of vital
physiological processes, unless they are specifically intended for monitoring of
vital physiological parameters, where the nature of variations is such that it could
result in immediate danger to the patient, for instance variations in cardiac
performance, respiration, activity of CNS in which case they are in Class IIb.
Active devices intended to emit ionising radiation and intended for diagnostic
and therapeutic interventional radiology including devices which control or monitor
such devices, or which directly influence their performance, are in Class IIb.
B
Rule 11
All active devices intended to administer and/or remove medicines, body liquids
or other substances to or from the body are in Class IIa, unless this is done in a manner
- that is potentially hazardous, taking account of the nature of the substances
involved, of the part of the body concerned and of the mode of application in
which case they are in Class IIb.
3.3
Rule 12
All other active devices are in Class I.
4. Special Rules
4.1
Rule 13
All devices incorporating, as an integral part, a substance which, if used separately,
can be considered to be a medicinal product, as defined in Article 1 of Directive 65/65/
EEC, and which is liable to act on the human body with action ancillary to that of the
devices, are in Class III.
All devices incorporating, as an integral part, a human blood derivative are in
Class III.
4.2
Rule 14
All devices used for contraception or the prevention of the transmission of sexually
transmitted diseases are in Class IIb, unless they are implantable or long term invasive
devices, in which case they are in Class III.
4.3
Rule 15
All devices intended specifically to be used for disinfecting, cleaning, rinsing or,
where appropriate, hydrating contact lenses are in Class IIb.
All devices intended specifically to be used for disinfecting medical devices are
in Class IIa.
This rule does not apply to products that are intended to clean medical devices
other than contact lenses by means of physical action.
4.4
Rule 16
Non-active devices specifically intended for recording of X-ray diagnostic images
are in Class IIa.
4.5
Rule 17
All devices manufactured utilising animal tissues or derivatives rendered non-
viable are Class III except where such devices are intended to come into contact with
intact skin only.
5. Rule 18
By derogation from other rules, blood bags are in Class IIb.
Schedule X
Based on Annex X of the Directive
CLINICAL EVALUATION
1. General provisions
1.1 As a general rule, confirmation of conformity with the requirements
concerning the characteristics and performances, referred to in sections 1 and 3 of
Annex I of the Directive, which is set out in Schedule I to these regulations, under the
normal conditions of use of the device and the evaluation of the undesirable side-
effects must be based on clinical data in particular in the case of implantable devices
and devices in Class III. Taking account of any relevant harmonized standards, where
appropriate, the adequacy of the clinical data must be based on
1.1.1 either a compilation of the relevant scientific literature currently available
on the intended purpose of the device and the techniques employed as well as, if
appropriate, a written report containing a critical evaluation of this compilation;
1.1.2 or the results of all the clinical investigations made, including those carried
out in conformity with section 2.
1.2 All the data must remain confidential, in accordance with the provisions of
regulation 7.5 of these regulations.
2. Clinical investigations
2.1 Objectives
The objectives of clinical investigation are
- to verify that, under normal conditions of use, the performance of the
devices conform to those referred to in section 3 of Annex I of the Directive,
which is set out in Schedule I to these regulations, and
B
- to determine any undesirable side-effects, under normal conditions of
use, and assess whether they constitute risks when weighed against the intended
performance of the device.
2.2
Ethical considerations
Clinical investigations must be carried out in accordance with the Helsinki
Declaration adopted by the 18th World Medical Assembly in Helsinki, Finland, in
1964, as last amended by the 41st World Medical Assembly in Hong Kong in 1989. It
is mandatory that all measures relating to the protection of human subjects are carried
out in the spirit of the Helinski Declaration. This includes every step in the clinical
investigation from first consideration of the need and justification of the study to
publication of the results.
2.3
Methods
2.3.1 Clinical investigations must be performed on the basis of an appropriate
plan of investigation reflecting the latest scientific and technical knowledge and defined
in such a way as to confirm or refute the manufacturer’s claims for the device; these
investigations must include an adequate number of observations to guarantee the
scientific validity of the conclusions.
2.3.2 The procedures used to perform the investigations must be appropriate to
the device under examination.
2.3.3 Clinical investigations must be performed in circumstances similar to the
normal conditions of use of the device.
2.3.4 All the appropriate features, including those involving the safety and
performances of the device, and its effect on patients must be examined.
2.3.5 All adverse incidents such as those specified in regulation 7.4 of these
regulations must be fully recorded and notified to the Director General Health and the
Director of Consumer Affairs.
2.3.6 The investigations must be performed under the responsibility of a medical
practitioner or another authorised qualified person in an appropriate environment.
The medical practitioner or other authorised person must have access to the
technical and clinical data regarding the device.
2.3.7 The written report, signed by the medical practitioner or other authorised
person responsible, must contain a critical evaluation of all the data collected during
the clinical investigation.
Schedule XI
Based on Annex XI of the Directive
CRITERIA TO BE MET FOR THE DESIGNATION OF NOTIFIED BODIES
1. The notified body, its Director and the assessment and verification staff shall
not be the designer, manufacturer, supplier, installer or user of the devices which they
inspect, nor the authorized representative of any of these persons. They may not be
directly involved in the design, construction, marketing or maintenance of the devices,
nor represent the parties engaged in these activities. This in no way precludes the
possibility of exchanges of technical information between the manufacturer and the
body.
2. The notified body and its staff must carry out the assessment and verification
operations with the highest degree of professional integrity and the requisite competence
in the field of medical devices and must be free from all pressures and inducements,
particularly financial, which might influence their judgment or the results of the
inspection, especially from persons or groups of persons with an interest in the results
of the verifications.
Should the notified body subcontract specific tasks connected with the
establishment and verification of the facts, it must first ensure that the subcontractor
meets the provisions of the regulations and, in particular, of this Schedule. The notified
body shall keep at the disposal of the national authorities the relevant documents
assessing the subcontractor’s qualifications and the work carried out by the subcontractor
under these regulations.
3. The notified body must be able to carry out all the tasks assigned to such bodies
by one of Annexes II to VI of the Directive, which are set out in Schedules II to VI
hereto and for which it has been notified, whether these tasks are carried out by the
body itself or on its responsibility. In particular, it must have the necessary staff and
possess the facilities needed to perform properly the technical and administrative tasks
entailed in assessment and verification. It must also have access to the equipment
necessary for the verifications required.
4. The notified body must have
- sound vocational training covering all the assessment and verification
operations for which the body has been designated,
- satisfactory knowledge of the rules on the inspections which they carry out
and adequate experience of such inspections,
- the ability required to draw up the certificates, records and reports to
demonstrate that the inspections have been carried out.
B
5. The impartiality of the notified body must be guaranteed. Their remuneration
must not depend on the number of inspections carried out, nor on the results of the
inspections.
6. The body must take out civil liability insurance, unless liability is assumed by
the State under domestic legislation or the Member State itself carries out the inspections
directly.
7. The staff of the notified body are bound to observe professional secrecy with
regard to all information gained in the course of their duties (except vis-à-vis the
competent administrative authorities of the State in which their activities are carried
out) pursuant to these regulations or any provision of national law putting it into effect.
Schedule XII
Based on Annex XII of The Directive
The CE conformity marking shall consist of the initials “CE” taking the following
form
For the avoidance of doubt, it is hereby declared that the grid providing the
background in the above graduated drawing is not part of the CE marking.
- If the CE marking is reduced or enlarged the proportions given in the
above graduated drawing must be respected.
- The various components of the CE marking must have substantially the
same vertical dimension, which may not be less than 5mm.
This minimum dimension may be waived for small-scale devices.
Mitbug[ fl-Istamperija tal-Gvern – Printed at the Government Printing Press